The OnControl Bone Marrow Biopsy Technique Is Superior to the Standard Manual Technique for Hematologists-In-Training: A Prospective, Rrandomized Comparison

by
Louis Juden Reed
1,*,
Radha Raghupathy
2,
Marianna Strakhan
1,
Thomas E. Philbeck
3,
Mimi Y. Kim
4,
Ramakrishna Barrini
1,2,
Zulfiqar Hussain
1,2,
Shaad Abdullah
1,2,
Sarah Schweber
1,2,
Kamalesh Bala
1,2 and
Thomas Pacello
1
1
Jacobi Medical Center, Albert Einstein College of Medicine, Division of Hematology/Oncology, 1400 Pelham Parkway South Suite 3N20, Bronx, NY 10461, USA
2
Montefiore Medical Center, Department of Oncology, Division of Hematology, Bronx, NY 10467, USA
3
Vidacare Corporation, Shavano Park, TX 78249, USA
4
Albert Einstein College of Medicine, Bronx, NY 10461, USA
*
Author to whom correspondence should be addressed.
Hematol. Rep. 2011, 3(3), e21; https://doi.org/10.4081/hr.2011.e21
Submission received: 23 August 2011 / Revised: 2 October 2011 / Accepted: 12 October 2011 / Published: 25 October 2011

Abstract

The purpose of this study was to compare a novel bone marrow device with the standard marrow needle in a prospective, randomized study in a teaching hospital employing hematologists- in-training. The new device, the OnControl Bone Marrow (OBM) Biopsy System, utilizes a battery-powered drill to insert the needle. Fifty-four bone marrows (27 standard and 27 OBM) were performed by 11 fellows under the observation and supervision of 3 attending hematologists and 1 research technologist. The primary endpoint of the study, the mean length of the marrow biopsy specimens, a surrogate for marrow quality, was determined by a pathologist in a blinded manner. The mean length of the marrow biopsy specimens was significantly longer (56%) for the OBM group (15.3 mm) than for the standard bone marrow (SBM) group (9.8 mm), p < 0.003. An objectively determined secondary endpoint; mean procedure time, skin-to-skin; also favored the OBM group (175 s) versus the SBM group (292 s), p < 0.007. Several subjective secondary endpoints also favored the OBM group. Only minor adverse events were encountered in the OBM and SBM study groups. It was concluded that bone marrow procedures (BMPs) performed by hematologists-in-training were significantly faster and superior in quality when performed with the OBM compared to the SBM. These data suggest that the OBM may be considered a new standard of care for adult hematology patients. OBM also appears to be a superior method for training hematology fellows.
Keywords: bone marrow biopsy; core biopsy; trephine biopsy; powered biopsy; hematologists-in-training bone marrow biopsy; core biopsy; trephine biopsy; powered biopsy; hematologists-in-training

Share and Cite

MDPI and ACS Style

Reed, L.J.; Raghupathy, R.; Strakhan, M.; Philbeck, T.E.; Kim, M.Y.; Barrini, R.; Hussain, Z.; Abdullah, S.; Schweber, S.; Bala, K.; et al. The OnControl Bone Marrow Biopsy Technique Is Superior to the Standard Manual Technique for Hematologists-In-Training: A Prospective, Rrandomized Comparison. Hematol. Rep. 2011, 3, e21. https://doi.org/10.4081/hr.2011.e21

AMA Style

Reed LJ, Raghupathy R, Strakhan M, Philbeck TE, Kim MY, Barrini R, Hussain Z, Abdullah S, Schweber S, Bala K, et al. The OnControl Bone Marrow Biopsy Technique Is Superior to the Standard Manual Technique for Hematologists-In-Training: A Prospective, Rrandomized Comparison. Hematology Reports. 2011; 3(3):e21. https://doi.org/10.4081/hr.2011.e21

Chicago/Turabian Style

Reed, Louis Juden, Radha Raghupathy, Marianna Strakhan, Thomas E. Philbeck, Mimi Y. Kim, Ramakrishna Barrini, Zulfiqar Hussain, Shaad Abdullah, Sarah Schweber, Kamalesh Bala, and et al. 2011. "The OnControl Bone Marrow Biopsy Technique Is Superior to the Standard Manual Technique for Hematologists-In-Training: A Prospective, Rrandomized Comparison" Hematology Reports 3, no. 3: e21. https://doi.org/10.4081/hr.2011.e21