Droplet Velocity Measurement Based on Dielectric Layer Thickness Variation Using Digital Microfluidic Devices

,
and
Department of Electrical and Electronic Engineering, Universiti Teknologi PETRONAS, Seri Iskander 32610, Malaysia
*
Author to whom correspondence should be addressed.
This article belongs to the Special Issue Micro and Nanoscale Biosensors

Abstract

In recent years, the number of interdisciplinary research works related to the development of miniaturized systems with integrated chemical and biological analyses is increasing. Digital microfluidic biochips (DMFBs) are one kind of miniaturized systems designed for conducting inexpensive, fast, convenient and reliable biochemical assay procedures focusing on basic scientific research and medical diagnostics. The role of a dielectric layer in the digital microfluidic biochips is prominent as it helps in actuating microliter droplets based on the electrowetting-on-dielectric (EWOD) technique. The advantages of using three different material layers of dielectric such as parafilm, polytetrafluoroethylene (PTFE) and ethylene tetrafluoroethylene (ETFE) were reported in the current work. A simple fabrication process of a digital microfluidic device was performed and good results were obtained. The threshold of the actuation voltage was determined for all dielectric materials of varying thicknesses. Additionally, the OpenDrop device was tested by utilizing a single-plate system to transport microliter droplets for a bioassay operation. With the newly proposed fabrication methods, these dielectric materials showed changes in contact angle and droplet velocity when the actuation voltage was applied. The threshold actuation voltage for the dielectric layers of 10–13 μm was 190 V for the open plate DMFBs.

1. Introduction

Microfluidic technology has been introduced as a part of lab-on-a-chip (LOCs) devices, in which several laboratory functions are integrated on a single chip [1,2]. Microfluidic technology has the ability to manipulate small volumes of samples or reagents from µL to pL range [3,4]. As microelectromechanical systems (MEMS) are scaled down, microfluidic technology becomes prominent in various medical diagnostics processes [3,5]. Microfluidic technology has the potential of replacing macroscale machines like biochemical analyzers in medical applications [6,7]. Microfluidic technology can be approached in two ways, i.e., continuous or digital. The physical structure of continuous flow-based microfluidic is less complex, as the liquid flow through channel and an additional hardware of micro valves, micro pumps are required to manipulate the liquid [6]. Unlike continuous flow microfluidics, the development of digital microfluidic biochips (DMFBs) is very advantageous in term of its feasibility, where only discrete droplets are manipulated independently to realize various liquid operations [8].
There are many applications of DMFBs in biochemistry, biotechnology and bioassays due to their small features. For example, in bioassays, clinical diagnostic analyses for glucose assays, blood sample preparation, and drug synthesis can be conducted [6,9], thanks to the flexibility, programmability, and reduced wastage of sample in digital microfluidic biochips [10,11]. Various liquid operations such as detection, actuation, merging, and separation of droplets can be performed on a DMFB [2,12]. A digital microfluidic biochip consists of a movable droplet on top of the electrode surface which is known as electrowetting [13]. The droplet behavior is determined by interfacial tensions as three-phase forces of liquid, solid and gas is pulled towards each contact line as denoted by Young’s equation:
c o s θ Y = σ s g − σ s l σ l g ,
where θ Y is the Young’s contact angle, σ s g is interfacial tension between solid–gas, σ s l is interfacial tension between solid–liquid, and σ l g is interfacial tension between liquid–gas [14,15]. The existence of an applied voltage to the electrode surface builds up an electric double layer between the dielectric and droplet interface, which will decrease the droplet contact angle [16]. However, the thickness of the dielectric layer, the dielectric material and the limit of applied voltage on the electrodes were major concerns for researchers working in the fabrication of DMFB devices over the past few years [17].
For EWOD, dielectric polarization takes place when voltage is applied at the bottom of the electrode. The effect of dielectric polarization creates an internal electric field which induces a charge accumulation on the droplet surface. Hence, the distribution of accumulating charges will create an electrostatic force acting on the droplet, which leading the droplet toward the active electrode [16]. The electrostatic force acting on the droplet is obtained by integrating the variables as follows:
F d = ∫ S 12 + S 13 T · n   d S ,
where F d is the electrostatic force, S 12 and S 13 are the subscripts of the droplet, surrounding fluid and dielectric layer, respectively, whereas, T is the Maxwell stress tensor. The forces due to the free electric charges and the forces due to the polarization of the material is known as Maxwell stress tensor [16]. An electrostatic force distribution increases due to the charge distribution near the contact line of three-phase forces.
However, to facilitate the movement of the droplet, a suitable thickness of the dielectric layer is necessary [18,19]. The electrode size also plays an important role in actuating the droplet because the top plate is absent in an open system. This research focused on determining the required applied voltage for the actuation of droplet with different dielectric materials. The dielectric thickness is the main key for droplet actuation because thinner layer of dielectric provides faster droplet actuation [8,20]. Nevertheless, a thin dielectric layer is susceptible to dielectric breakdown due to high voltage [2,18]. Thus, the required applied voltage for droplet actuation is dependent on the thickness of dielectric layer.
This paper presented the fundamental and experimental results of minimum applied voltage required to actuate the droplet using three dielectric materials such as parafilm, PTFE and ETFE. The material thickness would affect the droplet actuation. The velocity of the actuated droplet was measured experimentally. The paper was organized as follows. The related works detailing the dielectric materials used for the fabrication of digital microfluidic biochips (DMFBs) were reviewed in Section 2. The new methodology for fabricating the DMFB and the required applied voltage as a function of dielectric layer thickness were highlighted in Section 3. Then, the experimental results of contact angle and droplet velocity for all types of dielectric layers were reported in Section 4. Finally, the work was concluded and some future works were recommended in Section 5.

