How Does Ambient Light Affect the Image Quality of Phosphor Plate Digital Radiography? A Quantitative Analysis Using Contemporary Digital Radiographic Systems
Abstract
:1. Introduction
2. Materials and Methods
2.1. X-ray Exposure
2.2. Light Exposure and Scanning
2.3. Image Evaluation
2.4. Statistical Analysis
3. Results
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Conflicts of Interest
References
- Wenzel, A.; Møystad, A. Work Flow with Digital Intraoral Radiography: A Systematic Review. Acta Odontol. Scand. 2010, 68, 106–114. [Google Scholar] [CrossRef] [PubMed]
- Ramamurthy, R.; Canning, C.F.; Scheetz, J.P.; Farman, A.G. Impact of Ambient Lighting Intensity and Duration on the Signal-to-Noise Ratio of Images from Photostimulable Phosphor Plates Processed Using DenOptix® and ScanX® Systems. Dentomaxillofac. Radiol. 2004, 33, 307–311. [Google Scholar] [CrossRef] [PubMed]
- Erdelyi, R.A.; Duma, V.F.; Sinescu, C.; Dobre, G.M.; Bradu, A.; Podoleanu, A. Optimization of X-Ray Investigations in Dentistry Using Optical Coherence Tomography. Sensors 2021, 21, 4554. [Google Scholar] [CrossRef] [PubMed]
- Elkhateeb, S.M.; Aloyouny, A.Y.; Omer, M.M.S.; Mansour, S.M. Analysis of Photostimulable Phosphor Image Plate Artifacts and Their Prevalence. World J. Clin. Cases 2022, 10, 437–447. [Google Scholar] [CrossRef] [PubMed]
- Gulsahi, A.; Secgin, C.K. Assessment of Intraoral Image Artifacts Related to Photostimulable Phosphor Plates in a Dentomaxillofacial Radiology Department. Niger. J. Clin. Pract. 2016, 19, 248–253. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Deniz, Y.; Kaya, S. Determination and Classification of Intraoral Phosphor Storage Plate Artifacts and Errors. Imaging Sci. Dent. 2019, 49, 219–228. [Google Scholar] [CrossRef] [PubMed]
- Caliskan, A.; Sumer, A.P. Definition, Classification and Retrospective Analysis of Photostimulable Phosphor Image Artefacts and Errors in Intraoral Dental Radiography. Dentomaxillofac. Radiol. 2017, 46, 20160188. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Chiu, H.L.; Lin, S.H.; Chen, C.H.; Wang, W.C.; Chen, J.Y.; Chen, Y.K.; Lin, L.M. Analysis of Photostimulable Phosphor Plate Image Artifacts in an Oral and Maxillofacial Radiology Department. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endodontology 2008, 106, 749–756. [Google Scholar] [CrossRef] [PubMed]
- Ang, D.B.; Angelopoulos, C.; Katz, J.O. How Does Signal Fade on Photo-Stimulable Storage Phosphor Imaging Plates When Scanned with a Delay and What Is the Effect on Image Quality? Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endodontology 2006, 102, 673–679. [Google Scholar] [CrossRef] [PubMed]
- Terauchi, T.; Matsuda, Y.; Araki, K.; Seki, K.; Okano, T. Analysis of Signal Fading of Photostimulable Phosphor Plate System and Its Effect on the Accuracy in Detecting Proximal Caries. Dent. Med. Res. 2013, 33, 68–75. [Google Scholar] [CrossRef]
- Akdeniz, B.G.; Gröndahl, H.G. Degradation of Storage Phosphor Images Due to Scanning Delay. Dentomaxillofac. Radiol. 2006, 35, 74–77. [Google Scholar] [CrossRef] [PubMed]
- Akdeniz, B.G.; Gröndahl, H.G.; Kose, T. Effect of Delayed Scanning of Storage Phosphor Plates. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endodontol. 2005, 99, 603–607. [Google Scholar] [CrossRef] [PubMed]
