Flow-through Bulk Optode for Spectrophotometric Determination of Thiocyanate and Its Application to Water and Saliva Analysis
Department of Analytical Chemistry, Faculty of Chemistry, University of Murcia, 30071 Murcia, Spain
* Author to whom correspondence should be addressed.
Received: 23 June 2006 / Accepted: 3 October 2006 / Published: 9 October 2006
Abstract: A flow-through spectrophotometric bulk optode for the flow-injectiondetermination of thiocyanate is described. As active constituents, the optode incorporatesthe lipophilized pH indicator 5-octadecanoyloxy-2-(4-nitrophenylazo)phenol andmethyltridodecyl ammonium chloride, dissolved in a plasticized poly(vinyl)chloridemembrane entrapped in a cellulose support. The optode is applied, in conjunction with theflow injection technique, to the determination of thiocyanate at pH 7.5 (TRIS/H2SO4). Thesensor is readily regenerated with a 10-2 M NaOH carrier solution. The analyticalcharacteristics of this optode with respect to thiocyanate response time, dynamicmeasurement range, reproducibility and selectivity are discussed. The proposed FI methodis applied to the determination of thiocyanate in waters from different sources and in humansaliva samples in order to distinguish between smokers and non-smokers.
Keywords: Flow-through optode; spectrophotometry; thiocyanate; water; saliva.
Article Statistics
Click here to load and display the download statistics.
Notes: Multiple requests from the same IP address are counted as one view.
Cite This Article
MDPI and ACS Style
García, S., Mª; Ortuño, J.A.; Sánchez-Pedreño, C.; Albero, I., Mª; Fernández, J., Mª Flow-through Bulk Optode for Spectrophotometric Determination of Thiocyanate and Its Application to Water and Saliva Analysis. Sensors 2006, 6, 1224-1233.
AMA Style
García S, Mª, Ortuño JA, Sánchez-Pedreño C, Albero I, Mª, Fernández J, Mª. Flow-through Bulk Optode for Spectrophotometric Determination of Thiocyanate and Its Application to Water and Saliva Analysis. Sensors. 2006; 6(10):1224-1233.
Chicago/Turabian Style
García, Soledad, Mª; Ortuño, Joaquín A.; Sánchez-Pedreño, Concepción; Albero, Isabel, Mª; Fernández, José, Mª. 2006. "Flow-through Bulk Optode for Spectrophotometric Determination of Thiocyanate and Its Application to Water and Saliva Analysis." Sensors 6, no. 10: 1224-1233.