- freely available
- re-usable
Sensors 2013, 13(3), 3242-3251; doi:10.3390/s130303242
Article
Selective Serial Multi-Antibody Biosensing with TOPAS Microstructured Polymer Optical Fibers
1
DTU Fotonik, Department of Photonics Engineering, Technical University of Denmark, Ørsteds Plads, 2800 Kgs. Lyngby, Denmark
2
DAKO Denmark A/S, Produktionsvej 42, DK-2600 Glostrup, Denmark
3
Bioneer A/S, Kogle Allé 2, 2970 Hørsholm, Denmark
†
Present address: PCMA Rus, Kaluga Region, Russia.
* Author to whom correspondence should be addressed.
Received: 17 January 2013; in revised form: 22 February 2013 / Accepted: 6 March 2013 / Published: 8 March 2013
(This article belongs to the Special Issue Photonic Crystal Sensors)
Abstract: We have developed a fluorescence-based fiber-optical biosensor, which can selectively detect different antibodies in serial at preselected positions inside a single piece of fiber. The fiber is a microstructured polymer optical fiber fabricated from TOPAS cyclic olefin copolymer, which allows for UV activation of localized sensor layers inside the holes of the fiber. Serial fluorescence-based selective sensing of Cy3-labelled α-streptavidin and Cy5-labelled α-CRP antibodies is demonstrated.
Keywords: microstructured polymer optical fiber; fluorescence; antibodies
Article Statistics
Click here to load and display the download statistics.Cite This Article
MDPI and ACS Style
Emiliyanov, G.; Høiby, P.E.; Pedersen, L.H.; Bang, O. Selective Serial Multi-Antibody Biosensing with TOPAS Microstructured Polymer Optical Fibers. Sensors 2013, 13, 3242-3251.
AMA StyleEmiliyanov G, Høiby PE, Pedersen LH, Bang O. Selective Serial Multi-Antibody Biosensing with TOPAS Microstructured Polymer Optical Fibers. Sensors. 2013; 13(3):3242-3251.
Chicago/Turabian StyleEmiliyanov, Grigoriy; Høiby, Poul E.; Pedersen, Lars H.; Bang, Ole. 2013. "Selective Serial Multi-Antibody Biosensing with TOPAS Microstructured Polymer Optical Fibers." Sensors 13, no. 3: 3242-3251.