Next Article in Journal
Non-Invasive Device for Blood Pressure Wave Acquisition by Means of Mechanical Transducer
Previous Article in Journal
Analysis on the Live Coral Cover around Weizhou Island Using MODIS Data
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

A New Submersion Detection Sensor Using Two Resistance Temperature Detectors Operating on the Thermal Equilibrium Principle

Department of Electrical Engineering, 100 Inha-ro, Michuhol-gu, Inha University, Incheon 22212, Korea
*
Author to whom correspondence should be addressed.
Sensors 2019, 19(19), 4310; https://doi.org/10.3390/s19194310
Submission received: 19 August 2019 / Revised: 14 September 2019 / Accepted: 15 September 2019 / Published: 4 October 2019
(This article belongs to the Section Physical Sensors)

Abstract

:
In this study, a new submersion detection sensor with improved reliability and stability is proposed. The new sensor uses two Resistance Temperature Detectors (RTDs) and operates on the thermal equilibrium principle. The submersion detection sensor controls two RTDs that maintain a constant temperature difference between them in the surrounding environment. The first RTD is used as a reference sensor to measure ambient temperature and the second RTD is supplied with higher current than the reference sensor for self-heating. When submerged, because the thermal conductivity and convective heat transfer coefficient of water are higher than that of air, the temperature difference between the two RTDs is lower in water than in air based on the thermal equilibrium principle. Under these conditions, a submersion detector with a signal conditioning circuit detects these temperature differences. The static performance of the proposed sensor was evaluated by checking whether malfunctions occurred at varying ambient temperatures, differing humidities, and when there was rainfall. In addition, the dynamic performance was evaluated using the response time at varying ambient air temperatures before submersion and with changing water temperatures after submersion, as a metric. The proposed submersion detection sensor is expected to find useful application in aircrafts, ships, and various other industrial fields.

1. Introduction

In habitual flood areas, water level measuring systems are required to collect water level information for preventive measures [1]. Furthermore, in the case of fleets of ships, a system that prepares for problems caused by water or oil leakage and fuel shortage by determining the extent of the submersion of the ships, or by measuring the level of fuel in a fuel tank is required [2,3,4]. Therefore, submersion detection sensors are applied in various fields, and a lot of research has been carried out to enhance their reliability.
The submersion detection sensors studied and applied to date mainly utilize the detection of differences in capacitance, electrical resistance, and the refraction or reflection of optical fibers [5,6,7,8,9,10]. However, because submersion sensors detect differences in capacitance on the contact surfaces of the sensors, there may be malfunctions under conditions of high humidity and rainfall. Although the submersion sensor using two electrodes detects the reduced electrical resistance caused by an electrical short circuit when the electrode is submerged, it is difficult to guarantee the sensor’s electrical characteristics and reliability due to the problem of insulation [11,12]. Recently studied fiber optic sensors detect submersion using the difference in the refractive index of air or liquid at the surface of the sensor. Under high humidity or rainfall, however, there may be water on the fiber’s end and foreign objects attached to the end of the fiber that may cause malfunctions [13,14,15,16,17,18,19,20].
In this study, a submersion detection sensor that included a submersion sensor operating on the thermal equilibrium principle and a signal conditioning circuit were proposed in order to solve problems caused by conventional submersion sensors (malfunctions caused by high humidity and rainfall conditions, insulation issues caused by electrical short circuits, and malfunctions due to foreign objects on the sensor). The proposed submersion detection sensor uses two Resistance Temperature Detectors (RTDs) to sense temperature. The first RTD is used as a reference sensor to measure ambient temperature, and the second RTD is supplied with a higher current than the reference sensor for self-heating. The higher current also serves to create and maintain a temperature difference between the RTDs. Under submerged conditions, the thermal conductivity of water is higher than that of air, and the heat transfer coefficient of a fluid is significantly greater than that of air. Thus, if the temperature difference between the RTDs is reduced underwater, submersion is detected in the signal conditioning circuit. In this study, the heat transfer characteristics of a submersion sensor for some range of air temperature and submersion temperature were analyzed with MATLAB, and the static and dynamic characteristics of the sensor were evaluated through experiments [21].

