Article

Microwave Sensor for the Determination of DMSO Concentration in Water–DMSO Binary Mixture

by
Supakorn Harnsoongnoen
1,* and
Benjaporn Buranrat
2
1
The Biomimicry for Sustainable Agriculture, Health, Environment and Energy Research Unit, Department of Physics, Faculty of Science, Mahasarakham University, Kantarawichai District, Maha Sarakham 44150, Thailand
2
Faculty of Medicine, Mahasarakham University, Muang District, Maha Sarakham 44000, Thailand
*
Author to whom correspondence should be addressed.
Micromachines 2023, 14(7), 1378; https://doi.org/10.3390/mi14071378
Submission received: 7 June 2023 / Revised: 4 July 2023 / Accepted: 4 July 2023 / Published: 5 July 2023
(This article belongs to the Special Issue Recent Advances in Microwave Components and Devices)

Abstract

:
This research aims to develop a microwave sensor to accurately measure the concentration of dimethyl sulfoxide (DMSO) in water–DMSO binary mixtures. The proposed sensor will utilize microwave frequency measurements to determine the DMSO concentration, providing a non-invasive and efficient method for analyzing DMSO solutions. The research will involve the design, fabrication, and testing of the sensor, as well as the development of an appropriate calibration model. The outcomes of this study will contribute to improved monitoring and quality control in various fields, including pharmaceuticals, chemical synthesis, and biomedical research. The binary mixtures of dimethyl sulfoxide (DMSO) and water with varying concentrations were investigated in the frequency range of 1 GHz to 5 GHz at room temperature using a microwave sensor. The proposed microwave sensor design was based on an interdigital capacitor (IDC) microstrip antenna loaded with a hexagonal complementary ring resonator (HCRR). The performance of the sensor, fabricated using the print circuit board (PCB) technique, was validated through simulations and experiments. The reflection coefficient (S11) and resonance frequency (Fr) of binary mixtures of DMSO and water solutions were recorded and analyzed for DMSO concentrations ranging from 0% v/v to 75% v/v. Mathematical models were developed to analyze the data, and laboratory tests showed that the sensor can detect levels of DMSO/water binary mixtures. The sensor is capable of detecting DMSO concentrations ranging from 0% v/v to 75% v/v, with a maximum sensitivity of 0.138 dB/% for S11 and ΔS11 and 0.2 MHz/% for Fr and ΔFr at a concentration of 50% v/v. The developed microwave sensor can serve as an alternative for detecting DMSO concentrations in water using a simple and cost-effective technique. This method can effectively analyze a wide range of concentrations, including highly concentrated solutions, quickly and easily.

1. Introduction

Dimethyl sulfoxide (DMSO), also known as (CH3)2SO, is a colorless, water-soluble, hygroscopic, slightly alkaline organic non-toxic liquid with a slight odor that boils at 189 °C and freezes at 18.5 °C. DMSO is a widely used solvent in the fields of biology, chemistry, pharmacology, and medicine for various applications [1,2]. Its advantageous properties, including low toxicity and environmental compatibility, make it a highly valuable polar aprotic solvent for a variety of applications. Among its many uses, it serves as a cryopreservation agent for cells, a penetration enhancer in topical treatments, and a vital component in the fields of toxicology and pharmacology [3,4,5]. However, it is known that high concentrations of DMSO are cytotoxic. Therefore, it is important to measure and define threshold concentrations of DMSO for cells [6]. DMSO is often a principal additive in assay buffers, with a concentration normally ranging from 0.1% to 5% [7]. It is generally accepted to be non-toxic at concentrations below 10% [3,4,5]. Nevertheless, it should be noted that excessive concentrations of DMSO above 50% introduced into the bloodstream can lead to hemolysis, while direct intravenous injection of DMSO may cause local irritation and necrosis [8]. Therefore, measuring and monitoring the DMSO concentration is very important. Many methods exist for detecting the DMSO concentration, such as gas chromatography (GC) [9], high-performance liquid chromatography (HPLC) [10], mass spectrometry (MS) [11], near-infrared (NIR) spectroscopy [12], and the organic sulfur sequential chemical analysis robot (OSSCAR) [13]. In addition to measuring the DMSO concentration, research has also been conducted on the behavior resulting from the mixing of DMSO with water. The mixing behavior of the DMSO/water system is widely recognized for its strong non-ideal characteristics. Solutions comprising this solvent mixture deviate significantly from the ideal behavior and manifest physical and chemical properties that deviate from the norm of what is generally anticipated for an ideal solution. The non-ideal characteristics of the DMSO/water system are evident in various physical properties, including viscosity, density, calorimetry, refractive index, hygroscopicity, and the translational and rotational motion of DMSO–water mixtures [14,15,16]. Understanding the interaction between DMSO and water is crucial for elucidating the mechanisms underlying ice-blocking and biological toxicity. Furthermore, the distinct characteristics of DMSO–water mixtures have been extensively explored and analyzed through a variety of methods, including molecular dynamics (MD) simulations [17,18,19,20,21], soft X-ray spectra [22], neutron diffraction [23,24], vibrational spectroscopy [25], dielectric spectroscopy [26], Rydberg electron-transfer spectroscopy [27], a high-pressure spectroscopic probe [28], infrared absorption spectroscopy [29,30], Raman spectroscopy [31], and Fourier-transform microwave spectroscopy [32]. Another method that has gained interest today is the measurement of material properties using microwave sensors, which has many advantages, such as being easy to build, cheap, and small in size. Additionally, the results can be checked and analyzed in real time. The use of microwave sensors for measuring solution concentrations has been confirmed in various research applications, including the determination of concentrations of salt [33,34], sugar [33,35,36,37,38,39], nitrate and phosphate [40,41], alcohol [42], ethanol [43,44,45,46], methanol [47], glycerol [48], metal ions [49], and equol [50]. However, various research reviews have revealed that there has not been a study or proof of measuring the concentration of DMSO mixed with water using a microwave sensor. We are pleased to present the results of our study, which investigated the use of a microwave sensor to measure the DMSO concentration in water at room temperature. Our method is simple, low-cost, and capable of rapidly analyzing a wide range of concentrations, including high-concentration solutions. With easy operation, we tested concentrations from 0% v/v to 75% v/v.

