Development and Performance Characteristics of Personal Gamma Spectrometer for Radiation Monitoring Applications

Department of Physics, University of Myongji, Yongin 449-728, Korea
*
Author to whom correspondence should be addressed.
Sensors 2016, 16(6), 919; https://doi.org/10.3390/s16060919
Submission received: 9 April 2016 / Revised: 25 May 2016 / Accepted: 17 June 2016 / Published: 21 June 2016
(This article belongs to the Section Physical Sensors)

Abstract

:
In this study, a personal gamma (γ) spectrometer was developed for use in applications in various fields, such as homeland security and environmental radiation monitoring systems. The prototype consisted of a 3 × 3 × 20 mm3 Ce-doped Gd–Al–Ga–garnet (Ce:GAGG) crystal that was coupled to a Si photomultiplier (SiPM) to measure γ radiation. The γ spectrometer could be accessed remotely via a mobile device. At room temperature, the implemented Ce:GAGG-SiPM spectrometer achieved energy resolutions of 13.5%, 6.9%, 5.8%, and 2.3% for 133Ba at 0.356 MeV, 22Na at 0.511 MeV, 137Cs at 0.662 MeV, and 60Co at 1.33 MeV, respectively. It consumed only about 2.7 W of power, had a mass of just 340 g (including the battery), and measured only 5.0 × 7.0 cm2.

1. Introduction

Gamma (γ)-ray spectrometry analysis methods can be classified based on whether they involve direct or indirect detection, according to the manner in which the γ-rays are converted into electrical signals. The direct detection methods employ semiconductor devices that generate electrical signals immediately, with no intermediate stages, through γ-ray absorption. The indirect detection methods involve scintillation detection systems; in such a system, a photoelectric device detects electrical signals generated by photons with wavelengths in the visible range that are produced through interactions between the γ-rays and the scintillator.
CdZnTe, CdTe, and HgI2 are primarily used in room temperature semiconductor devices employing direct detection methods [1]. These materials are preferred in γ-ray spectrometry analysis because they yield energy resolutions superior to those achievable with scintillator crystals. However, the maintenance of these materials is more expensive than that of scintillator crystals, and their detection efficiencies and physical rigidities are very low [2]. Currently, compound semiconductor materials are also more expensive than scintillator crystals.
NaI(Tl), CsI(Tl), and LYSO(Ce) are the primary scintillator crystals used in indirect detection. These materials yield energy resolutions lower than those of semiconductors, but they are more sensitive. Because of the specific characteristics and limitations of each of these types of materials and the corresponding analysis methods, further development of detectors that combine scintillators and semiconductors is necessary and ongoing [3]. Thus, scintillator detectors remain promising candidates for use in homeland security (HLS) and environmental radiation monitoring systems (ERMSs).
In this study, a personal γ spectrometer was developed, and its feasibility was evaluated by analyzing its energy resolution using standard γ-ray sources. The device described herein, called the “Personal-Spect”, essentially consists of a scintillation detector, signal processing electronics, a voltage supply, signal analysis components, and display units.

