Next Article in Journal
Visual Tracking Based on Extreme Learning Machine and Sparse Representation
Next Article in Special Issue
An Electrochemical NO2 Sensor Based on Ionic Liquid: Influence of the Morphology of the Polymer Electrolyte on Sensor Sensitivity
Previous Article in Journal
Harvesting Entropy for Random Number Generation for Internet of Things Constrained Devices Using On-Board Sensors
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Sensors for Highly Toxic Gases: Methylamine and Hydrogen Chloride Detection at Low Concentrations in an Ionic Liquid on Pt Screen Printed Electrodes

by
Krishnan Murugappan
and
Debbie S. Silvester
*
Nanochemistry Research Institute, Department of Chemistry, Curtin University, GPO Box U1987, Perth 6845, WA, Australia
*
Author to whom correspondence should be addressed.
Present address: Department of Materials, Parks Rd, Oxford OX1-3PH, UK.
Sensors 2015, 15(10), 26866-26876; https://doi.org/10.3390/s151026866
Submission received: 24 September 2015 / Revised: 15 October 2015 / Accepted: 19 October 2015 / Published: 22 October 2015
(This article belongs to the Special Issue Ionic Liquids)

Abstract

:
Commercially available Pt screen printed electrodes (SPEs) have been employed as possible electrode materials for methylamine (MA) and hydrogen chloride (HCl) gas detection. The room temperature ionic liquid (RTIL) 1-ethyl-3-methylimidazolium bis(trifluoromethylsulfonyl)imide ([C2mim][NTf2]) was used as a solvent and the electrochemical behaviour of both gases was first examined using cyclic voltammetry. The reaction mechanism appears to be the same on Pt SPEs as on Pt microelectrodes. Furthermore, the analytical utility was studied to understand the behaviour of these highly toxic gases at low concentrations on SPEs, with calibration graphs obtained from 10 to 80 ppm. Three different electrochemical techniques were employed: linear sweep voltammetry (LSV), differential pulse voltammetry (DPV) and square wave voltammetry (SWV), with no significant differences in the limits of detection (LODs) between the techniques (LODs were between 1.4 to 3.6 ppm for all three techniques for both gases). The LODs achieved on Pt SPEs were lower than the current Occupational Safety and Health Administration Permissible Exposure Limit (OSHA PEL) limits of the two gases (5 ppm for HCl and 10 ppm for MA), suggesting that Pt SPEs can successfully be combined with RTILs to be used as cheap alternatives for amperometric gas sensing in applications where these toxic gases may be released.

1. Introduction

Miniaturization has been the driving force for many industries and it is likely that the electrochemical sensor industry is not far behind. Planar devices (e.g., screen printed electrodes, or SPEs) where all three electrodes are contained within a very small area have become increasingly more popular for use as electrode materials in amperometric gas sensors (AGSs) [1,2,3,4,5,6], but they have yet to become commercialised (in AGSs). SPEs contain various materials (e.g., graphite, carbon black and polymeric binders) that are squeezed through a mesh screen which defines the shape and size of the electrode [7]. They are relatively cheap and portable, and can be combined with non-volatile and highly thermally stable room temperature ionic liquid (RTIL) solvents to make more robust sensors. Previous work on ammonia [1], chlorine [8] and oxygen [2] gas-sensing on SPEs has already been reported and shows reasonable limits of detection (LODs), but not as low as the exposure limits for ammonia and chlorine, since lower concentration ranges were not studied. The real question is: can RTIL solvents be used with screen printed electrodes in amperometric gas sensors to detect analyte gases at levels at or below the exposure limits?
Many studies on gases in RTILs have so far focussed on higher concentrations with only one group to report ammonia gas-sensing in the ppm range (0 to 40 ppm), with LODs between 0.2 and 2 ppm in a range of RTILs [9]. The two gases that are the focus of this work (MA and HCl) are highly toxic at low ppm concentrations. For example, methylamine is a colourless and toxic gas with a distinct fish-like odour at concentrations above 100 ppm [10]. The current (USA) Occupational Safety and Health Administration Permissible Exposure Limit (OSHA PEL) is 10 ppm in the gas phase. HCl is highly corrosive and, in the presence of moisture, forms hydrochloric acid which causes burns and irritation to the skin, as well as corrosion to metals (e.g., iron) [11]. The current (USA) OSHA PEL of HCl is 5 ppm in the gas phase. Due to their high toxicity, it is therefore essential to be able to monitor and detect methylamine and hydrogen chloride gases at low ppm concentrations.
Current technology to detect methylamine and hydrogen chloride gases exists, but is limited. To the best of our knowledge, there are no commercial AGSs for methylamine, but colorimetric gas detectors (Sensidyne, 1–20 ppm, operating temperature up to 40 °C) and photoionization detectors (Mil-Ram Technology, 0–200 ppm, operating temperature up to 55 °C) are available. For HCl, companies such as Membrapor and Alphasense manufacture AGSs in the range of 0–20 ppm, but the operating temperatures do not go above 45 °C. This clearly shows that RTIL-based sensors have the possibility to be used in sensors over a much wider temperature range due to their non-volatility and high thermal stability.
In previous works the electrochemical behaviours of methylamine [12] and hydrogen chloride [13] gases have been investigated in room temperature ionic liquids (RTILs) on conventional Pt micro/macrodisk electrodes. In this work, the electrode reaction mechanism will first be investigated to see if the behaviour of these toxic gases on SPEs is similar to conventional electrodes. Then various concentrations will be studied to obtain analytical parameters (LODs and sensitivities) so that a direct comparison of LODs can be made with conventional Pt microdisk electrodes in the RTIL [C2mim][NTf2]. It is important to note that, in previous work, the mechanism for reduction of oxygen on SPEs was different from that on microelectrodes in this RTIL (reaction of the reduction product with the materials in the SPE paste) [2] and indicated that the detection of oxygen on Pt SPEs may not be feasible for long-term studies. This work will help determine whether RTILs and SPEs can be used in the design of cheap, portable and robust gas sensors for toxic gases such as methylamine and hydrogen chloride.

2. Experimental Section

2.1. Chemical Reagents

First 1-ethyl-3-methylimidazolium bis(trifluoromethylsulfonyl)imide ([C2mim][NTf2]) was purchased from Merck Millipore (Victoria, Australia). Ultra-pure water with a resistance of 18.2 MΩ·cm prepared by an ultra-pure laboratory water purification system (Millipore Pty Ltd., North Ryde, NSW, Australia) and acetonitrile (MeCN, Sigma-Aldrich, 99.8%) were used for washing the electrodes after use with RTILs. Methylamine (0.2% nitrogen fill) and hydrogen chloride gas (0.2% nitrogen fill) was purchased from CAC gases (Auburn, NSW, Australia). High purity nitrogen gas (99.9%) was from BOC gases (North Ryde, NSW, Australia) and used as a carrier gas.

2.2. Electrochemical Experiments

All electrochemical experiments were performed using a PGSTAT101 Autolab potentiostat (Eco Chemie, Utrecht, The Netherlands) interfaced to a PC with NOVA version 1.9 software. Pt SPEs employed in this study were purchased from DropSens (DRP-550, Oviedo, Spain). The SPEs consist of a three-electrode setup with a Pt working, Pt counter and Ag reference electrode. The SPEs were housed in a glass “T-cell” previously used to study ammonia [1] and oxygen [2] gases. Before the introduction of gas, the cell was purged under nitrogen to remove any impurities present in the RTIL (volume of 10 µL). Although it is widely accepted that vacuum techniques are superior to purge the RTIL of atmospheric impurities found in RTILs [14,15,16], it was not possible to use the vacuum due to the violent degassing in the current experimental set-up (where a layer of RTIL is spread over the flat surface of the SPE, in contrast with other work [14] where the RTIL is neatly contained within the cap of a pipette tip). The SPEs were used “as is” without any activation/precleaning as this represents the conditions required in the real world.

2.3. Gas-Mixing Setup

A similar gas-mixing setup as used by Lee et al. [2] was employed to obtain different concentrations of methylamine and hydrogen chloride. The % (or ppm) concentration of methylamine or hydrogen chloride that was introduced into the T-cell was calculated using the relative flow rates of the two flow meters. A digital flowmeter (0–1.2 L/min, John Morris Scientific, NSW, Australia) was used for the nitrogen gas and an analogue flowmeter (0–60 cm3/min, Dwyer, NSW, Australia) was used for hydrogen chloride/methylamine gas. Approximately 30 min was allowed for saturation to occur for each specific concentration and a stable response was determined when the current was identical in two cyclic voltammograms (CVs) taken 10 min apart. Response times were not investigated.

3. Results and Discussion

All experiments were conducted with the RTIL [C2mim][NTf2], which has a high intrinsic conductivity (8.8 mS·cm−1) and is one of the least viscous RTILs available (34 cP at 293 K) [17], suggesting a faster gas saturation time. It has also been observed previously (on a platinum microelectrode) that a linear current response vs. concentration can be obtained in [C2mim][NTf2] for one of the gases under study (methylamine [12]). We are currently performing mechanistic studies for HCl on a Pt microelectrode in a range of RTILs, which will be the focus of a future publication.

3.1. Methylamine Gas

3.1.1. Electrochemical Oxidation of Methylamine Gas on Pt SPEs

Figure 1 shows the electrochemical oxidation of 0.15% methylamine gas on a platinum SPE in the RTIL [C2mim][NTf2]. The mechanism for the oxidation reaction has been elucidated in our previous work [12] and therefore will not be discussed in detail here. In brief, methylamine is oxidised (at ca. 1.5 V) to form ammonia which itself undergoes oxidation at that potential. The combination of these two processes produces the oxidation peak (labelled peak I on Figure 1), which will be used as the analytical response. The oxidation peak (I) is followed by two reduction peaks (II and III) at −0.2 V and −0.9 V and a subsequent oxidation peak (IV) at −0.4 V. These processes correspond to: (II) the reduction of solvated protons, (III) the reduction of ammonium ion and (IV) the oxidation of adsorbed hydrogen.
Figure 1. CV for the oxidation of 0.15% methylamine gas on a Pt SPE in [C2mim][NTf2] at a scan rate of 0.1 V/s. Dotted line is the response in the absence of methylamine.
Figure 1. CV for the oxidation of 0.15% methylamine gas on a Pt SPE in [C2mim][NTf2] at a scan rate of 0.1 V/s. Dotted line is the response in the absence of methylamine.
Sensors 15 26866 g001
The blank scan (RTIL in the absence of methylamine, dotted line, Figure 1) also shows an electroactive reduction wave at ca. −1.2 V vs. Ag, which is believed to correspond to an impurity in the RTIL, possibly left over from the synthesis (e.g., unreacted imidazole). This does not affect the current for methylamine oxidation, since peak (I) occurs at much higher potentials (ca. 1.5 V), where the background current is low.
Figure 2 shows cyclic voltammetry for the oxidation of 0.15% methylamine at varying scan rates from 0.05 to 1 V/s when scanned from 0 to 2 V and back to −1.8 V. It can be seen that as the scan rates increase, the peak currents for all four peaks increase. The inset of Figure 2 shows a plot of the peak current for the oxidation peak (I) vs. the square root of the scan rate. A linear response is obtained, suggesting that the process occurring is diffusion-controlled, consistent with that observed previously on a Pt microelectrode [12]. These results are encouraging as it shows that the behaviour on a Pt SPE is very similar to a Pt microelectrode. It has been shown previously that oxygen voltammetry on SPEs in imidazolium RTILs [2] is not ideal, since the voltammetry was very different on a SPE compared to a microelectrode, suggesting that SPEs may not be suitable for oxygen gas detection in these RTILs. In the present case, it has been established that the behaviour on a SPE is similar to that on a microelectrode, so a study of the current response vs. concentration can be carried out for methylamine gas on SPEs.
Figure 2. CV for the oxidation of 0.15% methylamine gas on a Pt SPE in [C2mim][NTf2] at varying scan rates between 0.05 and 1 V/s. The inset shows the plot of peak current vs. the square root of scan rate for peak (I).
Figure 2. CV for the oxidation of 0.15% methylamine gas on a Pt SPE in [C2mim][NTf2] at varying scan rates between 0.05 and 1 V/s. The inset shows the plot of peak current vs. the square root of scan rate for peak (I).
Sensors 15 26866 g002

