Next Article in Journal
A Wireless Sensor Network Deployment for Soil Moisture Monitoring in Precision Agriculture
Next Article in Special Issue
Estimating Resting HRV during fMRI: A Comparison between Laboratory and Scanner Environment
Previous Article in Journal
A Systematic Design Optimization Approach for Multiphysics MEMS Devices Based on Combined Computer Experiments and Gaussian Process Modelling
Previous Article in Special Issue
Microneedle Array Electrode-Based Wearable EMG System for Detection of Driver Drowsiness through Steering Wheel Grip
 
 
Article
Peer-Review Record

SIMpLE: A Mobile Cloud-Based System for Health Monitoring of People with ALS†

Sensors 2021, 21(21), 7239; https://doi.org/10.3390/s21217239
by Arrigo Palumbo 1, Nicola Ielpo 1, Barbara Calabrese 1,*, Domenico Corchiola 2, Remo Garropoli 3, Vera Gramigna 4 and Giovanni Perri 5
Reviewer 1: Anonymous
Reviewer 2: Anonymous
Sensors 2021, 21(21), 7239; https://doi.org/10.3390/s21217239
Submission received: 20 September 2021 / Revised: 18 October 2021 / Accepted: 26 October 2021 / Published: 30 October 2021
(This article belongs to the Special Issue Neurophysiological Monitoring)

Round 1

Reviewer 1 Report

This article presented the development of the mobile cloud-based system for health monitoring systems for ALS people. The platform allows medical staff to monitor patients' health conditions at a distance, allowing them to intervene directly on the electronic instrumentation in remote mode. The developed system seems interesting to me. However, the paper quality from the reader's perspective is hard to read.

Below are my suggestions for the improvement of the paper's overall quality and its structures.

  1. The author should mention the novelty of the developed system and describe how it is different from existing systems in terms of performance or other factors.
  2. The structures of the paper are not in compliance with journal format; the authors are requested to restructures as follows:
    • Introduction
    • Method and Materials or Methodology
    • First, the author need to briefly describe the overall system in one workflow diagram, which has two central applications (i.e., a client application for data collection from a patient in real-time and a web application for physicians for real-time monitoring)
      1. SIMpLE: a mobile Application
        • design and implementation
      2. SIMpLE: Web Application cloud-based system for health monitoring
        • design and implementation

 

  1. The authors added unnecessary details in both application designs, such as DOSS, VPN, Servers. Instead of mentioning these details, It will be worth mentioning how individual system components work together in Algorithms.
  2. The system should be validated in terms of usability or User experience. For User experience evaluation, both patients and physicians must check the user perception towards the developed system.

Author Response

Please see the attachment. 

Author Response File: Author Response.pdf

Reviewer 2 Report

The paper focuses on the adoption of telemonitoring services during the pandemic for people affected by chronic diseases such as the Amyotrophic Lateral Sclerosis (ALS).

 

Remote monitoring and teleconsulting are surely essential services for fragile patients such as those with the ALS, especially during the COVID19 pandemy, but I would say even in less dangerous times.

Authors present an innovative secure medical monitoring and teleconsultation mobile cloud-based system for disabled people. The infrastructure is able to acquire and process vital signs and especially concerning ECG and EMG.

Medical staff is provided with functionalities to monitor patient’s conditions remotely and to intervene directly on the electronic instrumentation in case of need.

 

The paper is well written and the solution is interesting. Moreover, the article is certainly of interest for the readers of this journal.

 

I have only one main concernment that I would like to see addressed by the authors.
The system may fall under the regulatory framework for Medical Devices and Software as a Medical Device.

It is certainly a prototype obtained as a result of research activities, but this will not free the clinical staff from responsibilities in case of system failures.
I guess that this problem must be clearly introduced within the paper explaining that an engegnerised version of the prototype should be certified as compliant to the MD European regulations.
To conclude, I suggest the authors to add a paragraph that introduces the problem and refers to:

  1. The European medical device regulation (https://eur-lex.europa.eu/legal-content/EN/TXT/PDF/?uri=CELEX:32017R0745)
  2. The ISO 13485 quality standard for medical devices (https://www.iso.org/standard/59752.html)
  3. A methodology for risk assessment in Software as Medical Devices (https://ieeexplore.ieee.org/abstract/document/9194991)
  4. The ISO 14971 standard for risk management ( https://www.iso.org/standard/72704.html)

Author Response

Please see attachment.

Author Response File: Author Response.pdf

Round 2

Reviewer 1 Report

The authors have addressed all of my comments and concerns in the revised version.

Back to TopTop