2. Materials and Methods

An S12572-100C (photo-sensitive area: 3 × 3 mm2, Hamamatsu Photonics Co., Hamamatsu, Japan) Hamamatsu single Si photomultiplier (SiPM) was used in the Personal-Spect to minimize the overall system dimensions (Table 1).
The Personal-Spect prototype contained a 3 × 3 × 20 mm3 Ce-doped Gd–Al–Ga–garnet (Ce:GAGG) crystal, which was fabricated via the Czochralski crystal growth method using a seed crystal (Furukawa Denshi Co., Tokyo, Japan) and contained approximately 1% Ce. Ce:GAGG has a peak wavelength of 520–530 nm, which fits the spectral response range of the SiPM. Also, since this material has a density of 6.63 g·cm−3, its stopping power is high. Furthermore, its decay time (90 ns) is shorter than those of the conventional scintillators (CsI:Tl and NaI:Tl) [4]. In order to maximize the light output from the Ce:GAGG and to match it to the SiPM photosensitive area, the crystal geometry was optimized using Monte Carlo n-particle extended code [5]. Figure 1 shows the detector head of the Personal-Spect. The SiPM was coupled to a Ce:GAGG scintillator covered with five layers of a white diffusive (Teflon) reflector in order to optimize the scintillation light collection. The optical coupling was achieved by using optical grease (n = 1.465) as the coupling medium. The coupled detector was covered and sealed inside an Al tube to prevent background noise from external light.
Figure 2 shows the Personal-Spect structure and the overall spectroscopy system designed in this study. The fabricated Ce:GAGG-SiPM detector and electronics were installed in an Al shielding box to prevent background noise from external light.
Previously, charge-sensitive preamplifiers that convert scintillation detector charge signals into voltage signals, as well as shaping amplifiers that amplify converted voltage signals and simultaneously shape them into Gaussian forms, were designed for use in signal processing in γ-ray spectrometry [2]. However, such signal processing procedures cause signal attenuation due to impedance mismatch, decrease the signal-to-noise ratio through the amplification and shaping processes, and cause output signal loss due to detected signal overlap with the dead time.
Therefore, in this study, the signal processing stage was excluded due to the high electron amplification gains of SiPMs, and only the driver circuit of the scintillation detector was designed and built to analyze the output signal of the circuit. The left-hand side of Figure 2 shows a 700 mV pulse signal with a 2 μs pulse width that was generated by a 22Na source and measured using the driver circuit.
The voltage supply consisted of a main power supply and a voltage-boosting unit. The voltage-boosting unit was designed to increase an input voltage of 7.2 V up to 73 V ± 1.5 V, which is the input voltage of the SiPM. A Li-poly battery (7.4 V, 850 mAh) was used as the main power supply with a DC–DC converter module (UltraVolt, Inc., Ronkonkoma, NY, USA). This Personal-Spect had dimensions of 5.0 × 7.0 cm2, and it could run for more than 3 h with an average power consumption of 2.7 W.
The signal analysis was performed using a tablet computer, which constituted the display unit, and a miniature 4906-channel K102 Multichannel Analyser (Kromek Ltd., Sedgefield, UK).
In the experiments described herein, a standard disc-type radiation source was used to evaluate the Personal-Spect. Several standard γ-ray sources were used: 133Ba: 0.356 MeV, 22Na: 0.511 MeV and 1.27 MeV, 137Cs: 0.662 MeV, and 60Co: 1.17 MeV and 1.33 MeV (Spectrum Techniques, Oak Ridge, TN, USA). Each source had an activity of 1 μCi and was placed 10 mm from the surface of the developed device, and the live time was set to 300 s for each of the experiments.

3. Results and Discussion

Figure 3 shows the energy spectra obtained from the Personal-Spect. The 0.356 MeV peak of the 133Ba source exhibits an energy resolution of 13.5%. For the 0.511 MeV peak in the 22Na spectrum that was generated through positron annihilation during β decay, an energy resolution of 6.9% was measured. The 0.662 MeV peak in the 137Cs spectrum yielded an energy resolution of 5.8%, while for 60Co, an energy resolution of 2.3% was obtained at an energy of 1.33 MeV, and peaks at 1.17 MeV and 1.33 MeV are clearly observable [6].
Figure 4 depicts the Personal-Spect energy calibration line, which was used to evaluate the linearity of the γ-ray energy. The points on this graph were obtained and the line was estimated based on the measured 133Ba, 22Na, 137Cs, and 60Co photopeak energies and channels.
The fit shown in Figure 4 confirms the linearity of the relationship between the γ-ray energy and peak channel number for the energy range from 0.356 MeV to 1.33 MeV. The R2 value of the linear relationship was found to be 0.9924.
Figure 5 depicts the Personal-Spect detection efficiency curve, which was used to evaluate the full-energy peak efficiency (FEPE) for γ-rays of various energies. The FEPE is the efficiency with which γ-rays of a given energy produce full-energy peak pulses, rather than pulses of other sizes.
The efficiency curve in Figure 5 shows a relatively linear decrease within the high-energy range from 0.356 MeV to 1.33 MeV. Clearly, to be useful, the detector must be capable of absorbing a large fraction of the γ-ray energy. This capability can be realized by using a detector of suitable size or by choosing a scintillator material with a sufficiently high Z [7].
Figure 6 depicts the γ-ray energy spectrum for a mixed source containing 133Ba, 22Na, 137Cs, and 60Co [8]. The results presented in this figure confirm that γ-ray spectrometry analysis is possible for a mixed source based on the γ-ray energy peaks in the range from 0.3 MeV to 1.3 MeV.

4. Conclusions

A personal γ spectrometer was fabricated and evaluated to assess its usability in radiation monitoring applications. This device is low-power, requiring only 2.7 W from a single 7.4 V input, has a mass of 340 g (including the battery), and has dimensions of 5.0 × 7.0 cm2. It can be accessed remotely via a mobile device. This device exhibited higher energy resolutions, as confirmed by the spectra obtained for several standard γ sources. The features of the Personal-Spect were characterized, and the best full-width at half-maximum energy resolution was found to be 5.8% at 662 keV. Thus, the Personal-Spect is expected to be applicable in fields such as HLS and ERMS as a personal γ spectrometer and to replace expensive semiconductor detectors.

Acknowledgments

This work was supported by the 2015 Research Fund of Myongji University.

Author Contributions

Hye Min Park made substantial contributions in design of the gamma spectrometer, all experiments, data analysis, results interpretation and manuscript preparation. Koan Sik Joo made significant contributions in experimental design, data analysis, results interpretation, and manuscript preparation.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Sordo, S.D.; Abbene, L.; Caroli, E.; Mancini, A.M.; Zappettini, A.; Ubertini, P. Progress in the development of CdTe and CdZnTe semiconductor radiation detectors for astrophysical and medical applications. Sensors 2009, 9, 3491–3526. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Kim, H.S.; Ha, J.H.; Park, S.H.; Cho, S.Y.; Kim, Y.K. Fabrication and performance characteristics of a CsI(Tl)/PIN diode radiation sensor for industrial applications. Nucl. Instrum. Appl. Radiat. Isot. 2009, 67, 1463–1465. [Google Scholar] [CrossRef] [PubMed]
  3. Sanaei, B.; Baei, M.T.; Sayyed-Alangi, S.Z. Characterization of a new silicon photomultiplier in comparison with a conventional photomultiplier tube. J. Mod. Phys. 2015, 6, 425–433. [Google Scholar] [CrossRef]
  4. Yeom, J.Y.; Yamamoto, S.; Derenzo, S.E.; Spanoudaki, V.C.; Kamada, K.; Endo, T.; Levin, C.S. First performance results of Ce:GAGG scintillation crystals with silicon photomultipliers. IEEE Trans. Nucl. Sci. 2013, 60, 988–992. [Google Scholar] [CrossRef]
  5. Park, H.M.; Joo, K.S. Performance characteristics of a silicon photomultiplier based compact radiation detector for Homeland Security applications. Nucl. Instrum. Meth. Phys. Res. A 2015, 781, 1–5. [Google Scholar] [CrossRef]
  6. David, S.; Georgiou, M.; Fysikopoulos, E.; Loudos, G. Evaluation of a SiPM array coupled to a Gd3Al2Ga3O12:Ce (GAGG:Ce) discrete scintillator. Phys. Med. 2015, 31, 763–766. [Google Scholar] [CrossRef] [PubMed]
  7. Canberra. Gamma and X-ray Detection. Available online: http://www.canberra.com/literature/fundamental-principles/pdf/Gamma-Xray-Detection.pdf (accessed on 20 June 2016).
  8. Osovizky, A.; Ginzburg, D.; Manor, A.; Seif, R.; Ghelman, M.; Cohen-Zada, I.; Ellenbogen, M.; Bronfenmakher, V.; Pushkarsky, V.; Gonen, E.; et al. SENTIRAD—An innovative personal radiation detector based on a scintillation detector and a silicon photomultiplier. Nucl. Instrum. Meth. Phys. Res. A 2011, 652, 41–44. [Google Scholar] [CrossRef]
Figure 1. Structure of Ce-doped Gd–Al–Ga–garnet (Ce:GAGG) detector head (left); 3 × 3 mm2 Hamamatsu S12572-100C SiPM and 3 × 3 × 20 mm3 Ce:GAGG crystal (right).
Figure 1. Structure of Ce-doped Gd–Al–Ga–garnet (Ce:GAGG) detector head (left); 3 × 3 mm2 Hamamatsu S12572-100C SiPM and 3 × 3 × 20 mm3 Ce:GAGG crystal (right).
Sensors 16 00919 g001
Figure 2. (A) Personal-Spect structure and (B) overall spectroscopy system.
Figure 2. (A) Personal-Spect structure and (B) overall spectroscopy system.
Sensors 16 00919 g002
Figure 3. Measured energy spectra for 133Ba (0.356 MeV), 22Na (0.511 MeV), 137Cs (0.662 MeV), and 60Co (1.33 MeV).
Figure 3. Measured energy spectra for 133Ba (0.356 MeV), 22Na (0.511 MeV), 137Cs (0.662 MeV), and 60Co (1.33 MeV).
Sensors 16 00919 g003
Figure 4. Linear fit of measured 133Ba, 22Na, 137Cs, and 60Co photopeak energies and channels.
Figure 4. Linear fit of measured 133Ba, 22Na, 137Cs, and 60Co photopeak energies and channels.
Sensors 16 00919 g004
Figure 5. Full-energy peak efficiency in the 0.356–1.3 MeV energy range.
Figure 5. Full-energy peak efficiency in the 0.356–1.3 MeV energy range.
Sensors 16 00919 g005
Figure 6. Energy spectra emitted from 133Ba (0.303 MeV and 356 MeV), 22Na (0.511 MeV), 137Cs (0.662 MeV), and 60Co (1.33 MeV).
Figure 6. Energy spectra emitted from 133Ba (0.303 MeV and 356 MeV), 22Na (0.511 MeV), 137Cs (0.662 MeV), and 60Co (1.33 MeV).
Sensors 16 00919 g006
Table 1. Specifications of Si photomultiplier (SiPM) used in this study.
Table 1. Specifications of Si photomultiplier (SiPM) used in this study.
ParameterValue
Photosensitive area3 × 3 mm2
Number of pixels900
Spectral response range320–900 nm
Peak photon detection efficiency (at 450 nm)35%
Bias voltageVbr + 1.4 V
Breakdown voltage65 V ± 10 V
Gain2.8 × 106
Operating temperature−20 °C–40 °C

Share and Cite

MDPI and ACS Style

Park, H.M.; Joo, K.S. Development and Performance Characteristics of Personal Gamma Spectrometer for Radiation Monitoring Applications. Sensors 2016, 16, 919. https://doi.org/10.3390/s16060919

AMA Style

Park HM, Joo KS. Development and Performance Characteristics of Personal Gamma Spectrometer for Radiation Monitoring Applications. Sensors. 2016; 16(6):919. https://doi.org/10.3390/s16060919

Chicago/Turabian Style

Park, Hye Min, and Koan Sik Joo. 2016. "Development and Performance Characteristics of Personal Gamma Spectrometer for Radiation Monitoring Applications" Sensors 16, no. 6: 919. https://doi.org/10.3390/s16060919

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop