1. Introduction
Pressure tubes (PTs) are used to house uranium fuel bundles and pressurized heavy water coolant in the fuel channels of CANadian Deuterium Uranium (CANDU
®) nuclear reactors [
1]. These PTs are composed of Zr2.5%Nb, a zirconium alloy containing 2.4–2.8 percentage by weight of niobium (Nb) [
2]. Each fuel channel consists of a PT held within a larger diameter calandria tube (CT) by a series of four garter spring spacers. This CT separates the cooler (~70 °C) heavy water neutron moderator via a gas annulus from the pressurized heavy water heat transport system flowing through the PTs [
3]. The operating conditions that PTs are subject to include heavy water coolant temperatures between 250 and 300 °C and pressures between 10.5 and 9.9 MPa that vary from the inlet to the outlet, respectively [
3]. As a result of this pressure gradient, there is 65 MPa of axial stress in the PT wall and hoop stresses vary from 122 to 130 MPa from the inlet to the outlet, respectively [
3,
4]. In addition, PTs are also subject to neutron flux irradiation that varies from the inlet to the outlet, with a peak intensity halfway along the tube length of
, and with energies greater than 1 MeV [
4]. These operating conditions cause PTs to undergo various dimensional changes over time, including axial elongation, diametral expansion, sag, and wall thinning [
3].
Standards, published by the Canadian Standards Association (CSA), govern the PT design specifications, manufacturing process, operation, and inspection of fuel channels to mitigate the risks of heavy water coolant leaks forming in the PT during operation [
5,
6,
7,
8,
9,
10,
11]. One requirement, outlined in [
6], is to periodically inspect the PT-CT gap to ensure that the various PT deformation mechanisms do not result in the warmer PT coming into contact with the cooler CT. Due to the horizontal arrangement of the fuel channels, the location of the PT that is most at risk of contact with the CT is the bottom 6 o’clock position. PT-CT contact results in a temperature gradient in the PT, which accelerates deuterium diffusion into the PT at the heavy water coolant interface [
12]. Absorption of deuterium into the PT results in hydrogen embrittlement, and a loss of local ductility of the PT [
13]. If the solubility limit of deuterium in the PT is exceeded, then hydride blisters begin forming and growing until a critical size, which depends on the local tensile stress, is reached for crack initiation [
4]. This crack formation process is referred to as delayed hydride cracking (DHC), and if left unchecked, can lead to coolant leaks in a failure condition known as a loss of coolant accident (LOCA), requiring a costly fuel channel replacement [
6,
14]. As such, it is paramount to ensure that the PT-CT gap is maintained to prevent this failure mechanism from occurring. Assurance is provided by performing inspections, either by direct measurements of the gap [
15], or by the combination of the fuel channel sag measurements and garter spring detection [
6].
Non-destructive testing (NDT) techniques that are utilized for monitoring the PT-CT gap include a combination of multi-frequency eddy current (EC) testing (ECT) using a tool called a gap probe and ultrasound [
15]. ECT is used for extracting PT-CT gap data [
15], garter spring spacer location data for loaded spacers [
12,
16], and (LISS) nozzle proximity [
17,
18], while ultrasonic testing (UT) is primarily used for extracting PT wall thickness (WT) data, a crucial parameter for ECT measurements [
14]. In addition, ECT also has applications for the inspection of cracks in nuclear steam generator tubes, with Zhao et al. [
19] presenting the development of a new hybrid-spiral bobbin EC probe for the detection of arbitrarily oriented cracks.
For measurements of the PT-CT gap, inspection personnel are interested in being able to relate the gap probe’s voltage response to specific fuel channel parameters along a helical scan path with a high degree of precision and accuracy. To this end, there are currently two proposed approaches, with each having its own advantages and disadvantages. These approaches consist of a proposed deterministic inverse or error minimization algorithm (EMA) by Contant et al. [
20], which is based on an analytical model by Klein et al. [
14], and a machine learning model using a deep neural network trained on experimental data as presented by Purdy et al. [
21] and Darling et al. [
22]. The main advantage of using deterministic-based models over machine learning methods is that their accuracy is only dependent on the accuracy of the input parameters. However, a major limiting factor to the accuracy of current PT-CT gap predictions that assume a constant resistivity [
15] are uncertainties in the effect of fuel channel operating conditions on PT electrical resistivity. In a sensitivity analysis conducted by Klein et al. [
14], using possible ranges of input parameter variations, variations in probe lift-off (LO), PT WT, and PT resistivity were found to be the three largest sources of PT-CT gap uncertainty with PT resistivity variations producing about 0.1–0.2 mm of root mean square (RMS) gap prediction error per
of variation, depending on an EC frequency between 4 and 16 kHz. Therefore, an improved understanding of the effects of fuel channel operating conditions on axial and circumferential PT resistivity variations along the helical travel path of the gap probe is needed to compensate for this uncertainty, like the compensation performed by UT for PT WT variations [
6,
7]. However, there is currently a lack of published literature that examines or models the effects of fuel channel operating conditions on Zr2.5%Nb electrical resistivity. A good starting point for improving this understanding is characterizing, through a statistical study, the degree to which electrical resistivity can vary in the circumferential direction of as-manufactured PTs. This is a starting point for the next step, which would be to simulate the effects of circumferential variations in temperatures of up 20 °C between the top and bottom of the PT as reported by Rodgers et al. [
4] and that of irradiation-induced damage effects on resistivity.
The results of Thorpe et al. [
23], showed that there are indeed measurable radial variations in resistivity that appear to become more apparent at higher EC frequencies. Therefore, a natural follow-up question is whether variations in EC test parameters, like the inner or outer surface placement of the coil and inspection frequency, have any statistically significant effects on PT resistivity beyond just inherent variations created by variability in the conditions of the manufacturing process of PTs, which has been shown to lead to microstructural variations [
3,
24,
25,
26]. This question is relevant for the measurement of electrical resistivity using ECT, as it will indicate whether separate resistivities should be reported from either the PT surface or EC frequency. PT-CT gap inspections also rely on variations in resistivity in the circumferential direction being minimal, as there is no mechanism in the current gap algorithm to compensate for such variations [
15]. The accuracy of the gap values returned by this algorithm relies on empirical models obtained from calibrating the gap probe’s response to PT WT and PT-CT gap variations with resistivity held constant [
15]. Therefore, a statistical study seeking to characterize circumferential resistivity variations in as-manufactured PTs will be useful for personnel interested in estimating the amount of gap error that can be expected from this empirical model due to resistivity variations in PTs.
One novel objective of this paper is to characterize the electrical resistivity in the circumferential and radial directions of various independently manufactured PT samples using ECT to determine the testing parameters that have the most statistically significant effect on it. These testing parameters include EC frequency and the surface location of the probe (i.e., whether it is over the outer or inner PT surface). The advantages of using ECT to characterize PT electrical resistivity over other methods are its ability to obtain high-resolution data with minimal required sample preparation and low sensitivity to lift-off variations caused by variations in the zirconium oxide layer due to heat treatment (HT) as reported by Price [
27]. An additional novel objective is to determine, through experimental measurements, whether HT at 400 °C has a statistically significant impact on electrical resistivity variation in the circumferential direction or its measurement with the ECT parameters. HT has been shown by Bennett et al. [
28] to have an effect on resistivity in the axial direction, with reductions of up to 10% in the as-manufactured nominal PT resistivity of
prior to HT. Ref. [
28] also provides images that were obtained using scanning electron microscopy of the Zr2.5%Nb microstructure before and after various stages of HT at 400 °C. The effect on resistivity with time of HT has not been investigated for the circumferential direction. However, Price [
29] has reported that HT at 400 °C results in a reduction in the difference between axial and circumferential resistivities over time.
As mandated by CSA N285.4-09 [
11], cold-worked PTs can require up to 72 h of autoclave HT to accumulate a sufficient oxide layer thickness to inhibit heavy water-induced corrosion during service. Therefore, this paper also characterizes and tracks these resistivity variations at various stages of applied HT, including this 72-h time period. These results can then be used to determine an upper bound on the amount of error that can be expected to be introduced in PT-CT gap prediction algorithms from variations in the resistivity of newly installed PTs.
Based on the results obtained by Bennett et al. [
28], it is expected that HT at 400 °C should result in a resistivity change of
of applied HT time. As demonstrated in their paper [
28], this effect correlates strongly with the modeled changes in resistivity due to the decomposition of the
phase into
and
. Holt [
3] also reports that the extrusion process results in PTs exhibiting variations in circumferential texture with texture fractions that vary between 0.642 and 0.605 from the front to back ends of the PT, respectively. These textures describe grain orientations, and different lattice planes can have different atomic densities, thereby leading to increased electron scattering and resistivity in these directions [
30]. As a result, it is expected that there should be a statistically significant difference in resistivity between PT samples taken from the front and back ends of an extruded PT.
The novel results of this paper include high-resolution measurements of electrical resistivity variation of PTs in the circumferential direction after various stages of applied HT at 400 °C as well as a comparison of these results between different EC frequencies. This paper also explores the statistical significance of variations in EC test parameters like frequency and probe surface placement on the average circumferential electrical resistivity of PTs using a full-factorial experiment. These results are used to estimate the amount of PT-CT gap error that can be expected by not compensating for circumferential resistivity variations in PTs in the current gap algorithm [
15].
2. Model of Changes in Resistivity Due to Heat Treatment Using the Avrami Equation
In addition, the relationship between changes in electrical resistivity caused by the HT process and fraction complete of
decomposition in Zr2.5%Nb is modeled using a similar methodology as outlined by Bennett et al. [
28]. The intent of this section is to provide a more rigorous explanation for the determination of the Avrami coefficients reported by Bennett et al. [
28]. These results are later used to relate resistivity changes to variations in phase composition in the microstructure.
The relationship between phase transformation and HT time is given by the Avrami equation shown below [
31].
where,
is the volume fraction of a particular phase as a function of time, and
,
, and
are fitting constants that depend on the properties of the microstructural phases of a material.
A simple method for determining the fitting constants in Equation (1) is to linearize it by isolating the exponential terms and taking the double natural log of both sides of the equation. Linearizing Equation (1) allows these fitting constants to be determined using linear regression, which is simpler to perform than non-linear curve fitting, where the quality of the fit is dictated by the accuracy of a guess at the optimal values of the parameters [
32]. The resultant equation of this linearization process is given by:
From this equation, the slope of the line of best fit to the linearized data is , and the y-intercept is .
A necessary assumption for determining these fitting constants is that there is a linear correlation between wt. % Nb in the
phase and its volumetric phase fraction,
, during decomposition. This assumption is necessary because the literature on the decomposition process of the
phase in Zr2.5%Nb only gives the wt. % Nb concentrations in the
phase after various amounts of applied HT time. These results are given by Griffiths et al. [
26] in their temperature-time-transformation (TTT) diagram, which shows the relationship between wt. % Nb concentration in the
phase as a function of HT time in log hours at various temperatures. This assumption also follows from the derivation and extraction of Avrami fit parameters by Bennet et al. [
28].
Now, from the TTT diagram provided by Griffiths et al. [
26], the wt. % Nb concentrations in the
phase along the 400 °C line are given as ranges, which means that the actual wt. % Nb concentrations and uncertainties must be estimated using the standard practice of taking the true values as the midpoints of the ranges and the uncertainties as half of each range. Mathematically, this process is given by the following equations:
where
is the midpoint or average % wt. Nb concentration as a function of log time in hours,
, is the corresponding uncertainty, and
and
are the upper and lower bounds on the % wt. Nb concentrations for a particular log base 10 time value, respectively. These concentration values can then be substituted into Equation (2) for
as it is assumed that the rate of change of the wt. % Nb concentration in the
phase and the
phase fraction are equal, which therefore makes them interchangeable here for the purposes of extracting a rate of change in the
phase decomposition rate. These results are shown plotted in
Figure 1, where the time values were converted from the log base 10 time to the natural log time. The dashed line in this plot shows the results obtained using linear regression analysis to generate a line of best fit to the data points. The wt. % Nb concentrations in
Figure 1 range between 29% and 92.5%, between the lowest and highest data points.
From the line of best fit in
Figure 1, the optimal values for
and
are
and
, with
being determined using Equation (3) with this line of best fit’s y-intercept. The uncertainty in
was estimated using error propagation, while the uncertainty in
corresponds to the uncertainty in the slope from the weighted linear regression.
4. Results
Sample results obtained by applying the analysis procedure indicated at the end of
Section 3.2 (above) are shown in
Figure 4. The results in this figure were obtained from the probe positioned over the outer surface of PT Sample 5 after being heat-treated for 24 h using the EC frequencies listed in
Table 2 with the results in each subplot obtained from the EC frequencies of: (a) 74 kHz, (b) 120 kHz, (c) 200 kHz, (d) 460 kHz, (e) 900 kHz, and (f) 1500 kHz. Sample results from the calibration tests for each of the EC frequencies and PT sample surfaces are shown in
Figure 5 for the outer surface placement of the EC probe over the calibration PT samples listed in
Table 3. The corresponding linear regression analysis results from the sample plots in this figure are shown in
Table 5. These calibration results were used to convert the probe’s vertical voltage response into an absolute resistivity using the following equation:
for the corresponding inner or outer surface from which measurements were collected. In this equation,
corresponds to the circumferential position,
is the vertical voltage response,
is the resistivity-voltage (RV) calibration slope, and
is the y-intercept. The uncertainty,
, on the measured resistivity,
, as a function of circumferential position,
, and EC frequency,
, was estimated using the derivative error propagation method in the QExPy module in Python with the uncertainties on each of the parameters as inputs [
43].
A sensitivity analysis of the percentage contribution of each parameter’s uncertainty to the uncertainty of was also conducted using the QExPy module. From this analysis, it was found that the contributions of all three parameters were found to be evenly distributed with average uncertainty contributions and 2σ standard deviations of: for , for , and for .
The resistivities obtained from each data set were then averaged over all circumferential positions with the results being analyzed using a multi-parameter analysis of variance (MANOVA) methodology [
33], with the test factors outlined at the start of
Section 3.1. For a one-way analysis of variance (ANOVA), the relevant equations for this type of analysis are given by [
33]:
where
is the sample mean for test factor combination group
containing
resistivity values,
is the
resistivity for test factor combination
, and
is the overall average in resistivity across all
measurements. The variances of the measured values for each test factor are then given by Equation (13), while the variances between test factors are given by Equation (14), where
defines the number of test factors. Extending these equations for MANOVAs is simply a matter of adding additional equations to calculate the variances within and between each of the test factor combinations.
For this analysis, the
statsmodels module in Python [
44] was utilized for performing the MANOVAs on the full-factorial data. In performing a MANOVA, multiple repeat measurements are required for all test factor combinations, so that the true variance in the response data can be determined. Repeat measurements were gathered on a single PT sample at one of its applied HT states as being representative of the variations for the remaining PT samples and applied HT states. As a result, the true variances for the resistivity measurements for the other PT samples and HT states could not be determined. This issue was resolved by using the average circumferential electrical resistivities from each of the test factor combinations for the multiple PT samples as the repeats. As a result, each grouping of this response variable in the full-factorial data had four repeat measurements from each of the PT samples listed in
Table 1. PT Sample 5 was left out of this analysis, due to it having an incomplete data set for each of the applied HT levels, since this PT sample was being used as a control parameter for examining reproducibility.
The results of this MANOVA analysis are shown in
Figure 6 and
Figure 7. The plot in
Figure 6a shows the results of the predictive capabilities of a MANOVA model using only the four highest test factors and combinations shown in the Pareto plot of
Figure 6. In the Pareto plot of
Figure 7, groups of test factors are denoted by factor names separated by a “:”, with “Freq” and “Surf” being shorthand for the eddy current frequency and probe surface placement, respectively. The plot in
Figure 6b shows the predictive capabilities of a separate MANOVA model using all the test factors and combinations shown in
Figure 7.
The plots in
Figure 8 provide an illustration of the differences in circumferential resistivity trends between each of the PT samples in this experiment after being heat-treated for 24 h at 400 °C. These results were obtained with the EC probe positioned over the outer surface of each of these samples using EC frequencies of: (a) 120 kHz and (b) 900 kHz. Each data point in these plots corresponds to an average of all repeat measurements at the given circumferential position, with the error bars corresponding to the error propagated results in Equation (10) of the 95% confidence bounds, as determined using the data analysis steps on the reproducibility results from PT Sample 5. As shown, the trends in circumferential resistivity are fairly consistent between the EC frequencies for each PT sample, but the results do vary between each of the samples. These results are representative of the results obtained from the inner surfaces of the PT samples. The maximum 2σ standard deviation of all the circumferential resistivity measurements between each of the PT samples in this figure is
and
, for EC frequencies of 120 kHz and 900 kHz, respectively.
Furthermore, the plots in
Figure 9 provide an illustration of the effects of HT on the circumferential resistivities in each of the PT samples. The results in this figure were obtained for the EC probe located on the outer surface of PT Sample 4. The inner surface results are not shown in this plot, but they exhibit similar degrees of circumferential variation to the outer surface results. The different subplots in
Figure 9 correspond to the results that were obtained using EC frequencies of: (a) 120 kHz and (b) 900 kHz. The data points and error bars for the plots in
Figure 9 were determined using the same procedure as outlined for the plots in
Figure 8. From this figure, the maximum 2σ standard deviation of all circumferential resistivity measurements between each of the HT stages is
and
, for EC frequencies of 120 kHz and 900 kHz, respectively. Another result that can be seen from the plots in
Figure 9 is that the circumferential variation in resistivity in PT Sample 4 was similar between each of the applied HT stages.
In addition, the plots in
Figure 10 provide quantifications of the 2σ standard deviation of each set of circumferential resistivity measurements with these results being averaged across all PT samples for each EC frequency and probe surface placement. The error bars in each of these plots correspond to the 2σ standard deviation of 2σ circumferential variation values across all PT samples. The error bars in
Figure 10 quantify the amount of statistical variation within each of the PT samples and do not represent the experimental uncertainty. The results in subplot (a) were obtained with the probe placed over the inner PT surface, while the subplot (b) results were obtained with the outer PT surface placement. As shown, similar results were obtained between each of the EC frequencies and probe surface placements on the PT samples. There is also a weak correlation between variations in circumferential resistivity and applied HT time at 400 °C. The large error bars in these plots indicate that the size of the circumferential resistivity variations in each of the PT samples are mainly dictated by inherent differences between each of these samples, which could be a result of variability in manufacturing history.
The final set of results that were generated from this analysis are shown in
Figure 11, which plots the average electrical resistivity across all circumferential positions and EC frequencies for: (a), (b) the inner PT surface; and (b), (d) the outer PT surface as a function of: (a), (c) HT time in log hours; and (b), (d) fraction complete. This fraction complete scale corresponds to the results obtained using the Avrami fit parameters obtained at the end of
Section 2 in Equation (1) with
now corresponding to the fraction complete variable by the previously discussed assumptions. The vertical error bars in these plots correspond to the 2σ error estimates using the standard deviation formula [
33]. The horizontal error bars in
Figure 11a,c correspond to the results of propagating the estimated uncertainty in an HT time of 30 min, which corresponds to the amount of time necessary for the kiln to heat up to 400 °C. In
Figure 11c,d, no horizontal error bars are shown due to division by zero computation errors being encountered in attempting to propagate the uncertainties in the Avrami fit parameters in Equation (1). Line of best fits are shown in each of the subplots in
Figure 11, which were determined using linear regression analysis with the quantitative results given in
Table 6 and
Table 7, for the fits in
Figure 11a,c, and
Figure 11b,d, respectively. The plots in
Figure 11a,c and
Figure 11b,d show two different ways of presenting the same results. The difference in number of data points between each set of plots arises because the data points at zero HT time (zero fraction complete) are undefined on a logarithmic time scale.
A comparison of the results that were obtained from the two 4-point method measurement techniques on the non-heat-treated ring samples from each of the PT samples in
Table 1 and the results that were obtained from the ECT method are shown in
Table 8 below. The derivative error propagation method in the QExPy module [
43] in Python was used to calculate the uncertainties on each of the 4-point method results.
The results in
Table 8 for the ECT method were obtained by averaging all the resistivity measurements made with this method for each PT sample with the uncertainty calculated as the 2σ standard deviation of the measurements on each PT sample. Therefore, these results represent an average of all resistivity measurements through the WT of each PT sample with the uncertainties quantifying the amount of variation within a 95% confidence bound of each average.
As shown in
Table 8, the measured resistivities using ECT agree within error of the resistivities measured using the 4-point method. Therefore, it can be concluded that the as-built apparatus in
Figure 2 does indeed give accurate measurements of electrical resistivity in PTs. The results from both methods also show little to no variation, beyond the uncertainty confidence bounds between PT samples, which is an indicator of consistency in the as-manufactured PT sample properties for this set of PTs. The results are also consistent with those reported by Bennett et al. [
28] for a different PT sample.
5. Discussion
The results of the Pareto plot in
Figure 7 show that the set of test factors and combinations that capture the majority of the variance in the full-factorial average circumferential resistivity data are:
The effects of each of these test factor combinations on the average circumferential resistivity in PTs are also all statistically significant, as the F-test results for the comparison of variances between sample groups are all orders of magnitude less than the commonly used critical value of 0.05 [
33]. Therefore, the null hypothesis can be rejected for each of these listed test factor groups as there is a vanishingly low probability that these effects are the result of random influences.
These four test factor groups account for 72% of the total variance in the data, which is shown by the annotated correlation coefficient in
Figure 6a. If all test factors and combinations are considered as shown in
Figure 6b, the maximum amount of the total variance that can be captured is 79%. As a result, there is about 21% of unexplained variance in the resistivity data that cannot be attributed to any of the test factors or combinations. This 21% of variance is likely attributable to inherent differences in microstructure between each of the PT samples due to variations in the manufacturing conditions. One such parameter that was not accounted for as a test factor in this analysis was the effect of having PT samples cut from the different extrusion ends of the manufacturing process. This extrusion process is expected to introduce variations in the average grain texture between the front and back end samples listed in
Table 1, as indicated by Holt [
3]. The significance of the HT and EC frequency test factors on measurements of average circumferential electrical resistivity in PTs using ECT agree with the expectations outlined in the introduction as well as the through-thickness resistivity measurements obtained by Thorpe et al. [
23].
However, an anomalous result is the apparent significance of the combination of HT and EC frequency on the circumferential electrical resistivity in PTs as the interaction of these two test factors accounts for roughly 24% of all the variation in the data. Since different EC frequencies result in different penetration depths into a conductive material, as per the skin effect [
36], this result from the MANOVA indicates that the HT methodology used in this experiment generated variations in the radial resistivity trends in each of the tested PT samples, as shown in
Figure 7. Further evidence of this conclusion is shown in
Figure 8, with the different subplots showing that the HT process did not result in uniform rates of decrease in resistivity between the EC frequencies of 120 kHz and 900 kHz. This finding is unexpected, since the industrial kiln used for applying HTs should have been able to reach a uniform temperature everywhere within the HT compartment. It is possible that these radial resistivity variations in the PT samples are a result of the initial temperature gradient that forms between the PT surfaces and the midpoint in its WT. However, further investigation on the conditions immediately after the hot-extrusion process is needed to model the rate of heat transfer in PTs. Another limitation of this study is the lack of characterization of the microstructure of these PT samples before and after each stage of HT. As a result, the exact microstructural causes of these radial resistivity variations are unknown.
An unexpected result from the quantification of experimental uncertainties in the reproducibility plots shown in
Figure 4 is that the size of the error bars seems to randomly vary with the EC frequency, with the error bars being largest at the lowest frequency of 74 kHz. The only source of variation from this experiment that could explain this result is the lack of consistency in the resistivity-voltage (RV) calibration slope values as a function of EC frequency, as shown in
Table 5. As such, further investigations are required here to examine whether the reproducibility of circumferential resistivity scans in PTs can be made more consistent between EC frequencies using a consistent RV sensitivity for all frequencies. Another notable result from
Figure 7 is the apparent significance of the PT surface placement on the average circumferential resistivity measured by the EC probe. However, no statistically significant differences in average circumferential resistivity were observed between the different surface placements of the EC probe over each of the PT samples. Therefore, its likely that the significance of this parameter comes from Equation (6) under the null hypothesis not being satisfied for the variances in average circumferential resistivity between each of the PT surfaces. However, since in-situ microscopy examinations of the microstructure were not performed on each of these PT samples between HTs, it is unknown whether these effects have microstructural correlations with inner and outer surface differences in
ribbon thickness, as reported by Thorpe et al. [
23]. If these results are indeed an effect of the microstructure, then they would indicate that calibrations of EC probe resistivity measurements can be significantly improved by using PT samples with known resistivities, as measured from the outer and inner surfaces. To obtain such measurements, one would need a set of unalloyed and isotropic samples with the same geometric shape as the PTs to compare resistivities between the inner and outer surfaces in a given PT test sample.
From the plots in
Figure 11, the maximum 2σ standard deviations in electrical resistivity measurements in the circumferential direction of each of the PT samples is about
from the inner PT surface. Therefore, based on the sensitivity analysis of the concentric tube model by Klein et al. [
14], it can be inferred that circumferential variations in the resistivity of newly installed PT samples in the fuel channels of CANDU
® reactors introduce PT-CT gap errors of up to 0.12 mm and 0.18 mm for EC inspection data obtained using frequencies of 4 and 8 kHz from the inner PT surface, respectively.
The final set of results in
Figure 11a and
Table 6 show that HT at 400 °C consistently results in an average decrease in resistivity across all PT samples of about
and
from the inner and outer PT surfaces, respectively. The uncertainties on these values correspond to the 2σ standard deviation of each of the results. The outer surface results agree within error of the value reported by Bennett et al. [
28], but the inner surface results do not, which could be due differences in the
ribbon thickness near each of the surfaces as found from the metallographic analysis of the microstructure in the axial-transverse cross-section of a non-heat-treated PT sample by Thorpe et al. [
23]. However, the statistical significance of the results in
Figure 11 are low due to the 2σ error bars in these plots showing significant overlap with each other at each of the plotted data points. The ranking of test parameter influences on average circumferential resistivity in PTs in
Figure 7 also shows that the interaction term between the HT and PT surface has the least significant effect on the variance in these values.
Another result from
Figure 11b is that the decomposition of the
phase in Zr2.5%Nb PTs results in a change in resistivity of about
, where
is the change in fraction complete. The y-intercepts from these fitted lines are also generally in good agreement with those extrapolated from the fitted lines in the plots of
Figure 11a,c, which indicates that for each PT sample and probe surface placement, there is a strong correlation with the linear trend for all data points. The consistency of these results between each of the PT surfaces indicates that the phase transformation of
to
occurs at roughly equal rates for the two surfaces.
The results presented here demonstrate the utility of using ECT to perform high-resolution (radial and circumferential) resistivity measurements, even under conditions of growth of an insulating material oxide layer, relative to what would be possible by 4-point resistivity measurements. Future work would be to develop the inversion of multi-frequency EC measurements to extract depth dependence of resistivity, but this would require non-alloyed and isotropic calibration standards.