3. Our Approach

In this section, a simple method used to measure the contact angle of tested droplet is given. There are several methods available in coating the dielectric and hydrophobic layers, as explained in the upcoming section. The wetting phenomenon of tested droplet can be observed by measuring the contact angle of stationary droplet for the EWOD technique given by [35]:
θ 2 = t a n − 1 ( h / r )
where θ represents the contact angle, h is the height of apex, and r represents the droplet radius [36]. Generally, the contact angle can be measured in either side view or top view. Based on this equation, the side view method is more approachable as shown in Figure 3 as the top view method does not work on hydrophobic surfaces [35].
Figure 3. (a) Side view contact angle measurement. Change in contact angle: (b) first prototype and (c) second prototype [14]. Reproduced with permission from Siti Noor Idora Syafinaz Zulkepli, Nor Hisham Hamid and Vineeta Shukla, A low cost open droplet-based microfluidic devices on printed circuit board; published by IEEE, 2017.
A 2.5 μL of dionized (DI) water was placed on the electrode. PTFE and cooking oil layers were used as dielectric and hydrophobic layers as reported in [14]. A change in contact angle from 80.59° to 69.36° is observed from the side view of digital microscope as shown in Figure 3b for first prototype of DMFB as reported in [14], while Figure 3c shows a change in contact angle from 75.42° to 73° for second prototype of DMFB as reported in [14]. Therefore, the wetting behavior can be characterized via contact angle calculation.
Various techniques have been applied to develop DMFB on different platforms such as printed circuit board, inkjet printing, laser printed polyester film and photosensitized PCB [36]. For the inkjet printing technique, the device is printed onto paper, which leads to chemical reagent smearing. Therefore, this technique is limited only to certain types of chemical reagent. For laser printed polyester film, the microfluidic channel is formed by printing the device on a transparent film which is limited to continuous flow microfluidics. Meanwhile, plastic-based substrate is used in photosensitized PCB technique; unfortunately, the plastic-based substrate is not easily available. In this work, we have used printed circuit board (PCB) to develop our DMFB device as PCB substrate is easily available and inexpensive [37].
In our previous work, we have fabricated a low cost open EWOD DMFB device on a standard FR4 printed circuit board (PCB). An array of 2 × 4 copper-based electrodes (3 × 3 mm2) has been fabricated with a separation distance (between electrodes) of 203.2 µm. Polyethylene film and cooking oil have been used as the materials for dielectric and hydrophobic layers, respectively. A 2.5 µL of DI water was tested in air by using the EWOD technique. Then, the contact angle was measured to observe the phenomenon of EWOD. In this work, we have used the OpenDrop device provided by Gaudi Labs [3]. This DMFB device is shown in Figure 4a.
Figure 4. (a) OpenDrop device [3]; (b) 2 droplets of 2.5 μL size has been tested to observe the droplet movement.
In the current work, the EWOD DMFB was operated in air by using the OpenDrop device [3] to observe the threshold actuation voltage that can be applied on various dielectric materials. The electrodes were gold-coated and the DC to DC voltage regulator was integrated to actuate the droplet from 190 V to 330 V. As mentioned above, various methods of dielectric layer coatings are available in literature. However, the access to a clean room is quite challenging for certain researchers due to the lack of equipment and facilities [3,23]. Therefore, in order to overcome this problem, a simple and straightforward approach of dielectric layer coating is opted for based on the availability of the dielectric material. In this work, we have used three dielectric materials. The values of dielectric thickness (measured using digital vernier caliper) after performing hydrophobic coating are shown in Table 2. Additionally, the applied voltage required to actuate the droplet have been determined.
Table 2. Type of dielectric material used. ETFE: ethylene tetrafluoroethylene.
Regarding the coating of materials such as parafilm, PTFE film and ETFE film considered in the current work, these materials must be properly stretched so that the film is tensed evenly. These films were stretched on a clean glass, and then cut so that they can fit nicely on the top of electrodes. In order to prevent film misplacement, a frame was developed for our coating process. A kapton tape was used to hold the film firmly at the back of the frame as shown in Figure 5.
Figure 5. Preparing the dielectric layer coating: (a) frame without the dielectric coating; (b,c) kapton tape is used to stick the dielectric material; (d) frame with the dielectric layer.
The frame was then cleaned in the mixture of distilled water and isopropanol liquid using ultrasonic bath. After 15 min, the frame was rinsed with pure water and dried on the hot plate heated at around 80 °C.
In the dielectric film-coated frame, about 150 μL to 200 μL of fluoropel liquid was spun for 30 s at 3000 rpm. Then, the frame was dried on the hot plate heated from 80 °C to 150 °C for 20 min. This spin coating process is only necessary for fluoropel liquid. In the next phase, the hydrophobic layer was coated on top of the dielectric layer where silicon oil and cooking oil were used to prevent trapping of air bubbles inside the gap. Fluoropel liquid was used to reduce the surface energy between the droplet and the dielectric layer. These low-cost materials were attractive as they can be coated without accessing the clean room facilities. Unlike silicon oil and cooking oil, the spin coating process is necessary for coating the fluoropel liquid, as proper coating is needed.
The dielectric and hydrophobic layers were coated in order to actuate the stationary droplet for bioassay operation. A droplet can move easily to the adjacent electrodes when fluoropel is used as hydrophobic layer (instead of using silicon oil and cooking oil). The OpenDrop device was tested to observe the required applied voltage for the dielectric materials used such as parafilm, PTFE film and ETFE film as shown in Figure 4b.