2. Materials and Methods

2.1. Materials and Sensor Mechanical Design

Two RTDs were used in the proposed submersion detection sensor. The first RTD measured the air temperature and the second RTD was used as a heating sensor. The heating sensor maintains a temperature that is higher than the ambient air temperature by absorbing a higher current. The well-known self-heating characteristic of RTDs is utilized to heat the sensor.
The RTD used in this study was 1PT100GO1020, which was provided by Omega, and is made of glass wire wound platinum, Pt-100, Class B (IEC751), and α = 0.00385 (TCR: Temperature coefficient of resistance). The dependence of the resistance of the RTDs on changing temperature can be expressed in a linear approximation as follows:
R R T D = R 0 [ 1 + α ( T i ) ] ,
where R 0 is the resistance of RTD at 0 °C, 100 Ω, and α ( T i ) is the TCR at the measured temperature T i . Table 1 shows the response time and self-heating characteristics of the RTD sensor [22].
The submersion detection sensor shown in Figure 1 was designed with the heating RTD and the reference RTD facing one another and used a silicon bond to fix the RTDs. The fluid inflow holes consisted of four ∅ 3.0 mm holes at 90° intervals on the underside and twelve ∅ 1.8 mm holes at 30° intervals on the upper side. The reason behind why the fluid inflow holes were installed on the underside and the upper side was to prevent internal air pockets when submersion occurred vertically and to detect the submersion in all directions. Furthermore, the shield for preventing direct contact with water was structurally designed to prevent malfunctions caused by rainwater. The outer casing was made of aluminum and the total dimensions of the submersion detection sensor were ∅ 20.0 × 50.0 mm where the thickness of the sensor was 50.0 mm.

2.2. Transient Heat Transfer Analysis and Signal Conditioning Circuit Design

The operating principle of the submersion detection sensor proposed in this study can be explained theoretically through the analysis of transient heat transfer and a simulation performed with MATLAB.
In an electronic circuit, the temperature of an RTD was measured by the voltage across it. Normally, the RTD is supplied with a low current, several milliampere(mA), in order to minimize the errors that arise from self-heating due to the current flow. However, the submersion detection sensor presented in this study supplied tens of mA to the heated RTD for heating. Furthermore, the reference RTD was supplied with several mA to minimize self-heating. According to the first law of thermodynamics, the heat of the heating RTD can be expressed as follows:
Q ˙ = Q ˙ i + Q ˙ c v ,
where Q ˙ is the heat transfer rate of the heating RTD, Q ˙ i is the heat transfer rate of the heating element heated by the current applied to the resistance element (platinum) inside the RTD, and Q ˙ c v is the convective heat transfer rate that governs the amount of heat escaping into the fluid around the glass. In the case of a solid, a lumped system analysis can be applied to analyze the heat transfer in transient conditions if the Biot number is less than 0.1. The equation to obtain the Biot number is as follows [23]:
B i = h L C k ,
where B i is the Biot number, h is the heat transfer coefficient of fluid [W/m2 °C], k is thermal conductivity [W/(m °C)], and L C is the characteristic length of the heat transfer [m]. The value of L C was calculated using the following equation and was 0.0004545 m, assuming that the RTD and surrounding glass body was a single thermal mass.
L c = V A = V o l u m e t r i c   o f   h e a t i n g   e l e m e n t s u r f a c e   o f   h e a t i n g   e l e m e n t = π l r 2 2 π r l + 2 π r 2 .
The thermal conductivity k in Table 2 was obtained through experiments carried out in normal and submerged conditions. The value of k in the submerged condition was assumed to be 1.8 times larger than that under normal conditions because the heat was also transferred from the surface of the RTD to water. The calculated values of B i are as follows.
As shown in Table 2, the calculated values of B i were less than 0.1 under all conditions and could be analyzed by the lumped system.
Because the rate of the change in temperature calculated by the lumped system in the heating RTD was equal to the rate of the change in internal energy, it could be expressed as follows:
Q ˙ ( t ) = ϱ c V d T d t = m c d T d t ,
where ϱ is density, c is specific heat, V is volume, and m is mass. Using Equation (2), Q ˙ can be rewritten as follows:
Q ˙ = Q ˙ i + Q ˙ c v = I 2 R k L C + h A ( T i T ) ,
where I is current, and R is the resistance of the RTD. The heat loss of the RTD can be expressed by Newton’s law of cooling. In Equation (6), h is the convective heat transfer coefficient of fluids, A is the heat transfer area of the RTD, T i is the initial temperature (i.e., environmental temperature) of the RTD, and T is the temperature of the fluid at a sufficient distance from the RTD [24,25].
From Equations (5) and (6), we obtained the following differential equation:
m c d T d t = I 2 R k L C + h Α ( T i T )
The solution to Equation (7) could be written as follows:
T ( t ) = I 2 R k L C ( 1 e k L m c t ) + ( T i T ) e h Α m c t + T .
The value of the product of the mass, m [kg], and the specific heat, c [W/(kg°C)], was the inverse of 0.26 [°C /1 mW]. This value was the self-heating error value of the RTD specification presented in Table 1, and was equal to 0.00384615 [W/°C].
Figure 2 shows the diagram of the heat transfer occurring in the fluid of the heating RTD and the signal conditioning circuit.
In the signal conditioning circuit, the warm-up circuit consists of U 4 C (Op amp HA1-4742-2), R 11 , and C 1 . Together, they prevent malfunctions due to high humidity and moist conditions when the system is operating. V o , the voltage applied to U 4 C , was as follows:
V o ( t ) = 6 ( 1 e 1 R 11 C 1 t ) ,
From above, the time taken for the voltage to reach V o could be written as follows:
t = R 11 C 1 × ln ( 1 V o 6 ) .
The calculated t was about 1.52 s with V o set to 3.0 V, R 11 was 2.2 MΩ, and C 1 was 1.0 µF. In the case of t < 1.52, the relay R L 1 was activated by turning Q1 ON, which implies that R T D H was directly connected to a voltage of 12.0 V. In the case of t 1.52, the relay R L 1 was deactivated by turning Q1 OFF. In this case, R T D H was connected to 12.0 V through R 3 .
Equations (11) and (12) can be derived from the dependence of the resistance of the R T D H on temperature, taking into account the signal conditioning circuit in Equations (1) and (8).
In the case of t < 1.52, the temperature of R T D H was given as follows:
T ( t ) = ( V 2 R T D H × k L C ) × ( 1 e k L m c t ) + ( T i T ) e h Α m c t + T .
In the case of t 1.52, we had
T ( t ) = ( V R 3 + R T D H ) 2 R T D H k L C × ( 1 e k L m c t ) + ( T i T ) e h Α m c t + T .
The temperature of R T D A could be expressed as follows:
T ( t ) = ( V R 4 + R T D A ) 2 R T D A k L C × ( 1 e k L m c t ) + ( T i T ) e h Α m c t + T .
Based on Equations (11), (12), and (13), the results of the simulation performed using MATLAB for the temperature of each RTD when it was surrounded by air is presented in Figure 3.
In Figure 3, the purpose of heating the R T D H within the initial 1.52 s was to increase the ambient temperature by about 108 °C in order to prevent possible malfunctions when the RTD was powered ON in wet conditions, and to reduce warm-up time. After 1.52 s, the current supplied to the R T D H was reduced. This, in turn reduced the temperature difference between the R T D H and the surrounding air to about 35 °C. Next, the R T D A was simulated to be almost the same as the air temperature. The values of the convective heat transfer coefficient of air, h, ranging from 10 to 50 were simulated, and it took 25 and 5 s to reach thermal equilibrium for heat transfer coefficient values of 10 and 50, respectively. Figure 4 shows the changes in temperature between these RTDs under submerged conditions. This simulation result makes it easy to understand the principle behind the operation of the proposed sensor.
In order to understand easily the changing characteristics of the RTD with varying temperatures, the submersion temperature and the air temperature were both simulated under the same temperature of 20 °C. Submersion was carried out 10 s after the experiment commenced. The temperature difference between the RTDs decreased from 35 °C to about 19 °C (55%) 0.3 s after submersion. Although water has a convective heat transfer coefficient of about 600, three different heat transfer coefficient values of 500, 600, and 700 were simulated to allow for a sufficiently large margin of error. The response time of the submersion sensor was simulated to be 0.3 s. The results of the simulations of the submersion sensor performed with a submersion temperature of 20 °C and under atmospheric ambient temperatures varying from −40 °C and 100 °C are presented in Figure 5 and Figure 6 respectively. As shown in these figures, the temperature of R T D H was not affected by changes in the temperature of the environment.
Based on the simulation results, the simplified circuit for implementing the submersion detection is shown in Figure 7. In order to detect very small voltages from the temperature difference between the RTDs, the power supply circuit was designed using components LM7812 (±4 % at T=25 ℃), LM1117(± 1 % at T = 25 ℃) from Texas Instrument Inc and MC7806(±1.5 % at T = 25 ℃) from ON Semiconductor that were low drop voltage linear regulators with low noise. In addition, the component HA1-4741-2 manufactured by INTERSIL was used to precisely detect low voltages. HA1-4741-2 is an Op-amp that has a low input offset feature (3 mV at T=25 ℃). The reason for using VR1(Trimmer) was to solve the problem caused by device tolerance.
The simulation and measurement results of the temperature, resistance, and Op-amp output voltage of the RTDs under the same air and submersion temperature of 20 °C are presented in Table 3.
The test results in Table 3 show that the actual measurements were similar to the simulated values. From Table 3, the measured value of V 1 changed from 7.04 V to 6.92 V when submerged. This difference value of 0.11 V was too small to be used as a detection signal. Therefore, the signal monitoring circuit that uses an Op-amp was designed so that the voltage of V 5 can be amplified to 0.67 V for easy recognition. The comparator of U 5 C was used to generate the final output and the reference voltage of the U 5 C was configured to be 6.78 V. Under normal operating conditions (when surrounded by air), the output voltage is 7.04 V, and the sensing output is 0 V (“low”). In the submerged condition, the output voltage was 6.92 V, and the sensing output was 12 V (“high”). The error margins of the sensing voltages depend on the external conditions. They were configured to be 0.45 V under the normal condition of being surrounded by air and 0.21 V in the submerged condition, respectively. They were designed to have sufficient margins to prevent malfunctions caused by noise and the tolerance of the component. After the power was turned on, because the voltage difference between the RTDs raised slowly during the initial heating procedure, the sensing output circuit used the CD4081B AND gate from Texas Instruments to prevent submersion detection during the initial 1.52 s period. The reason behind why the design was based on logic gates instead of a μ-controller was to exclude the effect of EMI (Electromagnetic Interference).

