Sensors 2006, 6(11), 1555-1567; doi:10.3390/s6111555
Article
Design of a Flow-through Polarographic Sensor Based on Metal Films for Determining N-nitrosodiethanolamine Levels in Rabbit Biological Fluids
Department of Applied Chemistry, Chia Nan University of Pharmacy and Science, 60 Erh-Jen Road, Section 1, Jen Te, Tainan 71743, Taiwan
* Author to whom correspondence should be addressed.
Received: 25 February 2006; in revised form: 5 October 2006 / Accepted: 10 November 2006 / Published: 11 November 2006
(This article belongs to the Special Issue Sensors in Flow Analysis)
Abstract: The construction and characterization of a flow-through polarographic detectorfor catalyzing the electroreduction of N-nitrosodiethanolamine (NDELA), is discussed. Theflow-through cell is equipped with a gold wire electrode (a thin mercury film deposited on agold substance). The response is evaluated with respect to substance diameter, length,concentration of modifying film, operating potential, supporting electrolyte and pH, andflow rate in the DC mode. The system allows the determination of N-nitrosodiethanolaminein rabbit biological fluids with relatively inexpensive equipment.
Keywords: Flow-through polarographic detector; thin-film modified metal electrode; Nnitrosodiethanolamine; rabbit biological fluids.
Article Statistics
Click here to load and display the download statistics.Cite This Article
MDPI and ACS Style
Wang, L.-H.; Hsia, H.-C.; Lan, Y.-Z. Design of a Flow-through Polarographic Sensor Based on Metal Films for Determining N-nitrosodiethanolamine Levels in Rabbit Biological Fluids. Sensors 2006, 6, 1555-1567.
AMA StyleWang L-H, Hsia H-C, Lan Y-Z. Design of a Flow-through Polarographic Sensor Based on Metal Films for Determining N-nitrosodiethanolamine Levels in Rabbit Biological Fluids. Sensors. 2006; 6(11):1555-1567.
Chicago/Turabian StyleWang, Lai-Hao; Hsia, Hung-Chang; Lan, Yuan-Zhi. 2006. "Design of a Flow-through Polarographic Sensor Based on Metal Films for Determining N-nitrosodiethanolamine Levels in Rabbit Biological Fluids." Sensors 6, no. 11: 1555-1567.