Perioperative Assessment and Clinical Outcomes of Percutaneous Atrial Septal Defect Closure with Three-Dimensional Transesophageal Echocardiography
Abstract
:1. Introduction
2. Materials and Methods
2.1. Patients Selection
2.2. 2D TEE and 3D TEE Procedure and Image Optimization
2.3. Data Analysis
2.4. Transcatheter ASD Closure and Patient Follow-Up
2.5. Statistical Analysis
3. Results
4. Discussion
5. Conclusions
Limitations
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
References
- Brida, M.; Chessa, M.; Celermajer, D.; Li, W.; Geva, T.; Khairy, P.; Griselli, M.; Baumgartner, H.; Gatzoulis, M.A. Atrial septal defect in adulthood: A new paradigm for congenital heart disease. Eur. Heart J. 2022, 43, 2660–2671. [Google Scholar] [CrossRef] [PubMed]
- Fraisse, A.; Latchman, M.; Sharma, S.R.; Bayburt, S.; Amedro, P.; di Salvo, G.; Baruteau, A.E. Atrial septal defect closure: Indications and contra-indications. J. Thorac. Dis. 2018, 10, S2874–S2881. [Google Scholar] [CrossRef] [PubMed]
- Baumgartner, H.; De Backer, J.; Babu-Narayan, S.V.; Budts, W.; Chessa, M.; Diller, G.P.; Lung, B.; Kluin, J.; Lang, I.M.; Meijboom, F.; et al. 2020 ESC Guidelines for the management of adult congenital heart disease. Eur. Heart J. 2021, 42, 563–645. [Google Scholar] [CrossRef] [PubMed]
- Stout, K.K.; Daniels, C.J.; Aboulhosn, J.A.; Bozkurt, B.; Broberg, C.S.; Colman, J.M.; Crumb, S.R.; Dearani, J.A.; Fuller, S.; Gurvitz, M.; et al. 2018 AHA/ACC Guideline for the Management of Adults With Congenital Heart Disease: Executive Summary: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines. J. Am. Coll. Cardiol. 2019, 73, 1494–1563. [Google Scholar] [CrossRef] [PubMed]
- Suchon, E.; Pieculewicz, M.; Tracz, W.; Przewlocki, T.; Sadowski, J.; Podolec, P. Transcatheter closure as an alternative and equivalent method to the surgical treatment of atrial septal defect in adults: Comparison of early and late results. Med. Sci. Monit. 2009, 15, Cr612–Cr617. [Google Scholar] [PubMed]
- Seo, J.S.; Song, J.M.; Kim, Y.H.; Park, D.W.; Lee, S.W.; Kim, W.J.; Kim, D.H.; Kang, D.H.; Song, J.K. Effect of atrial septal defect shape evaluated using three-dimensional transesophageal echocardiography on size measurements for percutaneous closure. J. Am. Soc. Echocardiogr. 2012, 25, 1031–1040. [Google Scholar] [CrossRef]
- Carlson, K.M.; Justino, H.; O’Brien, R.E.; Dimas, V.V.; Leonard, G.T., Jr.; Pignatelli, R.H.; Mullins, C.E.; Smith, E.O.; Grifka, R.G. Transcatheter atrial septal defect closure: Modified balloon sizing technique to avoid overstretching the defect and oversizing the Amplatzer septal occluder. Catheter. Cardiovasc. Interv. 2005, 66, 390–396. [Google Scholar] [CrossRef] [PubMed]
- Quek, S.C.; Wu, W.X.; Chan, K.Y.; Ho, T.F.; Yip, W.C. Transcatheter closure of atrial septal defects—Is balloon sizing still necessary? Ann. Acad. Med. Singap. 2010, 39, 390–393. [Google Scholar] [CrossRef] [PubMed]
