Comparing Glucometer-Based and Laboratory-Based OGTT for Diabetes Diagnosis: A Narrative Review
Abstract
1. Introduction
2. Laboratory-Based OGTT: Method and Standards
3. Glucometer-Based Glucose Measurement: Method and Mechanism
4. Accuracy of Glucometer-Based OGTT: Evidence and Performance
5. Potential Benefits of Glucometer-Based OGTT
6. Advances in Glucometer Technology and Implications for OGTT Accuracy
7. Future Research Directions and Recommendations
8. Conclusions
Author Contributions
Funding
Acknowledgments
Conflicts of Interest
References
- Lin, X.; Xu, Y.; Pan, X.; Xu, J.; Ding, Y.; Sun, X.; Song, X.; Ren, Y.; Shan, P.-F. Global, regional, and national burden and trend of diabetes in 195 countries and territories: An analysis from 1990 to 2025. Sci. Rep. 2020, 10, 14790. [Google Scholar] [CrossRef] [PubMed]
- Forouhi, N.G.; Wareham, N.J. Epidemiology of diabetes. Medicine 2019, 47, 22–27. [Google Scholar] [CrossRef]
- Mayer-Davis, E.J.; Lawrence, J.M.; Dabelea, D.; Divers, J.; Isom, S.; Dolan, L.; Imperatore, G.; Linder, B.; Marcovina, S.; Pettitt, D.J.; et al. Incidence Trends of Type 1 and Type 2 Diabetes among Youths, 2002–2012. N. Engl. J. Med. 2017, 376, 1419–1429. [Google Scholar] [CrossRef] [PubMed]
- Manne-Goehler, J.; Geldsetzer, P.; Agoudavi, K.; Andall-Brereton, G.; Aryal, K.K.; Bicaba, B.W.; Bovet, P.; Brian, G.; Dorobantu, M.; Gathecha, G.; et al. Health system performance for people with diabetes in 28 low- and middle-income countries: A cross-sectional study of nationally representative surveys. PLoS Med. 2019, 16, e1002751. [Google Scholar] [CrossRef]
- American Diabetes Association Professional Practice Committee 2. Diagnosis and Classification of Diabetes: Standards of Care in Diabetes-2024. Diabetes Care 2024, 47, S20–S42. [Google Scholar] [CrossRef] [PubMed]
- Jagannathan, R.; Neves, J.S.; Dorcely, B.; Chung, S.T.; Tamura, K.; Rhee, M.; Bergman, M. The Oral Glucose Tolerance Test: 100 Years Later. Diabetes Metab. Syndr. Obes. 2020, 13, 3787–3805. [Google Scholar] [CrossRef]
- Manne-Goehler, J.; Atun, R.; Stokes, A.; Goehler, A.; Houinato, D.; Houehanou, C.; Hambou, M.M.S.; Mbenza, B.L.; Sobngwi, E.; Balde, N.; et al. Diabetes diagnosis and care in sub-Saharan Africa: Pooled analysis of individual data from 12 countries. Lancet Diabetes Endocrinol. 2016, 4, 903–912. [Google Scholar] [CrossRef]
- Dickson, L.M.; Buchmann, E.J.; Janse Van Rensburg, C.; Norris, S.A. The impact of differences in plasma glucose between glucose oxidase and hexokinase methods on estimated gestational diabetes mellitus prevalence. Sci. Rep. 2019, 9, 7238. [Google Scholar] [CrossRef]
- Vučić Lovrenčić, M.; Radišić Biljak, V.; Božičević, S.; Pape-Medvidović, E.; Ljubić, S. Validation of Point-of-Care Glucose Testing for Diagnosis of Type 2 Diabetes. Int. J. Endocrinol. 2013, 2013, 206309. [Google Scholar] [CrossRef]
- Gavin, J.R.; Bailey, C.J. Real-World Studies Support Use of Continuous Glucose Monitoring in Type 1 and Type 2 Diabetes Independently of Treatment Regimen. Diabetes Technol. Ther. 2021, 23, S-19. [Google Scholar] [CrossRef]
- St John, A.; Price, C.P. Existing and Emerging Technologies for Point-of-Care Testing. Clin. Biochem. Rev. 2014, 35, 155–167. [Google Scholar] [PubMed]
- Tran, N.K.; LaValley, C.; Bagley, B.; Rodrigo, J. Point of care blood glucose devices in the hospital setting. Crit. Rev. Clin. Lab. Sci. 2023, 60, 290–299. [Google Scholar] [CrossRef] [PubMed]
- Heller, A.; Feldman, B. Electrochemical Glucose Sensors and Their Applications in Diabetes Management. Chem. Rev. 2008, 108, 2482–2505. [Google Scholar] [CrossRef] [PubMed]
- Danne, T.; Nimri, R.; Battelino, T.; Bergenstal, R.M.; Close, K.L.; DeVries, J.H.; Garg, S.; Heinemann, L.; Hirsch, I.; Amiel, S.A.; et al. International Consensus on Use of Continuous Glucose Monitoring. Diabetes Care 2017, 40, 1631–1640. [Google Scholar] [CrossRef]
- King, F.; Ahn, D.; Hsiao, V.; Porco, T.; Klonoff, D.C. A Review of Blood Glucose Monitor Accuracy. Diabetes Technol. Ther. 2018, 20, 843–856. [Google Scholar] [CrossRef]
- Heidt, B.; Siqueira, W.F.; Eersels, K.; Diliën, H.; van Grinsven, B.; Fujiwara, R.T.; Cleij, T.J. Point of Care Diagnostics in Resource-Limited Settings: A Review of the Present and Future of PoC in Its Most Needed Environment. Biosensors 2020, 10, 133. [Google Scholar] [CrossRef]
- George, K.; Joy, S.; Kaladi Palliyalil, H.; Shaji, S.M.; Ajayakumar Sunanda, V.; Sreedevi, A. Benchmarking Point-of-Care Glucometers: A Comparative Study Using the Hexokinase Test and International Organization for Standardization (ISO) Standards. Cureus 2024, 16, e70981. [Google Scholar] [CrossRef]
- Fabre-Estremera, B.; Martínez-Chávez, E.; Manzano Ocaña, M.; Carcavilla Urquí, A.; de los Ángeles Morales Sánchez, M.; Pinilla Tejado, I.; González-Casado, I.; Losantos García, I.; Fernández-Calle, P.; Buño Soto, A.; et al. Use of point-of-care glucometers during an oral glucose tolerance test in children for prediabetes and diabetes diagnosis: A comparison study. Adv. Lab. Med. 2024, 5, 189–196. [Google Scholar] [CrossRef]
- Garabedian, L.F.; Ross-Degnan, D.; LeCates, R.F.; Wharam, J.F. Uptake and use of a diabetes management program with a mobile glucometer. Prim. Care Diabetes 2019, 13, 549–555. [Google Scholar] [CrossRef]
- Ginsberg, B.H. Factors Affecting Blood Glucose Monitoring: Sources of Errors in Measurement. J. Diabetes Sci. Technol. 2009, 3, 903–913. [Google Scholar] [CrossRef]
- Hassan, M.H.; Vyas, C.; Grieve, B.; Bartolo, P. Recent Advances in Enzymatic and Non-Enzymatic Electrochemical Glucose Sensing. Sensors 2021, 21, 4672. [Google Scholar] [CrossRef] [PubMed]
- Mandpe, P.; Prabhakar, B.; Gupta, H.; Shende, P. Glucose oxidase-based biosensor for glucose detection from biological fluids. Sens. Rev. 2020, 40, 497–511. [Google Scholar] [CrossRef]
- Sonagra, A.D.; Zubair, M.; Motiani, A. Hexokinase Method. In StatPearls [Internet]; StatPearls Publishing: Treasure Island, FL, USA, 2024. Available online: https://www.ncbi.nlm.nih.gov/sites/books/NBK587446/ (accessed on 16 June 2025).
- Huang, T.-Y.; Chang, H.-W.; Tsao, M.-F.; Chuang, S.-M.; Ni, C.-C.; Sue, J.-W.; Lin, H.-C.; Hsu, C.-T. Evaluation of accuracy of FAD-GDH- and mutant Q-GDH-based blood glucose monitors in multi-patient populations. Clin. Chim. Acta 2014, 433, 28–33. [Google Scholar] [CrossRef]
- St-Louis, P. Status of point-of-care testing: Promise, realities, and possibilities. Clin. Biochem. 2000, 33, 427–440. [Google Scholar] [CrossRef]
- Jamieson, E.L.; Dimeski, G.; Flatman, R.; Hickman, P.E.; Ross Dallas Jones, G.; Marley, J.V.; David McIntyre, H.; McNeil, A.R.; Nolan, C.J.; Potter, J.M.; et al. Oral glucose tolerance test to diagnose gestational diabetes mellitus: Impact of variations in specimen handling. Clin. Biochem. 2023, 115, 33–48. [Google Scholar] [CrossRef] [PubMed]
- Pfützner, A.; Schipper, C.; Ramljak, S.; Flacke, F.; Sieber, J.; Forst, T.; Musholt, P.B. Evaluation of the Effects of Insufficient Blood Volume Samples on the Performance of Blood Glucose Self-Test Meters. J. Diabetes Sci. Technol. 2013, 7, 1522–1529. [Google Scholar] [CrossRef] [PubMed]
- Dunseath, G.J.; Bright, D.; Luzio, S.D. Comparative Accuracy Evaluation of a Blood Glucose Meter with Novel Hematocrit Correction Technology, With Three Currently Used Commercially Available Blood Glucose Monitoring Systems. J. Diabetes Sci. Technol. 2019, 13, 568–574. [Google Scholar] [CrossRef]
- Lippi, G.; Ferrari, A.; Visconti, S.; Martini, L.; Demonte, D.; Cascio, C.L.; Capizzi, B. Screening fasting glucose before the OGTT: Near-patient glucometer- or laboratory-based measurement? Diagnosis 2025, 12, 262–267. [Google Scholar] [CrossRef]
- Wolde, M.; Tarekegn, G.; Kebede, T. Comparative Evaluations of Randomly Selected Four Point-of-Care Glucometer Devices in Addis Ababa, Ethiopia. J. Diabetes Sci. Technol. 2018, 12, 673–679. [Google Scholar] [CrossRef]
- Ogunbosi, B.O.; Jarrett, O.O.; Orimadegun, A.E.; Ayoola, O.O.; Osinusi, K. Comparison of point-of-care glucometers and laboratory based glucose oxidase test in determining blood glucose levels. Niger. J. Paediatr. 2022, 49, 266–271. [Google Scholar] [CrossRef]
- Baig, A.; Siddiqui, I.; Jabbar, A.; Azam, S.I.; Sabir, S.; Alam, S.; Ghani, F. Comparision between bed side testing of blood glucose by glucometer vs centralized testing in a tertiary care hospital. J. Ayub Med. Coll. Abbottabad 2007, 19, 25–29. [Google Scholar]
- Muktabhant, B.; Sanchaisuriya, P.; Sarakarn, P.; Tawityanon, W.; Trakulwong, M.; Worawat, S.; Schelp, F.P. Use of glucometer and fasting blood glucose as screening tools for diabetes mellitus type 2 and glycated haemoglobin as clinical reference in rural community primary care settings of a middle income country. BMC Public Health 2012, 12, 349. [Google Scholar] [CrossRef] [PubMed]
- Al-Hasani, W.; Ranasinghe, R.; Rogers, H.; Spanier, W.; Spears, K.; Gayle, C.; Long, L.; Dimitriadis, G.K.; Hunt, K.F.; Vincent, R.P. Clinical utility of point-of-care glucose testing in the assessment of gestational diabetes: Prospective cohort study. BJOG Int. J. Obstet. Gynaecol. 2024, 131, 1270–1278. [Google Scholar] [CrossRef] [PubMed]
- Tan, A.Y.S.; Tan, M.S.; Wu, A.; Seah, A.C.; Chong, C.; Koh, E.; Tan, N.C. Self-administered oral glucose tolerance test with capillary glucose measurements for the screening of diabetes mellitus in high-risk adults: A feasibility study. BMJ Open Diabetes Res. Care 2021, 9, e002556. [Google Scholar] [CrossRef] [PubMed]
- Jamieson, E.L.; Spry, E.P.; Kirke, A.B.; Atkinson, D.N.; Marley, J.V. Real-World Gestational Diabetes Screening: Problems with the Oral Glucose Tolerance Test in Rural and Remote Australia. Int. J. Environ. Res. Public Health 2019, 16, 4488. [Google Scholar] [CrossRef]
- Adam, S.; Rheeder, P. Evaluating the utility of a point-of-care glucometer for the diagnosis of gestational diabetes. Int. J. Gynaecol. Obstet. 2018, 141, 91–96. [Google Scholar] [CrossRef]
