3.1. Total Stopping Power Data Analysis
In
Figure 9,
Figure 10,
Figure 11,
Figure 12,
Figure 13,
Figure 14,
Figure 15,
Figure 16,
Figure 17,
Figure 18 and
Figure 19, the ratios of total SP values obtained in experiments and those obtained in SRIM calculations are shown for HIs from Ar to U. Some details on the
data used in the subsequent analysis should be mentioned. Available original data (with original errors) were preferably used in the analysis. In the absence of tabulated data in the original works, SP data were used from the database [
2].
The
values as a function of
for Ar to U ions were fitted using the correction function in the framework of the weighted LSM procedure:
where
,
, and
are fitting parameters.
The ratios of the
and
values for Ar and K ions are shown in
Figure 9. The Ar and K
data were taken from the respective tables [
9,
11,
20,
26,
31], whereas the Ar data obtained by Giessel et al. were taken from the database [
2]. The low-velocity Ar SP data [
22] were obtained using tabulated input and output energies (
and
, respectively), target thickness
, and the relationship:
As one can see in the figure, the Ar data [
26] obtained recently and the earlier data of Giessel et al. are in agreement with each other, whereas the data [
20] at
are inconsistent with them. The Ar and K data [
11] for a thin target at
are also in disagreement with the data [
9,
22]. The data [
11,
20] were excluded from the data fit. As one could expect, adding the data [
11,
20] led to an increase in fitted parameter errors and in the reduced
value determining fit quality. The fitting parameter values thus obtained for the Ar and K data are listed in
Table 1.
Figure 10 shows the Ca and Sc SP data [
9,
10,
11,
26,
27,
31,
32] in comparison to SRIM calculations. As one can see, the fitting curve does not provide the match of the Ca data [
9,
26,
27] using the function determined by Equation (
8), even when omitting the data [
31] at
. It is implied that the data [
9] at relatively high velocities and the data [
27] at relatively low velocities are reliable. In this regard, it is worth noting that the reduced
value obtained for the Ca data fit (1.66) is greater than the one obtained for Sc (0.18), indicating a good match, if the data [
11] were ignored. The last instances correspond to the noticeably lower SP values than those obtained in [
9]. The fitting parameter values thus obtained for the Ca and Sc data are listed in
Table 1.
Figure 9.
The same as in
Figure 8, but for the data from [
2,
9,
10,
11,
20,
22,
26,
31] for Ar and K ions only (
upper and
bottom panels, respectively). The results of data fits with Equation (
8) are shown by dashed lines (the data [
11,
20] were excluded from these fits). Relative velocity
is shown in the upper axes of the panels for orientation.
Figure 9.
The same as in
Figure 8, but for the data from [
2,
9,
10,
11,
20,
22,
26,
31] for Ar and K ions only (
upper and
bottom panels, respectively). The results of data fits with Equation (
8) are shown by dashed lines (the data [
11,
20] were excluded from these fits). Relative velocity
is shown in the upper axes of the panels for orientation.
Figure 10.
The same as in
Figure 9, but for the data from [
9,
10,
11,
26,
27,
31,
32] for Ca and Sc ions (
upper and
bottom panels, respectively). See the text for details.
Figure 10.
The same as in
Figure 9, but for the data from [
9,
10,
11,
26,
27,
31,
32] for Ca and Sc ions (
upper and
bottom panels, respectively). See the text for details.
Figure 11 shows the Ti and V SP data in comparison to the SRIM calculations. The Ti data [
20,
21,
23,
31,
32,
33] at relatively high velocities (
) and those of Giessel et al. are in satisfactory agreement with each other (the data [
21] and those of Giessel et al. were taken from the database [
2]). The data [
34] from range measurements (designated by R in the figure) agree with the low-velocity data [
10] and those obtained at relatively high velocities. These data [
34] were originally assigned to the electronic SP and were taken from the database [
2]. They were attributed to the total SP (it seems impossible to separate the inelastic and elastic components in range measurements, considering the dominance of elastic collisions at the end of the range). The V data [
31] were fitted with a constant value for the
ratios. The results of data fitting for the Ti and V ions are listed in
Table 1.
Figure 11.
The same as in
Figure 9 and
Figure 10, but for the data from [
2,
10,
20,
21,
23,
31,
32,
33] for Ti and V ions (
upper and
bottom panels, respectively). See the text for details.
Figure 11.
The same as in
Figure 9 and
Figure 10, but for the data from [
2,
10,
20,
21,
23,
31,
32,
33] for Ti and V ions (
upper and
bottom panels, respectively). See the text for details.
Figure 12.
The same as in
Figure 9,
Figure 10 and
Figure 11, but for the data from [
2,
7,
10,
11,
31,
32] for Cr and Mn ions (
upper and
bottom panels, respectively). See the text for details.
Figure 12.
The same as in
Figure 9,
Figure 10 and
Figure 11, but for the data from [
2,
7,
10,
11,
31,
32] for Cr and Mn ions (
upper and
bottom panels, respectively). See the text for details.
