Next Article in Journal
Next Article in Special Issue
Previous Article in Journal
Previous Article in Special Issue
Sensors 2013, 13(9), 11797-11817; doi:10.3390/s130911797
Article

Automatic Determination of Validity of Input Data Used in Ellipsoid Fitting MARG Calibration Algorithms

1,* , 2
, 2
, 1
 and 1
Received: 29 July 2013; in revised form: 29 August 2013 / Accepted: 3 September 2013 / Published: 5 September 2013
(This article belongs to the Special Issue State-of-the-Art Sensors Technology in Spain 2013)
View Full-Text   |   Download PDF [1903 KB, uploaded 21 June 2014]   |   Browse Figures
Abstract: Ellipsoid fitting algorithms are widely used to calibrate Magnetic Angular Rate and Gravity (MARG) sensors. These algorithms are based on the minimization of an error function that optimizes the parameters of a mathematical sensor model that is subsequently applied to calibrate the raw data. The convergence of this kind of algorithms to a correct solution is very sensitive to input data. Input calibration datasets must be properly distributed in space so data can be accurately fitted to the theoretical ellipsoid model. Gathering a well distributed set is not an easy task as it is difficult for the operator carrying out the maneuvers to keep a visual record of all the positions that have already been covered, as well as the remaining ones. It would be then desirable to have a system that gives feedback to the operator when the dataset is ready, or to enable the calibration process in auto-calibrated systems. In this work, we propose two different algorithms that analyze the goodness of the distributions by computing four different indicators. The first approach is based on a thresholding algorithm that uses only one indicator as its input and the second one is based on a Fuzzy Logic System (FLS) that estimates the calibration error for a given calibration set using a weighted combination of two indicators. Very accurate classification between valid and invalid datasets is achieved with average Area Under Curve (AUC) of up to 0.98.
Keywords: calibration; accelerometer; magnetometer; MARG; MEMS; automatic; validation; fuzzy; FLS; thresholding calibration; accelerometer; magnetometer; MARG; MEMS; automatic; validation; fuzzy; FLS; thresholding
This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Export to BibTeX |
EndNote


MDPI and ACS Style

Olivares, A.; Ruiz-Garcia, G.; Olivares, G.; Górriz, J.M.; Ramirez, J. Automatic Determination of Validity of Input Data Used in Ellipsoid Fitting MARG Calibration Algorithms. Sensors 2013, 13, 11797-11817.

AMA Style

Olivares A, Ruiz-Garcia G, Olivares G, Górriz JM, Ramirez J. Automatic Determination of Validity of Input Data Used in Ellipsoid Fitting MARG Calibration Algorithms. Sensors. 2013; 13(9):11797-11817.

Chicago/Turabian Style

Olivares, Alberto; Ruiz-Garcia, Gonzalo; Olivares, Gonzalo; Górriz, Juan M.; Ramirez, Javier. 2013. "Automatic Determination of Validity of Input Data Used in Ellipsoid Fitting MARG Calibration Algorithms." Sensors 13, no. 9: 11797-11817.



Sensors EISSN 1424-8220 Published by MDPI AG, Basel, Switzerland RSS E-Mail Table of Contents Alert