Next Article in Journal
LSPR-Based Label Free Sensing in Absorbance Mode: Sensing of Therapeutic Antibodies in Serum
Previous Article in Journal
Exploiting Brillouin Sensors for Refractive Index Measurements
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Abstract

Monitoring Hemolysis with a Parallel Plate Capacitor †

by
Anays Acevedo-Barrera
1,
Doris Atenea Cerecedo-Mercado
2 and
Augusto García-Valenzuela
1,*
1
Centro de Ciencia Aplicadas y Desarrollo Tecnológico, Universidad Nacional Autónoma de México, Apartado Postal 70-186, Ciudad de México 04510, Mexico
2
Laboratorio de Hematobiología, Escuela Nacional de Medicina y Homeopatía, Instituto Politécnico Nacional, Guillermo Massieu Helguera no. 239, La Escalera Ticomán, Ciudad de México 07320, Mexico
*
Author to whom correspondence should be addressed.
Presented at the 5th International Symposium on Sensor Science (I3S 2017), Barcelona, Spain, 27–29 September 2017.
Proceedings 2017, 1(8), 726; https://doi.org/10.3390/proceedings1080726
Published: 4 December 2017
A parallel plate capacitor can be used as a high resolution sensor of the electrical characteristics of a liquid sample contained by the capacitor. We are interested specifically in using parallel plate capacitors to characterize and monitor processes in biofluids. In this case, only dynamic measurements of the capacitance and resistance at a fixed frequency are generally convenient. In biofluids, the high polarizability of water and the electrical conduction by ions in solution, determine the electrical response of the system. When there are biological cells in suspension, the ion density is, in general, different inside the cell than outside it. Then, cells may act as strong polarizable entities, through the so-called Maxwell–Wagner effect. In addition, the formation of an electrical double layer at the interfaces between the biological fluid and the electrodes can influence both, the capacitance and resistance of the system. All these electrical effects combined must be taken into account to interpret measurements of the equivalent capacitance and resistance at a given frequency of the system. In this work, we combine all the mentioned effects into a practical model appropriate for frequencies above 10 kHz. Then, we present results of monitoring the process of hemolysis in human blood samples exposed to different values of osmotic stress. We use the model to interpret the obtained results. Overall, our conclusion is that electrical-impedance monitoring of a parallel plate capacitor containing blood is in fact sensitive to the hemolysis process and, more generally, can be a competitive way to characterize the osmotic fragility of biological cells. We discuss how to optimize measurements for future development of ad hoc sensors.

Share and Cite

MDPI and ACS Style

Acevedo-Barrera, A.; Cerecedo-Mercado, D.A.; García-Valenzuela, A. Monitoring Hemolysis with a Parallel Plate Capacitor. Proceedings 2017, 1, 726. https://doi.org/10.3390/proceedings1080726

AMA Style

Acevedo-Barrera A, Cerecedo-Mercado DA, García-Valenzuela A. Monitoring Hemolysis with a Parallel Plate Capacitor. Proceedings. 2017; 1(8):726. https://doi.org/10.3390/proceedings1080726

Chicago/Turabian Style

Acevedo-Barrera, Anays, Doris Atenea Cerecedo-Mercado, and Augusto García-Valenzuela. 2017. "Monitoring Hemolysis with a Parallel Plate Capacitor" Proceedings 1, no. 8: 726. https://doi.org/10.3390/proceedings1080726

APA Style

Acevedo-Barrera, A., Cerecedo-Mercado, D. A., & García-Valenzuela, A. (2017). Monitoring Hemolysis with a Parallel Plate Capacitor. Proceedings, 1(8), 726. https://doi.org/10.3390/proceedings1080726

Article Metrics

Back to TopTop