4. Results and Discussions

Measurement of Droplet Velocity

The droplet velocity was calculated via
V = d A E t A E
where d A E is the distance to reach the adjacent electrode, and t A E is the time taken to reach the adjacent electrode. For the applied voltage range of 190 V to 330 V, the velocity of droplet travelling across the 100 µm gap (between the electrodes) was calculated based on the captured video, as displayed graphically in Figure 6.
Figure 6. (a–e) The droplet is moving from one electrode array to the adjacent electrode array; (f) the droplet is merged together based on the captured video.
In the experiments where parafilm was used as the dielectric layer material, the droplet was stagnant (dielectric layer thickness of 120 µm). However, when the applied voltage was increased from 190 V to 330 V, the droplet was able to move in the cases of PTFE film and ETFE film as shown in Table 2. As previously discussed, droplet motion could be started only when a suitable dielectric layer thickness is chosen and a proper minimum voltage is applied. By varying the thickness values of parafilm, PTFE film and ETFE film, both droplet velocity and applied voltage increase as shown in Table 3 and Figure 7.
Table 3. Velocity measurement for various voltage values.
Figure 7. Experimental result for the droplet velocity measurement based on different dielectric materials.
For PTFE film, an average droplet velocity of 0.8 mm/s was observed when the voltage of 190 V was applied. Meanwhile, the droplet velocity of 1.23 mm/s was observed when the ETFE film was used. The droplet velocity using ETFE film was higher than that using PTFE film even though the thickness of PTFE film was less than 13 μm. This could be due to the fact that the fluoropel liquid used for hydrophobic coating of ETFE film was more durable as the frame was baked after spin coating. Besides the dielectric layer thickness, the coatings of dielectric and hydrophobic layers also affect the droplet velocity. While operating the OpenDrop device in air, the droplet speed was limited to 8.7 mm/s. On the other hand, the reported velocity of 80 mm/s was reported by the GaudiLabs team when the OpenDrop device was operated in the oil medium [3].
For this work, the unstretched parafilm was used because a thick parafilm (e.g., 120 μm) is unable to build a sufficiently strong electric field for moving the droplet. However, when the parafilm was stretched to 11 μm as shown in Figure 7, a significant droplet movement was observed. Therefore, these three different dielectric materials can be used to observe the velocity of the droplet. Based on the experimental results of OpenDrop device, the minimum DC voltage that can be applied is 190 V when the device is operated in air. The reliability of DMFB depends on the electrodes and the dielectric material used. Several electrodes became faulty after the DMFB device was operated repeatedly (due to breakdown of dielectric layer). Thus, for reliability purpose, the dielectric layer thickness and the dielectric material type should be chosen carefully. Also, droplet actuation is dependent on electrode size, electrode edges, and gap width between the inter electrode as reported in [14]. As shown in Figure 2, the threshold actuation voltage that can be applied is heavily dependent on the thickness of the dielectric material used.