3.1.2. Analytical Utility of Methylamine Gas on Pt SPEs

To study the analytical response, a lower concentration range for methylamine gas was employed and three different techniques were explored: linear sweep voltammetry (LSV), differential pulse voltammetry (DPV) and square wave voltammetry (SWV). DPV and SWV are often employed as more sensitive techniques compared to cyclic voltammetry/linear sweep voltammetry due to the lower impact of the charging current [18]. Figure 3 shows the oxidation of methylamine gas at five different concentrations (from 10 to 80 ppm) on a Pt SPE in [C2mim][NTf2] at a scan rate of 0.1 V/s using LSV, DPV and SWV for the oxidation peak (I). The optimised parameters for DPV were an amplitude of 0.05 V, modulation time of 0.025 s and interval time of 0.05 s. For SWV, the optimised parameters were an amplitude of 0.05 V and frequency of 20 Hz. It can be seen that as the concentration of methylamine increases, the current for all the oxidation peaks increases. The peak current was plotted against the respective concentrations to obtain calibration graphs, as shown in Figure 3b,d,f. Excellent linearity (R2 > 0.99) was obtained for the concentration range studied. LODs of 3.2, 3.0 and 3.6 ppm, and sensitivities of 4.6 × 10−8, 2.2 × 10−8 and 4.1 × 10−8 A/ppm were obtained for LSV, DPV and SWV, respectively. The LODs were calculated using three times the standard deviation (3σ) of the slope of the calibration line. The LODs are much better than those obtained on a Pt microelectrode (15.7, 14.2 and 15.9 ppm for LSV, DPV and SWV, respectively; these results are shown in Figure S1 of the supporting information). These results are highly encouraging and show that Pt SPEs can be used for methylamine gas detection at low ppm levels, allowing for cheaper and more miniaturised AGSs for methylamine. There has only been one report of LODs under 10 ppm for gases in pure RTILs, which was for ammonia gas on Pt printed electrodes [9], but this is the first time such low LODs for methylamine gas have been reported. The LOD is less than the OSHA PEL limit for methylamine (10 ppm), suggesting that RTIL/SPE-based sensors have the potential to be used in commercial sensors where humans might be exposed to methylamine.
Figure 3. (a) LSV, (c) DPV and (e) SWV for the oxidation of methylamine gas at different concentrations (10, 20, 40, 60 and 80 ppm in the gas phase) on a Pt SPE in [C2mim][NTf2]. Dashed line is the response in the absence of methylamine gas. The corresponding calibration plots obtained for (b) LSV, (d) DPV and (f) SWV using the peak currents from (a,c,e) are also shown.
Figure 3. (a) LSV, (c) DPV and (e) SWV for the oxidation of methylamine gas at different concentrations (10, 20, 40, 60 and 80 ppm in the gas phase) on a Pt SPE in [C2mim][NTf2]. Dashed line is the response in the absence of methylamine gas. The corresponding calibration plots obtained for (b) LSV, (d) DPV and (f) SWV using the peak currents from (a,c,e) are also shown.
Sensors 15 26866 g003

3.2. Hydrogen Chloride

3.2.1. Electrochemical Oxidation of Hydrogen Chloride Gas on Pt SPEs

The electrochemical reaction mechanism for HCl gas has been elucidated previously on a Pt macrodisk electrode [13] and will not be extensively discussed here. It is believed that HCl is in its dissociated form of [HCl2] and H+ when dissolved in RTILs [13].
2 HCl g a s H + + [ HCl 2 ]
Figure 4 shows the electrochemical behaviour of 0.15% hydrogen chloride gas on a Pt SPE scanned from 0 to 2.2 V to −0.9 V and back to 0 V for two CV cycles. On the first cycle, only one oxidation peak (peak (I), at ca. 1.65 V) is observed within the available potential window before solvent breakdown. This likely corresponds to the two-electron oxidation of [HCl2] to form H+ (solvated by the ionic liquid) and chlorine (Cl2) gas. [13] On the reverse sweep, two reduction peaks are observed, corresponding to the reduction of the two electrogenerated products: Cl2 (peak (II), at ca. 0.78 V) and the reduction of solvated protons (peak III, at ca. −0.38 V). Both of these processes have corresponding oxidation peaks: the oxidation of hydrogen (peaks IVa and IVb, two competing processes) [13] and the oxidation of chloride (peak (V), which is present only on the second CV cycle). It is noted that the reduction peak (III) is also present when first scanned in the negative direction (i.e., without first scanning over peak (I)), supporting the above hypothesis (Equation (1)) that solvated protons are present when HCl is dissolved in the RTIL.
Figure 4. CV (two cycles) for the oxidation of 0.15% hydrogen chloride gas (nitrogen fill) on a Pt SPE in [C2mim][NTf2] at a scan rate of 0.1 V/s. Dotted line is the response in the absence of hydrogen chloride gas.
Figure 4. CV (two cycles) for the oxidation of 0.15% hydrogen chloride gas (nitrogen fill) on a Pt SPE in [C2mim][NTf2] at a scan rate of 0.1 V/s. Dotted line is the response in the absence of hydrogen chloride gas.
Sensors 15 26866 g004
Figure 5 shows CV scans for the oxidation of 0.15% hydrogen chloride gas at varying scan rates between 0.05 to 3 V/s when scanned from 0 to 2.2 V to −0.9 V and back to 0 V (for clarity, only the second scan is shown). It is evident that the peak currents for all processes increase with increasing scan rates. Plots of peak current vs. square root of scan rate for peaks (I), (II) and (IVa) and (IVb) are linear, with R2 values > 0.99, suggesting that the electrochemical processes are most likely diffusion controlled. For the proton reduction peak (peak III), there is also a pre-peak that is especially evident at higher scan rates, which has previously been ascribed to a competing reaction for H+ reduction [13,19]. Plots of current vs. square root of scan rate for the proton reduction peak (peak III) and the pre-peak (at ca. 0 V) are approximately linear (R2 ~ 0.98–0.99), suggesting that they are also diffusion controlled, but the currents may be slightly influenced by the presence of other peaks.
Figure 5. CV (second scan) for the oxidation of 0.15% hydrogen chloride gas (nitrogen fill) on a Pt SPE in [C2mim][NTf2] at varying scan rates (0.05–3 V/s).
Figure 5. CV (second scan) for the oxidation of 0.15% hydrogen chloride gas (nitrogen fill) on a Pt SPE in [C2mim][NTf2] at varying scan rates (0.05–3 V/s).
Sensors 15 26866 g005

3.2.2. Analytical Utility of Hydrogen Chloride Gas on Pt SPEs

Since the electrochemical mechanism for HCl gas on a Pt SPE was similar to that on a Pt macrodisk [13] electrode, the current vs. concentration response for hydrogen chloride gas was studied on Pt SPEs. The proton reduction peak (III) was used as the analytical signal since the oxidation peak (I) is complicated by the appearance of peak (V) on the second scan which adds to the current. Figure 6 shows the results from LSV (6a), DPV (6c) and SWV (6e) experiments for five different concentrations of hydrogen chloride on a Pt SPE for the reduction of H+ (peak III). It can be seen that as the concentration increases, the current increases for all three electrochemical techniques. Figure 6b,d,f shows the calibration plots obtained for three techniques and it can be seen that excellent linearity is observed in all plots (R2 > 0.99). LODs of 3.4, 1.4 and 2.4 ppm with sensitivities of 2.9 × 10−8, 2.1 × 10−8 and 3.9 × 10−8 A/ppm were obtained for LSV, DPV and SWV, respectively. The LODs were calculated using three times the standard deviation (3σ) of the slope of the calibration line. This is comparable to LODs of 2.5, 3.7 and 3.7 ppm for LSV, DPV and SWV, respectively, on a Pt microelectrode (see Figure S2 in the supporting information). Again, the voltammetry on the microelectrode was very noisy due to the much lower currents measured. The LODs obtained are below the OSHA PEL limits of 5 ppm for HCl and suggest that cheap Pt SPEs can be used in conjunction with RTILs for the electrochemical detection of low concentrations of hydrogen chloride.
Figure 6. (a) LSV, (c) DPV and (e) SWV for the proton reduction peak at different concentrations of HCl (10, 20, 40, 60 and 80 ppm in the gas phase) on a Pt SPE in [C2mim][NTf2]. Dashed line is the response in the absence of hydrogen chloride gas. The corresponding calibration plots obtained for (b) LSV, (d) DPV and (f) SWV using the peak currents from (a,c,e) are also shown.
Figure 6. (a) LSV, (c) DPV and (e) SWV for the proton reduction peak at different concentrations of HCl (10, 20, 40, 60 and 80 ppm in the gas phase) on a Pt SPE in [C2mim][NTf2]. Dashed line is the response in the absence of hydrogen chloride gas. The corresponding calibration plots obtained for (b) LSV, (d) DPV and (f) SWV using the peak currents from (a,c,e) are also shown.
Sensors 15 26866 g006

