1. Introduction
Hidden caries are dentin lesions that cannot be easily detected via visual inspection because the occlusal enamel appears normal or only minimally demineralized [
1]. Even though hidden caries may be radiographically detected, X-ray radiation in routine examinations may sometimes cause side effects [
1,
2,
3]. Therefore, a method substituting for or assisting radiography and enabling lesions inside teeth to be diagnosed from the occlusal surface is needed.
Photoacoustic imaging is one imaging method used to noninvasively observe the subsurface morphology of living bodies [
4]. It detects ultrasonic waves generated by optical absorbing substances when they thermally expand after absorbing irradiated optical pulses [
4]. The penetration depth of photoacoustic imaging is greater than that of purely optical imaging techniques, which are usually affected by the strong optical scattering of tissue. Furthermore, photoacoustic imaging retains the high spatial resolution of optical methods, which cannot be achieved by conventional ultrasound imaging.
Photoacoustic imaging has been used in a variety of applications, including the imaging of blood vessels [
5], brain functions [
6], and other uses [
7,
8,
9]. Several groups have reportedly applied the photoacoustic method to the diagnosis of dental caries. Wang et al. used an ultraviolet laser light that is strongly absorbed by enamel to generate surface acoustic waves and optically measured the elasticity of the enamel layer [
10]. Blodgett used pulsed CO
2 and Nd:YAG lasers for excitation and detected photoacoustic waves using an optical interferometer [
11]. Sampathkumar et al. used 1064-nm and 532-nm Nd:YAG lasers for excitation and an interferometer based on a 532-nm laser light for detection and successfully obtained photoacoustic signals [
12,
13]. These photoacoustic systems have primarily been used for the detection of early caries formed on the enamel surface. Conversely, Cheng et al. applied photoacoustic tomography to detect caries lesions [
14]. In their experiments, a 532-nm Nd:YAG laser irradiated the occlusal surface of an extracted tooth sample and an ultrasonic transducer rotated around the sample. Even though they successfully obtained tomographic images via reconstruction calculations, the scanning of the transducer may be difficult for in vivo applications. Li et al. performed simulations to estimate changes in the temperature and pressure when dental tissues were irradiated with laser light. The results showed that those changes were in a margin of safety when applying photoacoustic method to clinical applications [
15]. El-Sharkawy et al. reported that they differentiated optically-induced surface acoustic waves from caries and normal parts by using a system based on Michelson interferometer [
16,
17]. Kim et al. found that the intensities of photoacoustic waves induced by visible and near-infrared lasers were different in the degree of progress of dental caries [
18]. These reports have shown the feasibility of the photoacoustic detection of dental caries; however, these methods have not been applied to the detection of hidden caries. Optical coherence tomography is also emerging as an optical method for detection of early dental caries. However, the method has a limitation in the imaging depth for detection of hidden caries lesions [
19,
20,
21].
The aim of the work reported in this article is to test the feasibility of a photoacoustic method for the detection of hidden caries. For this purpose, we propose a photoacoustic imaging system for in vivo applications.
3. Results
In our preliminary experiments, we found that the intensities of the photoacoustic signals varied with the surface conditions of the sample surface. Therefore, we tried to distinguish between caries and healthy (normal) parts of teeth using not the intensity but the shape of the frequency spectra of the photoacoustic signals.
Figure 5a shows the frequency spectra of normal and pigmented parts of a tooth calculated from the waveforms obtained from the simulated caries inside the tooth and the normal part. To suppress the effect of noise, we calculated the centered moving average values and these curves were normalized to the peak intensity. The difference between the intensities of the normalized frequency spectra from caries and the normal part was calculated and the result is shown in
Figure 5b, which shows that the difference spectrum has positive values from 0.5 MHz to 1.5 MHz.
In addition, we investigated the effects of stains or caries on the occlusal surface by comparing the frequency spectra generated in simulated caries lesions inside the teeth with those generated at occlusal surfaces colored with the pigment.
Figure 6 shows an appearance of occlusal surface stained with the pigment.
Figure 7a shows the frequency spectra from the pigment inside the teeth and the stained occlusal surface. The difference between these spectra is shown in
Figure 7b, which indicates that the frequency components between 0.4 MHz and 1.2 MHz are higher in the waves from the caries lesions inside the teeth.
Figure 8a shows the frequency spectra of the photoacoustic signals obtained from the 5% and 50% liquid samples with black inks. In the spectrum of the 5% concentration liquid, there is a peak at approximately 1 MHz with a spread to 4 MHz. The peak shifted to 2 MHz in the 50% liquid and the broad peak spanned 5 MHz. As shown in the differential spectrum in
Figure 8b, the difference had negative values in the range of 0–1.5 MHz and became positive in the range of 1.5–4.5 MHz.
We applied this method to the imaging of simulated hidden caries, and the result is shown in
Figure 9. In this experiment, the detected frequency spectra were integrated in the region from 0.5 MHz to 1.2 MHz and the integrated intensities were plotted so that darker spots showed stronger signals. In the experiments, laser light scattered from the dental tissue sometimes hits the acoustic probe directly and the photoacoustic waves from the probe itself are observed. We investigated waveforms with this type of noise and found that acoustic waves from the probe were not observed more than 1.5 μs after the laser radiation. Therefore, we analyzed the waveforms by only focusing on the signals after 1.5 μs.
Figure 9a shows the appearance of a sample with simulated caries formed 1.5 mm below the enamel surface. The diameter of the simulated caries was 2.5 mm. It was confirmed that no visible dark spot was observed on the enamel surface.
Figure 9b shows a photoacoustic image of the same sample obtained using our proposed system. One can see that the simulated caries inside the tooth is clearly observed and that the size of the dark spot is the same as the size of the simulated caries.
Figure 9c shows a photoacoustic image of the same tooth sample before applying the pigment.
In addition, we performed photoacoustic imaging experiments using model teeth with simulated caries of different sizes and depths.
Figure 10a shows images of a model tooth with a simulated caries 1.2 mm in diameter at a depth of 0.9 mm.
Figure 10b shows images of simulated caries 1.2 mm in diameter at a depth of 1.9 mm.
Figure 10c shows images of simulated caries 0.5 mm in diameter at a depth of 0.9 mm.
Finally, we performed photoacoustic imaging of an extracted human tooth with hidden caries. Before the photoacoustic experiment, we used a dental X-ray computed tomography (CT) device (SMX-100CT, Shimadzu, Kyoto, Japan) to obtain the exact shape and size of the hidden caries.
Figure 11a shows a photograph of the occlusal surface.
Figure 11b,c are CT cross sections along the A–B and C–D lines in
Figure 11a, respectively. These CT images indicate that there was a hidden caries at the boundary between the enamel and the dentin at a depth of approximately 0.8 mm from the enamel surface. The size of the caries was approximately 3.5 mm in the A–B direction and 2.5 mm in the C–D direction, and it was 0.7-mm thick. When there is such a large caries lesion below the occulsal surface, usually one would see pits and fissures with stains or large open areas. However, for the extracted tooth used in our experiment, we confirmed that no visible caries lesion was formed on the surface as shown in
Figure 11a and we used this sample to test the feasibility of photoacoustic method for detection of hidden caries lesions.
Figure 12 shows a photoacoustic image, taken using the proposed method, of the sample tooth shown in
Figure 11.
4. Discussion
From the above results of the experiments which we performed to show the feasibility of photoacoustic method for detection of hidden dental caries, we found that the photoacoustic method can be applied for detection of hidden dental caries in principle. However, in the current state of the imaging system and method which we proposed, only relatively large caries lesions with sizes larger than around 1 mm could be detected. In clinical practice, it is sometimes necessary to find lesions before they have reached this size so that one can use a preventive technique. To apply the proposed method for clinical applications, therefore, the sensitivity and resolution of the detected images should be drastically improved. This may be possible by using an image reconstruction technique such as circular back-propagation method [
24] because the current system uses only a simple B-mode imaging method.
In the photoacoustic methods for detecting dental caries reported so far, an optical interferometer or a piezoelectric transducer has been used to detect the ultrasonic waves. However, optical interferometers are usually too sensitive to fluctuations in the distance from the sample surface to be used in clinical applications. In addition, when piezoelectric transducers are used, usually the sample needs to be dipped in water to match the acoustic impedance, and this is difficult for in vivo applications. In our proposed method, in contrast, a composite ultrasonic probe with a soft polymer coating was used. The soft probe can be attached to the surface of the tooth crown, as shown in
Figure 2b, without water or gel and this makes the imaging system more feasible for clinical applications. However, we also found that this kind of composite probes have nonlinearity in the frequency response as shown in
Figure 8. Therefore, one should take this nonlinearity into account when performing photoacoustic imaging.
Even though beam scanning of the laser light is necessary to obtain photoacoustic images, typical systems based on a scanning mirror cannot be easily applied to clinical applications for dental caries detection because such systems are usually bulky and do not fit into oral cavities. To solve this problem, we developed a laser scanning system using a bundle of flexible hollow-optical fibers shown in
Figure 4. By using this probe, one can scan the laser beam using a simple linear stage at the input end and this makes the system simple and low in the cost.
The analysis of frequency spectra of photoacoustic waves from the extracted tooth samples with simulated caries inside showed that hidden caries lesions with a brown color may be detected by focusing the spectral differences. The difference between the spectral intensities from pigmented part and the normal part shown in
Figure 5b indicates that the intensities of the photoacoustic waves generated in the pigmented part are higher in the frequency range of 0.5 MHz to 1.5 MHz than those of waves generated in normal parts of a tooth. This may be caused by a resonance vibration in the cavity generated by the absorption of the laser pulse at the simulated caries lesion. We performed the same experiments with different caries model teeth and the majority of the model teeth showed similar results. Considering this and the spectral difference between the pigmented part and the stained occlusal surface shown in
Figure 7b, we found that we could detect simulated hidden caries lesions by focusing on the frequency range of 0.5–1.2 MHz. Then we could differentiate between photoacoustic waves generated from pigmented regions inside teeth from those generated from normal parts or stained occlusal surfaces. The result of the nonlinear response of the composite probe used in our system, which is shown in
Figure 8, indicates that the frequency response changed with the intensity of the photoacoustic wave due to the nonlinearity of the composite probe. This nonlinearity might enhance the change in the frequency responses shown in
Figure 5b and
Figure 7b, and therefore one can differentiate the strong photoacoustic signal of the pigmented parts from the normal parts by detecting the spectral changes.
The results of photoacoustic imaging of simulated caries showed that the image of brownish hidden caries could be obtained by the proposed photoacoustic imaging system when the size of lesions are relatively large. In the photoacoustic image of the tooth sample without pigment inside, which is shown in
Figure 9, no dark spot is evident, which indicates that fissures and pits on the occlusal surface do not induce photoacoustic waves sufficiently strong to be mistaken for the signal from the pigment. In the photoacoustic images of tooth samples with simulated caries lesions of different sizes, which are shown in
Figure 10a,b, the contrast of the observed image was worse than that of the simulated caries at shallower depths due to the optical scattering and absorption of the thicker dental tissues. However, we confirmed that the simulated caries was observed as a dark spot at the location of the caries. Note that in
Figure 10a,b the shapes of the obtained photoacoustic images do not coincide with the shapes of the fissures on the occlusal surface. This indicates that furcations and fissures do not affect the photoacoustic images. Furthermore, the size of the observed dark spot nearly coincides with that of the simulated caries when the simulated caries lesions are larger than 1.2 mm. However, no clear image was not obtained for the lesion with the size of 0.5 mm as shown in
Figure 10c.
As a result of photoacoustic imaging of an extracted human tooth with a real hidden caries lesion, the hidden caries with an ellipse shape was detected near the hidden caries lesion located by the CT observation, demonstrating the feasibility of the proposed method for the detection of real hidden caries. However, the shape of the dark spot in the observed photoacoustic image did not exactly coincide with the CT image. This is due to the non-uniformities in the enamel thickness and the color of real dental caries. The proposed method can only detect comparatively strong photoacoustic signals from caries with a brown color. At the present stage, therefore, the method cannot detect a demineralized area without a color change. However, this may become possible with a photoacoustic method using different polarization directions.
5. Conclusions
We proposed a photoacoustic imaging system for detecting caries lesions inside teeth. First, we prepared hidden caries models using a pigment with an absorption spectrum similar to that of real caries lesions and irradiated the occlusal surface of the model teeth containing the pigment with 532-nm-wavelength laser pulses. We found that the frequency spectra of the photoacoustic waves from the simulated caries lesions included components in the range of 0.5–1.2 MHz that were not seen in the spectra of waves from healthy parts of the teeth. This indicated that hidden caries could be detected using a photoacoustic imaging technique. We then fabricated an imaging system composed of a bundle of hollow-optical fibers for laser radiation and an acoustic probe that is attached to the tooth surface for clinical applications. The results of ex vivo imaging experiments using model teeth and an extracted tooth with hidden caries lesions demonstrated that caries lesions inside teeth that cannot be seen in a visual inspection can be detected by focusing on the above frequency components of the photoacoustic waves. By using this method, one may obtain images of hidden caries lesions non-invasively without X-ray radiation. We believe that this is the first report on the feasibility of photoacoustic method for imaging hidden caries lesion.
At this stage, our proposed method only detects the optical absorption caused by the color difference in the dentin and the detectable size of the lesion is limited to the one larger than around 1 mm. We are working on a photoacoustic method using different polarization directions to detect lesions without a brownish color. In addition, we are working on an image reconstruction technique such as circular back-propagation method to improve the imaging resolution for detection of much smaller caries lesions.