- Eskandarloo, A.; Yousefi, A.; Soheili, S.; Ghazikhanloo, K.; Amini, P.; Mohammadpoor, H. Evaluation of the Effect of Light and Scanning Time Delay on The Image Quality of Intra Oral Photostimulable Phosphor Plates. Open Dent. J. 2018, 11, 690–700. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Box, G.E.; Cox, D.R. An Analysis of Transformations Revisited, Rebutted. J. Am. Stat. Assoc. 1982, 77, 209–210. [Google Scholar] [CrossRef]
- Marinho-Vieira, L.E.; Martins, L.A.C.; Freitas, D.Q.; Haiter-Neto, F.; Oliveira, M.L. Revisiting Dynamic Range and Image Enhancement Ability of Contemporary Digital Radiographic Systems. Dentomaxillofac. Radiol. 2021, 51, 20210404. [Google Scholar] [CrossRef] [PubMed]
- Pauwels, R.; Silkosessak, O.; Jacobs, R.; Bogaerts, R.; Bosmans, H.; Panmekiate, S. A Pragmatic Approach to Determine the Optimal KVp in Cone Beam CT: Balancing Contrast-to-Noise Ratio and Radiation Dose. Dentomaxillofac. Radiol. 2014, 43, 20140059. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Nascimento, H.A.R.; Visconti, M.A.P.G.; Ferreira, L.M.; Suarez, M.A.; Neto, F.H.; Freitas, D.Q. Effect of Delayed Scanning on Imaging and on the Diagnostic Accuracy of Vertical Root Fractures in Two Photostimulable Phosphor Plates Digital Systems. Int. Endod. J. 2016, 49, 973–979. [Google Scholar] [CrossRef] [PubMed]
- Martins, M.G.B.Q.; Whaites, E.J.; Ambrosano, G.M.B.; Neto, F.H. What Happens If You Delay Scanning Digora Phosphor Storage Plates (PSPs) for up to 4 Hours? Dentomaxillofac. Radiol. 2006, 35, 143–146. [Google Scholar] [CrossRef] [PubMed]
X-ray Exposure Time (s) | Light Exposure Time (s) | |||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
0 | 5 | 10 | 30 | 60 | 90 | |||||||||||||
Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | |
0.1 | 0.38 | 0.22 | Aab | 1.82 | 0.31 | Bab | 4.75 | 0.64 | Cb | 24.89 | 2.86 | Db | 43.22 | 2.39 | Eb | 56.84 | 0.93 | Fb |
0.2 | 0.31 | 0.20 | Aab | 1.68 | 0.41 | Bab | 5.08 | 0.84 | Cb | 27.33 | 3.14 | Da | 46.37 | 2.39 | Eab | 58.97 | 1.72 | Fb |
0.32 | 0.25 | 0.13 | Aab | 1.37 | 0.46 | Bb | 4.94 | 0.57 | Cb | 27.42 | 2.99 | Da | 45.56 | 2.85 | Eb | 60.48 | 1.16 | Fb |
0.4 | 0.16 | 0.13 | Ab | 2.31 | 0.43 | Ba | 5.76 | 0.81 | Cab | 23.83 | 2.60 | Db | 46.59 | 0.76 | Eab | 61.60 | 0.90 | Fab |
0.5 | 0.65 | 0.50 | Aa | 2.50 | 0.70 | Ba | 6.95 | 0.92 | Ca | 30.44 | 3.08 | Da | 50.67 | 0.69 | Ea | 66.39 | 1.16 | Fa |
X-ray Exposure Time (s) | Light Exposure Time (s) | |||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
0 | 5 | 10 | 30 | 60 | 90 | |||||||||||||
Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | |
0.1 | 0.49 | 0.29 | Aa | 3.03 | 0.31 | Bab | 7.47 | 0.72 | Cab | 27.75 | 2.17 | Db | 61.34 | 1.35 | Eb | 94.18 | 1.18 | Fa |
0.2 | 0.66 | 0.64 | Aa | 3.31 | 0.41 | Ba | 8.51 | 0.82 | Ca | 33.80 | 1.23 | Da | 68.02 | 5.00 | Ea | 95.70 | 2.35 | Fa |
0.32 | 0.95 | 0.73 | Aa | 3.17 | 0.46 | Bab | 5.95 | 0.87 | Cb | 25.19 | 2.00 | Dc | 68.01 | 2.81 | Ea | 95.87 | 6.26 | Fa |
0.4 | 0.65 | 0.31 | Aa | 2.76 | 0.43 | Bab | 5.16 | 0.59 | Cc | 22.80 | 1.77 | Dc | 57.96 | 1.42 | Eb | 94.30 | 2.74 | Fa |
0.5 | 0.88 | 0.41 | Aa | 2.10 | 0.70 | Bb | 5.24 | 0.84 | Cc | 18.55 | 1.22 | Dd | 51.82 | 4.97 | Ec | 85.60 | 1.81 | Fb |
X-ray Exposure Time (s) | Light Exposure Time (s) | |||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
0 | 5 | 10 | 30 | 60 | 90 | |||||||||||||
Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | |
0.1 | 0.19 | 0.13 | Aa | 0.20 | 0.15 | Aa | 0.87 | 0.81 | Bab | 3.29 | 0.41 | Cb | 5.63 | 0.45 | Da | 7.39 | 0.20 | Eb |
0.2 | 0.18 | 0.12 | Aa | 0.36 | 0.31 | Aa | 0.76 | 0.26 | Bab | 3.56 | 0.42 | Cab | 6.32 | 0.57 | Da | 8.12 | 0.33 | Eab |
0.32 | 0.15 | 0.12 | Aa | 0.27 | 0.23 | Aa | 0.62 | 0.18 | Bb | 3.60 | 0.42 | Cab | 6.26 | 0.64 | Da | 8.13 | 0.23 | Eab |
0.4 | 0.14 | 0.09 | Aa | 0.27 | 0.23 | Aa | 0.83 | 0.25 | Bab | 3.13 | 0.35 | Cb | 6.20 | 0.15 | Da | 8.22 | 0.37 | Eab |
0.5 | 0.25 | 0.26 | Aa | 0.51 | 0.28 | Aa | 1.09 | 0.29 | Ba | 4.14 | 0.52 | Ca | 6.51 | 0.22 | Da | 8.75 | 0.20 | Ea |
X-ray Exposure Time (s) | Light Exposure Time (s) | |||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
0 | 5 | 10 | 30 | 60 | 90 | |||||||||||||
Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | Mean | SD | Sig. | |
0.1 | 0.35 | 0.24 | Ba | 0.32 | 0.22 | Ba | 0.64 | 0.31 | Ba | 2.84 | 0.31 | Ca | 5.72 | 0.18 | Da | 5.79 | 0.20 | Dab |
0.2 | 0.11 | 0.09 | Bab | 0.30 | 0.27 | CBa | 0.28 | 0.19 | CBb | 0.42 | 0.33 | Cc | 0.53 | 0.49 | Cb | 4.57 | 1.34 | Db |
0.32 | 0.15 | 0.11 | Bab | 0.29 | 0.29 | Ba | 0.27 | 0.17 | Bb | 1.60 | 0.21 | Cb | 3.28 | 0.36 | Db | 2.55 | 0.56 | DCc |
0.4 | 0.09 | 0.07 | Ab | 0.43 | 0.21 | Ba | 0.86 | 0.19 | Ba | 2.36 | 0.34 | Cab | 4.37 | 0.25 | Dab | 4.64 | 0.57 | Dab |
0.5 | 0.25 | 0.15 | Aab | 0.36 | 0.15 | Aa | 1.11 | 0.32 | Ba | 3.24 | 0.31 | Ca | 5.41 | 0.62 | Da | 6.57 | 0.52 | Da |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Sampaio-Oliveira, M.; Marinho-Vieira, L.E.; Wanderley, V.A.; Ambrosano, G.M.B.; Pauwels, R.; Oliveira, M.L. How Does Ambient Light Affect the Image Quality of Phosphor Plate Digital Radiography? A Quantitative Analysis Using Contemporary Digital Radiographic Systems. Sensors 2022, 22, 8627. https://doi.org/10.3390/s22228627
Sampaio-Oliveira M, Marinho-Vieira LE, Wanderley VA, Ambrosano GMB, Pauwels R, Oliveira ML. How Does Ambient Light Affect the Image Quality of Phosphor Plate Digital Radiography? A Quantitative Analysis Using Contemporary Digital Radiographic Systems. Sensors. 2022; 22(22):8627. https://doi.org/10.3390/s22228627
Chicago/Turabian StyleSampaio-Oliveira, Matheus, Luiz Eduardo Marinho-Vieira, Victor Aquino Wanderley, Gláucia Maria Bovi Ambrosano, Ruben Pauwels, and Matheus L. Oliveira. 2022. "How Does Ambient Light Affect the Image Quality of Phosphor Plate Digital Radiography? A Quantitative Analysis Using Contemporary Digital Radiographic Systems" Sensors 22, no. 22: 8627. https://doi.org/10.3390/s22228627
APA StyleSampaio-Oliveira, M., Marinho-Vieira, L. E., Wanderley, V. A., Ambrosano, G. M. B., Pauwels, R., & Oliveira, M. L. (2022). How Does Ambient Light Affect the Image Quality of Phosphor Plate Digital Radiography? A Quantitative Analysis Using Contemporary Digital Radiographic Systems. Sensors, 22(22), 8627. https://doi.org/10.3390/s22228627