Next Article in Journal
Colorimetric Methods of Magnesium Detection for Point-of-Care Heart Failure Management
Previous Article in Journal
Advancing Food Safety Sensing through Artificial Intelligence: Machine Learning-Enhanced Biosensors in Action
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Abstract

Investigation of Affordable Electrode Material Combinations in Electrochemical Biosensors †

Graduate School of Information, Production and Systems, Waseda University, Kitakyushu 808-0135, Japan
*
Author to whom correspondence should be addressed.
Presented at the 4th International Electronic Conference on Biosensors, 20–22 May 2024; Available online: https://sciforum.net/event/IECB2024.
Proceedings 2024, 104(1), 27; https://doi.org/10.3390/proceedings2024104027
Published: 28 May 2024
(This article belongs to the Proceedings of The 4th International Electronic Conference on Biosensors)
This study investigated carbon-based electrode materials for the application of wearable biosensors measuring uric acid concentration. This investigation focused on assessing various combinations of carbon-based electrodes, such as CVD-graphite, graphene ink, and carbon nanotubes, aiming to enhance sensor performance. These carbon materials were deposited on stainless steel (SUS304 SS) plates (2 mm × 10 mm), and their sensitivity for detecting uric acid was evaluated by utilizing cyclic voltammetry (CV) and employing differential pulse voltammetry (DPV). Among the nine electrode combinations tested with potassium ferricyanide, two were identified that generated reversible CV waveforms, indicating their potential for uric acid detection. The effective combinations consisted of one set with CVD graphite for both the working and counter electrodes (WE and CE) and another set combining CVD graphite for the WE with carbon nanotube ink for the CE. The assessments demonstrated that these electrodes could efficiently detect the electron current from the redox reaction of ferricyanide, indicating their potential capability to measure uric acid concentrations effectively. Subsequent analysis using uric acid concentrations ranging from 100 µM to 400 µM showed that CVD-graphite electrodes provided distinct, quantifiable responses in both CV and DPV tests. Notably, the DPV profile for CVD-graphite displayed a significantly high peak current, facilitating the creation of a reliable calibration curve for uric acid concentration detection. The findings concluded that CVD-graphite stands out as the optimal material for both the WE and the CE in uric acid detection within sweat, with a practical detection limit (LOD) of 5.25 µM. This LOD is notably significant, aligning well with the average uric acid concentration in normal sweat, approximately 60 µM, thereby affirming the viability of these carbon-based electrodes for effective disease diagnosis and monitoring through wearable biosensors.

Supplementary Materials

The conference presentation is available online at https://www.mdpi.com/article/10.3390/proceedings2024104027/s1.

Author Contributions

Conceptualization, T.N. and J.K.; methodology, T.N. and J.K.; software, T.N. and J.K.; validation, T.N. and J.K.; formal analysis, T.N. and J.K.; investigation, T.N. and J.K.; resources, S.J. and K.U.; data curation, T.N.; writing—original draft preparation, T.N.; writing—review and editing, T.N. visualization, T.N.; supervision, J.K.; project administration, T.N. and J.K. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Data associated with this work can be obtained by contacting the author.

Conflicts of Interest

The author declares no conflict of interest.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Nohgi, T.; Jia, S.; Ueda, K.; Kameoka, J. Investigation of Affordable Electrode Material Combinations in Electrochemical Biosensors. Proceedings 2024, 104, 27. https://doi.org/10.3390/proceedings2024104027

AMA Style

Nohgi T, Jia S, Ueda K, Kameoka J. Investigation of Affordable Electrode Material Combinations in Electrochemical Biosensors. Proceedings. 2024; 104(1):27. https://doi.org/10.3390/proceedings2024104027

Chicago/Turabian Style

Nohgi, Toru, Siyu Jia, Kenji Ueda, and Jun Kameoka. 2024. "Investigation of Affordable Electrode Material Combinations in Electrochemical Biosensors" Proceedings 104, no. 1: 27. https://doi.org/10.3390/proceedings2024104027

APA Style

Nohgi, T., Jia, S., Ueda, K., & Kameoka, J. (2024). Investigation of Affordable Electrode Material Combinations in Electrochemical Biosensors. Proceedings, 104(1), 27. https://doi.org/10.3390/proceedings2024104027

Article Metrics