Figure 12 shows a comparison of the Cr and Mn SP data [
7,
10,
11,
31,
32] to SRIM calculations (the data [
7] were taken from the database [
2]). The Cr data [
32] at
and the Mn data [
31] at
are in some disagreement with the data [
7] obtained with small errors later. The fitting curve does not provide a match of the Cr data [
7] with the data [
10] at low velocities (similarly to the Ca data). The reason for this is that in the fitting procedure, the precision data [
7] contributed more weights than the less precise ones [
10,
31,
32]. The data [
11] at
disagree with the data [
10] at the same velocity, as shown in the figure, and were ignored in the fitting procedure. The results of the fitting are listed in
Table 1.
Table 1.
Fitting parameter values
,
, and
as obtained for Ar to U ratios
fitted with Equation (
8). The ion symbols and ranges of Equation (
8) applicability for specified ions are listed in the first and last columns, respectively. The results of data fitting are shown in
Figure 9,
Figure 10,
Figure 11,
Figure 12,
Figure 13,
Figure 14,
Figure 15,
Figure 16,
Figure 17,
Figure 18 and
Figure 19.
Table 1.
Fitting parameter values
,
, and
as obtained for Ar to U ratios
fitted with Equation (
8). The ion symbols and ranges of Equation (
8) applicability for specified ions are listed in the first and last columns, respectively. The results of data fitting are shown in
Figure 9,
Figure 10,
Figure 11,
Figure 12,
Figure 13,
Figure 14,
Figure 15,
Figure 16,
Figure 17,
Figure 18 and
Figure 19.
Ion |
|
|
| Range |
---|
Ar 1 |
|
|
| 0.05–0.8 |
K 2 |
|
|
| 0.05–0.6 |
Ca 3 |
|
|
| 0.06–0.9 |
Sc 2 |
|
|
| 0.06–0.8 |
Ti |
|
|
| 0.06–0.8 |
V |
| | | 0.2–0.6 |
Cr 2 |
|
|
| 0.06–0.7 |
Mn 2 |
|
|
| 0.06–0.7 |
Fe |
|
|
| 0.04–0.7 |
Co |
|
|
| 0.04–0.5 |
Cu 2 |
|
|
| 0.05–0.8 |
Ge |
|
|
| 0.04–0.9 |
Br |
|
|
| 0.04–0.9 |
Kr 4 |
|
|
| 0.04–0.8 |
Y |
|
|
| 0.04–0.8 |
Ag |
|
|
| 0.06–0.7 |
I |
| | | 0.2–0.6 |
Xe |
|
|
| 0.05–0.8 |
Au |
|
|
| 0.04–0.4 |
Pb |
|
|
| 0.04–0.8 |
U |
|
|
| 0.04–0.8 |
Figure 13 shows the Fe and Co SP data [
7,
10,
32,
33,
35] in comparison to SRIM calculations. In contrast to the Ca and Cr data analysis (see
Figure 10 and
Figure 12), fitting curves provide an acceptable match of the Fe and Co data [
7] (taken from the database [
2]) at
with the data [
10] at low velocities (
and 0.546 were obtained for Fe and Co data fitting, respectively). The values of the fitting parameters are listed in
Table 1.
Figure 14 shows the Ni and Cu SP data [
7,
10,
21,
23,
31,
32,
33,
35] in comparison to SRIM calculations. The Ni data [
21,
35] at
are in good agreement with each other and exceed SRIM calculations within ∼5%. At
, the data [
7,
35] significantly exceed SRIM calculations (the data [
7] were taken from the database [
2]). Thus, the Ni data were not processed with Equation (
8). As for the Cu data, the fitting curve provides a match of the high velocity data [
7,
10,
21,
23,
31,
32,
35] with the low-velocity ones [
10]. Ignoring the data [
11],
was obtained. The results of the fitting are listed in
Table 1.
Figure 14.
The same as in
Figure 9,
Figure 10,
Figure 11 and
Figure 12, but for the data from [
7,
10,
21,
23,
31,
32,
33,
35] for Ni and Cu ions (
upper and
bottom panels, respectively). See the text for details.
Figure 14.
The same as in
Figure 9,
Figure 10,
Figure 11 and
Figure 12, but for the data from [
7,
10,
21,
23,
31,
32,
33,
35] for Ni and Cu ions (
upper and
bottom panels, respectively). See the text for details.
Figure 15 compares Ge and Br SP data [
7,
10,
12,
21] to SRIM calculations (the data [
7] were taken from the database [
2]). A lack of data for Ge ions at middle velocities (
) makes the fitting parameters (
and
values) somewhat questionable, despite a good data fit (
). The Br data [
7,
12,
21] at
are reasonably consistent. The data [
12] presented as the
values were corrected for
. The last was calculated with an approximate expression given in reduced values [
36]:
. This correction corresponded to ≃1.5% of the
value at the lowest velocities of the data [
12], presented with 10% accuracy. The results of the fitting are listed in
Table 1.
