Sensors 2009, 9(4), 3000-3010; doi:10.3390/s90403000
Article

Matrix Effects on the Microcystin-LR Fluorescent Immunoassay Based on Optical Biosensor

1email, 2email, 1email, 1email and 1,* email
Received: 12 March 2009; in revised form: 25 March 2009 / Accepted: 15 April 2009 / Published: 23 April 2009
(This article belongs to the Section Biosensors)
This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Abstract: Matrix effects on the microcystin-LR fluorescent immunoassay based on the evanescent wave all-fiber immunosensor (EWAI) and their elimination methods were studied. The results indicated that PBS and humic acid did not affect the monitoring of samples under the investigated conditions. When the pH was less than 6 or higher than 8, the fluorescence signals detected by immunosensor systems were obviously reduced with the decrease or increase of pH. When the pH ranged from 6 to 8, IC50 and the linear working range of MC-LR calculated from the detection curves were 1.01~1.04 μg/L and 0.12~10.5 μg/L, respectively, which was favourable for an MC-LR immunoassay. Low concentrations of Cu2+ rarely affected the detection performance of MC-LR. When the concentration of CuSO4 was higher than 5 mg/L, the fluorescence signal detected by EWAI clearly decreased, and when the concentration of CuSO4 was 10 mg/L, the fluorescence signal detected was reduced by 70%. The influence of Cu2+ on the immunoassay could effectively be compromised when chelating reagent EDTA was added to the pre-reaction mixture.
Keywords: Matrix effects; Microcystin-LR; fluorescent immunoassay; optical biosensor
PDF Full-text Download PDF Full-Text [211 KB, uploaded 21 June 2014 02:43 CEST]

Export to BibTeX |
EndNote


MDPI and ACS Style

Long, F.; Zhu, A.-N.; Sheng, J.-W.; He, M.; Shi, H.-C. Matrix Effects on the Microcystin-LR Fluorescent Immunoassay Based on Optical Biosensor. Sensors 2009, 9, 3000-3010.

AMA Style

Long F, Zhu A-N, Sheng J-W, He M, Shi H-C. Matrix Effects on the Microcystin-LR Fluorescent Immunoassay Based on Optical Biosensor. Sensors. 2009; 9(4):3000-3010.

Chicago/Turabian Style

Long, Feng; Zhu, An-na; Sheng, Jian-Wu; He, Miao; Shi, Han-Chang. 2009. "Matrix Effects on the Microcystin-LR Fluorescent Immunoassay Based on Optical Biosensor." Sensors 9, no. 4: 3000-3010.

Sensors EISSN 1424-8220 Published by MDPI AG, Basel, Switzerland RSS E-Mail Table of Contents Alert