- Adams, S.; Rheeder, P. Screening for gestational diabetes mellitus in a South African population: Prevalence, comparison of diagnostic criteria and the role of risk factors. S. Afr. Med. J. 2017, 107, 523–527. [Google Scholar] [CrossRef]
- Priya, M.; Mohan Anjana, R.; Pradeepa, R.; Jayashri, R.; Deepa, M.; Bhansali, A.; Mohan, V. Comparison of capillary whole blood versus venous plasma glucose estimations in screening for diabetes mellitus in epidemiological studies in developing countries. Diabetes Technol. Ther. 2011, 13, 586–591. [Google Scholar] [CrossRef]
- Suresh Babu, G.; Babu Rao, R.; Shwetha, K.; Imran, A.S.; Ratna Kumari, V. Comparing the Glucose Values by Glucometer and Laboratory Methods in the Diagnosis of Gestational Diabetes Mellitus: A Hospital Based Study. Int. J. Curr. Med. Appl. Sci. 2015, 7, 21–29. [Google Scholar]
- Gallardo, H.; Lomelin-Gascon, J.; Martinez, L.A.; Montoya, A.; Reyes-Muñoz, E.; Tapia-Conyer, R.C. 1358-P: Point of Care OGTT for the Screening of Gestational Diabetes: A Feasible Proposal for Low-Resource Settings. Diabetes 2020, 69 (Suppl. S1), 1358. [Google Scholar] [CrossRef]
- Eskandarifar, A.; Rasouli, M.A.; Mansouri, M.; Moosavi, S.; Fotoohi, A. Validity of glucose measurements in the blood by a glucometer reagent strip in critically ill infants. Diabetes Metab. Syndr. 2019, 13, 464–466. [Google Scholar] [CrossRef]
- Daly, N.; Carroll, C.; Flynn, I.; Harley, R.; Maguire, P.; Turner, M. Evaluation of point-of-care maternal glucose measurements for the diagnosis of gestational diabetes mellitus. BJOG Int. J. Obstet. Gynaecol. 2017, 124, 1746–1752. [Google Scholar] [CrossRef] [PubMed]
- O’Malley, E.G.; Reynolds, C.M.E.; O’Kelly, R.; Killalea, A.; Sheehan, S.R.; Turner, M.J. A Prospective Evaluation of Point-of-Care Measurements of Maternal Glucose for the Diagnosis of Gestational Diabetes Mellitus. Clin. Chem. 2020, 66, 316–323. [Google Scholar] [CrossRef] [PubMed]
- Khambule, L.; Chikomba, C.; Adam, Y.; Khan, L.; Haldane, C.; Vetter, B.; George, J. Performance of point-of-care glucose testing for the diagnosis of gestational diabetes in South Africa. Int. J. Gynecol. Obstet. 2025, 168, 812–821. [Google Scholar] [CrossRef] [PubMed]
- Pastakia, S.D.; Njuguna, B.; Onyango, B.A.; Washington, S.; Christoffersen-Deb, A.; Kosgei, W.K.; Saravanan, P. Prevalence of gestational diabetes mellitus based on various screening strategies in western Kenya: A prospective comparison of point of care diagnostic methods. BMC Pregnancy Childbirth 2017, 17, 226. [Google Scholar] [CrossRef]
- Zhang, P.; Engelgau, M.M.; Valdez, R.; Benjamin, S.M.; Cadwell, B.; Venkat Narayan, K.M. Costs of Screening for Pre-diabetes Among U.S. Adults: A comparison of different screening strategies. Diabetes Care 2003, 26, 2536–2542. [Google Scholar] [CrossRef]
- Pineda-Cortel, M.R.B.; Suratos, T.; Mamerto, T.P. Cut-off points of 75-gram oral glucose tolerance test as a diagnostic test for gestational diabetes mellitus in pregnant Filipino population. J. Lab. Precis. Med. 2022, 7, 31. [Google Scholar] [CrossRef]
- Malesker, M.A.; Hilleman, D.; Morrow, L.E.; Harris, J. Economic Evaluation of a Point-of-Care Blood Glucose Value. CHEST 2010, 138, 285A. [Google Scholar] [CrossRef]
- Blood Glucose Monitoring—Abstract—Europe PMC. Available online: https://europepmc.org/article/NBK/nbk555976 (accessed on 16 June 2025).
- Continuous Glucose Monitoring—NIDDK. National Institute of Diabetes and Digestive and Kidney Diseases. Available online: https://www.niddk.nih.gov/health-information/diabetes/overview/managing-diabetes/continuous-glucose-monitoring (accessed on 16 June 2025).
- Mathew, T.K.; Zubair, M.; Tadi, P. Blood Glucose Monitoring. In StatPearls; StatPearls Publishing: Treasure Island, FL, USA, 2025. Available online: http://www.ncbi.nlm.nih.gov/books/NBK555976/ (accessed on 16 June 2025).
- Larkins, M.C.; Thombare, A. Point-of-Care Testing. In StatPearls; StatPearls Publishing: Treasure Island, FL, USA, 2025. [Google Scholar]
- Nor, N.M.; Ridhuan, N.S.; Razak, K.A. Progress of Enzymatic and Non-Enzymatic Electrochemical Glucose Biosensor Based on Nanomaterial-Modified Electrode. Biosensors 2022, 12, 1136. [Google Scholar]
- Tan, X.C.; Tian, Y.X.; Cai, P.X.; Zou, X.Y. Glucose biosensor based on glucose oxidase immobilized in sol–gel chitosan/silica hybrid composite film on Prussian blue modified glass carbon electrode. Anal. Bioanal. Chem. 2005, 381, 500–507. [Google Scholar] [CrossRef]
- Fischback, M.; Kwon, K.Y.; Lee, I.; Shin, S.J.; Park, H.G.; Kim, B.C.; Kwon, Y.; Jung, H.-T.; Kim, J.; Ha, S. Enzyme precipitate coatings of glucose oxidase onto carbon paper for biofuel cell applications. Biotechnol. Bioeng. 2012, 109, 318–324. [Google Scholar] [CrossRef]
- Yadav, D.; Singh, S.P.; Dubey, P.K. Glucose Monitoring Techniques and Their Calibration. In Handbook of Metrology and Applications; Springer: Singapore, 2023; pp. 1855–1877. [Google Scholar] [CrossRef]
- Reddy, N.; Verma, N.; Dungan, K. Monitoring Technologies—Continuous Glucose Monitoring, Mobile Technology, Biomarkers of Glycemic Control. In Endotext [Internet]; MDText.com, Inc.: South Dartmouth, MA, USA, 2023. Available online: https://www.ncbi.nlm.nih.gov/sites/books/NBK279046/ (accessed on 16 June 2025).
- Katz, L.B.; Smith, A.; Cameron, H.; Setford, S.; Grady, M. 10/10 Accuracy Analysis of a Self-Monitoring Blood Glucose System. J. Diabetes Sci. Technol. 2021, 15, 717–718. [Google Scholar] [CrossRef]
Author, Year | Country | Sample Characteristics | Study Setting | Brief Study Protocol | Standard Criteria Used |
---|---|---|---|---|---|
Priya et al., 2011 [39] | India. | The sample size was 407. Participants were aged 20 years or older 54.1% were male With no known diabetes. | tertiary diabetes center. | CBG and VPG were assessed concurrently both in the fasting state and 2 h after a 75 g glucose load. | Diabetes, impaired fasting glucose (IFG) and impaired glucose tolerance (IGT) were defined using American Diabetes Association (ADA) and World Health Organization (WHO) criteria. |
Suresh Babu, G. et al., 2015 [40] | India | n = 182 pregnant women | Cross-sectional prospective study conducted at the Biochemistry laboratory of a hospital | Objective: To compare the blood glucose levels by glucometer and laboratory method in the diagnosis of GDM Both Venous and capillary blood were tested for glucose levels 2 h after 75 g Glucose load. | All testing was conducted by the same qualified and experienced lab technicians and under ambient conditions. |
Gallardo et al., 2020 [41] | Mexico | n = 328 pregnant women without diabetes | multicenter longitudinal cohort study in primary healthcare clinics. | The Objective was to compares the accuracy of two glucometers for GDM detection. | All participants were tested with OGTT for the diagnosis of GDM based on the ADA 2019 guidelines. |
Adam et al., 2018 [37] | South Africa | The sample size included 529 pregnant women without diabetes | Prospective cohort observational study at a primary healthcare clinic. | Objective: To evaluate the performance of the glucometer in the diagnosis of GDM. A 75 g 2 h OGTT was scheduled at 24–28 gestational weeks. Glucose was measured in venous and capillary blood | GDM was diagnosed via FIGO criteria. |
Lippi et al., 2025 [29] | Italy | n = 241 pregnant women | local phlebotomy center | Fasting glucose was measured in capillary blood using glucometer and in plasma with laboratory method using the hexokinase reference assay | NA |
Eskandarifar et al., 2019 [42] | Iran | The sample size included 130 critically ill infants less than 1 year of age | Besat hospital, Sanandaj at IRAN | Objective: To determine the accuracy of glucometer for early diagnosis of hypoglycemia in acutely ill infants. Blood sugar was measured by glucose oxidase method and glucometer reagent strip. | NA |
Wolde et al., 2018 [30] | Ethiopia | n = 200 100 with diabetes 100 healthy controls | Prospective cross-sectional study at Addis Ababa University | Four randomly selected POCG devices were evaluated against hexokinase method | ISO 15197:2003 and ISO 15197:2013 standards |
Ogunbosi et al., 2022 [31] | Nigeria | A total of 295 children between the ages of 0 and 15 years were included. | A cross-sectional study conducted in pediatric emergency and outpatient departments of two major hospitals. | Blood glucose levels were measured with One Touch® (LifeScan Inc., Milpitas, CA, USA) and Accu-Check® (Roche Diagnostics, Mannheim, Germany) glucometers and the glucose oxidase method at the same time and determined the effect of hematocrit on glucose readings. Study period: over a 6-month period. | NA |
Daly et al., 2017 [43] | Ireland | A total of 108 pregnant women between 24 and 28 weeks of gestation were enrolled | The study followed a prospective observational design at a perinatal facility |
Women screened selectively with a one-step 75 g OGTT recruited. At each OGTT time point two venous samples and one capillary sample taken A capillary sample was used for glucometer testing. | NA |
O’Malley et al., 2020 [44] | Ireland | The study included 202 female participants. | Prospective observational study |
The objective was to evaluate the use of POC measurements of maternal glucose to diagnose GDM with a 1-step 75 g OGTT. Maternal plasma and capillary glucose measured at fasting and at 1 and 2 h post glucose load. | Using updated laboratory standards as the reference |
Khambule et al., 2025 [45] | South Africa | n = 1076 pregnant women | Cross-sectional study in a prenatal clinic in Johannesburg. | Both venous and capillary blood evaluated for laboratory-based and POC glucose measurements | The up-to-date International Association of Diabetes and Pregnancy Study Groups diagnostic criteria were used. |
Pastakia et al., 2017 [46] | Sub-Saharan Africa (SSA) | n = 616 Pregnant women between 24 and 32 weeks of a singleton pregnancy | Prospective study at antenatal clinic |
Objective: To assess utility of various GDM POC screening strategies in a resource-constrained setting Testing over two days. Day 1: a POC 1 h 50 g glucose challenge test (GCT) and a POC glycated hemoglobin (HbA1c) assessed. Day 2: fasting blood glucose, 1 h and 2 h 75 g oral glucose tolerance test (OGTT) determined using both venous and POC tests, along with a venous HbA1c. | OGTTs conducted as per International Association of Diabetes and Pregnancy Study Groups (IADPSG) guidelines. |
Vučić Lovrenčić et al., 2013 [9] | Croatia | n = 237 participants with a previous history of dysglycemia | Prospective observational study at Vuk Vrhovac University Clinic |
Objective: to investigate the diagnostic accuracy of an innovative, interference-resistant POC glucose meter Venous and capillary blood sampling for the reference laboratory procedure and POC-glucose measurement was carried out at fasting and 2 h OGTT. | The International Association of Diabetes in Pregnancy Study Group (IADPSG) criteria was used to diagnose GDM. |
Al-Hasani et al., 2024 [34] | UK | n = 230 pregnant women | Prospective cohort study at antenatal clinics |
Objective:
To assess the clinical utility of POC CBG testing in the assessment of GDM during OGTT CBG was measured using the POC. VPG was measured by Roche analyzer The two methods were compared statistically | Categories of glucose tolerance were classified according to 2006 WHO diagnostic criteria. |
Fabre-Estremera et al., 2024 [18] | Spain | n = 98 pediatric patients | Prospective observational study at La Paz University Hospital. |
Objective: To evaluate the accuracy POCT glucometers during an OGTT for prediabetes and diabetes diagnosis in a comparison study. Glycaemia measured in venous blood using two glucometers with lab analysis as a reference | GDM was diagnosed based on the 2015 National Institute for Health and Clinical Excellence (NICE) Clinical Guideline criteria. |
Tan et al., 2021 [35] | Singapore | n = 30 patients with history of gestational diabetes or prediabetes | Prospective observational study at polyclinic |
Objective: to assess the feasibility and precision of a self-administered capillary OGTT for type-2 diabetes mellitus in high-risk individuals. Self-administered the capillary OGTT and concurrently their venous glucose samples were obtained. | NA |
NA |
Author, Year | Sensitivity and Specificity | Agreement Between Capillary and Venous Blood Glucose Measurement | Conclusion | Implications |
---|---|---|---|---|
Priya et al., 2011 [38] | NA | r = 0.681 (p < 0.001) (fasting state) r = 0.897 (p < 0.001) for the 2 h PG load Diabetes diagnosis rate: (capillary vs. venous) = 31.9% versus 21.1% (ADA) = 43.2% versus 38.6% (WHO) Accuracy of identifying diabetes = 83.3% (ADA) = 90.9% (WHO) | CBG measurement by Glucometer is a feasible alternative in developing countries. |
CBG measurement by Glucometer can be used for screening of diabetes and IGT in epidemiological studies It can be used where obtaining venous samples may be difficult. |
Suresh Babu, G. et al., 2015 [39] | sensitivity = 100% specificity = 98.8% | AUC = 0.994 Good agreement between glucometry and laboratory analysis, r = 0.9681 Bland—Altman (difference) Plot: a constant bias 1.7% with SD 4.3 (95% CI: −6.7 to 10) | Estimation of glucose using single step approach with a single finger prick capillary blood drop and instant results is a promising test. | Clinical judgment is a must for final decision. Safe to diagnose diabetes using a glucometer, i.e., CBG Usage requires high precision Continuous quality assurance procedures needed. |
Gallardo et al., 2020 [40] | First model POC venous OGTT sensitivity = 100% specificity = 62.8%. The second model, POC capillary OGTT; Sensitivity = 78.57% specificity = 74.1% | For the first model, POC venous OGTT; GDM incidence = 41.66% compared to 7.05% of the plasmatic test, The second model, POC capillary OGTT; GDM incidence = 30.23% compared to 8.13% of the plasmatic test ROC area under the curve for GDM prediction was 0.81 95% CI = 0.77–0.85 compared to the first model, ROC area under the curve = 0.76 95% CI = 0.65–0.88 |
Capillary OGTT is a valid alternative to the gold standard OGTT Especially important in low resource setting. The positive bias could be beneficial. As early treatment and control related to better perinatal health outcomes | POC OGTT can reduce diagnosis time. Reduce cost of diagnosis. Helpful in low resource setting. Further analysis needed to improve GDM, POC screening interventions. |
Adam et al., 2018 [37] | Sensitivity = 27.0% specificity = 89.4% |
Diagnosed with GDM by laboratory = 26.7% and glucometer measurements = 14.9%, CV = 15% to 17%. Bland–Altman plots: a positive bias of the glucometer results at 0 h, a negative bias at 1 and 2 h of the OGTT. | They had not recommended use of the Roche Accuchek Active glucometer for the diagnosis of GDM | The use of POC glucometers is not recommended. |
Lippi et al., 2025 [29] |
Sensitivity = 50.0% Specificity = 100.0% (For diagnosing fasting glucose values ≥ 7.0 mmol/L compared to the laboratory assay). |
The diagnostic accuracy = 99.2% The mean turnaround time (TAT) laboratory-based strategy vs. Glucometer = 32 min 8 vs. 8 s Imprecision of the glucometer vs. laboratory assay = 3.4% vs. 0.8% | Screening fasting glucose in capillary blood with a POC glucometer allows faster patient management but is associated with higher imprecision, inaccuracy, costs and avoidable finger pricks | POC capillary blood protocol is associated with higher imprecision |
Eskandarifar et al., 2019 [42] | sensitivity = 72%, specificity = 53%, |
Positive predictive value = 62% Negative predictive value = 7% correlation was significant (p < 0.001). Kappa statistics = 42%. | Glucometer based test cannot be regarded as a very suitable and reliable tool for diagnosis of hypoglycemia in critically ill infants. |
Glucometer cannot be used in diagnosing hypoglycemia in critically ill infants It may be appropriate for rapid screening in emergency situations |
Wolde et al., 2018 [30] | NA |
All four PoCG devices had strong positive relationship (>80%) with the reference method concentrations. None of the devices fulfilled the minimum accuracy measurement set by ISO standards. | Four PoCG measurements were poorly correlated with standard reference method. | The study highlighted the need for a standardized evaluation process before new glucometers are introduced to the Ethiopian market. |
Ogunbosi et al., 2022 [31] | NA |
Bland–Altman: acceptable level of bias (3.9 mg/dL) between the two glucometers. Correlation analysis: significant correlation between each of the glucometer methods and laboratory blood sugar | Though it can aid rapid decision-making, there is a need to periodically cross-check with the glucose oxidase method in the laboratory to optimize outcome | Use of a tested glucometer in clinical settings can aid in rapid decision-making |
Daly et al., 2017 [43] | Sensitivity = 92.5%, specificity = 76.5%, (based on adjustment of the POC fasting diagnostic threshold from ≥5.1 to ≥4.8 mol/L (aPOC)) | GDM was detected in using the reference standard, 47.2% (n = 51), customary practices, 17.6% (n = 19) and POC, 24.1% (n = 26) (p < 0.001). based on adjustment of the POC fasting diagnostic threshold from ≥5.1 to ≥4.8 mol/L (aPOC), PPV = 69.8%, NPV = 94.5% Accuracy = 94.5% | POC capillary maternal glucose tests were superior to customary laboratory practices for diagnosing GDM, particularly in low resource healthcare settings. | In low resource setting POC capillary maternal glucose test is useful in diagnosing GDM |
O’Malley et al., 2020 [44] | NA |
Based on the plasma measurements, 53.5% had GDM. As a predictor of GDM, diagnostic accuracy of POC measurement 83.0% Diagnostic accuracy of POC measurement = 83.0% | POC device can be used in low resource setting though not recommended in high resource setting. | If measures to inhibit glycolysis are available to implement, then use of POC device use may not be recommended. |
Khambule et al., 2025 [45] |
POC glucometers sensitivity = 17.6% to 87.18%, Specificity = 62.7% and 99.8%. (Laboratory-based fasting plasma glucose (FPG) sensitivity = 94%, specificity = 100%) | Bland–Altman plots: All POC glucometers showed moderate to poor reliability. The AUC = 0.59 to 0.79. (AUC = 0.98 for laboratory-based test) | They recommended laboratory-based FPG over POC glucometer as a diagnostic test for GDM. | Laboratory methods are recommended for the diagnosis of GDM |
Pastakia et al., 2017 [46] |
Compared to IADPSG testing, POC IADPSG had a sensitivity = 55.6% and specificity = 90.6%, respectively, while that of POC 1 h 50 g GCT was sensitivity = 55.6% and specificity = 63.9%. | GDM was diagnosed in 18 women, a prevalence of 2.9% | Though POC screening strategies feasible, it showed poor sensitivity for GDM detection in resource-constrained population of low GDM prevalence. | Studies required to identify sensitive and specific POC GDM screening strategies using adverse pregnancy outcomes as end points. |
Vučić Lovrenčić et al., 2013 [9] | NA |
Weighted Kappa = 0.858. Bland–Altman analysis: slight bias between the RLP- and POC-FPG | StatStrip POC glucose meter could serve as a reliable tool for the diabetes diagnosis, particularly in primary healthcare facilities with dispersed blood sampling services. | POC glucose could be used as an alternative to laboratory methods in low resource settings. |
Al-Hasani et al., 2024 [34] |
For the POC StatStrip® test, at 95% CI, Fasting Sensitivity = 88% (52–99%) Specificity = 97% (93–98%) at 2 h, Sensitivity = 97% (91–99%) Specificity = 79% (71–84%) |
POC StatStrip® test versus laboratory VPG measurement 15 (6.5%) versus eight (3.4%) at fasting and 105 (45.6%) versus 72 (31.1%) at 2 h Specificity and the NPV for the POC StatStrip® test for concentrations of ≤5.0 mmol/L at fasting or <7.5 mmol/L at 2 h were 100%, and Sensitivity and the PPV for concentrations of >9.5 mmol/L at 2 h were 100%. | Though POC measurement of CBG cannot entirely replace the laboratory method for the OGTT; it can be used to rule out/rule in GDM for glucose concentrations of ≤5.0 mmol/L at fasting or <7.5/>9.5 mmol/L at 2 h. | POC CBG can be used to aid in diagnosing GDM |
Fabre-Estremera et al., 2024 [18] | NA |
The diagnostic concordance between connected glucometer and the central laboratory was 71.1% Same clinical decision in the 92.8% of the cases, but treatment would have not been indicated in 4 patients (4.1%). | POCT glucometers show high correlation and accuracy compared with standard procedures but may not be used for diagnosis yet as severe clinical impact could happen. |
POCT glucometers can reduce time Reduce cost Highly correlated to central laboratory testing |
Tan et al., 2021 [35] | capillary OGTT sensitivity = 94.1% |
NPV = 91.7% r = 0.95; p < 0.001 (for fasting) and 2 h post-OGTT, r = 0.95; p < 0.001 The Fleiss’ Kappa Score (0.79, p < 0.0001) | Self-administered capillary OGTT is feasible and acceptable, especially among younger adults, with excellent sensitivity and NPV compared with plasma-based OGTT |
Capillary OGTT can be useful in identifying prediabetes and T2DM (comparing to venous sample) |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2025 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Ahmed, T.; Kishore, J.; Onamika, M.; Goswami, S.; Momo, I.R.; Sumon, H.; Bowden, R.G. Comparing Glucometer-Based and Laboratory-Based OGTT for Diabetes Diagnosis: A Narrative Review. Endocrines 2025, 6, 48. https://doi.org/10.3390/endocrines6030048
Ahmed T, Kishore J, Onamika M, Goswami S, Momo IR, Sumon H, Bowden RG. Comparing Glucometer-Based and Laboratory-Based OGTT for Diabetes Diagnosis: A Narrative Review. Endocrines. 2025; 6(3):48. https://doi.org/10.3390/endocrines6030048
Chicago/Turabian StyleAhmed, Tanvir, Jaimala Kishore, Mensila Onamika, Shuvratithi Goswami, Iffat Rahman Momo, Hanif Sumon, and Rodney G. Bowden. 2025. "Comparing Glucometer-Based and Laboratory-Based OGTT for Diabetes Diagnosis: A Narrative Review" Endocrines 6, no. 3: 48. https://doi.org/10.3390/endocrines6030048
APA StyleAhmed, T., Kishore, J., Onamika, M., Goswami, S., Momo, I. R., Sumon, H., & Bowden, R. G. (2025). Comparing Glucometer-Based and Laboratory-Based OGTT for Diabetes Diagnosis: A Narrative Review. Endocrines, 6(3), 48. https://doi.org/10.3390/endocrines6030048