4. Conclusions

The electrochemical behaviour of methylamine and hydrogen chloride gases was investigated on Pt SPEs in an ionic liquid. The mechanism of both gases on Pt SPEs was very similar to that on microelectrodes, i.e., no unusual behaviour on the SPEs was seen, which suggests that the materials in the paste/polymeric binder of SPES do not react with the toxic gases in a detrimental way (at least on the timescale of the electrochemical experiments). The low LODs obtained (3.2, 3.0 and 3.6 ppm for MA and 2.5, 3.7 and 3.7 ppm for HCl from LSV, DPV and SWV, respectively) suggest that the SPE/RTIL platform can be used to detect these highly toxic gases at low levels in real environments.

Supplementary Files

Supplementary File 1

Acknowledgments

Krishnan Murugappan thanks Curtin University for the award of a Curtin International Postgraduate Research Scholarship and Debbie S. Silvester thanks the Australian Research Council for funding via a Discovery Early Career Research Award (DE120101456). The authors thank Damien W.M. Arrigan and Junqiao Lee for reading through the manuscript before submission.

Author Contributions

Krishnan Murugappan performed the experiments and made the figures. Krishnan Murugappan and Debbie S. Silvester conceived and designed the experiments, analyzed the data and wrote the paper.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Murugappan, K.; Lee, J.; Silvester, D.S. Comparative study of screen printed electrodes for ammonia gas sensing in ionic liquids. Electrochem. Commun. 2011, 13, 1435–1438. [Google Scholar] [CrossRef]
  2. Lee, J.; Murugappan, K.; Arrigan, D.W.M.; Silvester, D.S. Oxygen reduction voltammetry on platinum macrodisk and screen-printed electrodes in ionic liquids: Reaction of the electrogenerated superoxide species with compounds used in the paste of Pt screen-printed electrodes? Electrochim. Acta 2013, 101, 158–168. [Google Scholar] [CrossRef]
  3. Hayat, A.; Marty, J.L. Disposable screen printed electrochemical sensors: Tools for environmental monitoring. Sensors 2014, 14, 10432–10453. [Google Scholar] [CrossRef] [PubMed]
  4. Xiong, L.; Compton, R.G. Amperometric gas detection: A review. Int. J. Electrochem. Sci. 2014, 9, 7152–7181. [Google Scholar]
  5. Metters, J.P.; Kadara, R.O.; Banks, C.E. New directions in screen printed electroanalytical sensors: An overview of recent developments. Analyst 2011, 136, 1067–1076. [Google Scholar] [CrossRef] [PubMed]
  6. Crowley, K.; O’Malley, E.; Morrin, A.; Smyth, M.R.; Killard, A.J. An aqueous ammonia sensor based on an inkjet-printed polyaniline nanoparticle-modified electrode. Analyst 2008, 133, 391–399. [Google Scholar] [CrossRef] [PubMed]
  7. Choudhry, N.A.; Kampouris, D.K.; Kadara, R.O.; Jenkinson, N.; Banks, C.E. Next generation screen printed electrochemical platforms: Non-enzymatic sensing of carbohydrates using copper (II) oxide screen printed electrodes. Anal. Methods 2009, 1, 183–187. [Google Scholar] [CrossRef]
  8. Murugappan, K.; Arrigan, D.W.M.; Silvester, D.S. Electrochemical behaviour of chlorine on platinum microdisk and screen printed electrodes in a room temperature ionic liquid. J. Phys. Chem. C 2015, 119, 23572–23579. [Google Scholar] [CrossRef]
