Next Article in Journal
Previous Article in Journal
Sensors 2010, 10(8), 7089-7098; doi:10.3390/s100807089
Article

Development of a Fully Automated Flow Injection Analyzer Implementing Bioluminescent Biosensors for Water Toxicity Assessment

1
, 1
, 2
, 2
 and 1,*
Received: 11 May 2010; in revised form: 2 June 2010 / Accepted: 12 July 2010 / Published: 27 July 2010
(This article belongs to the Section Biosensors)
View Full-Text   |   Download PDF [293 KB, updated 21 June 2014; original version uploaded 21 June 2014]
Abstract: This paper describes the development of an automated Flow Injection analyzer for water toxicity assessment. The analyzer is validated by assessing the toxicity of heavy metal (Pb2+, Hg2+ and Cu2+) solutions. One hundred μL of a Vibrio fischeri suspension are injected in a carrier solution containing different heavy metal concentrations. Biosensor cells are mixed with the toxic carrier solution in the mixing coil on the way to the detector. Response registered is % inhibition of biosensor bioluminescence due to heavy metal toxicity in comparison to that resulting by injecting the Vibrio fischeri suspension in deionised water. Carrier solutions of mercury showed higher toxicity than the other heavy metals, whereas all metals show concentration related levels of toxicity. The biosensor’s response to carrier solutions of different pHs was tested. Vibrio fischeri’s bioluminescence is promoted in the pH 5–10 range. Experiments indicate that the whole cell biosensor, as applied in the automated fluidic system, responds to various toxic solutions.
Keywords: Vibrio fischeri; toxicity; flow injection Vibrio fischeri; toxicity; flow injection
This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Export to BibTeX |
EndNote


MDPI and ACS Style

Komaitis, E.; Vasiliou, E.; Kremmydas, G.; Georgakopoulos, D.G.; Georgiou, C. Development of a Fully Automated Flow Injection Analyzer Implementing Bioluminescent Biosensors for Water Toxicity Assessment. Sensors 2010, 10, 7089-7098.

AMA Style

Komaitis E, Vasiliou E, Kremmydas G, Georgakopoulos DG, Georgiou C. Development of a Fully Automated Flow Injection Analyzer Implementing Bioluminescent Biosensors for Water Toxicity Assessment. Sensors. 2010; 10(8):7089-7098.

Chicago/Turabian Style

Komaitis, Efstratios; Vasiliou, Efstathios; Kremmydas, Gerasimos; Georgakopoulos, Dimitrios G.; Georgiou, Constantinos. 2010. "Development of a Fully Automated Flow Injection Analyzer Implementing Bioluminescent Biosensors for Water Toxicity Assessment." Sensors 10, no. 8: 7089-7098.


Sensors EISSN 1424-8220 Published by MDPI AG, Basel, Switzerland RSS E-Mail Table of Contents Alert