Industrial Calibration Procedure for Confocal Microscopes
Abstract
:1. Introduction
- It adds the Z-axis to traditional measuring optical microscopes, which only work in the XY plane.
- It allows analysis of the 3D geometry of the object surface and characterization of its quality from data points acquired while scanning it.
- Its lateral resolution is better than in traditional optical microscopy.
- It permits more precise 3D images of the objects being measured to be obtained that are of higher quality and in less time compared to other methods. This allows many useful measurements to be carried out in short intervals of time.
- Transparent specimens can be observed, as can sections with a certain thickness, without the need to section the object under study.
- Metrological characteristics of the instrument: measurement noise, flatness deviation, non-linearity errors, amplification coefficients, and perpendicularity errors between axes.
- Instrument geometry: alignment of components and the XY stage and rotary stage error motions.
- Source characteristics: focal spot size and drift.
- Detector characteristics: pixel response, uniformity and linearity, detector offset, and bad pixels.
- Reconstruction and data processing: surface determinations, data representation, and calculation approaches.
- Environmental conditions: temperature, humidity, and vibration.
2. Materials and Methods
- Leica DCM3D confocal microscope (Wetzlar, Germany) with a 10× objective (EPI-L, NA = 0,30). Field of view 1270 µm × 952 µm (768 × 576 pixels); 1.65 µm nominal voxel width. The overall range of the Z-axis is 944 µm using 2 µm axial steps (voxel height), but the instrument is used in a reduced working range of only 100 µm.
- SensoSCAN—LeicaSCAN DCM3D 3.41.0 software developed by Sensofar Tech Ltd. (Terrassa, Spain).
- Calibration of the X and Y scales, using a stage micrometer as a reference measurement standard.
- Estimation of perpendicularity error between X and Y axes.
- Estimation of the flatness deviation of the focal plane using an optical flat.
- Calibration of Z scale using a calibrated steel sphere.
- Calibration of the confocal microscope for the measurement of 2D roughness using periodic and aperiodic 2D roughness measurement standards.
- Easy to find.
- Easy to calibrate with low enough uncertainties in National Measurement Institutes (NMIs) or preferably in accredited calibration laboratories (ACLs).
- Stable mechanical artifacts that could guarantee long recalibration intervals.
- Common in the field of dimensional metrology in order to facilitate their acquisition, calibration, and correct use.
2.1. Flatness Verification
2.2. XY Plane Calibration
2.3. Z-Axis Calibration
2.4. Calibration for Roughness Measurements
- The way the surface is detected is totally different: microscopes use light, and stylus instruments use a mechanical tip. Usually, optical instruments present higher instrument noise than stylus instruments. Possible reasons are the effects of multiple scattering and discontinuities [45]. As a consequence, optical instruments tend to overestimate surface roughness.
- Stylus instruments permit evaluation lengths that are as long as necessary (see ISO 4288 [44]). Microscopes usually have small fields of view that limit the maximum length of the profile that can be scanned. For example, for samples with 0.1 µm 2 µm, ISO 4288 recommends using five sampling lengths mm for a total evaluation length mm. This is not a problem for stylus instruments, which can cope with longer evaluation lengths (up to 100 mm in some cases). However, the confocal microscope described at the beginning of Section 2 has a maximum evaluation length of 1.27 mm. Therefore, only one sampling length mm could be used. Using only one sampling length instead of five usually causes a bias toward lower values, which are accompanied by an increase in variability. The effect is considerably higher when even the sampling length has to be reduced.
2.5. Summary of Characteristics of Measurement Standards Used during Calibration
3. Results
3.1. Flatness Verification
3.2. XY Plane Calibration
3.3. Z-Axis Calibration
3.4. Calibration for Roughness Measurements
- Dominant contributions to uncertainties are the calibration uncertainties of the roughness standards.
- There is a high probability that all roughness standards were calibrated in the same calibration laboratory. Therefore, there will be strong correlation between them.