Figure 16 shows the Kr and Y SP data [
10,
11,
20,
26,
28] and those obtained by Geissel et al. (taken from the database [
2]) in comparison to SRIM calculations. As in the case of the Ar data, the Kr data [
20] at
lie noticeably below the data of Geissel et al. and the data [
26], which are in satisfactory agreement with each other. The last two, together with the low-velocity data [
10,
11], are well fitted with Equation (
8), as shown in the figure (
). Despite well matching the Y data [
28] to the low-velocity data [
10], fitting yielded a large value of
. Implying the data [
28] reliability, a bad data fit could be explained by a simplified fitting model using Equation (
8), which is unable to describe the data with small errors at
. The results of the fitting are listed in
Table 1.
Figure 16.
The same as in
Figure 9,
Figure 10,
Figure 11,
Figure 12,
Figure 13,
Figure 14 and
Figure 15, but for the data from [
2,
10,
11,
20,
26,
28] for Kr and Y ions (
upper and
bottom panels, respectively). See the text for details.
Figure 16.
The same as in
Figure 9,
Figure 10,
Figure 11,
Figure 12,
Figure 13,
Figure 14 and
Figure 15, but for the data from [
2,
10,
11,
20,
26,
28] for Kr and Y ions (
upper and
bottom panels, respectively). See the text for details.
Figure 17.
The same as in
Figure 9,
Figure 10,
Figure 11,
Figure 12,
Figure 13,
Figure 14,
Figure 15 and
Figure 16, but for the data from [
11,
12,
21,
23] for Ag and I ions (
upper and
bottom panels, respectively). See the text for details.
Figure 17.
The same as in
Figure 9,
Figure 10,
Figure 11,
Figure 12,
Figure 13,
Figure 14,
Figure 15 and
Figure 16, but for the data from [
11,
12,
21,
23] for Ag and I ions (
upper and
bottom panels, respectively). See the text for details.
Figure 17 shows the Ag and I SP data [
11,
12,
21,
23] in comparison to SRIM calculations. The Ag data [
11,
23] are the only ones available. The data [
23] were treated as the
values because nuclear stopping contributes only 0.8% of the total stopping at the lowest velocity, according to SRIM calculations. This value is much lower than the 5% total uncertainty assigned to the data. The I data [
12,
21,
23] at
are quite agreeable with each other, whereas the data [
12,
21] are varied at
. The
data [
12] were corrected for
in the same way as the Br data [
12] were. Equation (
8) was used to fit the Ag data, whereas the I data at
could be fitted with a constant for the
ratios, as was performed for the V data (see
Figure 11).
Figure 18 shows the Xe and Au SP data [
11,
20,
23,
26,
29,
37], which are compared to SRIM calculations. The data obtained by Geissel et al. for Xe, and those indicated as IAEA TECDOC for Au were taken from the database [
2]. The Xe data at
are close to each other. The data [
37] originally presented as
were limited to
. At these velocities,
values contribute less than 3% of
, which is less than the respective data errors assigned in [
37]. The Au data [
23] were assigned to the
values because nuclear stopping contributes only 1.5% of total stopping at the lowest energy, according to SRIM calculations. That is much less than the respective errors assigned in the work. For the Au data [
29],
estimates were based on the tabulated
data. These data corresponded to the Au input energies,
= 15–37 MeV. The average energies for the
values thus obtained corresponded to
. The Xe and Au data [
11] at
, corresponding to the different target thicknesses, were added, and both the data sets were fitted with Equation (
8). The steep fall from the Au SP data [
23,
29] to those obtained at
[
11] led to the large
and
fitted values (obtained with large errors), which noticeably exceeded those for other HIs (see
Table 1).
Figure 18.
The same as in
Figure 9,
Figure 10,
Figure 11,
Figure 12,
Figure 13,
Figure 14,
Figure 15,
Figure 16 and
Figure 17, but for the data from [
11,
20,
23,
26,
29,
37] for Xe and Au ions (
upper and
bottom panels, respectively). See the text for details.
Figure 18.
The same as in
Figure 9,
Figure 10,
Figure 11,
Figure 12,
Figure 13,
Figure 14,
Figure 15,
Figure 16 and
Figure 17, but for the data from [
11,
20,
23,
26,
29,
37] for Xe and Au ions (
upper and
bottom panels, respectively). See the text for details.
Figure 19 shows the Pb and U SP data [
11,
12,
20,
30], and those obtained by Geissel et al. (taken from the database [
2]) in comparison to SRIM calculations. Though the Pb data [
20] and those of Geissel et al. are in some disagreement with each other, they were fitted together with Equation (
8). A similar difference is seen for the U data of the same authors. These data, together with others [
12,
30], are in satisfactory agreement with each other. The U data [
12] presented as the
values were corrected for
in the same way as was performed for the Br and I data [
12]. The resulting
data are slightly higher than the data [
30] obtained later at the same velocities, but they are consistent within the error bars with the rest of the data, as seen in the figure. The Pb and U data were supplemented with the low-velocity Bi and U data [
11]. In doing so, a possible distinction in the SP values for Pb and Bi were neglected. The results of fitting are listed in
Table 1.
Figure 19.
The same as in
Figure 9,
Figure 10,
Figure 11,
Figure 12,
Figure 13,
Figure 14,
Figure 15,
Figure 16,
Figure 17 and
Figure 18, but for the data from [
2,
11,
12,
20,
30] for Pb and U ions only (
upper and
bottom panels, respectively). See the text for details.
Figure 19.
The same as in
Figure 9,
Figure 10,
Figure 11,
Figure 12,
Figure 13,
Figure 14,
Figure 15,
Figure 16,
Figure 17 and
Figure 18, but for the data from [
2,
11,
12,
20,
30] for Pb and U ions only (
upper and
bottom panels, respectively). See the text for details.
In
Figure 20, the fitting parameters listed in
Table 1 are shown as a function of the HI atomic number. As one can see in the figure, an amplitude
, corresponding to the ratio of
at high velocities, oscillates in a sporadic way within a magnitude range of 0.9–1.2. The exponent parameters (
and
), which determine decreasing
ratios at low velocities, correlate with each other to a certain extent in the region of
and, probably, for higher
(the Au parameters have been omitted from the consideration due to their enormous large values and errors). The parameterization with Equation (
8) using the parameter values listed in
Table 1 (along with their interpolation for
not listed in the table) seems to be useful for the estimates of the total carbon SP at low velocities, which can be applied as the correction to the SRIM calculations.
Figure 20.
Fitting parameters
,
, and
listed in
Table 1 are shown as functions of the HI atomic number (from upper to bottom panels, respectively). Solid lines are B-spline approximations.
Figure 20.
Fitting parameters
,
, and
listed in
Table 1 are shown as functions of the HI atomic number (from upper to bottom panels, respectively). Solid lines are B-spline approximations.
Now, using Equations (
1) and (
8) with the parameter values listed in
Table 1, one could estimate the
values as
where
is the electronic SP obtained in experiments. The
estimates are considered in the next section. An approximation similar to Equation (
8) can be applied to the available
data, assuming their reliability, together with
calculations. This reasoning takes into consideration that relatively small values of
were subtracted from
in order to obtain
in the low-velocity experiments [
9,
10,
11], whereas at relatively high velocities,
values do not differ from
ones with a good degree of accuracy. The inspection of the low-velocity data [
9,
10] showed that the contribution of the estimated
values does not exceed 25% as a whole (the only exception is about 29% for Fe at the lowest energy). This is the case for the energy-loss thin target data [
11], whereas for the thick target data [
11] for Xe and heavier ions, this contribution exceeds 29%.
3.2. Electronic Stopping Power Data Analysis
In
Figure 21,
Figure 22,
Figure 23,
Figure 24,
Figure 25,
Figure 26,
Figure 27,
Figure 28 and
Figure 29, electronic SP values derived from experiments for Ar to U ions are compared with those obtained with SRIM calculations. Some details about the
data used within this consideration should be mentioned, and they are discussed below. As with the
, the analysis favored the use of the available original data (with the original errors). The
values were fitted using the weighted LSM procedure with the exponential correction function of
, similar to Equation (
8):
where
,
, and
are fitting parameters.
The ratios of the
and
values for Ar and K ions are shown in
Figure 21. The Ar and K
data [
9,
11,
20,
22,
26,
31] were taken from the works’ respective tables, whereas the Ar data of Geissel et al. were from the database [
2]. High velocity data [
20,
31] and those of Geissel et al. are the same as shown in
Figure 9, whereas the data [
26] are limited to velocities
. The limit corresponds to the value from which
gives a lower contribution than the SP data errors assigned in [
26]. The Ar data [
20] are in contrast to similar ones [
26] and those of Geissel et al., which are in agreement with each other. The low-velocity data [
11] also contradict the data [
9,
22], as shown in
Figure 21. As a result, the data [
11,
20] were removed from the fitting of the
ratio. The fitting parameter values thus obtained for the Ar and K data are listed in
Table 2.
Figure 22 shows the Ca and Sc electronic SP data [
9,
10,
11,
26,
27,
31,
32] in comparison to the SRIM calculations. As in
Figure 10, the fitting curve does not provide Ca data for the match with Equation (
11) even after omitting the data [
31] at
. Note that the
value at the lowest
of the data [
27] corresponds to 2% of
, which is smaller than the SP data error (3.8%) assigned in the work. Thus, the data [
27] could be related to
values within the whole range of
. It is implied that the electronic SP data [
27] at relatively high velocities and the data [
9] at relatively low velocities are reliable, as for the total SP considerations. The best Sc data fit was obtained with a constant SP ratio when the data [
11] were ignored. The latter correspond to noticeably lower electronic SP values than those obtained in [
9]. The results of the Ca and Sc data fitting are listed in
Table 2.
Figure 21.
ratios for Ar and K data, [
9,
11,
20,
22,
26,
31] and Ar data [
2] of Geissel et al. (
upper and
bottom panels, respectively). The results of the data fitting with Equation (
11) are shown by dash-dotted lines (the data [
11,
20] were excluded). The upper axes of the panels correspond to relative velocity
, shown for orientation. See the text for details.