2. Materials and Methods

2.1. Designing and Fabricating Sensors

The proposed microstrip antenna sensor was designed based on an IDC loaded with an HCRR in the ground plane, as shown in Figure 1. The left and right sides of Figure 1a show the IDC and HSRR structures, respectively. A chamber tube is to be equipped in the region of high electric field strength, as shown in Figure 1a on the right side. The fabricated sensor, which is on a DiClad880 substrate with a dielectric constant of εr = 2.2 and a loss tangent of tan δ = 0.0009 and a thickness of 1.6 mm, is presented in Figure 1b. The layout and dimensions of the sensor structures are shown in Table 1. The equivalent circuit model of the proposed microwave sensor is shown in Figure 2. In our model, the IDC and feed line are represented as a series LC circuit (LIDC and CIDC), while the HCRR slot is represented as a parallel C circuit (CHCRR) with a series LR circuit (LHCRR and RMUT), where RMUT denotes the resistance of the material under test (MUT). The coupling capacitance between the patch with the IDC and the HCRR slot ground plane is represented by CC. The equation used to determine the resonance frequency (Fr) of the proposed device is:
F r = 1 2 π L H C R R C H C R R + C C
Figure 3 displays a comparison between the S11 spectra simulations and measurements for the microwave sensor in free space. The results of the simulations and actual measurements show that the Fr is 3.68 GHz and 3.69 GHz, respectively. A frequency difference of 10 MHz represents a 0.27% error. The simulations and actual measurements of S11, respectively, have values of −34.5 dB and −10.25 dB. The magnitude of S11 in the measurements was found to be lower than that in the simulation, and the reasons behind this discrepancy are not yet fully understood, necessitating further investigation. This discrepancy may be due to parasitic components and differences in size and location during the construction process. This leads to a disparity between the positions of the IDC and HSRR structures in terms of coupling compared to the simulations.

2.2. Preparation of Materials and Analyte Solutions

The analytical grade dimethyl sulfoxide (DMSO) that was used in the experiment was purchased from PanReac AppliChem (Cat. No. 67-68-5). Each concentration of DMSO was dissolved in DI water at concentrations of 0, 25, 50, and 75% v/v. Each concentration of DMSO solution was prepared in triplicate at each concentration.

2.3. Experimental Measurement Setup

Figure 4 illustrates the setup of the sensor and measuring device. The sensor is mounted on a foam base and connected to a Vector Network Analyzer (VNA) via a high-frequency cable. Prior to taking measurements, the short-open-load (SOL) calibration procedure is applied to Port 1 of the VNA. The S11 value is subsequently measured and meticulously recorded. The test solution is carefully loaded into the chamber tube using a micropipette. Subsequent to measuring the results, the chamber tube is thoroughly cleansed with DI water after each measurement is completed, ensuring accuracy and reliability. Multiple sample tests, comprising free space, an empty tube, DI water, and varying DMSO concentrations, were conducted. Each measurement was meticulously carried out using 1601 data points within the frequency range of 1–5 GHz to ensure precision and accuracy. The measurement results are plotted to compare the resonance frequency and S11 magnitude at different DMSO concentrations and DI water in the frequency range of 3.5–3.75 GHz. The frequency resolution in the measurements is 2.5 MHz, resulting from the 1601 data points used. A total of 0.2 mL of test solution was filled into a chamber tube, the temperature was kept constant at 25 ± 1 °C, and the relative humidity was maintained at 45 ± 1% to ensure a stable and consistent working environment.

3. Results

3.1. Reflection Coefficient of Sensor

The S11 spectra for different sample tests were recorded three times with a time interval of 2 min to ensure stability and reliability. Figure 5 shows the S11 spectra for free space, an empty tube, DI water, and different concentrations of DMSO. The magnitude of S11 for the sensor with the chamber tube installed has increased by 1.20 dB, and the Fr has decreased by 5 MHz compared to the case without the chamber tube installed. When DI water was filled into the chamber tube, the Fr dropped from 3.67 GHz to 3.64 GHz, a decrease of approximately 40 MHz, and the magnitude of S11 increased from −33.29 dB to −7.70 dB, an increase of 25.59 dB. The decrease in Fr is due to the permittivity constant of the DI water being greater than that of air, and the increase in S11 is a result of the electrical loss of the DI water being higher than that of air. However, when changing the assay from DI water to various concentrations of DMSO, the Fr and S11 magnitudes gradually decreased, as shown in Figure 5. We found that the signal of the S11 spectra in the frequency range of 3 to 4 GHz was uneven. To clarify the changes, we zoomed in on a narrower frequency range of 3.5 GHz to 3.75 GHz, which covers the notch range of the signal obtained from all the sample tests. Therefore, we applied a smoothing method based on a robust quadratic regression to smooth it out, as shown in Figure 6. However, even though the data was smoothed, the Fr and magnitude of S11 remained unchanged from the results obtained from the unsmoothed data.

3.2. Linearity of Binary Liquid Mixture Measurement

In this section, we use mixtures of DMSO and water to test the capability of our proposed sensor in characterizing binary liquid mixtures. To accomplish this goal, we carefully prepare DMSO/water mixtures with varying concentrations of 0%, 25%, 50%, and 75% v/v using a micropipette. This allows us to assess the sensor’s performance in accurately detecting and quantifying the composition of binary liquid mixtures. The magnitude of the S11 and Fr data at various DMSO concentrations were searched and analyzed to find the relationship between them. The horizontal axis displays the concentration of DMSO, while the left vertical axes show the values of S11 and Fr, as shown in Figure 7a,b, respectively. Scattered data points for both the S11 and Fr datasets were plotted and fitted to curved trendlines, allowing for the study of the resulting relationship between the variables. The change in S11 and Fr resulting from the alteration in the concentration of DMSO mixed with water is caused by a variation in the real and imaginary components of the complex permittivity of the DMSO/water binary mixture. The response of the microwave sensor was established by measuring the levels of S11 and Fr at various concentrations of DMSO. The relationship between the levels of S11 and Fr at different concentrations of DMSO was then determined using a mathematical model developed through linear regression analysis, which links DMSO concentrations with measurable S11 and Fr levels. Equation (2) was used to calculate the levels of S11 at various DMSO concentrations, which were plotted in Figure 7a. Likewise, the levels of Fr at various DMSO concentrations were calculated using Equation (3) and plotted in Figure 7b. The study revealed that the standard deviation (SD) value of Fr was lower than that of S11, indicating that the Fr values were more consistent and less dispersed compared to the S11 values.
S 11 d B = 0.0966 ρ 7.2069 with   R 2   =   0.9664
F r G H z = 14 × 10 4 ρ + 3.63 × 10 9 with   R 2   =   0.9800
where ρ represents the concentration of DMSO in units of % v/v. The relationship between the magnitude of S11 and Fr, obtained from the mathematical model, and the concentration of DMSO in the range of 0–75% v/v was found to be linear with a negative slope. The results indicated that the magnitude of both S11 and Fr altered when the concentration of DMSO varied between 0 to 75% v/v. This step is essential as fluctuations in the complex relative permittivity ( ε r ) of the sample have the potential to influence the sensor response, which is discernible from the changes in S11 and Fr measurements. The changes observed in the S11 at resonance can be attributed to variations in the sensor impedance, which, in turn, are influenced by the imaginary component ( ε r ) of the ε r . On the other hand, the change in Fr is primarily influenced by variations in CR, which is a consequence of changes in the real component ( ε r ) of the ε r of the mixed DMSO/water sample. The equation below represents the frequency (f) dependent behavior of the complex relative permittivity for mixed DMSO/water solutions:
ε r f = ε r f j ε r f
The ε r can be expressed as a function of the ε r and the ε r [21,51,52,53,54,55]. In conducting samples, the ε r has two contributions, as shown below:
ε r f = ε r d f ε r σ f
The ε r is composed of two contributions, namely the dielectric relaxation in dimethyl sulfoxide/loss ( ε r d ) and the loss due to ion drift ( ε r σ ). At low frequencies, the loss due to ion drift tends to obscure the dielectric contribution of ions.
ε r σ = σ ε 0 2 π f
Here, σ represents the ionic conductivity, and ε 0 denotes the permittivity of free space [55]. The dielectric constant of electrolyte–water solutions can be described by the following equation:
ε r = ε r w α c
Here, ε r w denotes the dielectric constant of DI water, c represents the concentration of the electrolyte solution, and α denotes the phenomenological, ion-specific parameter. Figure 7a,b obviously indicate that S11 and Fr decrease as the concentration of DMSO increases, respectively. It can be inferred that the complex relative permittivity of the mixtures is strongly dependent on the DMSO concentration. This phenomenon is expected to arise from the cooperative motion of DMSO–water molecules through hydrogen bonds [22,55]. As the concentration of DMSO increases within the range of 0–75% v/v, the ε r increases, as reported in [53].
This study utilized a newly proposed microwave sensor to measure the differences in S11 and Fr at various concentrations of DMSO and DI water. The resulting data were meticulously analyzed to investigate the relationship between these variables. Figure 8a,b depict the effect of the DMSO concentration on the shift in S11 (ΔS11) and Fr (ΔFr) from their reference values, in the range of 0–75% v/v, using DI water as the reference sample. A linear relationship was found between the S 11 and F r and the DMSO concentration, as shown in Equations (8) and (9), respectively.
S 11 d B = 0.0966 ρ 0.4947
With R2 = 0.9664
F r M H z = 14 × 10 4 ρ 0.25 × 10 6
With R2 = 0.9800
The S11 and Fr of the proposed sensor shift upward as the concentration of DMSO in the solution increases. The shift in Fr also results in a corresponding change in the S11 level at a fixed frequency of 3.64 GHz, which is the Fr for DI water. As a result, both S11 and Fr can be utilized for sensing purposes. The relationship between the S11 shift and the DMSO concentration is illustrated in Figure 8a, while the change in Fr with respect to the DMSO concentration is depicted in Figure 8b. It was observed that the relationship between the S11, Fr, ΔS11, and ΔFr is linear across the entire range of the 0–75% v/v change in the DMSO concentration. Furthermore, it was discovered that the R2 value for both S11 and ΔS11 was 0.9664, while the R2 value for both Fr and ΔFr was even higher at 0.98, indicating a stronger correlation compared to S11 and ΔS11.
The variations in S11 and ΔS11 illustrated in Figure 7a and Figure 8a are ascribed to alterations in the losses of binary liquid mixtures at different concentrations. In the context of mixed DMSO/water binary mixtures at microwave frequencies, the losses are from a polar origin, specifically from the reaction of water molecules with the incident field, as calculated using Equations (5) and (6) [56,57]. The variations in Fr and ΔFr illustrated in Figure 7b and Figure 8b are ascribed to alterations in the complex permittivity of binary liquid mixtures at different concentrations. To determine the complex permittivity of binary liquid mixtures at varying concentrations, it is possible to employ the dielectric mixture equation, as illustrated in (10) [58,59,60]:
ε r f = ε M U T f × 2 ε M U T f + ε W f + 2 V V F ε W f ε M U T f 2 ε M U T f + ε W f V V F ε W f ε M U T f
where ε M U T and ε W are the permittivities of DMSO and DI water, respectively. V V F is the volume fraction of water in the DMSO/water mixtures. This equation can be used to calculate theoretical values of permittivity, providing valuable insights for studying the dielectric behavior of these mixtures. It will enable us to gain a better understanding of the intermolecular interactions and structural properties of liquids, which is significant in various fields of study.

3.3. Sensitivity

Figure 9 shows the sensitivity (S) of the proposed microwave sensor obtained from the measurement of the S11, Fr, ΔS11, and ΔFr values. The proposed sensor demonstrated an exceptionally high sensitivity when the DMSO concentration was 50% v/v for the S11, Fr, ΔS11, and ΔFr parameters. The reason behind the proposed microwave sensor’s high sensitivity at a 50% v/v concentration is believed to be that this concentration falls within the range of the DMSO/water mixture, where the imaginary part of the complex permittivity, the Gibbs energy activation, ΔG, and relaxation time are high [53]. The sensitivity calculations, obtained from the S11, Fr, ΔS11, and ΔFr values, are presented in Equations (11)–(14), respectively.
S S 11 = S 11 ρ
S F r = F r ρ
S S 11 = ( S 11 ) ρ
S F r = ( F r ) ρ
Figure 9a demonstrates that the sensitivity for S11 and ΔS11 remains consistent across various concentrations of DMSO. However, the sensitivity derived from S11 will yield a negative value. Similarly, the sensitivity for Fr and ΔFr was found to be consistent across different concentrations of DMSO, with the sensitivity derived from Fr being negative, as shown in Figure 9b. When examining the sensitivity derived from all four parameters in Figure 9, it becomes evident that the sensitivity escalates as the DMSO concentration increases until it reaches a maximum at 50% v/v. Subsequently, the sensitivity gradually diminishes with further increments in the DMSO concentration.

3.4. Microwave Sensor Performance for Mixed DMSO/Water Detection

The measurements of the mixed DMSO/water concentration using planar microwave sensors have been presented for the first time. Nevertheless, we conducted a comparative analysis between the concentration range and parameters measured by each technique with the proposed method in order to present a more concise overview of the mixed DMSO/water measurements obtained through different techniques. This analysis is illustrated in Table 2. Unfortunately, due to a lack of available data on the sensitivity of other methods used to measure mixed DMSO/water, we are unable to compare our sensitivity findings. Our study allows for an evaluation of the electrical response of mixed DMSO/water in the microwave frequency range across a wide concentration range, including extremely high concentrations. The analysis of mixed DMSO/water commonly involves a variety of techniques, including GC [9], HPLC [10], MS [11], NIR [12,29,30], OSSCAR [13], soft X-ray [22], dielectric spectroscopy [26], and Raman spectroscopy [31]. GC, MS, and HPLC are renowned for their exceptional performance in measuring mixed DMSO/water, owing to their ability to offer high resolution, specificity, and sensitivity features. These techniques are particularly well suited to analyze low molecular weight compounds and mixtures, making them an excellent choice for such applications. Nevertheless, it is important to note that these methods involve intricate procedures, extended analysis times, frequently require labor-intensive sample pretreatment, and involve costly equipment in addition to necessitating highly specialized technical staff for operation. NIR spectroscopy is a useful and widely used technique for the analysis of mixed solutions, offering several advantages, including speed, efficiency, and non-destructiveness, but also has some limitations, including interferences and the need for specialized instrumentation and expertise. The OSSCAR technique has many advantages for mixed solution measurement, including high selectivity, high sensitivity, and high speed. However, it also has its disadvantages, including complexity, cost, limited scope, and demanding sample preparation. The soft X-ray technique offers several benefits for mixed solution measurement, including high sensitivity, element specificity, and non-destructiveness. Despite these advantages, the technique also has some limitations, including cost, complexity, demanding sample preparation, and limited scope of detection. The equipment required for soft X-ray analysis can be expensive, which may make it difficult for some laboratories to access. Additionally, the techniques involved can be complex and may require specialized technical staff to operate, and the sample preparation process can be demanding and time-consuming, which may impact the accuracy of the results. The advantages of using dielectric spectroscopy for mixed solution measurement include high sensitivity, non-invasiveness, and versatility. The technique can detect small changes in the electrical properties of a sample, making it possible to detect trace amounts of substances, and it does not require any modification of the sample, making it a non-invasive option. Additionally, it can be used for a wide range of applications, including liquids, solids, and suspensions. However, dielectric spectroscopy also has its disadvantages. The analysis of mixed solutions using this technique can be complex and may require specialized technical knowledge. Raman spectroscopy is a highly specific technique for mixed solution measurement, providing molecular-level information without altering the sample. However, its low sensitivity and complexity, as well as the cost and potential requirement for sample preparation, can be disadvantages. Nevertheless, we would like to emphasize that our proposed sensor presents a simple and cost-effective solution that provides rapid analysis, wide concentration measurement capabilities (including high concentrations), and user-friendly operation. In [43,44,45,46,47,48], the sensors exhibit high sensitivity and require a small sample size. However, these sensors necessitate the use of two ports for measurement. Moreover, the integration of microfluidics introduces a complex process of sensor fabrication and assembly. Consequently, the cost of building such sensors is also elevated. In [38], the sensor demonstrates a high level of sensitivity while utilizing only one measurement port. However, such sensors need to be combined with the lump elements. Furthermore, the integration of microfluidics adds complexity to the fabrication and assembly processes of these sensors. As a result, the construction cost of such sensors is also increased. However, the sensitivity of the sensor presented in this study is determined based on measurements of a mixed DMSO/water concentration. Previous studies that utilized microfluidic microwave sensors employed different substances and concentration units, which is frankly inconsistent with the parameters of this study. Consequently, making reasonable comparisons becomes challenging.

4. Conclusions

This research presents the design and evaluation of a microwave sensor for measuring DMSO levels in aqueous solutions. The sensor, using an IDC loaded with an HCRR, generates an intense electric field capable of detecting changes in the electrical properties of the specimens. The measured S11 spectra and simulation results agreed well within the frequency range of 1 to 5 GHz. The study illustrates a linear correlation between DMSO concentrations and four parameters, namely S11, Fr, ΔS11, and ΔFr, with maximum sensitivities of 0.138 dB/% (for S11 and ΔS11) and 0.2 MHz/% (for Fr and ΔFr), respectively. In conclusion, this research provides a thorough investigation of the microwave sensor’s potential for measuring DMSO levels in aqueous solutions. The proposed method is a simple and cost-effective technique that can effectively analyze a broad range of concentrations, including highly concentrated solutions, in a quick and straightforward manner.

Author Contributions

Conceptualization, S.H.; methodology, S.H. and B.B.; software, S.H.; validation, S.H.; formal analysis, S.H.; investigation, S.H.; resources, S.H. and B.B.; data curation, S.H.; writing—original draft preparation, S.H. and B.B.; writing—review and editing, S.H. and B.B.; visualization, S.H.; supervision, S.H.; project administration, S.H.; funding acquisition, S.H. All authors have read and agreed to the published version of the manuscript.

Funding

This research project was financially supported by Thailand Science Research and Innovation (TSRI).

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Verheijen, M.; Lienhard, M.; Schrooders, Y.; Clayton, O.; Nudischer, R.; Boerno, S.; Timmermann, B.; Selevsek, N.; Schlapbach, R.; Gmuender, H.; et al. DMSO induces drastic changes in human cellular processes and epigenetic landscape in vitro. Sci. Rep. 2019, 9, 4641. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Catalán, J.; Díaz, C.; García-Blanco, F. Characterization of binary solvent mixtures of DMSO with water and other cosolvents. J. Org. Chem. 2001, 66, 5846–5852. [Google Scholar] [CrossRef] [PubMed]
  3. Galvao, J.; Davis, B.; Tilley, M.; Normando, E.; Duchen, M.R.; Cordeiro, M.F. Unexpected low-dose toxicity of the universal solvent DMSO. FASEB J. 2014, 28, 1317–1330. [Google Scholar] [CrossRef] [PubMed]
  4. Yuan, Y.; Yang, Y.; Tian, Y.; Park, J.; Dai, A.; Roberts, R.M.; Liu, Y.; Han, X. Efficient long-term cryopreservation of pluripotent stem cells at −80 °C. Sci. Rep. 2016, 6, 34476. [Google Scholar] [CrossRef] [Green Version]
  5. Akkök, C.A.; Liseth, K.; Hervig, T.; Ryningen, A.; Bruserud, Ø.; Ersvaer, E. Use of different DMSO concentrations for cryopreservation of autologous peripheral blood stem cell grafts does not have any major impact on levels of leukocyte- and platelet-derived soluble mediators. Cytotherapy 2009, 11, 749–760. [Google Scholar] [CrossRef]
  6. Sumida, K.; Igarashi, Y.; Toritsuka, N.; Matsushita, T.; Abe-Tomizawa, K.; Aoki, M.; Urushidani, T.; Yamada, H.; Ohno, Y. Effects of DMSO on gene expression in human and rat hepatocytes. Hum. Exp. Toxicol. 2011, 30, 1701–1709. [Google Scholar] [CrossRef]
  7. Tjernberg, A.; Markova, N.; Griffiths, W.J.; Hallén, D. DMSO-related effects in protein characterization. J. Biomol. Screen. 2006, 11, 131–137. [Google Scholar] [CrossRef] [Green Version]
  8. Rubin, L.F. Toxicologic update of dimethyl sulfoxide. Ann. N. Y. Acad. Sci. 1983, 411, 6–10. [Google Scholar] [CrossRef]
  9. Luong, J.; Gras, R.; Shellie, R.A.; Cortes, H.J. Direct measurement of part-per-billion levels of dimethyl sulfoxide in water by gas chromatography with stacked injection and chemiluminescence detection. J. Sep. Sci. 2012, 35, 1486–1493. [Google Scholar] [CrossRef]
  10. Thumm, W.; Freitag, D.; Kettrup, A. Determination and quantification of dimethyl sulphoxide by HPLC. Chromatographia 1991, 32, 461–462. [Google Scholar] [CrossRef]
  11. Berresheim, H.; Tanner, D.J.; Eisele, F.L. Real-time measurement of dimethyl sulfoxide in ambient air. Anal. Chem. 1993, 65, 84–86. [Google Scholar] [CrossRef]
  12. Semin, D.J.; Malone, T.J.; Paley, M.T.; Woods, P.W. A novel approach to determine water content in DMSO for a compound collection repository. J. Biomol. Screen. 2005, 10, 568–572. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  13. Asher, E.C.; Dacey, J.W.H.; Jarniková, T.; Tortell, P.D. Measurement of DMS, DMSO, and DMSP in natural waters by automated sequential chemical analysis. Limnol. Oceanogr.-Methods 2015, 13, 451–462. [Google Scholar] [CrossRef]
  14. LeBel, R.G.; Goring, D.A.I. Density, viscosity, refractive index, and hygroscopicity of mixtures of water and dimethyl sulfoxide. J. Chem. Eng. Data. 1962, 7, 100–101. [Google Scholar] [CrossRef]
  15. Cowie, J.M.G.; Toporowski, P.M. Association in the binary liquid system dimethyl sulfoxide-water. Can. J. Chem. 1961, 39, 2240–2243. [Google Scholar] [CrossRef] [Green Version]
  16. Packer, K.J.; Tomlinson, D.J. Nuclear spin relaxation and self-diffusion in the binary system, dimethyl sulfoxide (DMSO)+ Water. Trans. Faraday Soc. 1971, 67, 1302–1314. [Google Scholar] [CrossRef]
  17. Fahy, G.M. Cryoprotectant toxicity neutralization. Cryobiology 2010, 60, S45–S53. [Google Scholar] [CrossRef]
  18. Luzar, A.; Chandler, D. Structure and hydrogen-bond dynamics of water-dimethyl sulfoxide mixtures by computer-simulations. J. Chem. Phys. 1993, 98, 8160–8173. [Google Scholar] [CrossRef] [Green Version]
  19. Kirchner, B.; Hutter, J. The structure of a DMSO-Water mixture from car-parrinello simulations. Chem. Phys. Lett. 2002, 364, 497–502. [Google Scholar] [CrossRef] [Green Version]
  20. Idrissi, A.; Marekha, B.; Barj, M.; Jedlovszky, P. Thermodynamics of mixing water with dimethyl sulfoxide, as seen from computer simulations. J. Phys. Chem. B 2014, 118, 8724–8733. [Google Scholar] [CrossRef] [Green Version]
  21. Yang, L.J.; Yang, X.Q.; Huang, K.M.; Jia, G.Z.; Shang, H. Dielectric properties of binary solvent mixtures of dimethyl sulfoxide with water. Int. J. Mol. Sci. 2009, 10, 1261–1270. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  22. Engel, N.; Atak, K.; Lange, K.M.; Gotz, M.; Soldatov, M.; Golnak, R.; Suljoti, E.; Rubensson, J.E.; Aziz, E.F. DMSO-Water Clustering in Solution Observed in Soft X-ray Spectra. J. Phys. Chem. Lett. 2012, 3, 3697–3701. [Google Scholar] [CrossRef] [PubMed]
  23. Soper, A.K.; Luzar, A. A Neutron Diffraction Study of Dimethyl Sulfoxide-Water Mixtures. J. Chem. Phys. 1992, 97, 1320–1331. [Google Scholar] [CrossRef]
  24. Soper, A.K.; Luzar, A. Orientation of water molecules around small polar and nonpolar groups in solution: A neutron diffraction and computer simulation study. J. Phys. Chem. 1996, 100, 1357–1367. [Google Scholar] [CrossRef]
  25. Wulf, A.; Ludwig, R. Structure and dynamics of water confined in dimethyl sulfoxide. ChemPhysChem 2006, 7, 266–272. [Google Scholar] [CrossRef]
  26. Lu, Z.; Manias, E.; Macdonald, D.D.; Lanagan, M. Dielectric relaxation in dimethyl sulfoxide/water mixtures studied by microwave dielectric relaxation spectroscopy. J. Phys. Chem. A 2009, 113, 12207–12214. [Google Scholar] [CrossRef]
  27. Carles, S.; Desfrancçois, C.; Schermann, J.P.; Berges, J.; Houée-Levin, C. Rydberg electron-transfer spectroscopy and Ab initio studies of dimethylsulfoxide-water neutral and anion dimers. Int. J. Mass Spectrom. 2001, 205, 227–232. [Google Scholar] [CrossRef]
  28. Chang, H.; Jiang, J.; Feng, C.; Yang, Y.; Su, C.; Chang, P.; Lin, S. High-pressure spectroscopic probe of hydrophobic hydration of the methyl groups in dimethyl sulfoxide. J. Chem. Phys. 2003, 118, 1802–1807. [Google Scholar] [CrossRef]
  29. Oh, K.I.; Rajesh, K.; Stanton, J.F.; Baiz, C.R. Quantifying hydrogen-bond populations in dimethyl sulfoxide/water mixtures. Angew. Chem. 2017, 129, 11533–11537. [Google Scholar] [CrossRef]
  30. Wong, D.B.; Sokolowsky, K.P.; El-Barghouthi, M.I.; Fenn, E.E.; Giammanco, C.H.; Sturlaugson, A.L.; Fayer, M.D. Water dynamics in water/DMSO binary mixtures. J. Phys. Chem. B 2012, 116, 5479–5490. [Google Scholar] [CrossRef]
  31. Yang, B.; Cao, X.; Wang, C.; Wang, S.; Sun, C. Investigation of hydrogen bonding in water/DMSO binary mixtures by Raman spectroscopy, Spectrochim. Spectrochim. Acta Part A Mol. Biomol. Spectrosc. 2020, 228, 117704. [Google Scholar] [CrossRef] [PubMed]
  32. Lv, D.; Evangelisti, L.; Maris, A.; Song, W.; Salvitti, G.; Melandri, S. Characterizing the interactions of dimethyl sulfoxide with water: A rotational spectroscopy study. J. Phys. Chem. A 2022, 126, 6882–6889. [Google Scholar] [CrossRef] [PubMed]
  33. Harnsoongnoen, S.; Wanthong, A.; Charoen-In, U.; Siritaratiwat, A. Planar microwave sensor for detection and discrimination of aqueous organic and inorganic solutions. Sens. Actuators B Chem. 2018, 271, 300–305. [Google Scholar] [CrossRef]
  34. Harnsoongnoen, S.; Wanthong, A. A non-contact planar microwave sensor for detection of high-salinity water containing NaCl, KCl, CaCl2, MgCl2 and Na2CO3. Sens. Actuators B Chem. 2021, 331, 129355. [Google Scholar] [CrossRef]
  35. Harnsoongnoen, S.; Wanthong, A. Real-time monitoring of sucrose, sorbitol, d-glucose and d-fructose concentration by electromagnetic sensing. Food Chem. 2017, 232, 566–570. [Google Scholar] [CrossRef] [PubMed]
  36. Harnsoongnoen, S.; Wanthong, A. Coplanar Waveguide Transmission Line Loaded with Electric-LC Resonator for Determination of Glucose Concentration Sensing. IEEE Sens. J. 2017, 17, 1635–1640. [Google Scholar] [CrossRef]
  37. Harnsoongnoen, S.; Wanthong, A. Coplanar waveguides loaded with a split ring resonator-based microwave sensor for aqueous sucrose solutions. Meas. Sci. Technol. 2016, 27, 015103. [Google Scholar] [CrossRef]
  38. Ebrahimi, A.; Scott, J.; Ghorbani, K. Microwave reflective biosensor for glucose level detection in aqueous solutions. Sens. Actu. A Phys. 2020, 301, 2020. [Google Scholar] [CrossRef]
  39. Harnsoongnoen, S.; Buranrat, B. Advances in a Microwave Sensor-Type Interdigital Capacitor with a Hexagonal Complementary Split-Ring Resonator for Glucose Level Measurement. Chemosensors 2023, 11, 257. [Google Scholar] [CrossRef]
  40. Harnsoongnoen, S.; Wanthong, A.; Charoen-In, U.; Siritaratiwat, A. Microwave sensor for nitrate and phosphate concentration sensing. IEEE Sens. J. 2019, 19, 2950–2955. [Google Scholar] [CrossRef]
  41. Harnsoongnoen, S. Metamaterial-Inspired microwave sensor for detecting the concentration of mixed phosphate and nitrate in water. IEEE Trans. Instrum. Meas. 2021, 70, 9509906. [Google Scholar] [CrossRef]
  42. Mohammadi, S.; Wiltshire, B.; Jain, M.C.; Nadaraja, A.V.; Clements, A.; Golovin, K.; Roberts, D.J.; Johnson, T.; Foulds, I.; Zarifi, M.H. Gold coplanar waveguide resonator integrated with a microfluidic channel for aqueous dielectric detection. IEEE Sens. J. 2020, 20, 9825–9833. [Google Scholar] [CrossRef]
  43. Ebrahimi, A.; Withayachumnankul, W.; Al-Sarawi, S.; Abbott, D. High-sensitivity metamaterial-inspired sensor for microfluidic dielectric characterization. IEEE Sens. J. 2014, 14, 1345–1351. [Google Scholar] [CrossRef] [Green Version]
  44. Zhao, W.S.; Wang, B.X.; Wang, D.W.; You, B.; Liu, Q.; Wang, G. Swarm intelligence algorithm-based optimal design of microwave microflfluidic sensors. IEEE Trans. Ind. Electron. 2022, 69, 2077–2087. [Google Scholar] [CrossRef]
  45. Gan, H.Y.; Zhao, W.S.; Liu, Q.; Wang, D.W.; Dong, L.; Wang, G.; Yin, W.Y. Differential microwave microflfluidic sensor based on microstrip complementary split-ring resonator (MCSRR) structure. IEEE Sens. J. 2020, 20, 5876–5884. [Google Scholar] [CrossRef]
  46. Fan, L.-C.; Zhao, W.-S.; Wang, D.-W.; Liu, Q.; Chen, S.; Wang, G. An ultrahigh sensitivity microwave sensor for microflfluidic applications. IEEE Microw. Wirel. Compon. Lett. 2020, 30, 1201–1204. [Google Scholar] [CrossRef]
  47. Ebrahimi, A.; Scott, J.; Ghorbani, K. Ultrahigh-sensitivity microwave sensor for microfluidic complex permittivity measurement. IEEE Trans. Microw. Theory. Tech. 2019, 67, 4269–4277. [Google Scholar] [CrossRef]
  48. Ebrahimi, A.; Tovar-Lopez, F.J.; Scott, J.; Ghorbani, K. Difffferential microwave sensor for characterization of glycerol–water solutions. Sens. Actuators B Chem. 2020, 321, 128561. [Google Scholar] [CrossRef]
  49. Liang, Y.; Ma, M.; Zhang, F.; Liu, F.; Lu, T.; Liu, Z.; Li, Y. Wireless microfluidic sensor for metal ion detection in water. ACS Omega. 2021, 6, 9302–9309. [Google Scholar] [CrossRef]
  50. Loutchanwoot, P.; Harnsoongnoen, S. Microwave microfluidic sensor for detection of high equol concentrations in aqueous solution. IEEE Trans. Biomed. Circuits Syst. 2022, 16, 244–251. [Google Scholar] [CrossRef]
  51. Kaatze, U.; Pottel, R.; Schaefer, M. Dielectric spectrum of dimethyl sulfoxide/water mixtures as a function of composition. J. Phys. Chem. 1989, 93, 5623–5627. [Google Scholar] [CrossRef]
  52. Puranik, S.M.; Kumbharkhane, A.C.; Mehrotra, S.C. Dielectric study of dimethyl sulfoxide-water mixtures using the time-domain technique. J. Chem. Soc. Faraday Trans. 1992, 88, 433–435. [Google Scholar] [CrossRef]
  53. Yang, X.Q.; Yang, L.J.; Huang, K.M.; Tian, W.Y.; Shang, H. Experimental and the theoretical studies of the dielectric properties of DMSO–H2O mixtures. J. Solution. Chem. 2010, 39, 849–856. [Google Scholar] [CrossRef]
  54. Krishnamoorthy, A.N.; Zeman, J.; Holm, C.; Smiatek, J. Preferential solvation and ion association properties in aqueous dimethyl sulfoxide solutions. Phys. Chem. Chem. Phys. 2016, 18, 31312–31322. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  55. Chang, K.; Geise, G.M. Dielectric permittivity properties of hydrated polymers: Measurement and connection to ion transport properties. Ind. Eng. Chem. Res. 2020, 59, 5205–5217. [Google Scholar] [CrossRef]
  56. Gavish, N.; Promislow, K. Dependence of the dielectric constant of electrolyte solutions on ionic concentration: A microfield approach. Phys. Rev. E 2016, 94, 012611. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  57. Sihvola, A. Mixing rules with complex dielectric coefficients. Subsurf. Sens. Technol. Appl. 2000, 1, 393–415. [Google Scholar] [CrossRef]
  58. Steeman, P.A.M.; Maurer, F.H.J. An interlayer model for the complex dielectric constant of composites. Colloid Polym. Sci. 1990, 268, 315–325. [Google Scholar] [CrossRef]
  59. Tuncer, E.; Gubanski, S.M.; Nettelblad, B. Dielectric relaxation in dielectric mixtures: Application of the finite element method and its comparison with dielectric mixture formulas. J. Appl. Phys. 2001, 89, 8092–8100. [Google Scholar] [CrossRef]
  60. Saeed, K.; Pollard, R.D.; Hunter, I.C. Substrate integrated waveguide cavity resonators for complex permittivity characterization of materials. IEEE Trans. Microw. Theory Tech. 2008, 56, 2340–2347. [Google Scholar] [CrossRef]
Figure 1. The proposed planar microwave sensor (a) layout and (b) sensor fabrication.
Figure 1. The proposed planar microwave sensor (a) layout and (b) sensor fabrication.
Micromachines 14 01378 g001
Figure 2. Modeling the proposed sensor using an equivalent circuit.
Figure 2. Modeling the proposed sensor using an equivalent circuit.
Micromachines 14 01378 g002
Figure 3. Comparison of simulated and measured S11 spectra.
Figure 3. Comparison of simulated and measured S11 spectra.
Micromachines 14 01378 g003
Figure 4. Measurement setup.
Figure 4. Measurement setup.
Micromachines 14 01378 g004
Figure 5. The S11 spectra in frequency range of 1 GHz–5 GHz for free space, empty tube, DI water and different concentrations of DMSO/water binary mixture.
Figure 5. The S11 spectra in frequency range of 1 GHz–5 GHz for free space, empty tube, DI water and different concentrations of DMSO/water binary mixture.
Micromachines 14 01378 g005
Figure 6. The S11 spectra and smoothed data spectra were obtained from measurements of DI water and different concentrations of DMSO/water binary mixture samples versus the 3.5 GHz–3.75 GHz frequency range.
Figure 6. The S11 spectra and smoothed data spectra were obtained from measurements of DI water and different concentrations of DMSO/water binary mixture samples versus the 3.5 GHz–3.75 GHz frequency range.
Micromachines 14 01378 g006
Figure 7. The linearity of (a) S11 and (b) Fr with the different concentrations of DMSO.
Figure 7. The linearity of (a) S11 and (b) Fr with the different concentrations of DMSO.
Micromachines 14 01378 g007aMicromachines 14 01378 g007b
Figure 8. The linearity of (a) ΔS11 and (b) ΔFr with the different concentrations of DMSO.
Figure 8. The linearity of (a) ΔS11 and (b) ΔFr with the different concentrations of DMSO.
Micromachines 14 01378 g008aMicromachines 14 01378 g008b
Figure 9. The sensitivity of the sensor and the parameter sensing of (a) S11 and ΔS11 and (b) Fr and ΔF vary with different concentrations of DMSO.
Figure 9. The sensitivity of the sensor and the parameter sensing of (a) S11 and ΔS11 and (b) Fr and ΔF vary with different concentrations of DMSO.
Micromachines 14 01378 g009
Table 1. Layout and geometrical parameters of a microwave sensor.
Table 1. Layout and geometrical parameters of a microwave sensor.
ParameterW1W2W3W4W5L1L2L3L4L5SRCDθ
Value (mm)40200.81.2545307.39.69.250.66.40.68.860°
Table 2. The comparisons of microwave sensors for DMSO concentration detection.
Table 2. The comparisons of microwave sensors for DMSO concentration detection.
Ref.MethodSpecimenConcentrationDataSensitivity
[9]GCMixed DMSO/water0.0001–0.01% v/vVoltageNA
[10]HPLCMixed DMSO/water0–0.005% v/vAbsorbanceNA
[11]MSMixed DMSO/
water/NH3
0–80 × 10−12% v/vPhase with solar
radiation peaks
NA
[12]NIRMixed DMSO/water0–20% v/vTransmissionNA
[13]OSSCARDMSO/DMSP0–1.23% v/vVoltageNA
[22]Soft X-rayMixed DMSO/water0–100% v/vEmission intensityNA
[26]Dielectric spectroscopyMixed DMSO/water0–47.62% v/vS21NA
[29]Infrared absorption spectroscopyMixed DMSO/water0–0.24% v/vIntensityNA
[30]Infrared absorption spectroscopyMixed DMSO/water0–47.62% v/vAbsorbanceNA
[31]Raman spectroscopyMixed DMSO/water0–90% v/vRaman intensityNA
[38]Open-ended
microstrip transmission line loaded CSRR
Glucose0−5 mg/mL S 11 , F r 0.5 (dB/(mg/mL))
0.5 × 10−3 (MHz/(mg/mL))
[43]Microstrip coupled
CSRR
Ethanol0–100% v/v S 21 , F r NA
[44]Microstrip
transmission line loaded series LC
Ethanol0–100% v/v F r 0.695%
[45]Microstrip complementary split-ring resonator (MCSRR)Ethanol0–100% v/v F r 0.626%
[46]Microstrip line loaded CSRREthanol0–100% v/v F r 0.98%
[47]Microstrip
transmission line loaded a shunt-connected series LC
resonator
Methanol0–100% v/v S 21 , F r 0.9%
[48]Microstrip
transmission line terminated with a series RLC resonator
Glycerol0–90% v/v S 11 D C 0.446 (dB/%)
This proposesIDC loaded HCRRMixed DMSO/water0–75% v/v S 11 , F r 0.138 d B / % 0.2 M H z / %
NA—data not available.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Harnsoongnoen, S.; Buranrat, B. Microwave Sensor for the Determination of DMSO Concentration in Water–DMSO Binary Mixture. Micromachines 2023, 14, 1378. https://doi.org/10.3390/mi14071378

AMA Style