4. Discussion
4.1. Expanded Uncertainty Estimation for Length Measurements in the XY Plane
4.2. Expanded Uncertainty Estimation for Height Measurements along the Z-Axis
4.3. Expanded Uncertainty for Roughness Measurements
4.4. Propagation of Uncertainty When Measuring the Radius of a Cylindrical Surface
5. Conclusions
- Amplification coefficients , , and .
- Non-linearity errors.
- Perpendicularity error between X and Y axes.
- Relative difference in pixel dimensions along X and Y axes.
- Repeatabilities when measuring lengths or heights.
- Flatness deviations in the XY plane.
- Bias deviation when measuring roughness.
- Instrument noise when measuring roughness.
- Repeatability when measuring roughness.
Author Contributions
Funding
Conflicts of Interest
Appendix A. Algorithm for Spherical Cap Fitting
References
- Carmignato, S.; Voltan, A.; Savio, E. Metrological performance of optical coordinate measuring machines under industrial conditions. CIRP Ann. Manuf. Technol. 2010, 59, 497–500. [Google Scholar] [CrossRef]
- European Committee for Standardization (CEN). ISO 3274:1996 Geometrical Product Specifications (GPS) Surface Texture: Profile Method. Nominal Characteristics of Contact (stylus) Instruments; CEN: Bruxelles, Belgium, 1996. [Google Scholar]
- Leach, R.K. Towards a complete framework for calibration of optical surface and coordinate measuring instruments. In Proceedings of the SPIE Optical Metrology Plenary Session, ICM, Munich, Germany, 26 June 2019. [Google Scholar]
- European Committee for Standardization (CEN). ISO 25178-6:2010 Geometrical Product Specification (GPS) Surface Texture: Areal—Part 6: Classification Methods for Measuring Surface Texture; CEN: Bruxelles, Belgium, 2010. [Google Scholar]
- Giusca, C.; Leach, R.K. Measurement Good Practice Guide (No. 128): Calibration of the Metrological Characteristics of Imaging Confocal Microscopes (ICMs); National Physical Laboratory (NPL): Teddington, UK, 2012. [Google Scholar]
- Minsky, M. Memoir on Inventing the Confocal Scanning Microscope. Scanning 1988, 10. [Google Scholar] [CrossRef]
- Claxton, N.S.; Fellers, T.J.; Davidson, M.W. Laser scanning confocal microscopy. In Encyclopedia of Medical Devices and Instrumentation, 2nd ed.; Webster, J.G., Ed.; John Wiley & Sons, Inc.: Hoboken, NJ, USA, 2006; pp. 1–37. [Google Scholar] [CrossRef]
- Watson, T. Fact and Artefact in Confocal Microscopy. Adv. Dent. Res. 1997, 11, 128–138. [Google Scholar] [CrossRef]
- Cheng, C.; Wang, J.; Leach, R.K.; Lu, W.; Liu, X.; Jiang, X. Corrected parabolic fitting for height extraction in confocal microscopy. Opt. Express 2019, 27, 3682–3697. [Google Scholar] [CrossRef]
- Alburayt, A.; Syam, W.P.; Leach, R.K. Lateral scale calibration for focus variation microscopy. Meas. Sci. Technol. 2019, 29. [Google Scholar] [CrossRef]
- Joint Committee for Guides in Metrology (JCGM). International Vocabulary of Metrology (VIM)—Basic and General Concepts and Associated Terms, 3rd ed.; JCGM: Paris, France, 2012. [Google Scholar]
- Leach, R.K.; Giusca, C.; Haitjema, H.; Evans, C.; Jiang, X. Calibration and verification of areal surface texture measuring instruments. CIRP Ann. Manuf. Technol. 2015, 64, 797–813. [Google Scholar] [CrossRef]
- Caja, J.; Sanz, A.; Maresca, P.; Fernández, T.; Wang, C. Some Considerations about the Use of Contact and Confocal Microscopy Methods in Surface Texture Measurement. Materials 2018, 11, 1484. [Google Scholar] [CrossRef] [Green Version]
- Wang, C.; Gómez, E.; Yu, Y. Characterization and correction of the geometric errors in using confocal microscope for extended topography measurement. Part I: Models, Algorithms Development and Validation. Electronics 2019, 8, 733. [Google Scholar] [CrossRef] [Green Version]
- Wang, C.; Gómez, E.; Yu, Y. Characterization and Correction of the Geometric Errors Using a Confocal Microscope for Extended Topography Measurement, Part II: Experimental Study and Uncertainty Evaluation. Electronics 2019, 8, 1217. [Google Scholar] [CrossRef] [Green Version]
- Webb, R.H. Confocal optical microscopy. Rep. Prog. Phys. 1996, 59, 427–471. [Google Scholar] [CrossRef]
- Webb, R.H. Theoretical Basis of Confocal Microscopy. Meth. Enzymol. 1999, 307, 3–20. [Google Scholar] [CrossRef] [PubMed]
- Salerni, G. Uso de la microscopía confocal de reflectancia en dermatología. Dermatol. Argent. 2011, 17, 230–235. [Google Scholar]
- Confocal Microscopy, Molecular Expressions. Available online: https://micro.magnet.fsu.edu/primer/techniques/confocal/index.html (accessed on 26 October 2019).
