Line-Monitoring, Hyperspectral Fluorescence Setup for Simultaneous Multi-Analyte Biosensing
1
Laboratory of Optical Imaging and Sensing, Graduate School at Shenzhen, Tsinghua University, Shenzhen 518055, China
2
MOE Key laboratory of Laser Life Science & Institute of Laser Life Science, South China Normal University, Guangzhou 510631, China
* Author to whom correspondence should be addressed.
Received: 5 September 2011; in revised form: 5 October 2011 / Accepted: 17 October 2011 / Published: 25 October 2011
(This article belongs to the Special Issue Biochips)
Abstract: Conventional fluorescence scanners utilize multiple filters to distinguish different fluorescent labels, and problems arise because of this filter-based mechanism. In this work we propose a line-monitoring, hyperspectral fluorescence technique which is designed and optimized for applications in multi-channel microfluidic systems. In contrast to the filter-based mechanism, which only records fluorescent intensities, the hyperspectral technique records the full spectrum for every point on the sample plane. Multivariate data exploitation is then applied to spectra analysis to determine ratios of different fluorescent labels and eliminate unwanted artifacts. This sensor is designed to monitor multiple fluidic channels simultaneously, providing the potential for multi-analyte biosensing. The detection sensitivity is approximately 0.81 fluors/μm2, and this sensor is proved to act with a good homogeneity. Finally, a model experiment of detecting short oligonucleotides has demonstrated the biomedical application of this hyperspectral fluorescence biosensor.
Keywords: hyperspectral fluorescence detection; microfluidic system; biosensor
Article Statistics
Click here to load and display the download statistics.Cite This Article
MDPI and ACS Style
Liu, Z.; Shi, H.; Liu, L.; Deng, S.; Ji, Y.; Ma, S.; Ma, H.; He, Y. Line-Monitoring, Hyperspectral Fluorescence Setup for Simultaneous Multi-Analyte Biosensing . Sensors 2011, 11, 10038-10047.
AMA StyleLiu Z, Shi H, Liu L, Deng S, Ji Y, Ma S, Ma H, He Y. Line-Monitoring, Hyperspectral Fluorescence Setup for Simultaneous Multi-Analyte Biosensing . Sensors. 2011; 11(11):10038-10047.
Chicago/Turabian StyleLiu, Zhiyi; Shi, Heng; Liu, Le; Deng, Sunan; Ji, Yanhong; Ma, Suihua; Ma, Hui; He, Yonghong. 2011. "Line-Monitoring, Hyperspectral Fluorescence Setup for Simultaneous Multi-Analyte Biosensing ." Sensors 11, no. 11: 10038-10047.