2.3. Self-Diagnosis Circuit

The submersion sensor and detector were composed of RTDs and electronic circuits, and a self-diagnosis circuit was configured to detect malfunctions caused by bad RTDs or problems with electronic circuits. The circuit illustrated in Figure 8 shows a portion of the self-diagnosis circuit. It checks the voltages in each major circuit to verify that the voltage at each terminal is in the normal range by processing the signals using the AND gate.
The ambient temperature of the operational environment of the submersion detector is required to be in the range −40 °C to +100 °C. Because the R T D H is always 35 °C higher than the R T D A , the minimum temperature of the R T D H should be −5 °C but was assigned a value of −20 °C to allow for a margin of error. Furthermore, the maximum temperature was fixed as +220 °C considering the margin of error based on the simulation performed at an ambient temperature of +100 °C when surrounded by air. In addition, the voltage V 1 that included the margin ratio, ranged from 6.42 V (minimum) to 7.96 V (maximum). The voltages of V 7 and V 8 were chosen so that the circuit could determine whether the voltage of V 1 was in the normal range. As shown in Table 4, the output of the submersion detector was disabled when the RTDs exhibited a fault condition (open, short, defective circuit, etc.) and the self-diagnosis circuit output 0 V to inform the system of the fault.

2.4. Test Equipment Design

As shown in Figure 9, the Arduino MEGA, which is a well-known open source hardware platform, was used to test the response time of the submersion detection. The test equipment used two separate submersion electrodes to get the exact time of submersion. It also had a signal conditioning circuit and a comparator in order to detect the submersion signal. The response time was displayed at a resolution of 1 ms on an LCD after the submersion signal was detected from the submersion electrodes. A UART (Universal Asynchronous Receiver Transmitter) was used as an interface with the LCD. Because the self-diagnosis output and final-sensing output was 12 V, a level shifter was used to step the 12 V signals down to 5 V signals for the Arduino MEGA.

3. Results

The four samples of the submersion detector configured by the proposed submersion detection sensor and signal conditioning circuit were fabricated. The static characteristics of the samples were tested to verify malfunctions in high humidity and rainfall conditions as shown in Figure 10. The submersion response time under dynamically varying conditions were tested in the directions of 0° (vertical), 45°, and 90° (horizontal) using fresh-water and salt-water.

3.1. Static Test

3.1.1. Humidity Test

MIL-STD-810G is the U.S. military standard for environmental testing applied to military equipment and commercial products. Humidity tests were performed in accordance with Procedure II of the MIL-STD-810G Method 507.5 [26]. Furthermore, PSL-4J temperature and humidity chamber (ESPEC Corporation) was used as a humidity test system [27]. Under our test conditions, the temperature was 25−60 °C and the humidity was 95 %. The results of the test conducted with the profile shown in Figure 11 for a total of 240 h showed no malfunctions, and no submersion was detected as shown in Table 5.

3.1.2. Rain Test

Rainfall tests were conducted in accordance with Procedure I and III of the MIL-STD-810G Method 506.4, and the results of the test showed no malfunctions in normal rain, blowing rain, and drip conditions as shown in Table 6.

3.2. Dynamic Characteristics: Submersion Sensing Response Time

The submersion sensing response time of the dynamic characteristic was tested in accordance with MIL-STD-801G in an ambient temperature range of −40 to +71 °C. The temperature of fresh-water and salt-water was examined as shown in Table 7 and tested in the range of −1.7–32 °C. The test results are presented in Table 8 and Table 9 [28,29]. There was no change in the submersion sensing response time due to environmental temperature and submersion temperature, and the submersion sensing response time was measured in the range of 0.4–0.6 s. The reason the submersion sensing response time in the horizontal direction was about 0.1 s shorter than that for the vertical direction was thought to be that the inflow speed of water into the submersion detector was faster due to its structure. The response time obtained from the simulation with MATLAB was about 0.3 s. The difference of 0.1–0.3 s between the simulation and the experiment was thought to stem from the fact that the inflow speed of water to the submersion sensor and the time taken to fill the inside with water were not considered in the simulation.

3.3. Self-Diagnosis Test

The sensors attached to expensive systems may cause catastrophic effects on the system if a fault or malfunction occurs. Therefore, it is essential to have a self-diagnosis function to determine whether the sensor is in the normal condition. The test results of the proposed self-diagnosis function of the submersion detector is shown in Table 10. When the RTD sensor failed either because the circuit was open, or because of a short circuit, the error was detected under all conditions.

3.4. Repeatability Test

Figure 12 shows the experimental setup that was prepared for the repeatability test. The test equipment shown in Figure 12 displays the response time, pass/fail result of submersion detection, and output voltages of the signal conditioning circuit for each test. The repeatability was tested through 1000 experiments with fresh water and salt water. There was no malfunction observed as shown in Table 11.

4. Discussion

We verified that the proposed submersion detection sensor operated on the thermal equilibrium principle, and thus had a very strong dependence on environmental factors, i.e., environmental temperature, submersion temperature, humidity, and rainfall conditions. Furthermore, it turned out that the submersion sensing response time was about 0.4–0.6 s, which was a little longer than those of existing submersion detection sensors. The malfunctions caused by external environmental conditions in practical applications are unacceptable, and the submersion sensing response time of 0.4–0.6 s is not an issue in most practical applications. Table 12 shows the comparison between the proposed submersion detection sensor and the existing submersion detection sensors.

5. Conclusions

In this study, two RTDs were used to propose a new submersion detection sensor that operates based on the thermal equilibrium principle. A signal conditioning circuit and a self-diagnosis circuit were also proposed. The transient heat transfer mechanism of the RTDs used in this submersion detection sensor was analyzed with MATLAB. Although the response time is slower than other submersion detection sensors, it is particularly robust against malfunctions caused by external environmental factors. It was verified that there were no malfunctions under the humidity and rainfall test conditions of MIL-STD-801G. Therefore, it can be useful for detecting the submersion of ships and aircraft exposed to the external environment, detecting submersion in low areas, and detecting submersion of major facilities. Because the RTD sensor used for the proposed submersion sensor was produced at a cost of tens of U.S. dollars, economic feasibility can be improved by using thin film type RTDs in the sensor. As future research work, we will conduct studies with the goal of reducing the size of the sensor and improving the submersion sensing response time.

Author Contributions

Conceptualization, Y.L.; methodology, Y.L.; software, Y.L.; validation, Y.L.; formal analysis, Y.L.; investigation, Y.L.; writing—original draft preparation, Y.L.; writing—review and editing, Y.L. and Y.S.L.; supervision, Y.S.L.; funding acquisition, Y.S.L. and Y.L.

Funding

This work was supported by Inha University Research Grant.

Acknowledgments

The authors would like to thank Lee Jae-hoon for helping with the test and data collection.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Abdullahi, S.I.; Habaebi, M.H.; Gunawan, T.S.; Islam, M.R. Miniaturized Water Flow and Level Monitoring System for Flood Disaster Early Warning. In Proceedings of the IOP Conference Series: Materials Science and Engineering, Kuala Lumpur, Malaysia, 8–9 August 2017; IOP Publishing: Bristol, UK, 2017; Volume 260, p. 012019. [Google Scholar]
  2. Vázquez, C.; Gonzalo, A.B.; Vargas, S.; Montalvo, J. Multi-sensor system using plastic optical fibers for intrinsically safe level measurements. Sens. Actuators A Phys. 2004, 116, 22–32. [Google Scholar] [CrossRef] [Green Version]
  3. Sohn, K.R.; Shim, J.H. Liquid-level monitoring sensor systems using fiber Bragg grating embedded in cantilever. Sens. Actuators A Phys. 2009, 152, 248–251. [Google Scholar] [CrossRef]
  4. Lomer, M.; Arrue, J.; Jauregui, C.; Aiestaran, P.; Zubia, J.; López-Higuera, J.M. Lateral polishing of bends in plastic optical fibres applied to a multipoint liquid-level measurement. Sens. Actuators A Phys. 2007, 137, 68–73. [Google Scholar] [CrossRef]
  5. Manut, A.; Zoolfakar, A.S.; Muhammad, N.A.; Zolkapli, M. Characterization of inter digital capacitor for water level sensor. In Proceedings of the 2011 IEEE Regional Symposium on Micro and Nano Electronics, Kota Kinabalu, Malaysia, 28–30 September 2011; IEEE: Piscataway, NJ, USA, 2011; pp. 359–363. [Google Scholar]
  6. Chetpattananondh, K.; Tapoanoi, T.; Phukpattaranont, P.; Jindapetch, N. A self-calibration water level measurement using an interdigital capacitive sensor. Sens. Actuators A Phys. 2014, 209, 175–182. [Google Scholar] [CrossRef]
  7. Quoc, T.V.; Dac, H.N.; Quoc, T.P.; Dinh, D.N.; Duc, T.C. A printed circuit board capacitive sensor for air bubble inside fluidic flow detection. Microsyst. Technol. 2015, 21, 911–918. [Google Scholar] [CrossRef]
  8. Gong, C.S.A.; Chiu, H.K.; Huang, L.R.; Lin, C.H.; Hsu, Z.D.; Tu, P.H. Low-cost comb-electrode capacitive sensing device for liquid-level measurement. IEEE Sens. J. 2016, 16, 911–918. [Google Scholar] [CrossRef]
  9. Paczesny, D.; Tarapata, G.; Michał, M.; Jachowicz, R. The capacitive sensor for liquid level measurement made with ink-jet printing technology. Procedia Eng. 2015, 120, 911–918. [Google Scholar] [CrossRef]
  10. Babu, C.S.; Manohar, P. Design of a low-cost signal conditioning circuit for self-compensated non-contact capacitive type multi threshold liquid level sensor. In Proceedings of the International Conference on Circuits, Communication, Control and Computing, Bengaluru, India, 20–22 November 2014; IEEE: Piscataway, NJ, USA, 2014. [Google Scholar]
  11. Wang, X.; Wang, Y.; Leung, H.; Mukhopadhyay, S.C.; Tian, M.; Zhou, J. Mechanism and experiment of planar electrode sensors in water pollutant measurement. IEEE Trans. Instrum. Meas. 2014, 62, 516–523. [Google Scholar] [CrossRef]
  12. Casanella, R.; Casas, O.; Pallàs-Areny, R. Oil-water interface level sensor based on an electrode array. In Proceedings of the 2006 IEEE Instrumentation and Measurement Technology Conference, Sorrento, Italy, 24–27 April 2006; IEEE: Piscataway, NJ, USA, 2006; pp. 710–713. [Google Scholar]
  13. Tomita, S.; Tachino, H.; Kasahara, N. Water sensor with optical fiber. J. Lightwave Technol. 1990, 8, 1829–1832. [Google Scholar] [CrossRef]
  14. Michie, W.C.; Culshaw, B.; McKenzie, I.; Konstantakis, M.; Graham, N.B.; Moran, C.; Santos, F.; Bergqvist, E.; Carlstrom, B. Distributed sensor for water and pH measurements using fiber optics and swellable polymeric systems. Opt. Lett. 1995, 20, 103–105. [Google Scholar] [CrossRef] [PubMed]
  15. Arregui, F.J.; Matías, I.R.; Cooper, K.L.; Claus, R.O. Simultaneous measurement of humidity and temperature by combining a reflective intensity-based optical fiber sensor and a fiber Bragg grating. IEEE Sens. J. 2002, 2, 482–487. [Google Scholar] [CrossRef]
  16. Sohn, K.R.; Kim, K.T.; Song, J.W. Optical fiber sensor for water detection using a side-polished fiber coupler with a planar glass-overlay-waveguide. Sens. Actuators A Phys. 2002, 101, 137–142. [Google Scholar] [CrossRef]
  17. Nath, P.; Singh, H.K.; Datta, P.; Sarma, K.C. All-fiber optic sensor for measurement of liquid refractive index. Sens. Actuators A Phys. 2008, 148, 16–18. [Google Scholar] [CrossRef]
  18. Kazemi, A.A.; Yang, C.; Chen, S. Fiber optic cryogenic liquid level detection system for space applications. In Photonics in the Transportation Industry: Auto to Aerospace II; International Society for Optics and Photonics: Bellingham, WA, USA, 2009. [Google Scholar]
  19. Key, K.H.; Kim, H.P.; Sohn, K.R. Development of optical dual-sensors for submersion monitoring using zigbee-based wireless sensor networks. J. Sens. Sci. Technol. 2010, 19, 184–190. [Google Scholar] [CrossRef] [Green Version]
  20. Shao, M.; Qiao, X.; Zhao, X.; Zhang, Y.; Fu, H. Liquid level sensor using fiber Bragg grating assisted by multimode fiber core. IEEE Sens. J. 2016, 16, 2374–2379. [Google Scholar] [CrossRef]
  21. MathWorks. Available online: https://www.mathworks.com/products/MATLAB.html (accessed on 28 July 2019).
  22. OMEGA™. Available online: https://www.omega.co.uk/pptst/1PT100G_RTD_ELEMENTS.html (accessed on 28 July 2019).
  23. Wikipedia. Available online: https://en.wikipedia.org/wiki/Lumped element model (accessed on 28 July 2019).
  24. Wikipedia. Available online: https://en.wikipedia.org/wiki/Newton%27s_law_of_cooling (accessed on 28 July 2019).
  25. Wikipedia. Available online: https://en.wikipedia.org/wiki/Thermal_conduction (accessed on 28 July 2019).
  26. Wikipedia. Available online: http://everyspec.com/MIL-STD/MIL-STD-0800-0899/MIL-STD-810G_12306/ (accessed on 28 July 2019).
  27. ESPEC CORP. Available online: https://www.espec.co.jp/english/ (accessed on 28 July 2019).
  28. World Sea Temperatures. Available online: https://www.seatemperature.org (accessed on 28 July 2019).
  29. K-water. Available online: http://english.kwater.or.kr/eng/main.do (accessed on 28 July 2019).
  30. NASA’s ‘Salt of the Earth’ Aquarius Reveals First Map. Available online: https://www.nasa.gov/mission_pages/aquarius/multimedia/gallery/pia14786.html (accessed on 28 July 2019).
Figure 1. Mechanical design of the proposed submersion sensor.
Figure 1. Mechanical design of the proposed submersion sensor.
Sensors 19 04310 g001
Figure 2. (a) Heat transfer diagram of a heating resistance temperature detector; (b) signal conditioning circuit.
Figure 2. (a) Heat transfer diagram of a heating resistance temperature detector; (b) signal conditioning circuit.
Sensors 19 04310 g002
Figure 3. Temperature of resistance temperature detectors (RTDs) when surrounded by air.
Figure 3. Temperature of resistance temperature detectors (RTDs) when surrounded by air.
Sensors 19 04310 g003
Figure 4. Temperature of resistance temperature detectors (RTDs) under submerged conditions (atmosphere + 20 °C / water + 20 °C).
Figure 4. Temperature of resistance temperature detectors (RTDs) under submerged conditions (atmosphere + 20 °C / water + 20 °C).
Sensors 19 04310 g004
Figure 5. Temperature of resistance temperature detectors(RTDs) under submerged conditions (atmosphere −40 °C/ water +20 °C).
Figure 5. Temperature of resistance temperature detectors(RTDs) under submerged conditions (atmosphere −40 °C/ water +20 °C).
Sensors 19 04310 g005
Figure 6. Temperature of resistance temperature detectors (RTDs) under submerged conditions (atmosphere +100 °C / water +20 °C).
Figure 6. Temperature of resistance temperature detectors (RTDs) under submerged conditions (atmosphere +100 °C / water +20 °C).
Sensors 19 04310 g006
Figure 7. Simplified circuit to detect submersion.
Figure 7. Simplified circuit to detect submersion.
Sensors 19 04310 g007
Figure 8. Part of the self-diagnosis circuit for detecting malfunctions.
Figure 8. Part of the self-diagnosis circuit for detecting malfunctions.
Sensors 19 04310 g008
Figure 9. Test equipment of the submersion sensor.
Figure 9. Test equipment of the submersion sensor.
Sensors 19 04310 g009
Figure 10. Test setup of the submersion sensor.
Figure 10. Test setup of the submersion sensor.
Sensors 19 04310 g010
Figure 11. Test profile of humidity.
Figure 11. Test profile of humidity.
Sensors 19 04310 g011
Figure 12. Experimental setup for the repeatability test.
Figure 12. Experimental setup for the repeatability test.
Sensors 19 04310 g012
Table 1. Specification of the resistance temperature detectors (1PT100GO1020).
Table 1. Specification of the resistance temperature detectors (1PT100GO1020).
Operating Temperature Range [°C,F]Shape: Cylinder Size: [mm]Self-Heating Error [°C/mW] @ Air Flow V = 1 [m/s]Response Time [Second]
Flowing Water
V = 0.4 [m/s]
Flowing Air
V = 1 [m/s]
Response 50%Response 90%Response 50%Response 90%
−220–400,
(−365–750)
Radius: 1
Length: 10
0.260.140.357.021.0
Table 2. Values of the Biot number in air and submersion conditions.
Table 2. Values of the Biot number in air and submersion conditions.
ConditionAir = 10
[W/m2 °C]
Air = 30
[W/m2 °C]
Air = 50
[W/m2 °C]
Water = 500
[W/m2 °C]
Water = 700
[W/m2 °C]
k [W/m2 °C]8.58.58.51515
B i 0.00053470.00160410.00267350.015150.02121
Table 3. Simulation and measurement results of the temperature and voltage of resistance temperature detectors (RTDs) with the signal conditioning circuit in Figure 6.
Table 3. Simulation and measurement results of the temperature and voltage of resistance temperature detectors (RTDs) with the signal conditioning circuit in Figure 6.
ConditionSimulation ResultsMeasurement Results
Before SubmersionAfter SubmersionDifferenceBefore SubmersionAfter SubmersionDifference
temperature of R T D H [°C]55391655.639.316.3
resistance of R T D H [Ω]121.17115.016.16121.40115.136.27
voltage of R T D H [V] @   V 1 7.026.900.127.046.920.11
temperature of R T D A [°C]20.020.0020.620.30.3
resistance of R T D A [Ω]107.7107.70107.9107.80.1
voltage of R T D A [V] @   V 2 3.253.2503.253.250
output of U 4 D [V] @ V 3 5.755.7505.775.770
output of U 5 A [V] @ V 4 1.261.140.121.261.150.11
output of U 5 B [V] @ V 5 7.206.541.127.236.560.67
reference of U 5 B [V]6.86.806.786.780
sensing margin [V]0.400.26-0.450.21-
output of U 5 c [V] @ V 6 0 V (“low”)12 V (“high”)-0 V (“low”)12 V (“high”)-
Sensing output [V]0 V (“low”)12 V (“high”)-0 V (“low”)12 V (“high”)-
Table 4. Logic table of the self-diagnosis circuit.
Table 4. Logic table of the self-diagnosis circuit.
Condition Temp. of RTDH
[°C]
Resistanceof RTDH
[Ω]
V1
[V]
V7
[V]
V8
[V]
Self-Diagnosis OutputSensing OutputFinal Sensing Output
Normallow temp−2092.36.428.356.1312 V0 V0 V
high temp+220184.76.428.356.1312 V12 V12 V
malfunctionRTD open--128.356.130 V12 V0 V
RTD short--08.356.130 V12 V0 V
Table 5. Humidity test result.
Table 5. Humidity test result.
Test ConditionSample 1Sample 2Sample 3Sample 4
MIL-STD-801G
Method 507.5
Procedure Ⅱ
submersion not detectedsubmersion not detectedsubmersion not detectedsubmersion not detected
Table 6. Rainfall test results.
Table 6. Rainfall test results.
Test ConditionDescriptionSample 1Sample 2Sample 3Sample 4
MIL-STD-801G
Method 506.4
Procedure Ⅰ
rain and blowing rainsubmersion not detectedsubmersion not detectedsubmersion not detectedsubmersion not detected
MIL-STD-801G
Method 506.4
Procedure Ⅱ
Dripsubmersion not detectedsubmersion not detectedsubmersion not detectedsubmersion not detected
Table 7. Water Temperature.
Table 7. Water Temperature.
WaterMinimum Temperature [°C]Maximum Temperature [°C]psu (Practical Salinity Unit)
Fresh water129-
Salt water−1.731.7maximum 40 psu [30]
Table 8. Submersion sensing response time of fresh-water.
Table 8. Submersion sensing response time of fresh-water.
Air Temperature
[°C]
Fresh-Water Temperature
[°C]
Submersion DirectionSample 1
[Second]
Sample 2
[Second]
Sample 3
[Second]
Sample 4
[Second]
−4020° (vertical)0.520.560.510.50
45°0.340.500.520.55
90° (horizontal)0.400.380.490.42
320° (vertical)0.470.460.530.43
45°0.480.390.320.45
90° (horizontal)0.390.420.430.39
020° (vertical)0.410.590.540.40
45°0.330.450.470.44
90° (horizontal)0.350.420.400.39
320° (vertical)0.560.460.510.53
45°0.450.430.470.39
90° (horizontal)0.370.390.440.37
7120° (vertical)0.490.550.500.58
45°0.420.440.490.45
90° (horizontal)0.350.420.380.41
320° (vertical)0.510.590.580.53
45°0.490.450.480.49
90° (horizontal)0.360.440.430.38
Table 9. Submersion sensing response time of salt-water (40 psu).
Table 9. Submersion sensing response time of salt-water (40 psu).
Air Temperature
[°C]
Salt-Water Temperature
[°C]
Submersion DirectionSample 1
[Second]
Sample 2
[Second]
Sample 3
[Second]
Sample 4
[Second]
−4020° (vertical)0.550.510.440.57
45°0.430.490.380.44
90° (horizontal)0.320.360.350.35
320° (vertical)0.520.480.560.55
45°0.440.470.390.45
90° (horizontal)0.390.410.380.40
020° (vertical)0.500.470.580.41
45°0.430.550.460.44
90° (horizontal)0.390.400.300.35
320° (vertical)0.540.490.460.54
45°0.480.410.390.48
90° (horizontal)0.450.330.380.32
7120° (vertical)0.560.470.490.56
45°0.510.420.530.37
90° (horizontal)0.480.350.360.35
320° (vertical)0.380.520.510.54
45°0.400.430.460.43
90° (horizontal)0.350.380.370.35
Table 10. Self-diagnosis test results.
Table 10. Self-diagnosis test results.
Test ConditionSelf-Diagnosis Output (Normal: 12 [V], Detected: 0 [V])
Sample1Sample2Sample3Sample4
Normal12 V12 V12 V12 V
R T D H open0 V0 V0 V0 V
R T D H short0 V0 V0 V0 V
R T D A open0 V0 V0 V0 V
R T D A short0 V0 V0 V0 V
Table 11. Repeatability test result (atmosphere 20 °C).
Table 11. Repeatability test result (atmosphere 20 °C).
WaterTemperature [°C]Submersion DirectionRepeatability Test ConditionTest Result
Fresh-water20°, 90°1000 timesPass
320°, 90°1000 timesPass
Salt-water with 40 psu20°, 90°1000 timesPass
320°, 90°1000 timesPass
Table 12. Comparison with existing submersion sensors.
Table 12. Comparison with existing submersion sensors.
Malfunction PossibilitySelf-DiagnosisResponse Time [sec]Electro Magnetic Interference (EMI) Influence
Proposed submersion sensornonepossible0.4–0.6none
Resistive electrode submersion sensormediumimpossibleless than 0.1none
Capacitive submersion sensorhighpossibleless than 0.1yes
Optical submersion sensorhighpossibleless than 0.1none

Share and Cite

MDPI and ACS Style

Lee, Y.; Lee, Y.S. A New Submersion Detection Sensor Using Two Resistance Temperature Detectors Operating on the Thermal Equilibrium Principle. Sensors 2019, 19, 4310. https://doi.org/10.3390/s19194310

AMA Style

Lee Y, Lee YS. A New Submersion Detection Sensor Using Two Resistance Temperature Detectors Operating on the Thermal Equilibrium Principle. Sensors. 2019; 19(19):4310. https://doi.org/10.3390/s19194310

Chicago/Turabian Style

Lee, Youngjun, and Young Sam Lee. 2019. "A New Submersion Detection Sensor Using Two Resistance Temperature Detectors Operating on the Thermal Equilibrium Principle" Sensors 19, no. 19: 4310. https://doi.org/10.3390/s19194310

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop