Theoretical Applicability of Different Occluder Systems for Entry Closure in Type B Aortic Dissection: An Image-Morphological Study
Abstract
1. Introduction
2. Materials and Methods
2.1. Image Analysis
2.2. Devices
2.3. Statistical Analysis
3. Results
3.1. ASD-Occluder from GORE®
3.2. Septal-Occluder from GORE®
3.3. Amplatzer™-Occluder from Abbott
4. Discussion
5. Limitations
6. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
Abbreviations
TBAD | Type B Aortic Dissection |
dSINE | Distal Stent Graft-Induced New Entry |
CA | Celiac Artery |
LSA | Left Subclavian Artery |
References
- Lombardi, J.V.; Hughes, G.C.; Appoo, J.J.; Bavaria, J.E.; Beck, A.W.; Cambria, R.P.; Charlton-Ouw, K.; Eslami, M.H.; Kim, K.M.; Leshnower, B.G.; et al. Society for Vascular Surgery (SVS) and Society of Thoracic Surgeons (STS) reporting standards for type B aortic dissections. J. Vasc. Surg. 2020, 71, 723–747. [Google Scholar] [CrossRef] [PubMed]
- Kotelis, D.; Bischoff, M.S.; Jobst, B.; von Tengg-Kobligk, H.; Hinz, U.; Geisbüsch, P.; Böckler, D. Morphological risk factors of stroke during thoracic endovascular aortic repair. Langenbeck’s Arch. Surg. 2012, 397, 1267–1273. [Google Scholar] [CrossRef] [PubMed]
- Oberhuber, A.; Raddatz, A.; Betge, S.; Ploenes, C.; Ito, W.; Janosi, R.A.; Ott, C.; Langheim, E.; Czerny, M.; Puls, R.; et al. Interdisciplinary German clinical practice guidelines on the management of type B aortic dissection. Gefässchirurgie 2023, 28 (Suppl. 1), 1–28. [Google Scholar] [CrossRef]
- Argyriou, C.; Spiliopoulos, S.; Katsanos, K.; Papatheodorou, N.; Lazarides, M.K.; Georgiadis, G.S. Safety and Efficacy of Intentional Celiac Artery Coverage in Endovascular Management of Thoracoabdominal Aortic Diseases: A Systematic Review and Meta-analysis. J. Endovasc. Ther. 2022, 29, 646–658. [Google Scholar] [CrossRef] [PubMed]
- Goremedical ASD-Occluder. Available online: https://www.goremedical.com/en-emea/products/cardioform-asd-occluder (accessed on 22 September 2025).
- Goremedical Septal-Occluder. Available online: https://www.goremedical.com/en-emea/products/cardioform-septal-occluder (accessed on 22 September 2024).
- Abbott. Available online: https://www.cardiovascular.abbott/us/en/hcp/products/structural-heart/structural-interventions/amplatzer-septal.html (accessed on 17 September 2025).
- Springermedizin. 2022. Available online: https://www.springermedizin.de/emedpedia/operative-und-interventionelle-gefaessmedizin/aneurysmen-der-aorta-descendens-und-der-thorakoabdominellen-aorta-endovaskulaere-therapie-und-hybridverfahren?epediaDoi=10.1007%2F978-3-662-45856-3_68 (accessed on 24 July 2025).
- USZ. 2023. Available online: https://www.usz.ch/fachbereich/kardiologie/angebot/verschluss-des-offenen-foramen-ovale-pfo-verschluss-verschluss-des-vorhofseptumdefekts/ (accessed on 24 July 2025).
- Heart Center. Brandenburg. 2023. Available online: https://herzzentrum.immanuel.de/herzzentrum-brandenburg-bei-berlin-leistungen/therapiemoeglichkeiten/korrektur-von-vorhofscheidewanddefekten/ (accessed on 24 July 2025).
- Erasmi, A.; Sievers, H.-H.; Scharfschwerdt, M.; Eckel, T.; Misfeld, M. In Vitro hydrodynamics, cusp bending deformation and root distensibility for different types of aortic valve sparing operations: Remodeling, sinus prosthesis and reimplantation. J. Thorac. Cardiovasc. Surg. 2005, 130, 1044–1049. [Google Scholar] [CrossRef] [PubMed][Green Version]
- Berdajs, D.; Mosbahi, S.; Strano, F.; Forro, Z.; Burki, M.; von Segesser, L.K. Impact of synthetic elements on aortic root haemodynamics: Computed fluid dynamics of root reconstruction and valve reimplantation. Eur. J. Cardiothorac. Surg. 2017, 51, 432–441. [Google Scholar] [CrossRef] [PubMed][Green Version]
- Morita, S.; Asou, T.; Kuboyama, I.; Harasawa, Y.; Sunagawa, K.; Yasui, H. Inelastic vascular prosthesis for proximal aorta increases pulsatile arterial load and causes left ventricular hypertrophy in dogs. J. Thorac. Cardiovasc. Surg. 2002, 124, 768–774. [Google Scholar] [CrossRef] [PubMed]
- Kolh, P.; D’Orio, V.; Lambermont, B.; Gerard, P.; Gommes, C.; Limet, R. Increased aortic compliance maintains left ventricular performance at lower energetic cost. Eur. J. Cardiothorac. Surg. 2000, 17, 272–278. [Google Scholar] [CrossRef] [PubMed][Green Version]
- Scharfschwerdt, M.; Sievers, H.-H.; Greggersen, J.; Hanke, T.; Misfeld, M. Prosthetic replacement of the ascending aorta increases wall tension in the residual aorta. Ann. Thorac. Surg. 2007, 83, 954–957. [Google Scholar] [CrossRef] [PubMed]
- Yeh, C.-H.; Chen, M.-C.; Wu, Y.-C.; Wang, Y.-C.; Chu, J.-J.; Lin, P.J. Risk factors for descending aortic aneurysm formation in medium-term follow-up of patients with type A aortic dissection. Chest 2003, 124, 989–995. [Google Scholar] [CrossRef] [PubMed]
- Geirsson, A.; Bavaria, J.E.; Swarr, D.; Keane, M.G.; Woo, Y.J.; Szeto, W.Y.; Pochettino, A. Fate of the residual distal and proximal aorta after acute type A dissection repair using a contemporary surgical reconstruction algorithm. Ann. Thorac. Surg. 2007, 84, 1955–1964. [Google Scholar] [CrossRef] [PubMed]
- Yuan, X.; Kan, X.; Dong, Z.; Xu, X.Y.; Nienaber, C.A. Nonsurgical Repair of the Ascending Aorta: Why Less Is More. J. Clin. Med. 2023, 12, 4771. [Google Scholar] [CrossRef] [PubMed] [PubMed Central]
- Sandström, C.; Roos, H.; Henrikson, O.; Fagman, E.; A Johnsson, Å.; Jeppsson, A.; Falkenberg, M. Endovascular plugs to occlude proximal entries in chronic aortic dissection. Interact. Cardiovasc. Thorac. Surg. 2022, 35, ivac201. [Google Scholar] [CrossRef] [PubMed]
ASD-Occluder from GORE® | Septal-Occluder from GORE® | Amplatzer™ from Abbott | |
---|---|---|---|
Recommended treatable defect size (mm) | 8–35 | 5–17 | 4–56 |
Available sizes (mm) | 1–5 | 1–3 | 1–27 |
Material | Platinum-filled nitinol wire framework, PTFE | Platinum-filled nitinol wire framework, PTFE | Nitinol braid and PTFE |
Introducer Sheath (Fr) | 10–14 | 10 | 6–13 |
Total (n = 102) | TBAD (n = 87) | dSINE (n = 15) | p-Value | |
---|---|---|---|---|
Demographic aspects | ||||
Age ± standard deviation | 63.8 ± 13.5 | 63.5 ± 13.7 | 65.5 ± 12.1 | 0.511 |
Men (n, %) | 84 (82.4%) | 72 (82.7%) | 12 (80%) | 0.531 |
Classification in terms of acuity (n, %) | ||||
Acute | 63 (61.8%) | 63 (72.4%) | 0.411 | |
Subacute | 12 (11.8%) | 12 (13.8%) | 0.553 | |
Chronic | 27 (26.4%) | 12 (13.8%) | 15 (100%) | 0.421 |
Pre-existing conditions, n (%) | ||||
Arterial hypertension | 95 (93.1%) | 81 (93.1%) | 14 (93.3%) | 0.605 |
Diabetes mellitus | 12 (11.7%) | 81 (93.1%) | 2 (13.3%) | 0.481 |
Coronary artery disease | 16 (15.6%) | 15 (17.2%) | 1 (6.6%) | 0.437 |
Stroke | 14 (13.7%) | 13 (14.9%) | 1 (6.6%) | 0.701 |
Marfan syndrome | 6 (5.8%) | 6 (6.8%) | 0 | 0.371 |
Nicotine consumption | 13 (12.7%) | 11 (12.6%) | 2 (13.3%) | 0.511 |
Chronic kidney disease | 15 (14.7%) | 14 (16%) | 3 (20%) | 0.478 |
Total (n = 102) | TBAD (n = 87) | dSINE (n = 15) | p-Value | |
---|---|---|---|---|
Applicability (n, %) | ||||
Yes | 83 (81.4%) | 72 (82.8%) | 11 (73.3%) | 0.293 |
No | 19 (18.6%) | 15 (17.2%) | 4 (26.7%) | |
Various ASD-Occluder sizes 8–35 mm, (n, %) | ||||
Nr. 1 (8–15 mm) | 49 (48%) | 40 (46%) | 9 (60%) | |
Nr. 2 (13–20 mm) | 65 (63.7%) | 54 (62%) | 11 (73.3%) | |
Nr. 3 (18–25 mm) | 72 (70.5%) | 61 (70.1%) | ||
Nr. 4 (23–30 mm) | 78 (76.4%) | 67 (77%) | ||
Nr. 5 (28–35 mm) | 83 (81.4%) | 72 (82.8%) | ||
Vascular coverage (n, %) | LSA | CA | ||
Partial | 16 (15.7%) | 5 (5.7%) | 11 (73.3%) | 0.288 |
Complete | 9 (8.8%) | 9 (10.3%) | 0 |
Total (n = 102) | TBAD (n = 87) | dSINE (n = 15) | p-Value | |
---|---|---|---|---|
Applicability (n, %) | ||||
Yes | 59 (57.8%) | 50 (57.5%) | 9 (60%) | 0.544 |
No | 43 (42.1%) | 37 (42.5%) | 6 (40%) | |
Various Septal-Occluder sizes (5–17 mm) (n, %) | ||||
Nr. 1 (5–11 mm) | 23 (22.5%) | 20 (23%) | 3 (20%) | |
Nr. 2 (11–14 mm) | 38 (37.2%) | 33 (38%) | 5 (33.3%) | |
Nr. 3 (14–17 mm) | 59 (57.8%) | 50 (57.5%) | 9 (60%) | |
Vascular coverage (n, %) | LSA | CA | ||
Partial | 11 (10.7%) | 2 (2.3%) | 9 (60%) | 0.506 |
Complete | 4 (3.9%) | 4 (4.6%) |
Total (n = 102) | TBAD (n = 87) | dSINE (n = 15) | p-Value | |
---|---|---|---|---|
Applicability (n, %) | ||||
Yes | 92 (90.1%) | 78 (89.7%) | 14 (93.3%) | 0.549 |
No | 10 (9.8%) | 9 (10.3%) | 1 (6.7%) | |
Vascular coverage (n, %) | LSA | CA | ||
Partial | 24 (23.5%) | 12 (13.8%) | 12 (80%) | 0.201 |
Complete | 23 (22.5%) | 23 (26.4%) | 0 |
Type B Aortic Dissection (n = 87) | ||||
---|---|---|---|---|
ASD-Occluder from GORE® | Septal-Occluder from GORE® | Amplatzer™ from Abbott | p-Value | |
Use without LSA/CA coverage | 58/87 (66.6%) | 44/87 (50.5%) | 43/87 (49.4%) | <0.001 |
Use with possible partial LSA/CA coverage | 5/87 (5.7%) | 2/87 (2.3%) | 12/87 (13.8%) | 0.002 |
Use with possible LSA/CA complete coverage | 9/87 (10.3%) | 4/87 (4.6%) | 23/87 (26.4%) | 0.007 |
Technically not possible | 15/87 (17.2%) | 37/87 (42.5%) | 9/87 (10.3%) | 0.001 |
TBAD | dSINE | p-Value | |
---|---|---|---|
Entry length in mm (mean value ± standard deviation) | 15.4 ± 13.2 | 8.9 ± 5 | 0.02 |
Entry width in mm (mean value ± standard deviation) | 15 ± 11.1 | 9.9 ± 7 | 0.071 |
Maximum diameter in mm (mean value ± standard deviation) | 20.03 ± 13.7 | 12.5 ± 7.8 | 0.197 |
dSINE (n = 15) | ||||
---|---|---|---|---|
ASD-Occluder from GORE® | Septal-Occluder from GORE® | Amplatzer™ from Abbott | p-Value | |
Use without LSA/CA coverage | 0/15 (0%) | 0/15 (0%) | 2/15 (13.3%) | 0.40 |
Use with possible partial LSA/CA coverage | 11/15 (73.3%) | 9/15 (60%) | 12/15 (80%) | 0.368 |
Technically not possible | 4/15 (26.7%) | 6/15 (40%) | 1/15 (6.7%) | 0.042 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2025 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Yordanov, M.; Oberhuber, A.; Schäfers, J.F.; Anzz, R.; Ibrahim, A. Theoretical Applicability of Different Occluder Systems for Entry Closure in Type B Aortic Dissection: An Image-Morphological Study. Biomedicines 2025, 13, 2338. https://doi.org/10.3390/biomedicines13102338
Yordanov M, Oberhuber A, Schäfers JF, Anzz R, Ibrahim A. Theoretical Applicability of Different Occluder Systems for Entry Closure in Type B Aortic Dissection: An Image-Morphological Study. Biomedicines. 2025; 13(10):2338. https://doi.org/10.3390/biomedicines13102338
Chicago/Turabian StyleYordanov, Miroslav, Alexander Oberhuber, Johannes Frederik Schäfers, Raman Anzz, and Abdulhakim Ibrahim. 2025. "Theoretical Applicability of Different Occluder Systems for Entry Closure in Type B Aortic Dissection: An Image-Morphological Study" Biomedicines 13, no. 10: 2338. https://doi.org/10.3390/biomedicines13102338
APA StyleYordanov, M., Oberhuber, A., Schäfers, J. F., Anzz, R., & Ibrahim, A. (2025). Theoretical Applicability of Different Occluder Systems for Entry Closure in Type B Aortic Dissection: An Image-Morphological Study. Biomedicines, 13(10), 2338. https://doi.org/10.3390/biomedicines13102338