Sensors 2012, 12(1), 1002-1013; doi:10.3390/s120101002
Article

FBG Sensor for Contact Level Monitoring and Prediction of Perforation in Cardiac Ablation

1,* email, 2email, 2email and 1email
Received: 25 November 2011; in revised form: 3 January 2012 / Accepted: 14 January 2012 / Published: 17 January 2012
(This article belongs to the Special Issue Optical Fiber Sensors)
This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Abstract: Atrial fibrillation (AF) is the most common type of arrhythmia, and is characterized by a disordered contractile activity of the atria (top chambers of the heart). A popular treatment for AF is radiofrequency (RF) ablation. In about 2.4% of cardiac RF ablation procedures, the catheter is accidently pushed through the heart wall due to the application of excessive force. Despite the various capabilities of currently available technology, there has yet to be any data establishing how cardiac perforation can be reliably predicted. Thus, two new FBG based sensor prototypes were developed to monitor contact levels and predict perforation. Two live sheep were utilized during the study. It was observed during operation that peaks appeared in rhythm with the heart rate whenever firm contact was made between the sensor and the endocardial wall. The magnitude of these peaks varied with pressure applied by the operator. Lastly, transmural perforation of the left atrial wall was characterized by a visible loading phase and a rapid signal drop-off correlating to perforation. A possible pre-perforation signal was observed for the epoxy-based sensor in the form of a slight signal reversal (12–26% of loading phase magnitude) prior to perforation (occurring over 8 s).
Keywords: cardiac ablation; radiofrequency ablation; ablation catheter; transmural perforation; contact monitoring
PDF Full-text Download PDF Full-Text [696 KB, uploaded 21 June 2014 04:15 CEST]

Export to BibTeX |
EndNote


MDPI and ACS Style

Ho, S.C.M.; Razavi, M.; Nazeri, A.; Song, G. FBG Sensor for Contact Level Monitoring and Prediction of Perforation in Cardiac Ablation. Sensors 2012, 12, 1002-1013.

AMA Style

Ho SCM, Razavi M, Nazeri A, Song G. FBG Sensor for Contact Level Monitoring and Prediction of Perforation in Cardiac Ablation. Sensors. 2012; 12(1):1002-1013.

Chicago/Turabian Style

Ho, Siu Chun Michael; Razavi, Mehdi; Nazeri, Alireza; Song, Gangbing. 2012. "FBG Sensor for Contact Level Monitoring and Prediction of Perforation in Cardiac Ablation." Sensors 12, no. 1: 1002-1013.


Sensors EISSN 1424-8220 Published by MDPI AG, Basel, Switzerland RSS E-Mail Table of Contents Alert