Sensors 2007, 7(8), 1462-1479; doi:10.3390/s7081462
Article

Piezoelectric Sensor for Determination of Genetically Modified Soybean Roundup Readyâ in Samples not Amplified by PCR

1 Institute of Animal Reproduction and Food Research, Polish Academy of Sciences, Tuwima 10, 10-474 Olsztyn, Poland 2 Genetic Modifications Analysis Laboratory, Institute of Biochemistry and Biophysics, Polish Academy of Sciences, Pawińskiego 5a, 02-106 WARSAW, Poland 3 Institute of Genetics and Biotechnology, Warsaw University, Pawińskiego 5a, 02-106 Warsaw, Poland
* Author to whom correspondence should be addressed.
Received: 19 July 2007; Accepted: 10 August 2007 / Published: 14 August 2007
PDF Full-text Download PDF Full-Text [983 KB, uploaded 20 June 2008 16:52 CEST]
Abstract: The chemically modified piezoelectrodes were utilized to develop relativelycheap and easy to use biosensor for determination of genetically modified Roundup Readysoybean (RR soybean). The biosensor relies on the immobilization onto goldpiezoelectrodes of the 21-mer single stranded oligonucleotide (probes) related to5-enolpyruvylshikimate-3-phosphate synthase (EPSPS) gene, which is an active componentof an insert integrated into RR soybean genome. The hybridization reaction between theprobe and the target complementary sequence in solution was monitored. The system wasoptimized using synthetic oligonucleotides, which were applied for EPSPS gene detectionin DNA samples extracted from animal feed containing 30% RR soybean amplified by thePCR and nonamplified by PCR. The detection limit for genomic DNA was in the range of4.7·105 numbers of genom copies contained EPSPS gene in the QCM cell. The propertiessuch as sensitivity and selectivity of piezoelectric senor presented here indicated that it could be applied for the direct determination of genetically modified RR soybean in the samples non-amplified by PCR.
Keywords: Piezoelectrodes; Roundup Ready soybean; Genomic DNA.

Article Statistics

Load and display the download statistics.

Citations to this Article

Cite This Article

MDPI and ACS Style

Stobiecka, M.; Cieśla, J.M.; Janowska, B.; Tudek, B.; Radecka, H. Piezoelectric Sensor for Determination of Genetically Modified Soybean Roundup Readyâ in Samples not Amplified by PCR. Sensors 2007, 7, 1462-1479.

AMA Style

Stobiecka M, Cieśla JM, Janowska B, Tudek B, Radecka H. Piezoelectric Sensor for Determination of Genetically Modified Soybean Roundup Readyâ in Samples not Amplified by PCR. Sensors. 2007; 7(8):1462-1479.

Chicago/Turabian Style

Stobiecka, Magdalena; Cieśla, Jarosław M.; Janowska, Beata; Tudek, Barbara; Radecka, Hanna. 2007. "Piezoelectric Sensor for Determination of Genetically Modified Soybean Roundup Readyâ in Samples not Amplified by PCR." Sensors 7, no. 8: 1462-1479.

Sensors EISSN 1424-8220 Published by MDPI AG, Basel, Switzerland RSS E-Mail Table of Contents Alert