1. Introduction
With the increasing bacteria hazards in environment, food system, and health care, biomedical sensors have raised a lot of attention since its vast application in chemical and biomedical sensing. An ideal biosensor should be able to provide fast, accurate, sensitive and cost-effective analysis [
1], with the potential of easily operation and integration ability at the same time. Waveguide based biomedical sensor stands out for its merits of high sensitivity, integration ability, free from electromagnetic influence, material selection flexibility, etc. Furthermore, optical systems have the merit of fast response which is a crucial factor in bedside diagnosis. The waveguide sensor in this paper is proposed for monitoring bacteria in fluid, e.g., the existence and concentration of
E. coli in patients’ drain fluid after colon anastomosis. The waveguide utilizes evanescent wave for sensing, which is sensitive to the targets on the surface of the waveguide. A vertical waveguide structure is used in this design to increase the waveguide sensitivity. The surface of the waveguide can then be functionalized with antibodies to fix targets on it for detection.
2. Material and Design
The waveguide was designed vertically to generate evanescent waves on both left and right surfaces for sensing, as shown in
Figure 1a. This increases the sensing area and optical energy distributed in the cladding to interact with immobilized targets. Both aspects will increase waveguide sensitivity. The fundamental mode is simulated and shown in
Figure 1b. SU-8 is chosen as waveguide core material because of its suitable optical properties: transparent above 400 nm and with a refractive index of 1.57 at 1.3 um wavelength. It is a photoresist that can be fabricated by E-beam lithography, a maskless direct writing method which can guarantee fast prototyping, sub-micron and high precision fabrication [
2]. Compared with traditional materials in the fabrication of this proposed vertical large height-width ratio structure, SU-8 outstands other materials as no dry etching is needed, which guarantees the side wall surface quality. Moreover, it is very stable after polymerization [
3], which is crucial as a chemical or biomedical sensor material. After chemical treatment, the surface of SU-8 can be functionalized with antibodies [
4,
5,
6,
7,
8,
9] and this enables its applications in biomedical field.
SU-8 waveguide structure parameters were designed by COMSOL simulation in previous research [
10], optimizing optical energy for sensing and at the same time considering fabrication feasibility. When fabricating waveguides, SiO2 was first thermally grown on Si wafer as an isolation layer. Afterwards, the waveguides were fabricated by E-beam lithography in Kavli by Raith EBPG-5200. Then, a slow ramp-up and cool-down baking process were used to improve the adhesion of SU-8 to the substrate. After development, a protection photoresist layer was coated on the top and then the wafer was diced into chips ready for the next step.
3. Surface Functionalization
Surface functionalization is the process of attaching antigen/antibody to the surface of a waveguide. This includes surface modification and antibody/antigen immobilization [
11]. Surface modification is aimed at creating a surface with the property that can react with the antibodies afterwards. The general procedure starts with creating –OH groups on the surface and then creating amine groups on the previous layer. Afterwards, by applying glutaraldehyde on the chip, the amine groups and antibodies can be bridged. After the antibodies are immobilized on the surface, the wafer is successfully functionalized and is ready for antigen immobilization.
A functionalization method was designed and demonstrated to be effective to fix antibodies and capture
E. coli on the SU-8 surface. The procedure is illustrated in
Figure 2.
The SU-8 surface is activated by oxygen plasma for 5 min at a pressure of 200 mTorr and a power of 30 W [
4,
12]. Through plasma treatment, the epoxy group is opened and –OH groups are created on the SU-8 surface.
This step creates amine groups on the surface. APTES is diluted in 75% ethanol and 25% water at a 1:100 ratio. Then the chips are incubated overnight at room temperature.
The wafer is activated using a 1.2% (v/v) glutaraldehyde solution in 1× phosphate buffer saline (1× PBS) at room temperature for 2 h. Then it is washed excessively with DI water and 1× PBS. Glutaraldehyde is a di-aldehyde, which contains 2 active aldehyde moieties (=CHO). These aldehydes react with amine groups which are present on the APTES and on the antibody. In this way, glutaraldehyde bridges the APTES with the antibody.
The original antibody solution is diluted in 1× PBS to make an antibody solution of 10 µg/mL. The solution on the surface is dropped and incubated for 1 h at room temperature.
This step is to create an E. coli-antibody association, through the process of applying E. coli on the functionalized specimens and incubating the bacteria for a defined time period (2 h). Then, wash the incubated specimens with DI water.
After the steps designated above, the chips are functionalized with E. coli and ready for the following optical measurements.
4. Measurements
After functionalization, the treated SU-8 waveguide chips were incubated with
E. coli solutions with concentrations of 10
2 CFU/mL, 10
3 CFU/mL, 10
4 CFU/mL, 10
5 CFU/mL, 10
6 CFU/mL for 2 h to immobilize
E. coli on the waveguide sensing surface. A negative control group was also set up with the same chemical treatment but without antibody treatment. The incubation is shown in
Figure 3. When incubated with
E. coli solutions of different concentrations, different numbers of
E. coli will be fixed on the surface. After flush away the excess uncaptured
E. coli, the waveguide with captured bacteria on the surface will absorb more optical energy than the reference waveguide without antibody treatment. In this way, a bacteria concentration-absorption relationship can be obtained. After calibration, the waveguide can be used to detect unknown concentration solutions.
5. Results
After incubation, the chips were fixed and then measured. The absorption of waveguides with different
E. coli treatments was measured. The results are shown in
Figure 4. This figure shows an increasing trend with the increase in
E. coli concentration, which demonstrates its sensitivity to
E. coli. The waveguide is able to detect bacteria when the concentration is 10
2 CFU/mL. Moreover, it becomes more significant when the concentration reaches 10
4 CFU/mL. The sign of leakage in patients is normally when the
E. coli concentration quickly ramps up to 10
5 CFU/mL in the drain fluid after surgery [
13]. The proposed system meets therefore the medical requirements and further improvements will make such a system vital for early diagnostics.
Author Contributions
Y.X. conceived and designed the experiments, performed the experiments; G.P. and P.F. discussed and gave suggestions on the design and experiments; Y.X. and Y.Z. conducted the simulations; Y.X. and L.O. did the biomedical treatments; Y.X. wrote the paper and P.F. and G.P. revised the paper.
Acknowledgments
The authors would like to acknowledge to staff from EKL, Kavli Lab and chemical department for their assistance during the fabrication and the discussion during that time. The first author Yu Xin would like to acknowledge the Chinese Scholarship Council (CSC) and the Micro-nano System Lab in National University of Defense Technology to support her doing research in TU Delft.
Conflicts of Interest
The authors declare no conflict of interest.
References
- Estevez, M.C.; Alvarez, M.; Lechuga, L.M. Integrated optical devices for lab-on-a-chip biosensing applications. Laser Photonics Rev. 2012, 6, 463–487. [Google Scholar] [CrossRef]
- Sum, T.C.; Bettiol, A.A.; van Kan, J.A.; Watt, F.; Pun, E.Y.B.; Tung, K.K. Proton beam writing of low-loss polymer optical waveguides. Appl. Phys. Lett. 2003, 83, 1707–1709. [Google Scholar] [CrossRef]
- Microchem. SU-8 3000 Permanent Epoxy Negative Photoresist. Available online: http://www.microchem.com/pdf/SU-83000DataSheet.pdf (accessed on 5 October 2017).
- Grimaldi, I.A.; Testa, G.; Persichetti, G.; Loffredo, F.; Villani, F.; Bernini, R. Plasma functionalization procedure for antibody immobilization for SU-8 based sensor. Biosens. Bioelectron. 2016, 86, 827–833. [Google Scholar] [CrossRef] [PubMed]
- Deepu, A.; Sai, V.V.R.; Mukherji, S. Simple surface modification techniques for immobilization of biomolecules on SU-8. J. Mater. Sci. Mater. Med. 2009, 20 (Suppl. 1), 25–28. [Google Scholar] [CrossRef] [PubMed]
- Shew, B.Y.; Cheng, Y.C.; Tsai, Y.H. Monolithic SU-8 micro-interferometer for biochemical detections. Sens. Actuators A Phys. 2008, 141, 299–306. [Google Scholar] [CrossRef]
- Tao, S.L.; Popat, K.C.; Norman, J.J.; Desai, T.A. Surface modification of SU-8 for enhanced biofunctionality and nonfouling properties. Langmuir 2008, 24, 2631–2636. [Google Scholar] [CrossRef] [PubMed]
- Jiang, L.; Gerhardt, K.P.; Myer, B.; Zohar, Y.; Pau, S. Evanescent-Wave Spectroscopy Using an SU-8 Waveguide for Rapid Quantitative Detection of Biomolecules. J. Microelectromech. Syst. 2008, 17, 1495–1500. [Google Scholar] [CrossRef]
- Joshi, M.; Pinto, R.; Rao, V.R.; Mukherji, S. Silanization and antibody immobilization on SU-8. Appl. Surf. Sci. 2007, 253, 3127–3132. [Google Scholar] [CrossRef]
- Xin, Y.; Pandraud, G.; van Langen-Suurling, A.; French, P. Tapered SU8 waveguide for evanescent sensing by single-step fabrication. In Proceedings of the 2017 IEEE SENSORS, Glasgow, UK, 29 October–1 November 2017; pp. 1–3. [Google Scholar]
- Boiragi, I.; Makkar, R.; Choudhury, B.D.; Mallik, A.; Chalapathi, K.; Sebastian, J. SU-8 Polymer Based Waveguide Biochemical Sensor for Medical Diagnostic Application. In Proceedings of the ICOP 2009-International Conference on Optics and Photonics CSIO, Chandigarh, India, 30 October–1 November 2009; pp. 2–5. [Google Scholar]
- Hamdi, F.S.; Woytasik, M.; Couty, M.; Francais, O.; le Pioufle, B.; Dufour-Gergam, E. Low temperature irreversible poly(DiMethyl) Siloxane packaging of silanized SU8 microchannels: Characterization and lab-on-chip application. J. Microelectromech. Syst. 2014, 23, 1015–1024. [Google Scholar] [CrossRef]
- Komen, N.; Slieker, J.; Willemsen, P.; Mannaerts, G.; Pattyn, P.; Karsten, T.; de Wilt, H.; van der Harst, E.; van Leeuwen, W.; Decaestecker, C.; et al. Polymerase chain reaction for Enterococcus faecalis in drain fluid: The first screening test for symptomatic colorectal anastomotic leakage. The Appeal-study: Analysis of Parameters Predictive for Evident Anastomotic Leakage. Int. J. Colorectal Dis. 2013, 29, 15–21. [Google Scholar] [CrossRef]
| Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2018 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).