1. Introduction
Spectrometry is a general term used for analyzing the specific spectrum and employed for the analysis of materials. Raman spectrometry [
1] is used to study molecular structures, identify molecules in unknown samples and give spectral information [
2]. Jin Xinghuan et al. designed the modern grating-based Raman spectrometer. The spectral resolution, wavelength range and the simple spectrometer’s structural parameters are the contemporary way of thinking geometrical models [
3].
Different designs of the Raman spectrometer have been published. Wang et al. designed a crossed Czerny–Turner spectrometer using convergent illumination of the grating for the fluorescence spectrum of the organic particles [
4]. Yinchao Zhang and Chen Wang proposed a double-grating monochromator with a different fiber arrangement [
5]. The Offner system with a convex grating is exceptional in aberration correction and distortion control [
5,
6]. Czerny and Turner first showed that the coma aberration introduced by the off-axis reflection from a spherical mirror could be corrected by a symmetrical but oppositely oriented spherical mirror in spectrometer design [
7]. The Czerny–Turner imaging spectrometer, a plane grating and two spherical mirrors are configured in a coma-free geometry with the Shafer equation used to resolve spectral intensity [
8].
Here, we introduce the design simulation of the Raman spectrometer’s optical structure using plain, reflecting grating and a focusing spherical mirror, shortening the overall system’s length to obtain higher resolution, thus enabling the development of a robust and low-cost Raman spectrometer for optical measurements in the laboratory.
2. Design Parameters for the Simulation of the Raman Spectrometer
The different designing parameters, such as the surface type of each lens, the radius of curvature, thickness and focal length, were chosen in the lens data editor window in Zemax [
9], as shown in
Table 1. We adopted the physical optics propagation (POP) algorithm over geometrical ray-tracing because geometrical ray-tracing can only be used when diffraction limits are negligible.
A typical spectrometer consists of a source light, an entrance slit, spherical collimating and focusing mirrors, grating and CCD (charge-coupled device) detector [
10]. Diffraction gratings are extraordinary due to their imaging efficiency in terms of resolution and lowest wavefront aberration. However, they are usually unable to achieve a flat focus curve [
11], so a spherical mirror is needed to concentrate the subject on the image surface.
Figure 1A shows the schematic diagram representing the layout for the Raman spectrometer’s optical configuration. Raman scattered light from the sample is focused on the entrance slit with the focusing lens of numerical aperture 0.16, further collimated with collimating mirrors and incident on the diffraction grating. The Raman signal is diffracted from the grating and split up into its components focused on the CCD camera through the focusing mirror (
Figure 1A). The radius of the curvature of the collimating and focusing mirror is
and
, respectively.
A sample is usually defined before the entrance slit of the spectrometer. The upper and the lower energy levels of the sample have certain vibrational levels in an electronic state which can be represented by horizontal lines in an energy state. The vibrational transition can occur when molecules move from one vibrational level to another energy level. By monitoring the vibrational transition, one can identify and study the functional groups or chemical substances.
We chose a 527 nm center wavelength laser as the light source, whose energy was equal to the energy bandgap between the sample’s upper and lower energy levels. The sample absorbs the incident light and excites to the upper energy level. Electrons in the excited state decay to a lower vibrational level through a non-radiative process and then decay to a lower energy level and emit photons of different energy called the Raman spectrum. The Raman spectrum of a sample is usually distributed in the range from 200 to 2500 cm
[
3] and the Raman spectrum wavelength range observed in the experiment was from 530 nm to 630 nm. A 1024 × 1024 pixels charge-coupled device (CCD) was chosen to record the spatial profile. The scale of each pixel was (5 × 5)
m and the receiving surface of CCD was 1.25 cm
. The numerical aperture (NA) of the object side was chosen to be 0.16, while the grating constant was 0.100 lines/
m and the width of the slit was 30
m. We chose the CCD size, which was comparable to the image plane [
12].
The initial parameters of the system design are shown in
Table 2.
The ray tracing of the Raman spectrometer is shown in
Figure 1B. The angles
and
are the diffraction angles of the two edge wavelengths after passing through the diffraction grating; they intersect the focusing mirror at A and B.
and
are the angles between the reflected light of the two edge wavelengths on the focusing mirror and the horizontal line. Finally, they gather on points C and D of the linear CCD. For the references of the mathematical formulas, we refer the reader to the references [
4,
5,
8,
10]. Using the diffraction grating formula
where
d is the grating constant, the diffraction angles
and
can be calculated as
Taking the diffraction order equal to 1 (
n = 1), the minimum size (length) of the focusing mirror to reflect the two edge wavelengths according to the geometrical structure is given as
After reflection from the focusing mirror, the horizontal angle of the diffracted rays is changed by 2
as given below:
Finally, the size of the image is calculated using the geometrical structure
Equation (
6) shows the minimum length of the CCD detector surface. Based on the original configuration and using the Equations (1)–(6), all the necessary parameters to design the Raman spectrometer can be determined.
3. Simulation Results and Analysis
After optimization,
x increased to 3 mm. The light diffracted by the grating had a certain convergence angle, corresponding to
x, and it was reduced to 4.5 mm, whereas other parameters, such as detector size and pixels, and mirror focal length remained unchanged. The optimized spot diagram of the adjacent wavelengths is shown in
Figure 2A. The image formed on the spectrometer’s image plane showed that the adjacent wavelengths at different positions were well separated and distinguishable, as shown in
Figure 2B.
It can be seen, from
Figure 2B, that the energy spectrum of the neighboring wavelengths at various wavelengths could be easily distinguished. The modulation transfer function (MTF) diagram, which describes the spatial response of our Raman spectrometer at the center wavelength (575 nm), is shown in
Figure 3. It gives the contrast ratio between the input and output images. The MTF curve was approximately the same in the whole band. Therefore, the curve of the center wavelength is demonstrated in
Figure 3. The MTF shows how the spatial frequency content of the entity is correctly transferred to the image and describes the performance of the optical system. The higher the value of the MTF, the greater the image quality of the device is. As shown in
Figure 3, when the spatial frequency was 10 per mm
, the optical transfer function (OTF) was more significant than 0.8, which means that the efficiency of the designed spectrometer system was higher [
13].
All results were obtained in the Sequential Mode of Zemax. In this mode, the slit width can be measured only by increasing the height of the point light source. Following the above settings to trace the light, the results obtained in this mode are closer to the real-world observations.
The intensity distribution of adjacent wavelengths shown in
Figure 4A shows the intensity distribution of all the different wavelengths expressed by the geometric linear response function. The linear response function is the cross-sectional representation of the image density distribution [
14]. The length and width of the receiving surface of the detector were 10 mm and 5 mm, respectively. The wavelength band imaged on the detector and the images produced by the two edge rays were similar to the two ends of the detector (
Figure 4A). Using the CCD, the light of the neighboring wavelengths of separate bands could also be separated, which indicates that the illumination of the adjacent wavelengths was easily separable and distinguishable with the present detector pixels. This finding further shows the functional viability of this system. The radiant flux received by the detector surface per unit area and beam spot size at the detector surface for a 580 nm wavelength is shown in
Figure 4B.
Illuminations at longer wavelengths result in a decrease in Raman signals. There is less illumination for a higher wavelength (total irradiance) than the lower wavelength in the Raman spectrometer, so the above relation worked correctly in our designed Raman spectrometer. The optical term encircled energy describes the energy concentration in the optical picture or predicted laser in a defined area. We observed the energy concentration variations concerning the change in the size of the image on the detector; as the point image’s size increased, the fraction of enclosed energy increased, which is consistent with what reported in [
15].
4. Conclusions
The Raman spectrometer utilizing diffraction grating incorporates the standard Czerny–Turner system features with the flat-field grating spectrometer system. The spectrometer system’s modeling approach is suggested by introducing the corresponding relationship among pixel and wavelength, linear CCD receiving surface length and image surface dimension. The excitation light with a center wavelength of 527 nm was used for simulation and optimization in the Zemax optical modeling program. Spot diagram, irradiance plot, MTF and geometric encircled energy for the Raman spectrum of the sample, which lie in the visible range, were simulated. The analysis showed that the image efficiency of the system was higher and Raman spectra could be obtained from a sample using the above-designed Raman spectrometer system. This study could pave the way for developing a robust, miniature, compact and low-cost Raman spectrometer for optical measurements.