Sensors 2010, 10(11), 10014-10026; doi:10.3390/s101110014
Article

Effects of the Intermittent Pneumatic Circulator on Blood Pressure during Hemodialysis

1
, 2
, 3
 and 4,5,*
Received: 12 July 2010; in revised form: 10 September 2010 / Accepted: 23 October 2010 / Published: 9 November 2010
(This article belongs to the Special Issue Sensors in Biomechanics and Biomedicine)
View Full-Text   |   Download PDF [241 KB, uploaded 21 June 2014]   |   Browse Figures
Abstract: Hypotension is frequently reported during hemodialysis. This study aimed to examine the effect of the intermittent pneumatic circulator on blood pressure during hemodialysis. Sixteen subjects with chronic hemodialysis were recruited. Each subject randomly received two test conditions on separate days, hemodialysis with and without the circulator. The circulator was applied to the subject on lower extremities during 0.5–1 hr, 1.5–2 hr, 2.5–3 hr, and 3.5–4 hr of hemodialysis. Systolic and diastolic blood pressures (SBP and DBP) and heart rate (HR) were analyzed at pre-dialysis, 1 hr, 2 hr, and 3 hr of hemodialysis. Stroke volume (SV) and cardiac output (CO) were evaluated between 2.5 and 3.0 hr of hemodialysis. Blood chemicals (sodium, calcium, potassium, and phosphorous) and Kt/V before and after each hemodialysis session were analyzed. The number of episodes of hypotension was also recorded. The circulator intervention significantly improved SBP and DBP across all time points (P = 0.002 for SBP; P = 0.002 for DBP). The frequency of hypotension was significantly decreased (P = 0.028). SV and CO were significantly improved with the circulator intervention (P = 0.017 for SV; P = 0.026 for CO) and no statistical significances were found on blood chemicals or Kt/V analyses. The results suggested that the circulator intervention helps stabilize blood pressure and appears to be a practical treatment. Future studies are suggested to develop new circulator innovations with sensor feedback systems to enhance safety and maximize treatment efficiency.
Keywords: hemodialysis; dialysis; mechanical pump; blood pressure; circulator hemodialysis; dialysis; mechanical pump; blood pressure; circulator
This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Export to BibTeX |
EndNote


MDPI and ACS Style

Hsu, T.-C.; Chang, Y.-J.; Huang, Y.-Y.; Hsu, M.-J. Effects of the Intermittent Pneumatic Circulator on Blood Pressure during Hemodialysis. Sensors 2010, 10, 10014-10026.

AMA Style

Hsu T-C, Chang Y-J, Huang Y-Y, Hsu M-J. Effects of the Intermittent Pneumatic Circulator on Blood Pressure during Hemodialysis. Sensors. 2010; 10(11):10014-10026.

Chicago/Turabian Style

Hsu, Tzu-Chao; Chang, Ya-Ju; Huang, Yu-Yao; Hsu, Miao-Ju. 2010. "Effects of the Intermittent Pneumatic Circulator on Blood Pressure during Hemodialysis." Sensors 10, no. 11: 10014-10026.



Sensors EISSN 1424-8220 Published by MDPI AG, Basel, Switzerland RSS E-Mail Table of Contents Alert