5. Conclusions

The electrowetting-on-dielectric (EWOD) technique has been widely employed in droplet actuation for microfluidic applications. However, it requires a high operating voltage to realize droplet actuation. In order to alleviate this issue, the optimum thickness values of several types of dielectric layers have been identified. First, the surface wettabilities of cooking oil and PTFE have been observed by measuring the contact angles of DI water droplets on hydrophobic and dielectric layers, respectively. The results have indicated that for both cooking oil and PTFE layers, the contact angle decreased from 80.59 to 69.36° (measured from the side view method of contact angle measurement) as reported in [14]. The average droplet velocities in cases using parafilm, PTFE and ETFE dielectric layers of different thickness values have been investigated by varying the operating voltage. The current experiment has shown that the average droplet velocities in cases using parafilm (11 μm), PTFE (10 μm), and ETFE (13 μm) increased rapidly as the applied voltage increased from 190 V to 330 V. However, for unstretched parafilm of thickness 120 μm, no droplet movement has been observed, as the critical thickness of dielectric layer that causes droplet actuation should be less than 14 μm. Even though the thicknesses of both parafilm and PTFE were less than that of ETFE, the droplet velocity in ETFE was the highest (i.e., 8.7 mm/s) at 330 V operating voltage. This is because both dielectric and hydrophobic coatings play their roles in droplet actuation as well. Based on the experimental results, the dielectric layer of optimum thickness 11 μm with tolerable coating method has been recommended in order to reduce the operating voltage. Finally, the effect of the optimum thickness of the dielectric layer on the operating voltage in EWOD actuation has been tested and confirmed. This study contributes to the understanding of the effect of dielectric layer thickness on the droplet actuation. Also, a simple coating method has been proposed for future EWOD applications.

Author Contributions

S.N.I.S.Z. performed the experiments and collected data; S.N.I.S.Z. and N.H.H. analyzed the data; S.N.I.S.Z. and V.S. wrote the paper.

Funding

This research work was supported by Universiti Teknologi PETRONAS and Ministry of Higher Education (MOHE) under E-science grant No. 0153AB-B65.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Gupta, S.; Ramesh, K.; Ahmed, S.; Kakkar, V. Lab-on-Chip Technology: A Review on Design Trends and Future Scope in Biomedical Applications. Int. J. Bio-Sci. Bio-Technol. 2016, 8, 311–322. [Google Scholar] [CrossRef] [Scilit]
  2. Sheikh, M.A.; Ali, N.B.Z.; Hamid, N.H.; Hussin, F.A.; Shukla, V. On enhancing the reliability of digital microfluidic biochips (DMFB) through electrode cells health classification. In Proceedings of the 2015 6th Asia Symposium on Quality Electronic Design (ASQED), Kula Lumpur, Malaysia, 4–5 August 2015; pp. 186–191. [Google Scholar]
  3. Alistar, M.; Gaudenz, U. OpenDrop: An Integrated Do-It-Yourself Platform for Personal Use of Biochips. Bioengineering 2017, 4, 45. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  4. Jain, V.; Devarasetty, V.; Patrikar, R. Effect of electrode geometry on droplet velocity in open EWOD based device for digital microfluidics applications. J. Electrostat. 2017, 87, 11–18. [Google Scholar]
  5. Basha, I.; Ho, E.; Yousuff, C.; Hamid, N. Towards Multiplex Molecular Diagnosis—A Review of Microfluidic Genomics Technologies. Micromachines 2017, 8, 266. [Google Scholar] [CrossRef] [Scilit]
  6. Samiei, E.; Tabrizian, M.; Hoorfar, M. A review of digital microfluidics as portable platforms for lab-on a-chip applications. Lab Chip 2016, 16, 2376–2396. [Google Scholar] [PubMed]
  7. Yousuff, C.M.; Ho, E.T.W.; Hussain, I.K.; Hamid, N.H.B. Microfluidic Platform for Cell Isolation and Manipulation Based on Cell Properties. Micromachines 2017, 8, 15. [Google Scholar] [CrossRef] [Scilit]
  8. Choi, K.; Ng, A.H.; Fobel, R.; Wheeler, A.R. Digital microfluidics. Annu. Rev. Anal. Chem. 2012, 5, 413–440. [Google Scholar]
  9. Su, W.; Gao, X.; Jiang, L.; Qin, J. Microfluidic platform towards point-of-care diagnostics in infectious diseases. J. Chromatogr. A 2015, 1377, 13–26. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  10. Sukhatme, S.; Agarwal, A. Digital microfluidics: Techniques, their applications and advantages. J. Bioeng. Biomed. Sci. 2012, 8, 2. [Google Scholar] [CrossRef]
  11. Bender, B.F.; Garrell, R.L. Digital microfluidic system with vertical functionality. Micromachines 2015, 6, 1655–1674. [Google Scholar] [CrossRef] [Scilit]
  12. Hu, K.; Hsu, B.-N.; Madison, A.; Chakrabarty, K.; Fair, R. Fault detection, real-time error recovery, and experimental demonstration for digital microfluidic biochips. In Proceedings of the Conference on Design, Automation and Test in Europe, Grenoble, France, 18–22 March 2013; pp. 559–564. [Google Scholar]
  13. Lin, Y.Y.; Evans, R.D.; Welch, E.; Hsu, B.N.; Madison, A.C.; Fair, R.B. Low Voltage Electrowetting-on-Dielectric Platform using Multi-Layer Insulators. Sens. Actuators B Chem. 2010, 150, 465–470. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  14. Zulkepli, S.N.I.S.; Hamid, N.H.; Shukla, V. A low cost open droplet-based microfluidic devices on printed circuit board. In Proceedings of the 2017 7th IEEE International Conference on Control System, Computing and Engineering (ICCSCE), Penang, Malaysia, 24–26 November 2017; pp. 204–209. [Google Scholar]
  15. Jain, V.; Hole, A.; Deshmukh, R.; Patrikar, R. Dynamic capacitive sensing of droplet parameters in a low-cost open EWOD system. Sens. Actuators A Phys. 2017, 263, 224–233. [Google Scholar] [CrossRef] [Scilit]
  16. Mugele, F.; Baret, J.-C. Electrowetting: From basics to applications. J. Phys. Condens. Matter 2005, 17, R705. [Google Scholar]
  17. Lin, Y.-Y.; Welch, E.R.; Fair, R.B. Low voltage picoliter droplet manipulation utilizing electrowetting-on-dielectric platforms. Sens. Actuators B Chem. 2012, 173, 338–345. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  18. Shen, H.-H.; Chung, L.-Y.; Yao, D.-J. Improving the dielectric properties of an electrowetting-on-dielectric microfluidic device with a low-pressure chemical vapor deposited Si3N4 dielectric layer. Biomicrofluidics 2015, 9, 022403. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  19. Rodrigues, V.A.V. Digital Microfluidic Devices: The role of the Dielectric Layer. Ph.D. Thesis, Universidade Nova de Lisboa, Lisbon, Portugal, 2014. [Google Scholar]
  20. Freire, S.L. Perspectives on digital microfluidics. Sens. Actuators A Phys. 2016, 250, 15–28. [Google Scholar] [CrossRef] [Scilit]
  21. Pollack, M.G.; Fair, R.B.; Shenderov, A.D. Electrowetting-based actuation of liquid droplets for microfluidic applications. Appl. Phys. Lett. 2000, 77, 1725–1726. [Google Scholar] [CrossRef] [Scilit]
  22. Yu, Y.; Chen, J.; Li, J.; Yang, S.; Fan, S.-K.; Zhou, J. Microfabrication of a digital microfluidic platform integrated with an on-chip electrochemical cell. J. Micromech. Microeng. 2013, 23, 095025. [Google Scholar] [CrossRef] [Scilit]
  23. Abdelgawad, M.; Wheeler, A.R. Low-cost, rapid-prototyping of digital microfluidics devices. Microfluid. Nanofluid. 2008, 4, 349. [Google Scholar] [CrossRef] [Scilit]
  24. Chae, J.B.; Kwon, J.O.; Yang, J.S.; Kim, D.; Rhee, K.; Chung, S.K. Optimum thickness of hydrophobic layer for operating voltage reduction in EWOD systems. Sens. Actuators A Phys. 2014, 215, 8–16. [Google Scholar]
  25. Li, Y.; Baker, R.J.; Raad, D. Improving the performance of electrowetting on dielectric microfluidics using piezoelectric top plate control. Sens. Actuators B Chem. 2016, 229, 63–74. [Google Scholar] [CrossRef] [Scilit]
  26. Banerjee, A.N.; Qian, S.; Joo, S.W. High-speed droplet actuation on single-plate electrode arrays. J. Colloid Interface Sci. 2011, 362, 567–574. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  27. Jain, V.; Raj, T.P.; Deshmukh, R.; Patrikar, R. Design, fabrication and characterization of low cost printed circuit board based EWOD device for digital microfluidics applications. Microsyst. Technol. 2017, 23, 389–397. [Google Scholar] [CrossRef] [Scilit]
  28. Gong, M.M.; Sinton, D. Turning the page: Advancing paper-based microfluidics for broad diagnostic application. Chem. Rev. 2017, 117, 8447–8480. [Google Scholar] [CrossRef] [Scilit] [PubMed]
  29. Jenkins, G.; Wang, Y.; Xie, Y.L.; Wu, Q.; Huang, W.; Wang, L.; Yang, X. Printed electronics integrated with paper-based microfluidics: New methodologies for next-generation health care. Microfluid. Nanofluid. 2015, 19, 251–261. [Google Scholar]
  30. Krauss, S.T.; Remcho, T.P.; Monazami, E.; Thompson, B.L.; Reinke, P.; Begley, M.R.; Landers, J.P. Inkjet printing on transparency films for reagent storage with polyester–toner microdevices. Anal. Methods 2016, 8, 7061–7068. [Google Scholar]
  31. Yu, Y.; Chen, J.; Zhou, J. Parallel-plate lab-on-a-chip based on digital microfluidics for on-chip electrochemical analysis. J. Micromech. Microeng. 2013, 24, 015020. [Google Scholar] [CrossRef] [Scilit]
  32. Khodayari, M.; Carballo, J.; Crane, N.B. A material system for reliable low voltage anodic electrowetting. Mater. Lett. 2012, 69, 96–99. [Google Scholar]
  33. Chabal, Y.J. Fundamental Aspects of Silicon Oxidation; Springer Science & Business Media: Berlin, Germany, 2012; Volume 46. [Google Scholar]
  34. Liu, H.; Dharmatilleke, S.; Maurya, D.K.; Tay, A.A. Dielectric materials for electrowetting-on-dielectric actuation. Microsyst. Technol. 2010, 16, 449–460. [Google Scholar]
  35. Yuan, Y.; Lee, T.R. Contact angle and wetting properties. In Surface Science Techniques; Springer: Berlin, Germany, 2013; pp. 3–34. [Google Scholar]
  36. Guo, J.; Li, H.; Chen, Y.; Kang, Y. A microfluidic impedance cytometer on printed circuit board for low cost diagnosis. IEEE Sens. J. 2014, 14, 2112–2117. [Google Scholar]
  37. Ren, C.; Zhang, S.; Song, D.; Guo, J. Lab on dielectric film deposited PCB device for characterization of electrical property of biological cells. IEEE Trans. Dielectr. Electr. Insul. 2016, 23, 1895–1897. [Google Scholar] [CrossRef] [Scilit]

Article Metrics

Citations

Article Access Statistics

Multiple requests from the same IP address are counted as one view.