- Leach, R.K.; Bourell, D.; Carmignato, S.; Donmez, A.; Senin, N.; Dewulf, W. Geometrical metrology for metal additive manufacturing. CIRP Ann. Manuf. Technol. 2019, 68, 677–700. [Google Scholar] [CrossRef]
- Introduction to Confocal Microscopy. Available online: https://www.olympus-lifescience.com/en/microscope-resource/primer/techniques/confocal/confocalintro/ (accessed on 26 October 2019).
- Tata, B.V.R.; Raj, B. Confocal laser scanning microscopy: Applications in material science and technology. Bull. Mater. Sci. 1998, 21, 263–278. [Google Scholar] [CrossRef] [Green Version]
- Cohen-Or, D.; Kaufmann, A. Fundamentals of Surface Voxelization. CVGIP: Graph. Model. Image Process. 1995, 57, 453–461. [Google Scholar] [CrossRef]
- Joint Committee for Guides in Metrology, Working Group 1 (JCGM/WG 1). JCGM 100:2008 Evaluation of Measurement Data—Guide to the Expression of Uncertainty in Measurement (GUM), 1st ed.; JCGM: Paris, France, 2008. [Google Scholar]
- European Accreditation (EA). EA-4/02—Evaluation of the Uncertainty of Measurement in Calibration; EA: London, UK, 2013. [Google Scholar]
- European Committee for Standardization (CEN). ISO/IEC 17025:2017 General Requirements for the Competence of Testing and Calibration; CEN: Bruxelles, Belgium, 2017. [Google Scholar]
- SENSOFAR-TECH LTD. Application Note: Flatness Error on Imaging Confocal Microscopes; Sensofar: Tarrasa, Spain, 2009. [Google Scholar]
- Centro Español de Metrología (CEM). Procedimiento DI-035 para la Calibración de Patrones de Planitud de Vidrio; CEM: Madrid, Spain, 2004. [Google Scholar]
- NIST/SEMATECH e-Handbook of Statistical Methods. Available online: http://www.itl.nist.gov/div898/handbook/ (accessed on 27 November 2019).
- De Vicente, J.; Molpeceres, C.; Guarneros, O.; García-Ballesteros, J. Calibración de Microscopios Confocales; XVII National Congress of Mechanical Engineering: Gijón, Spain, 2008. [Google Scholar]
- Guarneros, O.; De Vicente, J.; Maya, M.; Ocaña, J.L.; Molpeceres, C.; García-Ballesteros, J.; Rodríguez, S.; Duran, H. Uncertainty Estimation for Performance Evaluation of a Confocal Microscope as Metrology Equipment. MAPAN—J. Metrol. Soc. I 2014, 29, 29–42. [Google Scholar] [CrossRef] [Green Version]
- De Vicente, J.; Sánchez-Pérez, A.M.; Maresca, P.; Caja, J.; Gómez, E. A model to transform a commercial flatbed scanner into a two-coordinates measuring machine. Measurement 2015, 73, 304–312. [Google Scholar] [CrossRef]
- European Committee for Standardization (CEN). ISO 25178-70 Geometrical Product Specification (GPS) Surface Texture: Areal—Part 70: Material Measures; CEN: Bruxelles, Belgium, 2014. [Google Scholar]
- Centro Español de Metrología (CEM). Procedimiento DI-004 para la Calibración de Medidoras de una Coordenada Vertical; CEM: Madrid, Spain, 2013. [Google Scholar]
- Wang, C.; Caja, J.; Gómez, E.; Maresca, P. Procedure for Calibrating the Z-axis of a Confocal Microscope: Application for the Evaluation of Structured Surfaces. Sensors 2019, 19, 527. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- European Committee for Standardization (CEN). ISO 5436-1:2000 Geometrical Product Specipications (GPS) Surface Texture: Profile Method. Measurement Standards—Part 1: Material Measures; CEN: Bruxelles, Belgium, 2000. [Google Scholar]
- Calibration standards, HALLE Präzisions-Kalibriernormale GmbH. Available online: http://halle-normale.de/framesets/englisch/products/products.html (accessed on 29 October 2019).
- Balcon, M.; Carmignato, S.; Savio, E. Performance verification of a confocal microscope for 3D metrology tasks. Qual.—Access Success 2012, 13, 63–66. [Google Scholar]
- De Vicente, J.; Raya, F. Simplified Statistical Method for Uncertainty Estimation in Coordinate Metrology. In Proceedings of the 9th International Metrology Congress, Bordeaux, France, 18–21 October 1999. [Google Scholar]
- Deutscher Kalibrierdienst (DKD). Guideline DKD 4-2 Calibration of Measuring Instruments and Standards for Roughness Measuring Technique (Sheet 1: Calibration of Standards for Roughness Measuring Technique); DKD: Braunschweig, Germany, 2011. [Google Scholar]
- Deutscher Kalibrierdienst (DKD). Guideline DKD-R 4-2 Calibration of Devices and Standards for Roughness Metrology (Sheet 2: Calibration of the Vertical Measuring System of Stylus Instrument); DKD: Braunschweig, Germany, 2011. [Google Scholar]
- Deutscher Kalibrierdienst (DKD). Guideline DKD-R 4-2 Calibration of Devices and Standards for Roughness Metrology (Sheet 3: Calibration of Standards with Periodic Profiles in Horizontal Direction by Means of Stylus Instrument); DKD: Braunschweig, Germany, 2011. [Google Scholar]
- European Committee for Standardization (CEN). ISO 4287:1999 Geometrical Product Specifications (GPS) Surface Texture: Profile Method. Terms, Definitions and Surface Texture Parameters; CEN: Bruxelles, Belgium, 1999. [Google Scholar]
- European Committee for Standardization (CEN). ISO 4288:1996 Geometrical Product Specifications (GPS) Surface Texture: Profile Method. Rules and Procedures for the Assessment of Surface Texture; CEN: Bruxelles, Belgium, 1996. [Google Scholar]
- Leach, R.K.; Haitjema, H. Bandwidth characteristics and comparisons of surface texture measuring instruments. Meas. Sci. Technol. 2010, 21. [Google Scholar] [CrossRef]
- Joint Committee for Guides in Metrology, Working Group 1 (JCGM/WG 1). JCGM 101:2008 Evaluation of Measurement Data—Supplement 1 to the “Guide to the Expression of Uncertainty in Measurement”—Propagation of Distributions using a Monte Carlo Method (GUM-S1), 1st ed.; JCGM: Paris, France, 2008. [Google Scholar]
Reference Measurement Std. | Parameter | Certified Value (µm) | Std. Uncertainty (μm) |
---|---|---|---|
Optical flat | Total flatness error | 0.118 | 0.025 |
RMS flatness | 0.028 | 0.007 | |
Stage Micrometer | Average pitch | 9.980 | 0.005 |
Sphere | Diameter | 4 001.08 | 0.25 |
Roughness std. #1 metallic, aperiodic | () 1 | 0.183 48 | 0.039 |
Roughness std. #2 metallic, aperiodic | () 1 | 0.512 185 | 0.041 |
Roughness std. #3 metallic, aperiodic | () 1 | 1.677 176 | 0.057 |
Roughness std. #4 glass, periodic | () 1 | 0.460 100 | 0.030 |
Roughness std. #5 metallic, aperiodic | () 1 | 0.850 120 | 0.030 |
Roughness std. #6 glass, periodic | () 1 | 2.440 200 | 0.080 |
Position | RMS Flatness (μm) |
---|---|
0° | 0.48 |
90° | 0.59 |
Position | (μm) | Non-Linearity RMS (μm) | |||
---|---|---|---|---|---|
1 | 0° | 9.892 34 | 0.000 53 | 0.34 | 0.71 |
2 | 45° | 9.897 33 | 0.000 57 | 0.38 | 0.60 |
3 | 90° | 9.891 56 | 0.000 49 | 0.42 | 0.69 |
4 | 135° | 9.889 50 | 0.000 47 | 0.34 | 0.61 |
Position | Illumination | RMS Error (μm) | |
---|---|---|---|
0° | Blue | 0.86 | 3.9740 |
45° | Blue | 1.08 | 3.9562 |
90° | Blue | 1.08 | 3.9638 |
0° | White | 0.86 | 3.9740 |
45° | White | 0.89 | 3.9828 |
90° | White | 0.87 | 3.9766 |
Reference Meas. Std. | Average Ra (μm) | Repeatability | Corrected Ra | Bias Estimation (μm) | |
---|---|---|---|---|---|
Roughness std. #1 | 0.43 | 0.06 | 0.43 | 0.25 | 0.04 |
Roughness std. #2 | 0.59 | 0.06 | 0.60 | 0.08 | 0.05 |
Roughness std. #3 | 1.70 | 0.11 | 1.71 | 0.04 | 0.07 |
Roughness std. #4 | 0.51 | 0.04 | 0.52 | 0.06 | 0.03 |
Roughness std. #5 | 0.95 | 0.05 | 0.96 | 0.11 | 0.03 |
Roughness std. #6 1 | 2.50 | 0.06 | 2.53 | 0.09 | 0.08 |
Parameter | Value | Units | Standard Uncertainty |
---|---|---|---|
0.008 83 | - | 0.00050 | |
−0.000040 | - | 0.000036 | |
−0.000798 | - | 0.000074 | |
0.0101 | - | 0.0014 | |
0.72 | - | - | |
0.54 | µm | - | |
0.70 | µm | - | |
0.40 | µm | - | |
0.80 1 | µm | - | |
0.11 | µm | 0.05 | |
0 | µm | 0.06 | |
0.07 | µm | ||
0.09 | µm |
Variable | Mean Value | Units | Standard Uncertainty | Distribution Type |
---|---|---|---|---|
0.00883 | - | 0.00050 | Normal | |
0 1 | - | 0.000036 | Normal | |
0 | Uniform | |||
0 1 | - | 0.000074 | Normal | |
0 | Uniform | |||
0.0101 | - | 0.0014 | Normal | |
0.72 | - | - | - | |
0 | µm | Normal | ||
0 | µm | Normal | ||
© 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
Share and Cite
Mínguez Martínez, A.; de Vicente y Oliva, J. Industrial Calibration Procedure for Confocal Microscopes. Materials 2019, 12, 4137. https://doi.org/10.3390/ma12244137
Mínguez Martínez A, de Vicente y Oliva J. Industrial Calibration Procedure for Confocal Microscopes. Materials. 2019; 12(24):4137. https://doi.org/10.3390/ma12244137
Chicago/Turabian StyleMínguez Martínez, Alberto, and Jesús de Vicente y Oliva. 2019. "Industrial Calibration Procedure for Confocal Microscopes" Materials 12, no. 24: 4137. https://doi.org/10.3390/ma12244137