Figure 21.
ratios for Ar and K data, [
9,
11,
20,
22,
26,
31] and Ar data [
2] of Geissel et al. (
upper and
bottom panels, respectively). The results of the data fitting with Equation (
11) are shown by dash-dotted lines (the data [
11,
20] were excluded). The upper axes of the panels correspond to relative velocity
, shown for orientation. See the text for details.
Figure 22.
The same as in
Figure 21, but for the data from [
9,
10,
11,
26,
27,
31,
32] for Ca and Sc ions (
upper and
bottom panels, respectively). See the text for details.
Figure 22.
The same as in
Figure 21, but for the data from [
9,
10,
11,
26,
27,
31,
32] for Ca and Sc ions (
upper and
bottom panels, respectively). See the text for details.
Figure 23 shows the Ti and Cr electronic SP data [
7,
10,
11,
20,
21,
23,
31,
32,
33], together with the Ti data of Geissel et al. in comparison to SRIM calculations. The Ti data [
21] and those of Geissel et al., as well as Cr data [
7] were taken from the database [
2]. The Ti data at relatively high velocities (
) are in satisfactory agreement with each other. The Cr data [
7] at the lowest velocities, for which the
values exceeded data errors (2.5%), were excluded from fitting. The best Cr data fit was obtained with a constant SP ratio, when the datum [
11] was disregarded. The last corresponds to the significantly lower electronic SP value than the one obtained in [
9] at the same velocity. The results of the data fitting are given in
Table 2.
Figure 24 compares the Mn and Fe electronic SP data [
7,
10,
11,
31,
32,
33,
35] to SRIM calculations (the data [
7] were taken from the database [
2]). The Mn data [
31,
32] at
differ from the data [
7] obtained later with minor errors. The Mn datum [
11] at
does not agree with the data [
10] at the same velocity and was excluded from fitting. The Mn and Fe data [
7] at the lowest velocities, for which the
values exceeded data errors (2.5%), were also excluded from fitting. The results of the fitting are listed in
Table 2.
Figure 23.
The same as in
Figure 21 and
Figure 22, but for the data from [
7,
10,
11,
20,
21,
23,
31,
32,
33] for Ti and Cr ions (
upper and
bottom panels, respectively). See the text for details.
Figure 23.
The same as in
Figure 21 and
Figure 22, but for the data from [
7,
10,
11,
20,
21,
23,
31,
32,
33] for Ti and Cr ions (
upper and
bottom panels, respectively). See the text for details.
Figure 25 compares the Co and Cu electronic SP data [
7,
10,
11,
23,
32,
33,
35] to the SRIM calculations (the data [
7] were taken from the database [
2]). The SP data [
7] at the lowest velocities, at which the
values exceeded the data errors (2.5%), were excluded from fitting. The Cu data [
11] at
contradicted the data [
10] at the same velocity and were excluded from fitting. As a result, the fitting curves provide rather good matches for the Co and Cu data [
7] with the data [
10] obtained at low velocities. At the same time, these curves vary markedly. The results of the fitting are listed in
Table 2.
Figure 26 compares the Ge and Br electronic SP data [
7,
10,
12,
21] to the SRIM calculations (the data [
7] were taken from the database [
2]). The Br data [
7,
12,
21] at
are reasonably consistent. As in previous cases, the Br data [
7] at the lowest velocities, for which the
values exceeded the data errors (2.5%), were excluded from fitting. As in the case of the Co and Cu ratios, fitting curves obtained for the Ge and Br ratios vary markedly. The results of the fitting are listed in
Table 2.
Figure 25.
The same as in
Figure 21,
Figure 22,
Figure 23 and
Figure 24, but for the data from [
7,
10,
11,
23,
32,
33,
35] for Co and Cu ions (
upper and
bottom panels, respectively). See the text for details.
Figure 25.
The same as in
Figure 21,
Figure 22,
Figure 23 and
Figure 24, but for the data from [
7,
10,
11,
23,
32,
33,
35] for Co and Cu ions (
upper and
bottom panels, respectively). See the text for details.
Figure 27 shows the Kr and Y electronic SP data and those obtained by Geissel et al. (taken from the database [
2]) in comparison to the SRIM calculations. As for the Ar data, the Kr data [
20] at
lie noticeably below the data of Geissel et al. and the data [
26], which are in satisfactory agreement with each other. Note that the errors in the Kr data [
26] at the lowest velocities exceed the respective
values. The data [
26], together with the data of Geissel et al. and the low-velocity data [
10,
11], are well fitted with Equation (
11), as shown in the figure. Fitting the Y data was restricted by the velocities at which the
values did not exceed data errors when the data [
28] are considered. These data do not match the low-velocity data [
10] (
was obtained using Equation (
11)) Implying the reliability of the data [
28], a bad data fit could be explained by the oversimplified fitting model, which cannot describe the specific behavior of the SP ratios obtained with these data in contrast to the others considered here. The results of the fitting are listed in
Table 2.
Figure 27.
The same as in
Figure 21,
Figure 22,
Figure 23,
Figure 24,
Figure 25 and
Figure 26, but for the data from [
2,
10,
11,
20,
26,
28] for Kr and Y ions (
upper and
bottom panels, respectively). See the text for details.
Figure 27.
The same as in
Figure 21,
Figure 22,
Figure 23,
Figure 24,
Figure 25 and
Figure 26, but for the data from [
2,
10,
11,
20,
26,
28] for Kr and Y ions (
upper and
bottom panels, respectively). See the text for details.
Table 2.
Fitting parameter values
,
, and
, as obtained for the Ar to U
data fitted with Equation (
11). The ion symbols and ranges of the applicability of Equation (
11) for specified ions are listed in the first and last columns, respectively. The results of data fitting are shown in
Figure 21,
Figure 22,
Figure 23,
Figure 24,
Figure 25,
Figure 26,
Figure 27,
Figure 28 and
Figure 29.
Table 2.
Fitting parameter values
,
, and
, as obtained for the Ar to U
data fitted with Equation (
11). The ion symbols and ranges of the applicability of Equation (
11) for specified ions are listed in the first and last columns, respectively. The results of data fitting are shown in
Figure 21,
Figure 22,
Figure 23,
Figure 24,
Figure 25,
Figure 26,
Figure 27,
Figure 28 and
Figure 29.
Ion |
|
|
| Range |
---|
Ar 1 |
|
|
| 0.05–0.8 |
K 2 |
|
|
| 0.05–0.6 |
Ca 3 |
|
|
| 0.06–0.9 |
Sc 2 |
| | | 0.06–0.8 |
Ti |
|
|
| 0.06–0.8 |
V |
| | | 0.2–0.6 |
Cr 2,4 |
| | | 0.06–0.7 |
Mn 2,4 |
|
|
| 0.06–0.7 |
Fe 4 |
|
|
| 0.04–0.7 |
Co 4 |
|
|
| 0.04–0.5 |
Cu 2,4 |
|
|
| 0.05–0.8 |
Ge |
|
|
| 0.04–0.9 |
Br |
|
|
| 0.04–0.9 |
Kr 5 |
|
|
| 0.04–0.8 |
Y |
|
|
| 0.04–0.8 |
Ag |
|
|
| 0.06–0.7 |
I |
| | | 0.2–0.6 |
Xe |
|
|
| 0.05–0.8 |
Pb |
|
|
| 0.04–0.8 |
U |
|
|
| 0.04–0.8 |
The iodine stopping power data [
12,
21,
23] at
(considered in
Section 3.1) are in rather good agreement with each other, whereas at
, the data [
12,
21] are varied (see
Figure 17). At
, the
values are less than 3% of
and less than the data errors (5–10%). Thus, the electronic SP ratios at
for I ions are the same as the ones obtained by a constant fit to the iodine
ratios (see
Table 2).
Figure 28 shows the Ag and Xe electronic SP data [
11,
12,
20,
21,
23,
26,
37], and those obtained by Geissel et al. (taken from the database [
2]) in comparison to SRIM calculations. For the Ag data [
23] at the lowest velocity, nuclear stopping accounts for 0.8% of total stopping (which is much lower than the 5% uncertainty assigned to the data). These data were attributed to the
ones. The Xe data [
26] at the lowest velocities, for which the
values exceeded the data errors, were excluded from fitting. The original
data [
37] at
were only used for Xe data fitting. At these velocities,
values contribute less than 3% of
(which is less than the respective data errors assigned in [
37]). TRIM simulations were used for the estimates of the
values in [
37], with their subsequent subtraction from the measured
. The low-velocity data [
37] at
, shown in the figure, disagree with the data [
11] due to the overestimation of nuclear stopping in the TRIM simulations. At the same time, all the Xe data at
are in satisfactory agreement with each other. The results of the fitting with Equation (
11) are listed in
Table 2.
Figure 28.
The same as in
Figure 21,
Figure 22,
Figure 23,
Figure 24,
Figure 25,
Figure 26 and
Figure 27, but for the data from [
2,
11,
12,
20,
21,
23,
26,
37] for Ag and Xe ions (
upper and
bottom panels, respectively). See the text for details.
Figure 28.
The same as in
Figure 21,
Figure 22,
Figure 23,
Figure 24,
Figure 25,
Figure 26 and
Figure 27, but for the data from [
2,
11,
12,
20,
21,
23,
26,
37] for Ag and Xe ions (
upper and
bottom panels, respectively). See the text for details.
Figure 29 shows the
Pb and U data [
11,
12,
20,
30], and those obtained by Geissel et al. (taken from the database [
2]) in comparison to the SRIM calculations. Though the Pb data [
20] and those of Geissel et al. are in some disagreement with each other, they were fitted with Equation (
11) together, as for the
ratios. As in previous cases, the data of Geissel et al. at the lowest velocities, for which the
values exceeded 5% data errors, were excluded from the fitting procedure. These data, together with others [
12,
30], seem to be in satisfactory agreement with each other. The U data [
30] at the lowest velocities, for which the
values exceeded 6% errors assigned in the work, were excluded from fitting. The Pb and U data were supplemented by the Bi and U data [
11] at
. In doing so, the possible distinction in the SP values for Pb and Bi was neglected. The results of the fitting are listed in
Table 2.
As mentioned above, the contribution of nuclear stopping is more than 29% for Xe and heavier ions according to the data [
11] obtained for a thick target at
. At the same time, the derived values of
for thin and thick targets differ from each other by less than 5% in magnitude, which value allowed one to consider these thin and thick target data sets together, as shown in
Figure 28 and
Figure 29.
In
Figure 30, the fitting parameters listed in
Table 2 are shown as a function of the HI atomic number. Amplitude
, corresponding to the
ratio at high velocities, oscillates in a sporadic way within 0.8–1.2, i.e., similarly to the
ratio. The exponent parameter values (
and
), which determine decreasing
at low velocities, correlate with each other to a certain extent within
and probably at higher
. These correlations are similar to the
ones (see
Figure 20).
Figure 29.
The same as in
Figure 21,
Figure 22,
Figure 23,
Figure 24,
Figure 25,
Figure 26,
Figure 27 and
Figure 28, but for the data from [
2,
11,
12,
20,
30] for Pb and U ions (
upper and
bottom panels, respectively). See the text for details.
Figure 29.
The same as in
Figure 21,
Figure 22,
Figure 23,
Figure 24,
Figure 25,
Figure 26,
Figure 27 and
Figure 28, but for the data from [
2,
11,
12,
20,
30] for Pb and U ions (
upper and
bottom panels, respectively). See the text for details.
Figure 30.
The same as in
Figure 20, but for fitting parameters
,
, and
listed in
Table 2.
Figure 30.
The same as in
Figure 20, but for fitting parameters
,
, and
listed in
Table 2.
As one might expect, amplitudes
and
listed in
Table 1 and
Table 2, respectively, do not differ significantly within their errors and express the difference in experimental electronic and nuclear stopping powers as compared to SRIM calculations at high velocities. Thus, in further consideration, these values can be replaced by their average value,
a.
In
Figure 31, some examples of the electronic stopping-power SRIM values corrected with Equation (
11) are shown in comparison to the original ones [
1]. As one can see, the corrections become significant for the most heavy Pb and U ions at velocities
. Clearly, such behavior is determined by the availability of the
data [
11] at
.
Figure 31.
The electronic stopping-power SRIM values corrected with Equation (
11) (lines) are shown in comparison to the original SRIM values [
1] for Ar, Br, Xe, Pb, and U ions (small symbols) at low velocities.
Figure 31.
The electronic stopping-power SRIM values corrected with Equation (
11) (lines) are shown in comparison to the original SRIM values [
1] for Ar, Br, Xe, Pb, and U ions (small symbols) at low velocities.
Now, with the correction functions for
and
, which correspond to Equation (
8) and Equation (
11), respectively; Equation (
10) can be rewritten as
and thus, the nuclear stopping power could be empirically estimated for the specified HI. The
values are determined by the parameter values of the
and
functions listed in
Table 1 and
Table 2, respectively, and the average value
a.
3.3. Nuclear Stopping Power Estimates
For the orientation, the different nuclear SP approximations mentioned in
Section 2.2 are shown in
Figure 32 in the common form of reduced
functions. These functions are compared with the “universal” nuclear stopping power used in SRIM calculations:
where the reduced energy is determined as
An empirical formula [
14] is also shown in
Figure 32, which corresponds to the data obtained in measurements made in a forward direction along the beam axis within a narrow acceptance angle (as mentioned in
Section 1):
where
is determined by Equation (
4). The approximation expressed by Equation (
6) (as “the best” one considered in [
19]) is also added to the figure. It uses
values corresponding to a screening length
where
is the Bohr radius, whereas
values determined using Equation (
4) correspond to the respective screening length definition [
18]. For
determination, approximate scaling was earlier proposed [
36], with the use of the integtation of scaling function
. The last, in turn, is determined by the parameters of the specified interaction potential (see, for example, [
18]). In the present work,
, corresponding to the Thomas-Fermi potential was integrated, and the results thus obtained are also shown in
Figure 32.
Figure 32.
Some approximations for reduced nuclear stopping power
are shown: the one used in the SRIM calculation [
1] (solid line), and those considered in this section and in
Section 2.2. The last instances are designated as “emp. (forward dir.)” [
14] “LSSZ” [
17], “Thomas-Fermi” [
18,
36] (different symbols connected by solid lines), and “Kr-C” [
19] (dotted line). See the text for details.
Figure 32.
Some approximations for reduced nuclear stopping power
are shown: the one used in the SRIM calculation [
1] (solid line), and those considered in this section and in
Section 2.2. The last instances are designated as “emp. (forward dir.)” [
14] “LSSZ” [
17], “Thomas-Fermi” [
18,
36] (different symbols connected by solid lines), and “Kr-C” [
19] (dotted line). See the text for details.
As one can see in the figure, all approximations give close
values at
, whereas at
, they are varied, and the difference from
reaches a factor of ∼3 at
. At the same time, the low-velocity approximations considered here correspond to
, which may restrict the opportunity for the
extrapolation by values of
. Further, the
values converted from
obtained with Equation (
12) are compared with the SRIM and empirical approximations [
1,
14] given by Equations (
13) and (
15), respectively.
Different behaviors of nuclear stopping, arising as a result of the application of Equation (
12) application and its subsequent conversion into reduced
values, could be reproduced by three different approximations, corresponding to three groups of HIs. These seem to be quite unexpected results, which are shown in
Figure 33,
Figure 34 and
Figure 35.
The
functions for the Ar, Ca, Sc, Ti, Mn, Y, and Ag ions are shown in
Figure 33. These functions have similar behaviors, differing by a factor of ∼2. At
, the
values for Ar, Ca, Sc, Ti, and Y approach those given by Equation (
15). At the same time, all the functions show a steep fall in the
values at
, as compared with SRIM and empirical approximations [
1,
14]. In view of uncertainties in the
estimates, this data set and two others could be approximated by the “universal” expressions [
19] fitted with their respective parameters:
where
A,
B,
C, and
D are the fitting parameters. The unweighted LSM for fitting was applied to the obtained
functions relating to the specific set of HIs. The
changes were limited by ranges of
and
. The 3p1 and 4p expressions (Equations (
17) and (
19), respectively) correspond to the best fits as compared to the 3p2 expression (Equation (
18)), according to the
criterion. The results of fitting are shown in
Figure 33, and the parameter values obtained for the best Ar–Ag fitting are listed in
Table 3.
Figure 33.
The
functions obtained for Ar, Ca, Sc, Ti, Mn, Y, and Ag with Equation (
12) are shown by respective intermittent lines. These are compared to SRIM and empirical approximations [
1,
14], expressed by Equations (
13) and (
15) (symbols connected by respective solid lines).
obtained with Equations (
17)–(
19) fittings to this HI set are shown by thick solid lines denoted as 3p1, 3p2, and 4p.
Figure 33.
The
functions obtained for Ar, Ca, Sc, Ti, Mn, Y, and Ag with Equation (
12) are shown by respective intermittent lines. These are compared to SRIM and empirical approximations [
1,
14], expressed by Equations (
13) and (
15) (symbols connected by respective solid lines).
obtained with Equations (
17)–(
19) fittings to this HI set are shown by thick solid lines denoted as 3p1, 3p2, and 4p.
In
Figure 34, the
functions for the K, Fe, Co, and Xe ions are shown. They differ from each other by a factor of ∼3 at
and show a larger difference in the
values at
. The behaviors of the
values for these HIs differs from that obtained for the previous HI group and from calculations according to the SRIM and empirical approximations [
1,
14]. The approximations with Equations (
17)–(
19) did not show a good fit to the
dependencies for these HIs. The unexpected behaviors of the fitted
dependencies implies that nuclear stopping is about the same in the energy range under consideration and plays a minor role for this HI set at low energies, as compared to the previous Ar–Ag case and to the approximations [
1,
14]. The best fit
thus obtained approaches the approximations [
1,
14] at
, and gives lower values than those by a factor of 4–10 at
. The best fit is with the 3p1 expression (Equation (
17)) using a fixed value of the B parameter. The parameter values corresponding to this fit are listed in
Table 3.
The
functions for Cu, Ge, Br, Kr, Pb, and U ions are shown in
Figure 35. In contrast to the previous cases, nuclear stopping is only manifested at
. The
approximations fitted to these functions have a maximum at
. The maximum value is lower than the values given by the SRIM and empirical approximations [
1,
14] by a factor of 5–7 at this energy. The
functions obtained with the 3p1 and 4p expressions (Equations (
18) and (
19), respectively) close to these functions. The best-fitting parameter values for this HI set are listed in
Table 3.
Figure 34.
The same as in
Figure 33, but for K, Fe, Co, and Xe.
Figure 34.
The same as in
Figure 33, but for K, Fe, Co, and Xe.
Table 3.
The
A,
B,
C, and
D parameter values are listed as the result of the best-fitting
functions using Equations (
17)–(
19) applied to each HI set. The
dependencies obtained for different HIs were combined into three sets (indicated in the first column) corresponding to the similarity in the
behavior. The results of data fitting are also shown in
Figure 33,
Figure 34 and
Figure 35.
Table 3.
The
A,
B,
C, and
D parameter values are listed as the result of the best-fitting
functions using Equations (
17)–(
19) applied to each HI set. The
dependencies obtained for different HIs were combined into three sets (indicated in the first column) corresponding to the similarity in the
behavior. The results of data fitting are also shown in
Figure 33,
Figure 34 and
Figure 35.
HI set | Equation | A | B | C | D |
---|
Ar–Ag | (17) | 0.229 ± 0.023 | 0.127 ± 0.080 | 2.48 ± 0.34 | |
K–Xe | (19) | 20.89 ± 1.47 | −0.065 ± 0.029 | 3.94 1 | 3.94 1 |
Cu–U | (18) | 0.0728 ± 0.0033 | 1.06 ± 0.12 | 3.38 ± 0.89 | |
In
Figure 36, the best-fit
functions obtained from the analysis of the Ar–Ag, K–Xe, and Cu–U sets, which correspond to the parameter values listed in
Table 3, are shown in comparison to the approximations [
1,
14].