Article
Design and Manufacture of a Test Device for Radiosynthesizer Vacuum Pumps
by Victor Amador Diaz, Scott E. Snyder and Amy L. Vavere
Instruments 2023, 7(2), 15; https://doi.org/10.3390/instruments7020015 - 6 Apr 2023
Viewed by 3200
Abstract
Vacuum pump wear is the most prevalent failure mode of the IBA Synthera® automated radiochemistry system. Rebuilding or replacing the pump causes equipment downtime and increases the radiation exposure of the service personnel. We built a dedicated test device to assess new [...] Read more.
Vacuum pump wear is the most prevalent failure mode of the IBA Synthera® automated radiochemistry system. Rebuilding or replacing the pump causes equipment downtime and increases the radiation exposure of the service personnel. We built a dedicated test device to assess new or rebuilt pumps prior to installation, thus reducing downtime and radiation exposure during repairs. The Testbed incorporates a microprocessor that actuates the pump, valves, and pressure sensor; communicates with the user through lights, buttons, and an alphanumeric screen; and outputs test results to a laptop. The Testbed increases productivity and safety in the radiochemistry laboratory. Full article
Show Figures

Figure 1

Back to TopTop