Cardiovascular Medicine is published by MDPI from Volume 28 Issue 1 (2025). Previous articles were published by another publisher in Open Access under a CC-BY (or CC-BY-NC-ND) licence, and they are hosted by MDPI on mdpi.com as a courtesy and upon agreement with Editores Medicorum Helveticorum (EMH).
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Interesting Images

Coronary Stent Fracture: Problem or Blessing?

Cardiovascular Department, University Hospital, CH-3010 Bern, Switzerland
Cardiovasc. Med. 2008, 11(2), 66; https://doi.org/10.4414/cvm.2008.01308
Submission received: 22 November 2007 / Revised: 22 December 2007 / Accepted: 22 January 2008 / Published: 22 February 2008

Case report

A 57-year-old man underwent coronary angiography for angina and had a Cypher® stent (eluting sirolimus; Cordis Corporation, Miami, Florida USA) implanted into his tortuous right coronary artery (Figure 1A). The long stent straightened out the bends in the middle segment of the right coronary artery but created a kink at its proximal end (Figure 1B). The patient remained asymptomatic but had a follow-up coronary angiogram 20 months later because of palpitations (left ventricular bigeminus). The stent was found fractured in its mid-segment at the site of an initial bend of the coronary artery (Figure 1C). This fracture had relaxed the artery and the proximal kink had disappeared. Overall the artery had again a more natural appearance.
Stent fractures have been reported as rare complications. They are germane to long coronary stents that became more prevalent with the advent of drug-eluting stents curbing the fear of long in-stent restenosis. The topic has raised considerable interest in Asian countries [1] but is all but ignored in the rest of the world. Long stents impact unphysiological long straight segments on naturally tortuous coronary arteries engendering considerable strain and often kinks at both ends. A stent fracture may relax the strain on a coronary artery and thereby be more of a blessing than a problem.

Conflicts of Interest

The author declares no conflict of interest.

References

  1. Lee, M.S.; Jurewitz, D.; Aragon, J.; Forrester, J.; Makkar, R.R.; Kar, S. Stent fracture associated with drug-eluting stents: clinical characteristics and implications. Catheter. Cardiovasc. Interv. 2007, 69, 387–94. [Google Scholar] [PubMed]
Figure 1. A Right coronary artery before stenting. B Right coronary artery immediately after stenting. C Right coronary artery at follow-up.
Figure 1. A Right coronary artery before stenting. B Right coronary artery immediately after stenting. C Right coronary artery at follow-up.
Cardiovascmed 11 00066 g001

Share and Cite

MDPI and ACS Style

Meier, B. Coronary Stent Fracture: Problem or Blessing? Cardiovasc. Med. 2008, 11, 66. https://doi.org/10.4414/cvm.2008.01308

AMA Style

Meier B. Coronary Stent Fracture: Problem or Blessing? Cardiovascular Medicine. 2008; 11(2):66. https://doi.org/10.4414/cvm.2008.01308

Chicago/Turabian Style

Meier, Bernhard. 2008. "Coronary Stent Fracture: Problem or Blessing?" Cardiovascular Medicine 11, no. 2: 66. https://doi.org/10.4414/cvm.2008.01308

APA Style

Meier, B. (2008). Coronary Stent Fracture: Problem or Blessing? Cardiovascular Medicine, 11(2), 66. https://doi.org/10.4414/cvm.2008.01308

Article Metrics

Back to TopTop