  9. Carter, M.T.; Stetter, J.R.; Findlay, M.W.; Patel, V. Rational design of amperometric gas sensors with ionic liquid electrolytes. ECS Trans. 2014, 64, 95–103. [Google Scholar] [CrossRef]
  10. Compressed Gas Association, Inc. Handbook of Compressed Gases; Springer US: Dordrecht, The Netherlands, 1999. [Google Scholar]
  11. Pohanish, R.P. Sittig’s Handbook of Toxic and Hazardous Chemicals and Carcinogens: A-K; Elsevier: Amsterdam, The Netherlands, 2012. [Google Scholar]
  12. Murugappan, K.; Kang, C.; Silvester, D.S. Electrochemical oxidation and sensing of methylamine gas in room temperature ionic liquids. J. Phys. Chem. C 2014, 118, 19232–19237. [Google Scholar] [CrossRef]
  13. Aldous, L.; Silvester, D.S.; Pitner, W.R.; Compton, R.G.; Lagunas, M.C.; Hardacre, C. Voltammetric studies of gold, protons, and [HCl2]- in ionic liquids. J. Phys. Chem. C 2007, 111, 8496–8503. [Google Scholar] [CrossRef]
  14. O’Mahony, A.M.; Silvester, D.S.; Aldous, L.; Hardacre, C.; Compton, R.G. Effect of water on the electrochemical window and potential limits of room-temperature ionic liquids. J. Chem. Eng. Data 2008, 53, 2884–2891. [Google Scholar] [CrossRef]
  15. Silvester, D.S.; Compton, R.G. Electrochemistry in room temperature ionic liquids: A review and some possible applications. Z. Phys. Chem. 2006, 220, 1247–1274. [Google Scholar] [CrossRef]
  16. Zhao, C.; Bond, A.M.; Compton, R.G.; O’Mahony, A.M.; Rogers, E.I. Modification and implications of changes in electrochemical responses encountered when undertaking deoxygenation in ionic liquids. Anal. Chem. 2010, 82, 3856–3861. [Google Scholar] [CrossRef] [PubMed]
  17. Barrosse-Antle, L.E.; Bond, A.M.; Compton, R.G.; O’Mahony, A.M.; Rogers, E.I.; Silvester, D.S. Voltammetry in room temperature ionic liquids: Comparisons and contrasts with conventional electrochemical solvents. Chem. Asian J. 2010, 5, 202–230. [Google Scholar] [CrossRef] [PubMed]
  18. Compton, R.G.; Banks, C.E. Understanding Voltammetry; World Scientific: Singapore, 2007. [Google Scholar]
  19. Silvester, D.S.; Aldous, L.; Hardacre, C.; Compton, R.G. An electrochemical study of the oxidation of hydrogen at platinum electrodes in several room temperature ionic liquids. J. Phys. Chem. B 2007, 111, 5000–5007. [Google Scholar] [CrossRef] [PubMed]

Share and Cite

MDPI and ACS Style

Murugappan, K.; Silvester, D.S. Sensors for Highly Toxic Gases: Methylamine and Hydrogen Chloride Detection at Low Concentrations in an Ionic Liquid on Pt Screen Printed Electrodes. Sensors 2015, 15, 26866-26876. https://doi.org/10.3390/s151026866

AMA Style

Murugappan K, Silvester DS. Sensors for Highly Toxic Gases: Methylamine and Hydrogen Chloride Detection at Low Concentrations in an Ionic Liquid on Pt Screen Printed Electrodes. Sensors. 2015; 15(10):26866-26876. https://doi.org/10.3390/s151026866

Chicago/Turabian Style

Murugappan, Krishnan, and Debbie S. Silvester. 2015. "Sensors for Highly Toxic Gases: Methylamine and Hydrogen Chloride Detection at Low Concentrations in an Ionic Liquid on Pt Screen Printed Electrodes" Sensors 15, no. 10: 26866-26876. https://doi.org/10.3390/s151026866

Article Metrics

Back to TopTop