- Alsaileek, A.A.; Omran, A.; Godman, M.; Najm, H.K. Echocardiographic visualization of laceration of atrial septum during balloon sizing of atrial septal defect. Eur. J. Echocardiogr. 2007, 8, 155–157. [Google Scholar] [CrossRef]
- Song, B.G.; Park, S.W.; Lee, S.C.; Choi, J.O.; Park, S.J.; Chang, S.A.; Oh, J.K.; Yang, J.H. Real-time 3D TEE for multiperforated interatrial septum. JACC Cardiovasc. Imaging 2010, 3, 1199. [Google Scholar] [CrossRef]
- Lang, R.M.; Badano, L.P.; Tsang, W.; Adams, D.H.; Agricola, E.; Buck, T.; Faletra, F.F.; Franke, A.; Hung, J.; de Isla, L.P.; et al. EAE/ASE recommendations for image acquisition and display using three-dimensional echocardiography. Eur. Heart J. Cardiovasc. Imaging 2012, 13, 1–46. [Google Scholar] [CrossRef] [PubMed]
- Roberson, D.A.; Cui, V.W. Three-dimensional transesophageal echocardiography of atrial septal defect device closure. Curr. Cardiol. Rep. 2014, 16, 453. [Google Scholar] [CrossRef]
- Taha, F.; Elshedoudy, S. Role of 3D transesophageal echocardiography in transcatheter closure of atrial septal aneurysms. Echocardiography 2019, 36, 1884–1894. [Google Scholar] [CrossRef] [PubMed]
- Kitakata, H.; Itabashi, Y.; Kanazawa, H.; Miura, K.; Kimura, M.; Shinada, K.; de Jongh, M.C.; Endo, J.; Tsuruta, H.; Murata, M.; et al. Appropriate device selection for transcatheter atrial septal defect closure using three-dimensional transesophageal echocardiography. Int. J. Cardiovasc. Imaging 2021, 37, 1159–1168. [Google Scholar] [CrossRef]
- Hascoet, S.; Hadeed, K.; Marchal, P.; Dulac, Y.; Alacoque, X.; Heitz, F.; Acar, P. The relation between atrial septal defect shape, diameter, and area using three-dimensional transoesophageal echocardiography and balloon sizing during percutaneous closure in children. Eur. Heart J. Cardiovasc. Imaging 2015, 16, 747–755. [Google Scholar] [CrossRef] [PubMed]
- Cho, E.J. Clinical Implication (Application) of Measurement of Device Size in Patients With Atrial Septal Defects (ASD) by 3-Dimensional Transesophageal Echocardiography (3D TEE). J. Cardiovasc. Imaging 2023, 31, 188–190. [Google Scholar] [CrossRef] [PubMed]
- Johri, A.M.; Witzke, C.; Solis, J.; Palacios, I.F.; Inglessis, I.; Picard, M.H.; Passeri, J.J. Real-time three-dimensional transesophageal echocardiography in patients with secundum atrial septal defects: Outcomes following transcatheter closure. J. Am. Soc. Echocardiogr. 2011, 24, 431–437. [Google Scholar] [CrossRef]
- Deng, B.; Chen, K.; Huang, T.; Wei, Y.; Liu, Y.; Yang, L.; Qiu, Q.; Zheng, S.; Lv, H.; Wang, P.; et al. Assessment of atrial septal defect using 2D or real-time 3D transesophageal echocardiography and outcomes following transcatheter closure. Ann. Transl. Med. 2021, 9, 1309. [Google Scholar] [CrossRef]
- Silvestry, F.E.; Cohen, M.S.; Armsby, L.B.; Burkule, N.J.; Fleishman, C.E.; Hijazi, Z.M.; Lang, R.M.; Rome, J.J.; Wang, Y. Guidelines for the Echocardiographic Assessment of Atrial Septal Defect and Patent Foramen Ovale: From the American Society of Echocardiography and Society for Cardiac Angiography and Interventions. J. Am. Soc. Echocardiogr. 2015, 28, 910–958. [Google Scholar] [CrossRef] [PubMed]
- Mahmoud, H.M.; Al-Ghamdi, M.A.; Ghabashi, A.E.; Anwar, A.M. A Proposed Maneuver to Guide Transseptal Puncture Using Real-Time Three-Dimensional Transesophageal Echocardiography: Pilot Study. Cardiol. Res. Pract. 2015, 2015, 174051. [Google Scholar] [CrossRef]
- Song, J.; Lee, S.Y.; Baek, J.S.; Shim, W.S.; Choi, E.Y. Outcome of transcatheter closure of oval shaped atrial septal defect with amplatzer septal occluder. Yonsei Med. J. 2013, 54, 1104–1109. [Google Scholar] [CrossRef] [PubMed]
- Kijima, Y.; Akagi, T.; Nakagawa, K.; Takaya, Y.; Oe, H.; Ito, H. Three-dimensional echocardiography guided closure of complex multiple atrial septal defects. Echocardiography 2014, 31, E304–E306. [Google Scholar] [CrossRef]
- Majunke, N.; Bialkowski, J.; Wilson, N.; Szkutnik, M.; Kusa, J.; Baranowski, A.; Heinisch, C.; Ostermayer, S.; Wunderlich, N.; Sievert, H. Closure of atrial septal defect with the Amplatzer septal occluder in adults. Am. J. Cardiol. 2009, 103, 550–554. [Google Scholar] [CrossRef] [PubMed]
- Shrivastava, S.; Shrivastava, S.; Allu, S.V.V.; Schmidt, P. Transcatheter Closure of Atrial Septal Defect: A Review of Currently Used Devices. Cureus 2023, 15, e40132. [Google Scholar] [CrossRef] [PubMed]
- Boon, I.; Vertongen, K.; Paelinck, B.P.; Demulier, L.; Van Berendoncks, A.; De Maeyer, C.; Marchau, F.; Panzer, J.; Vandekerckhove, K.; De Wolf, D. How to Size ASDs for Percutaneous Closure. Pediatr. Cardiol. 2018, 39, 168–175. [Google Scholar] [CrossRef]
- Harikrishnan, S.; Narayanan, N.K.; Sivasubramonian, S. Sizing balloon-induced tear of the atrial septum. J. Invasive Cardiol. 2005, 17, 546–547. [Google Scholar]
- Jang, J.Y.; Heo, R.; Cho, M.S.; Bae, J.; Hong, J.A.; Lee, S.; Ahn, J.M.; Park, D.W.; Kim, D.H.; Kang, D.H.; et al. Efficacy of 3D transoesophageal echocardiography for transcatheter device closure of atrial septal defect without balloon sizing. Eur. Heart J. Cardiovasc. Imaging 2018, 19, 684–689. [Google Scholar] [CrossRef] [PubMed]
- Kim, B.J.; Song, J.; Huh, J.; Kang, I.S. Outcomes After Transcatheter Closure of Atrial Septal Defect Without Using a Balloon-Sizing Technique: A Randomized Controlled Comparison with Closures Using a Balloon-Sizing Technique. Heart Surg. Forum 2022, 25, E288–E293. [Google Scholar] [CrossRef]
- Wang, J.K.; Tsai, S.K.; Lin, S.M.; Chiu, S.N.; Lin, M.T.; Wu, M.H. Transcatheter closure of atrial septal defect without balloon sizing. Catheter. Cardiovasc. Interv. 2008, 71, 214–221. [Google Scholar] [CrossRef]
- Gupta, S.K.; Sivasankaran, S.; Bijulal, S.; Tharakan, J.M.; Harikrishnan, S.; Ajit, K. Trans-catheter closure of atrial septal defect: Balloon sizing or no Balloon sizing—Single centre experience. Ann. Pediatr. Cardiol. 2011, 4, 28–33. [Google Scholar] [CrossRef]
- Sobrino, A.; Basmadjian, A.J.; Ducharme, A.; Ibrahim, R.; Mercier, L.A.; Pelletier, G.B.; Marcotte, F.; Garceau, P.; Burelle, D.; O’Meara, E.; et al. Multiplanar transesophageal echocardiography for the evaluation and percutaneous management of ostium secundum atrial septal defects in the adult. Arch. Cardiol. Mex. 2012, 82, 37–47. [Google Scholar] [PubMed]
- Simpson, J.; Lopez, L.; Acar, P.; Friedberg, M.K.; Khoo, N.S.; Ko, H.H.; Marek, J.; Marx, G.; McGhie, J.S.; Meijboom, F.; et al. Three-dimensional Echocardiography in Congenital Heart Disease: An Expert Consensus Document from the European Association of Cardiovascular Imaging and the American Society of Echocardiography. J. Am. Soc. Echocardiogr. 2017, 30, 1–27. [Google Scholar] [CrossRef] [PubMed]
- Mani, A.; Harikrishnan, S.; Sasidharan, B.; Ganapathi, S.; Valaparambil, A.K. Utility of 3D Echocardiography for Device Sizing During Transcatheter ASD Closure: A Comparative Study. J. Cardiovasc. Imaging 2023, 31, 180–187. [Google Scholar] [CrossRef] [PubMed]
- Roushdy, A.; El Sayegh, A.; Ali, Y.A.; Attia, H.; El Fiky, A.; El Sayed, M. A novel three-dimensional echocardiographic method for device size selection in patients undergoing ASD trans-catheter closure. Egypt. Heart J. 2019, 72, 1. [Google Scholar] [CrossRef] [PubMed]
- Perone, F.; Bernardi, M.; Redheuil, A.; Mafrica, D.; Conte, E.; Spadafora, L.; Ecarnot, F.; Tokgozoglu, L.; Santos-Gallego, C.G.; Kaiser, S.E.; et al. Role of Cardiovascular Imaging in Risk Assessment: Recent Advances, Gaps in Evidence, and Future Directions. J. Clin. Med. 2023, 12, 5563. [Google Scholar] [CrossRef] [PubMed]
- Egidy Assenza, G.; Spinardi, L.; Mariucci, E.; Balducci, A.; Ragni, L.; Ciuca, C.; Formigari, R.; Angeli, E.; Vornetti, G.; Gargiulo, G.D.; et al. Transcatheter Closure of PFO and ASD: Multimodality Imaging for Patient Selection and Perioperative Guidance. J. Cardiovasc. Dev. Dis. 2021, 8, 78. [Google Scholar] [CrossRef]
Age/year (mean ± standard deviation) | 35.05 ± 13.87 |
Follow-up period/month(mean ± standard deviation) | 15.30 ± 9.18 |
Women (n, %) | 41 (67.2) |
CAD (n, %) | 7 (11.4) |
DM (n, %) | 5 (8.19) |
AF (n, %) | 2 (3.2) |
HT (n, %) | 8 (13.1) |
TIA/Stroke (n, %) | 4 (6.55) |
PH (n, %) (MPAP > 20 mmHg via RHC) | 26 (42.6) |
Symptoms with RHD (n, %) | 48 (78.6) |
Symptoms without RHD (n, %) (Qp/Qs > 1.5) | 9 (14.7) |
RHC (n, %) | 38 (63.9) |
COPD (n, %) | 5 (8.19) |
CKD (n, %) | 4 (6.55) |
Circular | Oval | Complex | |||||
---|---|---|---|---|---|---|---|
Residual Rims (mm) | Imaging | Mean ± SD | p | Mean ± SD | p | Mean ± SD | p |
AV valve | 2D | 9.36 ± 2.51 | 0.550 | 7.04 ± 2.04 | 0.259 | 5.97 ± 0.91 | 0.626 |
3D | 9.80 ± 2.70 | 7.45 ± 1.66 | 6.12 ± 1.02 | ||||
Aortic | 2D | 7.18 ± 2.51 | 0.563 | 5.21 ± 2.66 | 0.540 | 0.95 ± 1.67 | 0.012 |
3D | 7.47 ± 2.79 | 5.81 ± 2.34 | 2.29 ± 2.23 | ||||
SVC | 2D | 14.63 ± 3.68 | 0.680 | 13.45 ± 3.31 | 0.683 | 12.90 ± 1.54 | 0.115 |
3D | 15.08 ± 3.91 | 13.94 ± 3.65 | 13.21 ± 1.65 | ||||
IVC | 2D | 10.27 ± 1.88 | 0.695 | 8.15 ± 1.77 | 0.215 | 6.66 ± 1.11 | <0.001 |
3D | 10.48 ± 1.92 | 8.94 ± 1.88 | 8.31 ± 0.99 | ||||
Posterior | 2D | 12.74 ± 3.03 | 0.422 | 11.61 ± 2.58 | 0.586 | 10.82 ± 2.49 | 0.121 |
3D | 13.26 ± 3.21 | 12.12 ± 2.77 | 12.13 ± 2.24 | ||||
Min ASD Dimension (mm) | 2D | 14.51 ± 2.52 | 0.733 | 13.18 ± 2.52 | 0.091 | 18.77 ± 3.45 | 0.037 |
3D | 14.15 ± 3.21 | 11.85 ± 1.88 | 21.18 ± 3.57 |
Defect Type | Imaging and Device | Mean ± SD | p |
---|---|---|---|
Circular | Device size (mm) | 18.38 ± 4.12 | 0.446 |
2D size (mm) | 16.96 ± 3.68 | ||
3D size (mm) | 17.50 ± 3.78 | ||
Oval | Device size (mm) | 21.35 ± 3.26 | 0.380 |
2D size (mm) | 19.14 ± 2.92 | ||
3D size (mm) | 20.28 ± 3.01 | ||
Complex | Device size (mm) | 28.95 ± 3.86 | 0.069 |
2D size (mm) | 24.52 ± 3.72 | ||
3D size (mm) | 27.20 ± 3.53 | ||
Comparison of the difference between maximum dimension measurements obtained by imaging methods and implanted device size. | |||
Defect type | Device size-measured dimension | Mean ± SD | p |
Circular | Device-2D (mm) | 1.41 ± 0.86 | 0.549 |
Device-3D (mm) | 0.88 ± 0.60 | ||
Oval | Device-2D (mm) | 2.21 ± 0.93 | 0.170 |
Device-3D (mm) | 1.07 ± 0.55 | ||
Complex | Device-2D (mm) | 4.44 ± 1.04 | 0.025 |
Device-3D (mm) | 1.76 ± 0.65 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Saraç, İ.; Birdal, O. Perioperative Assessment and Clinical Outcomes of Percutaneous Atrial Septal Defect Closure with Three-Dimensional Transesophageal Echocardiography. Diagnostics 2024, 14, 1755. https://doi.org/10.3390/diagnostics14161755
Saraç İ, Birdal O. Perioperative Assessment and Clinical Outcomes of Percutaneous Atrial Septal Defect Closure with Three-Dimensional Transesophageal Echocardiography. Diagnostics. 2024; 14(16):1755. https://doi.org/10.3390/diagnostics14161755
Chicago/Turabian StyleSaraç, İbrahim, and Oğuzhan Birdal. 2024. "Perioperative Assessment and Clinical Outcomes of Percutaneous Atrial Septal Defect Closure with Three-Dimensional Transesophageal Echocardiography" Diagnostics 14, no. 16: 1755. https://doi.org/10.3390/diagnostics14161755
APA StyleSaraç, İ., & Birdal, O. (2024). Perioperative Assessment and Clinical Outcomes of Percutaneous Atrial Septal Defect Closure with Three-Dimensional Transesophageal Echocardiography. Diagnostics, 14(16), 1755. https://doi.org/10.3390/diagnostics14161755