Next Article in Journal
Evaluating Mortality Predictors in COVID-19 Intensive Care Unit Patients: Insights into Age, Procalcitonin, Neutrophil-to-Lymphocyte Ratio, Platelet-to-Lymphocyte Ratio, and Ferritin Lactate Index
Next Article in Special Issue
A Lateral-Flow Device for the Rapid Detection of Scedosporium Species
Previous Article in Journal
Metabolic Fingerprint in Childhood Acute Lymphoblastic Leukemia
Previous Article in Special Issue
The Role and Value of Professional Rapid Testing of Acute Respiratory Infections (ARIs) in Europe: A Special Focus on the Czech Republic, Poland, and Romania
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Communication

Clinical Evaluation of Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex Assay

Department of Pathology, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China
*
Author to whom correspondence should be addressed.
Diagnostics 2024, 14(7), 683; https://doi.org/10.3390/diagnostics14070683
Submission received: 6 March 2024 / Accepted: 22 March 2024 / Published: 24 March 2024
(This article belongs to the Special Issue Microbiology Laboratory: Sample Collection and Diagnosis Advances)

Abstract

:
The performance of the Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex Assays were evaluated using 167 specimens, including 158 human respiratory specimens and 9 external quality assessment program (EQAP) samples. For respiratory specimens, CoV-2/Flu/RSV plus exhibited perfect agreement with the standard-of-care (SOC) methods (Cohen’s κ: 1, 100% agreement). The overall positive and negative percent agreement (PPA and NPA) were 100%, with 95% confidence intervals of 96.50 to 100% and 85.70 to 100%, respectively. On the other hand, Resp-4-Plex revealed an almost perfect agreement with the SOC methods (Cohen’s κ: 0.92, 97.71% agreement). The overall PPA and NPA were 100% (95.76 to 100%) and 88.46% (70.20 to 96.82%), respectively. For EQAP samples, the results of CoV-2/Flu/RSV plus (9/9) and Resp-4-Plex (4/4) were concordant with the expected results. The experimental limit of detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) was the lowest (25 copies/mL for both methods), and that of the respiratory syncytial virus was the highest (400 copies/mL for CoV-2/Flu/RSV plus and 100 copies/mL for Resp-4-Plex). Threshold cycle (Ct) value correlation showed a large positive linear association between CoV-2/Flu/RSV plus and Resp-4-Plex, with R-squared values of 0.92–0.97, and on average, the Ct values of CoV-2/Flu/RSV plus were higher than that of Resp-4-Plex by 1.86–2.78, except for Flu A1 target (−0.66). To conclude, the performance of both assay was comparable to the SOC methods for both upper and lower respiratory specimens. Implementation of these rapid assay may reinforce the diagnostic capacity for the post-pandemic co-circulation of SARS-CoV-2 and other respiratory viruses.

1. Introduction

The coronavirus disease 2019 (COVID-19) pandemic has lasted more than 3 years and accumulated 774,469,939 reported cases across six World Health Organization regions (as of 10 February 2024) [1]. With the availability of antiviral treatments and the build-up of herd immunity barrier, the global strategy against COVID-19 has been shifting towards ‘living with the virus’ [2,3,4]. Hong Kong has likewise lifted all pandemic control measures since 1 March 2023 [5]. Recently, the resurgence of respiratory viruses other than severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has been reported worldwide [6,7,8,9,10,11]. Similarities in clinical presentation make the identification of the culprit and coinfection difficult, posing challenges to diagnosis and patient management. In this regard, rapid, multiplex nucleic acid amplification testing (NAAT) of clinically important respiratory viruses may facilitate timely clinical decision making.
The Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex Assay are two commercial in vitro diagnostic NAATs under the U.S. Food and Drug Administration’s Emergency Use Authorization. They are designed for the qualitative detection of SARS-CoV-2, influenza A and B viruses, and respiratory syncytial virus (RSV). Xpert Xpress CoV-2/Flu/RSV plus targets the nucleocapsid (N), envelope and RNA-dependent RNA polymerase (RdRP) genes of SARS-CoV-2, the matrix (M), basic polymerase, and acidic protein genes of influenza A virus, the M and non-structural (NS) protein genes of influenza B virus and the N genes of RSV A and B [12]. The assay is run on the GeneXpert System which allows for random access of tests with a runtime of about 40 min. The Resp-4-Plex Assay, on the other hand, targets the RdRP and N genes of SARS-CoV-2, the M genes of influenza A virus and RSV, and the NS1 gene of influenza B virus. The assay is run on the Alinity m System, which allows for batch testing of up to 1080 specimens in 24 h, with the first result available in less than 2 h [13].
In this study, we evaluated the performance of the Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex Assay with a collection of human respiratory specimens, external quality assessment program (EQAP) samples, and an external run control of target viruses.

2. Materials and Methods

2.1. Specimens

This study was pursued in the Department of Pathology, Hong Kong Sanatorium & Hospital. A total of 167 specimens were tested, including 158 human respiratory specimens collected between October 2022 and February 2024, as well as 9 samples from 2 EQAPs. The respiratory specimens included 121 from the upper and 37 from the lower respiratory tract (URT and LRT). URT specimens comprised 29 nasopharyngeal swabs, 37 combined nasopharyngeal and throat swabs, and 55 combined nasal and throat swabs preserved in a viral transport medium. LRT specimens included 2 endotracheal aspirate, 5 bronchoalveolar lavage fluid, 5 tracheal aspirate, and 25 sputum specimens.

2.2. Standard-of-Care (SOC) Methods

The specimens were routinely tested with any of or combination of the SOC methods, including BIOFIRE Respiratory 2.1 plus Panel (RP2.1plus) or Pneumonia plus Panel (PNplus) on FilmArray 2.0 or Torch System (BioFire Diagnostics, Salt Lake City, UT, USA); Xpert Xpress SARS-CoV-2 on GeneXpert System (Cepheid, Sunnyvale, CA, USA); and LightMix SarbecoV E-gene plus EAV control (Tib MolBiol, Berlin, Germany) on LightCycler 480 II (Roche, Basel, Switzerland).

2.3. Evaluation of Clinical and Analytical Performance

The remnant specimens were tested with both the Xpert Xpress CoV-2/Flu/RSV plus (Cepheid, Sunnyvale, CA, USA) and Alinity m Resp-4-Plex Assay (Abbott Molecular, Desplaines, IL, USA) as far as possible, depending on residual quantity. All experimental procedures were performed according to the manufacturers’ recommendations. Cohen’s κ was calculated to measure the overall agreement between CoV-2/Flu/RSV plus (or Resp-4-Plex) and the SOC methods [14].
Positive and negative percent agreement (PPA and NPA) of each pathogen target were calculated by using the following formulae:
PPA = TP/(TP + FN) × 100%
NPA = TN/(TN + FP) × 100%
FN, FP, TN, and TP stand for a number of false-negative, false-positive, true-negative, and true-positive specimens, respectively. The 95% confidence intervals (CI) for PPA and NPA were calculated using the online version of GraphPad software (Version 2024) (modified Wald method) [15].
The experimental limit of detection (LoD) was estimated by testing at least quintuple replicates of NATtrol Flu/RSV/SARS-CoV-2 External Run Control & Verification Panel (ZeptoMetrix, Buffalo, NY, USA) at 7 dilutions (10, 25, 50, 100, 200, 300, and 400 copies/mL). This control comprises a mixture of equal concentration (15,000 copies/mL) of influenza A virus (A/Brisbane/10/07), influenza B virus (B/Florida/02/06), RSV B (CH93(18)-18), and SARS-CoV-2 (USA-WA1/2020) [16]. The correlation of threshold cycle (Ct) values between CoV-2/Flu/RSV plus and Resp-4-Plex was assessed by considering the R-squared value of respective linear regression.

3. Results

3.1. Clinical Performance

The results are summarized in Table 1.
For CoV-2/Flu/RSV plus, a total of two specimens were excluded from analysis due to inadequacy (specimen 22) and invalid results (specimen 52). For specimen 47, error 5007 was encountered in the first run due to Flu B probe check failure and the repeated run was valid. There were 2 invalid runs (0.83%) out of a total of 240.
CoV-2/Flu/RSV plus exhibited perfect agreement with the SOC methods (Cohen’s κ: 1, 100% agreement). The overall PPA and NPA were 100%, with a 95% CI of 96.50 to 100% and 85.70 to 100%, respectively.
For Resp-4-Plex, a total of 27 specimens were excluded from analysis due to inadequacy (25 specimens) and invalid results (2 specimens). The error rate was 1.07% (2 out of 187 runs).
Resp-4-Plex displayed almost perfect agreement with the SOC methods, with Cohen’s κ of 0.92 and 97.71% overall agreement. The overall PPA and NPA were 100% (95.76 to 100%) and 88.46% (70.20 to 96.82%), respectively.
Specimens 30, 32, 33, and 34 were positive for SARS-CoV-2 variant XBB.1 with C26270T and A26275G mutations detected in the E gene. The E gene Ct values of the SOC method Xpert Xpress SARS-CoV-2 were 0, and N2 Ct values ranged from 25.8 to 34.3. SARS-CoV-2 was detected by both CoV-2/Flu/RSV plus and Resp-4-Plex in these four specimens.
For the three specimens positive for two viruses (specimen 106, SARS-CoV-2 and influenza A virus; specimen 107, SARS-CoV-2 and RSV; specimen 108, influenza A virus and RSV), the results of CoV-2/Flu/RSV plus and Resp-4-Plex were concordant with that of the SOC methods.
For Resp-4-Plex, the results of three specimens were discordant with CoV-2/Flu/RSV plus and the SOC method RP2.1plus. These included specimens 117 (RSV detected by Resp-4-Plex), 122, and 124 (SARS-CoV-2 detected by Resp-4-Plex). The Ct values ranged from 35.96 to 38.34. CoV-2/Flu/RSV plus and RP2.1plus were negative for all four virus targets.

3.2. EQAP Samples

Table 2 summarizes the results of EQAP samples. The samples were obtained from two EQAPs. Nine samples were tested by CoV-2/Flu/RSV plus. Only four samples were tested by Resp-4-Plex, as the other five samples were not sufficient for testing. The results of both assay were concordant with the expected answers. The Ct values of Resp-4-Plex appeared to be lower than those of CoV-2/Flu/RSV plus.

3.3. Experimental LoD and Precision

Table 3 summarizes the detection rate at each dilution of the reference material and precision in terms of the coefficient of variation (CV) in Ct values. The LoD for simultaneous detection of all four viruses was 400 copies/mL for CoV-2/Flu/RSV plus and 100 copies/mL for Resp-4-Plex. For individual viruses, the LoD of SARS-CoV-2, influenza A and B viruses, and RSV were 25, 100, 200, and 400 copies/mL, respectively, for CoV-2/Flu/RSV plus, and 25, 50, 50, and 100 copies/mL, respectively, for Resp-4-Plex. The average CV at or above the LoD was 1.61% (0.70–2.63%) for CoV-2/Flu/RSV plus and 1.56% (0.57–3.76%) for Resp-4-Plex.

3.4. Inter-Assay Correlation of Ct Values

Figure 1 shows the correlation of Ct values between CoV-2/Flu/RSV plus and Resp-4-Plex. The data points were generated from clinical specimens, EQAP samples, and dilutions of reference material at or above the experimental LoD of individual viruses. R-squared values for SARS-CoV-2, the two targets of influenza A virus (Flu A1 or Flu A2 of CoV-2/Flu/RSV plus versus Flu A of Resp-4-Plex), influenza B virus, and RSV, were 0.96, 0.96, 0.96, 0.97, and 0.92 with average Ct value differences of 2.78 ± 1.60 (−6.87 to 6.03), −0.66 ± 1.20 (−3.77 to 2.45), 2.38 ± 1.19 (−0.66 to 5.21), 1.86 ± 1.28 (−0.36 to 4.93), and 2.37 ± 2.05 (−0.99 to 9.62), respectively.

4. Discussion

Our data showed that the clinical performance of CoV-2/Flu/RSV plus and Resp-4-Plex were comparable to the SOC assay. Commercial PCR panels like RP2.1plus and PNplus are the choices for broad-range detection; CoV-2/Flu/RSV plus and Resp-4-Plex, on the other hand, focus on the four clinically important respiratory viruses. In our laboratory, the positive rate of CoV-2/Flu/RSV plus during the winter surge of influenza and COVID-19 was 38.38% (January and February 2024). The time lapse between specimen receipt by laboratory staff and result reporting was 1.67 h on average. The data suggest that the rapid PCR testing of the four viruses may deliver actionable results to physicians within a reasonable timeframe for a significant portion of symptomatic patients, and this may, in turn, lessen the test burden on broad-range PCR instruments. Notably, Resp-4-Plex is less favorable than CoV-2/Flu/RSV plus in turnaround time, yet its high sample throughput is a merit for laboratories with massive test volumes.
Interestingly, from our CoV-2/Flu/RSV plus data, the SARS-CoV-2-influenza virus combination comprised the majority (80.00%) of codetections in the post-COVID-19 era. From the literature, the coinfection of SARS-CoV-2 and influenza viruses significantly increased the odds of receiving invasive mechanical ventilation and in-hospital mortality in hospitalized populations [17]. Another study by Pawlowski and coworkers revealed that ‘flurona’ was rare and mostly observed in relatively young and healthy patients without an increased risk of hospitalization, intensive care unit (ICU) admission, or death, yet with more emblematic viral symptoms [18]. The authors remarked that the clinical outcome of coinfection in patients with risk factors for severe COVID-19 required further observation. A systematic review and meta-analysis of 95 studies revealed that the odds ratios for ICU admission, mechanical ventilation support, and mortality were significantly higher among COVID-19 patients coinfected with influenza A virus than with mono-infections [19]. These findings underscore the importance of rapid PCR testing as well as seasonal COVID-19 and influenza vaccination for high-risk individuals.
There were three discordant results for Resp-4-Plex. The specimens revealed late amplification, and the comparator assay were negative for the targets detected. We had insufficient clinical and epidemiological data to interpret the validity of these positive signals. This reflects the need for a retest algorithm and clinician advice for weak positive results in real-life practice.
An interesting observation was identified about the positive cutoff values of CoV-2/Flu/RSV plus. Based on the LoD data, it appeared that the positive cutoff value of RSV was much higher than other targets. For instance, we encountered a ‘negative’ RSV result with a Ct value of 38.1, fluorescence intensity (EndPt) of 257 with a sigmoidal amplification curve, and ‘positive’ results of other targets with Ct values of 39.3 to 44.1 and EndPt of 46 to 150. In this regard, checking the amplification curves can help avoid false negative results.
The performance of CoV-2/Flu/RSV plus and Resp-4-Plex have also been evaluated in other studies. For instance, Noble and coworkers compared the performance of CoV-2/Flu/RSV plus with various SOC assay [20]. A total of 97 respiratory specimens were tested. The error rate was 0.85% (1/118) which was similar to this study (0.83%). The overall PPA and NPA were 100% and 83.3%, respectively. The authors commented that the NPA was negatively influenced by the low number of negative specimens (n = 12). The authors also reported that CoV-2/Flu/RSV plus was compatible with SARS-CoV-2 variants, including Omicron BA.1. Our data also showed that both CoV-2/Flu/RSV plus and Resp-4-Plex were compatible with SARS-CoV-2 XBB.1 variant. In another study by Johnson and coworkers, CoV-2/Flu/RSV plus was compared with its prior version, Xpert Xpress CoV-2/Flu/RSV, with 171 clinical specimens, including 11 mixed specimens, to assess target interaction [21]. The authors reported that the percentage of agreement for SARS-CoV-2, influenza A and B viruses, and RSV were 100%, 100%, 99.4%, and 100%, respectively. CoV-2/Flu/RSV plus was also inclusive for SARS-CoV-2 variants Omicron BA.1 and BA.2. Zhen and colleagues evaluated the analytical performance of Resp-4-Plex and compared its performance to various assay [22]. Resp-4-Plex exhibited 100% PPA for influenza A virus and RSV and 95% for SARS-CoV-2 and influenza B virus. The NPA for all four viruses was 100%. The LoD was 26 ± 21 copies/mL for SARS-CoV-2, 36 ± 22 copies/mL for influenza A virus, 22 ± 22 copies/mL for influenza B virus, and 22 ± 23 copies/mL for RSV, which were lower than the LoDs estimated in this study except for SARS-CoV-2. Quinton and colleagues compared the performance of Resp-4-Plex and NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage in a study involving 181 remnant specimens [23]. They reported that both assay showed 100% total agreement with the SOC assay. The overall precision of both assay was 100%. All in all, the performance of CoV-2/Flu/RSV plus and Resp-4-Plex was verified in a number of studies.
Our study had two major constraints. First, not all remnant specimens were sufficient for parallel testing by both methods. In particular, the number of influenza B virus-positive specimens for Resp-4-Plex was suboptimal (n = 4). Second, more LRT specimens should be tested to thoroughly assess the performance of both assay.

5. Conclusions

We evaluated two representative types of ‘random-access’ molecular diagnostic assay for multiplex detection of four clinically important respiratory viruses, each of which is merited by short assay time or high sample throughput. These merits may allow for more flexibility to cope with the sharp rise in test burden and stat specimens during the co-circulation of SARS-CoV-2 and other respiratory viruses.

Author Contributions

Conceptualization, B.S.-F.T. and W.-S.C.; data curation, W.-S.C.; writing—original draft preparation, W.-S.C.; writing—review and editing, B.S.-F.T., E.S.-K.M., T.S.-K.W., C.-H.A., C.L.-N.W., C.-P.L., K.T.-W.L., K.-P.W., S.-K.Y., C.-Y.W., T.-C.C., J.H. and W.-S.C.; supervision, B.S.-F.T. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki, and approved by the HKSH Medical Group Research Committee. The reference number is RC-2023-02.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study are available upon request from the corresponding author.

Acknowledgments

We thank the colleagues of Department of Pathology, Hong Kong Sanatorium & Hospital for their dedicated and professional work during this ongoing COVID-19 pandemic.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. World Health Organization. WHO Coronavirus (COVID-19) Dashboard. Available online: https://covid19.who.int/table (accessed on 10 February 2024).
  2. GOV.UK. Guidance—COVID-19 Response: Living with COVID-19. Available online: https://www.gov.uk/government/publications/covid-19-response-living-with-covid-19/covid-19-response-living-with-covid-19 (accessed on 2 March 2023).
  3. Bloomberg. Singapore Confronts the Division and Fear That Come from Living with COVID. Available online: https://www.bloomberg.com/news/articles/2021-10-14/singapore-confronts-division-and-fear-bred-by-living-with-covid?leadSource=uverify%20wall (accessed on 2 March 2023).
  4. SaxInstitute Public Health Research & Practice. Learning to Live with COVID-19 in Australia: Time for a New Approach. Available online: https://www.phrp.com.au/issues/september-2021-volume-31-issue-3/learning-to-live-with-covid-19/ (accessed on 2 March 2023).
  5. The Government of the Hong Kong Special Administrative Region. Latest Highlights. Available online: https://www.coronavirus.gov.hk/eng/index.html (accessed on 2 March 2023).
  6. The Lancet Respiratory Medicine. Patterns of respiratory infections after COVID-19. Lancet Respir. Med. 2024, 12, 1. [Google Scholar] [CrossRef] [PubMed]
  7. Hui, D.S.; Zumla, A.; Memish, Z.A. De ja vu? Post-COVID-19 Surge in Respiratory Illnesses Among Children in China Emphasizes Need for Proactive Surveillance, Openness, Early Detection and Reporting of Causative Pathogen(s), and Their AMR Status. J. Epidemiol. Glob. Health 2023. advance online publication. [Google Scholar] [CrossRef] [PubMed]
  8. Weidmann, M.D.; Green, D.A.; Berry, G.J.; Wu, F. Assessing respiratory viral exclusion and affinity interactions through co-infection incidence in a pediatric population during the 2022 resurgence of influenza and RSV. Front. Cell. Infect. Microbiol. 2023, 13, 1208235. [Google Scholar] [CrossRef] [PubMed]
  9. Radhi, A.H.; Ibrahim, Z.H.; Alhilifi, R. The epidemiological pattern of seasonal influenza in four sentinel sites in Iraq. Influenza Other Respir. Viruses 2023, 17, e13134. [Google Scholar] [CrossRef] [PubMed]
  10. Temte, J.L.; Goss, M.; Bell, C.; Barlow, S.; Temte, E.; Bateman, A.; Uzicanin, A. Changing pattern of respiratory virus detections among school-aged children in a small community—Dane County, Wisconsin, September to December 2022. Influenza Other Respir. Viruses 2023, 17, e13171. [Google Scholar] [CrossRef]
  11. Chan, W.S.; Yau, S.K.; To, M.Y.; Leung, S.M.; Wong, K.P.; Lai, K.C.; Wong, C.Y.; Leung, C.P.; Au, C.H.; Wan, T.S.; et al. The Seasonality of Respiratory Viruses in a Hong Kong Hospital, 2014–2023. Viruses 2023, 15, 1820. [Google Scholar] [CrossRef] [PubMed]
  12. GeneXpert. Xpert Xpress CoV-2/Flu/RSV plus Package Insert (302-6992, Rev. D, 2023–10). Available online: https://www.cepheid.com/content/dam/www-cepheid-com/documents/package-insert-files/xpert-xpress-cov-2-flu-rsv-plus/Xpert%20Xpress%20CoV-2%20FLU%20RSV%20plus%20EUA%20IFU%20302-6992%20Rev%20D.pdf (accessed on 27 February 2024).
  13. Abbott. Alinity m Resp-4-Plex Assay. Available online: https://www.molecular.abbott/int/en/alinity-m-resp-4-plex-assay (accessed on 27 February 2024).
  14. IDOSTATISTICS. Cohen’s Kappa Free Calculator. Available online: https://idostatistics.com/cohen-kappa-free-calculator/#calcolobox (accessed on 13 February 2024).
  15. GraphPad by Dotmatics. Confidence Interval of a Proportion. Available online: https://www.graphpad.com/quickcalcs/confInterval1/ (accessed on 13 February 2024).
  16. ZeptoMetrix. NATtrol Flu/RSV/SARS-CoV-2 External Run Control & Verification Panel. Available online: https://www.zeptometrix.com/products/flu-rsv-sars-cov-2-external-run-control-6-x-05ml (accessed on 27 February 2023).
  17. Swets, M.C.; Russell, C.D.; Harrison, E.M.; Docherty, A.B.; Lone, N.; Girvan, M.; Hardwick, H.E.; ISARIC4C Investigators; Visser, L.G.; Openshaw, P.J.M.; et al. SARS-CoV-2 co-infection with influenza viruses, respiratory syncytial virus, or adenoviruses. Lancet 2022, 399, 1463–1464. [Google Scholar] [CrossRef] [PubMed]
  18. Pawlowski, C.; Silvert, E.; O’Horo, J.C.; Lenehan, P.J.; Challener, D.; Gnass, E.; Murugadoss, K.; Ross, J.; Speicher, L.; Geyer, H.; et al. SARS-CoV-2 and influenza coinfection throughout the COVID-19 pandemic: An assessment of coinfection rates, cohort characteristics, and clinical outcomes. PNAS Nexus 2022, 1, pgac071. [Google Scholar] [CrossRef] [PubMed]
  19. Yan, X.; Li, K.; Lei, Z.; Luo, J.; Wang, Q.; Wei, S. Prevalence and associated outcomes of coinfection between SARS-CoV-2 and influenza: A systematic review and meta-analysis. Int. J. Infect. Dis. IJID Off. Publ. Int. Soc. Infect. Dis. 2023, 136, 29–36. [Google Scholar] [CrossRef] [PubMed]
  20. Noble, L.D.; Scott, L.E.; Munir, R.; Du Plessis, M.; Steegen, K.; Hans, L.; Marokane, P.; Da Silva, P.; Stevens, W.S. Rapid Evaluation of the Xpert® Xpress CoV-2 plus and Xpert® Xpress CoV-2/Flu/RSV plus Tests. Diagnostics 2022, 13, 34. [Google Scholar] [CrossRef] [PubMed]
  21. Johnson, G.; Gregorchuk, B.S.J.; Zubrzycki, A.; Kolsun, K.; Meyers, A.F.A.; Sandstrom, P.A.; Becker, M.G. Clinical evaluation of the GeneXpert® Xpert® Xpress SARS-CoV-2/Flu/RSV PLUS combination test. Can. J. Microbiol. 2023, 69, 146–150. [Google Scholar] [CrossRef] [PubMed]
  22. Zhen, W.; Manji, R.; Smith, E.; Wuitschick, J.; Lucic, D.; Berry, G.J. Evaluation of the Alinity m Resp-4-Plex Assay for the Detection of Severe Acute Respiratory Syndrome Coronavirus 2, Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus. Microbiol. Spectr. 2022, 10, e0109021. [Google Scholar] [CrossRef] [PubMed]
  23. Quinton, M.; Geahr, M.; Gluck, L.; Jarrett, J.; Mostafa, H.H. Evaluation of the respiratory NeuMoDx™ Flu A-B/RSV/SARS-CoV-2 Vantage and Alinity m Resp-4-Plex assay. J. Clin. Virol. Off. Publ. Pan Am. Soc. Clin. Virol. 2022, 150–151, 105164. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Correlation of Ct values between Xpert Xpress CoV-2/Flu/RSV plus (x axis) and Resp-4-Plex Assay (y axis) for (A) SARS-CoV-2, (B) influenza A virus, (C) influenza B virus, and (D) RSV.
Figure 1. Correlation of Ct values between Xpert Xpress CoV-2/Flu/RSV plus (x axis) and Resp-4-Plex Assay (y axis) for (A) SARS-CoV-2, (B) influenza A virus, (C) influenza B virus, and (D) RSV.
Diagnostics 14 00683 g001
Table 1. Performance of Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex on clinical specimens.
Table 1. Performance of Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex on clinical specimens.
Pathogen TargetsFNFPTNTPPPA95% CI for
PPA (%)
NPA95% CI for
NPA (%)
Remarks
Xpert Xpress CoV-2/Flu/RSV plusFor CoV-2/Flu/RSV plus, 1 specimen was insufficient for testing and the result of 1 specimen was invalid. Both specimens were excluded from analysis.

For Resp-4-Plex, 25 specimens were insufficient for testing, and the results of 2 specimens were invalid. A total of 27 specimens were excluded from analysis.

Specimens 30, 32, 33, and 34 were positive for SARS-CoV-2 variant XBB.1, with C26270T and A26275G mutations detected in the E gene. The E gene Ct values of the SOC Xpert Xpress SARS-CoV-2 were 0. The N2 Ct values ranged from 25.8 to 34.3.

For CoV-2/Flu/RSV plus, error 5007 was encountered in the first run of specimen 47, and the repeated run was valid.

SARS-CoV-2 and influenza A virus were codetected in specimen 106.

SARS-CoV-2 and RSV were codetected in specimen 107.

Influenza A virus and RSV were codetected in Specimen 108.

For Resp-4-Plex, specimen 117 was positive for RSV (Ct value: 35.96). Specimens 122 and 124 were positive for SARS-CoV-2 (Ct values: 38.34 and 37.74). CoV-2/Flu/RSV plus and the SOC method RP2.1plus were negative for all 4 virus targets.

CoV-2/Flu/RSV plus:
Percentage of agreement: 100%
Cohen’s κ: 1 (perfect agreement)

Resp-4-Plex:
Percentage of agreement: 97.71%
Cohen’s κ: 0.92 (almost perfect agreement)
SARS-CoV-2001173910089.32 to 100.0010096.18 to 100.00
Influenza A virus001302610084.76 to 100.0010096.55 to 100.00
Influenza B virus001332310083.09 to 100.0010096.63 to 100.00
RSV001164010089.56 to 100.0010096.15 to 100.00
Overall002812810096.50 to 100.0010085.70 to 100.00
Alinity m Resp-4-Plex
SARS-CoV-202913810089.07 to 100.0097.8592.03 to 99.88
Influenza A virus001052610084.76 to 100.0010095.76 to 100.00
Influenza B virus00127410045.41 to 100.0010096.47 to 100.00
RSV01933710088.80 to 100.0098.9493.64 to >99.99
Overall032310510095.76 to 100.0088.4670.20 to96.82
CI, confidence interval; Ct, threshold cycle; FN, false negative; FP, false positive; NPA, negative percent agreement; PPA, positive percent agreement; RSV, respiratory syncytial virus; SARS-CoV-2, severe acute respiratory syndrome coronavirus 2; SOC, standard-of-care; TN, true negative; TP, true positive.
Table 2. Performance of Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex on EQAP samples.
Table 2. Performance of Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex on EQAP samples.
SamplesExpected ResultsXpert Xpress CoV-2/Flu/RSV plus Ct ValuesResp-4-Plex Ct Values
SARS-CoV-2Flu A1Flu A2Flu BRSVCOV2FLUAFLUBRSV
EQAP 1, sample 1SARS-CoV-226000025.09NDNDND
EQAP 1, sample 2Not detected00000NDNDNDND
EQAP 1, sample 3SARS-CoV-228.8000028.1NDNDND
EQAP 1, sample 4SARS-CoV-233.1000032.05NDNDND
EQAP 2, sample 1Influenza A virus019.121.600The samples were not sufficient
for testing by Resp-4-Plex
EQAP 2, sample 2RSV000022.7
EQAP 2, sample 3Not detected00000
EQAP 2, sample 4SARS-CoV-228.60000
EQAP 2, sample 5Influenza B virus00018.60
ND, not detected.
Table 3. Detection rate and precision for each dilution of the reference material.
Table 3. Detection rate and precision for each dilution of the reference material.
Virus Concentration (Copies/mL)TargetXpert Xpress CoV-2/Flu/RSV plusResp-4-Plex
# Proportion DetectedCt Value Range* CV# Proportion DetectedCt Value Range* CV
10SARS-CoV-24/541.2–44.1N/A3/536.18–36.85N/A
Flu A1/5FluA1: 38.2–44.5
FluA2: N/A
N/A3/536.99–39.44N/A
Flu B0/5N/AN/A2/537.55–38N/A
RSV0/538.7 for 1 replicateN/A3/537.68–37.98N/A
25SARS-CoV-25/537.8–40.22.21%5/535.77–39.523.76%
Flu A4/5FluA1: 34.9–38.7
FluA2: 37.2–41.8
N/A3/536.76–38.14N/A
Flu B2/537.5–43.4N/A1/537.97 for 1 replicateN/A
RSV2/536.7–41.5N/A1/536.05 for 1 replicateN/A
50SARS-CoV-25/537.8–39.51.72%5/534.46–34.940.58%
Flu A4/5FluA1: 34.1–38.8
FluA2: 37.4–42.1
N/A5/535.55–37.12N/A
Flu B2/538.6–42.9N/A5/535.47–37.74N/A
RSV0/539.9–43.7N/A4/535.55–36.2N/A
100SARS-CoV-25/536.1–37.31.07%20/2033.24–35.881.71%
Flu A5/5FluA1: 33–35.8
FluA2: 36.9–39.9
2.60%
2.63%
20/2033.98–38.012.91%
Flu B4/535.9–37.7N/A20/2034.42–36.791.82%
RSV2/537.7–43.2N/A20/2033.71–37.963.17%
200SARS-CoV-220/2035.2–36.71.04%5/532.95–33.720.87%
Flu A20/20FluA1: 32.8–34.4
FluA2: 36.5–38.8
1.28%
1.46%
5/534.04–35.010.97%
Flu B20/2035.9–39.52.21%5/534.37–35.10.75%
RSV15/2036.3–39.6N/A5/533.71–34.410.78%
300SARS-CoV-213/1335–35.70.70%5/532.18–33.361.21%
Flu A13/13FluA1: 32.2–34.2
FluA2: 35.6–37.6
1.55%
1.72%
5/533.4–34.981.66%
Flu B13/1335.1–37.21.54%5/533.97–34.560.57%
RSV11/1336.1–38.2N/A5/532.23–34.292.03%
400SARS-CoV-220/2034.4–35.70.85%5/532.37–32.950.64%
Flu A20/20FluA1: 32.3–33.3
FluA2: 35.4–37.8
0.87%
1.68%
5/533.87–34.70.89%
Flu B20/2034.6–37.11.59%5/533.32–34.961.59%
RSV20/2035.2–382.30%5/531.98–33.612.18%
* CV, coefficient of variation. #, Proportion detected is represented by ‘number of detected replicates/total number of replicates tested’. N/A, not applicable.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Chan, W.-S.; Wong, K.-P.; Yau, S.-K.; Wong, C.-Y.; Chan, T.-C.; Hung, J.; Lai, K.T.-W.; Leung, C.-P.; Wang, C.L.-N.; Au, C.-H.; et al. Clinical Evaluation of Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex Assay. Diagnostics 2024, 14, 683. https://doi.org/10.3390/diagnostics14070683

AMA Style

Chan W-S, Wong K-P, Yau S-K, Wong C-Y, Chan T-C, Hung J, Lai KT-W, Leung C-P, Wang CL-N, Au C-H, et al. Clinical Evaluation of Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex Assay. Diagnostics. 2024; 14(7):683. https://doi.org/10.3390/diagnostics14070683

Chicago/Turabian Style

Chan, Wai-Sing, Kan-Pui Wong, Siu-Kei Yau, Ching-Yan Wong, Tsz-Ching Chan, Jeffrey Hung, Kristi Tsz-Wan Lai, Chin-Pang Leung, Candy Ling-Na Wang, Chun-Hang Au, and et al. 2024. "Clinical Evaluation of Xpert Xpress CoV-2/Flu/RSV plus and Alinity m Resp-4-Plex Assay" Diagnostics 14, no. 7: 683. https://doi.org/10.3390/diagnostics14070683

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop