<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD Journal Publishing DTD v2.3 20070202//EN" "journalpublishing.dtd">
<article xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" xml:lang="en" article-type="research-article">
<front>
<journal-meta>
<journal-id journal-id-type="nlm-ta">Sensors</journal-id>
<journal-title>Sensors</journal-title>
<issn pub-type="epub">1424-8220</issn>
<publisher>
<publisher-name>Molecular Diversity Preservation International (MDPI)</publisher-name></publisher></journal-meta>
<article-meta>
<article-id pub-id-type="doi">10.3390/s121115063</article-id>
<article-id pub-id-type="publisher-id">sensors-12-15063</article-id>
<article-categories>
<subj-group>
<subject>Article</subject></subj-group></article-categories>
<title-group>
<article-title>Indirect Determination of Mercury Ion by Inhibition of a Glucose Biosensor Based on ZnO Nanorods</article-title></title-group>
<contrib-group>
<contrib contrib-type="author">
<name><surname>Chey</surname><given-names>Chan Oeurn</given-names></name><xref ref-type="corresp" rid="c1-sensors-12-15063"><sup>*</sup></xref></contrib>
<contrib contrib-type="author">
<name><surname>Ibupoto</surname><given-names>Zafar Hussain</given-names></name></contrib>
<contrib contrib-type="author">
<name><surname>Khun</surname><given-names>Kimleang</given-names></name></contrib>
<contrib contrib-type="author">
<name><surname>Nur</surname><given-names>Omer</given-names></name></contrib>
<contrib contrib-type="author">
<name><surname>Willander</surname><given-names>Magnus</given-names></name></contrib>
<aff id="af1-sensors-12-15063">Physical Electronics and Nanotechnology Division, Department of Science and Technology, Campus Norrköping, Linköping University, SE-60174 Norrköping, Sweden; E-Mails: <email>zafar.hussain.ibupoto@liu.se</email> (Z.H.I.); <email>kimleang.khun@liu.se</email> (K.K.); <email>omer.nour@liu.se</email> (O.N.); <email>magnus.willander@liu.se</email> (M.W.)</aff></contrib-group>
<author-notes>
<corresp id="c1-sensors-12-15063">
<label>*</label> Author to whom correspondence should be addressed; E-Mail: <email>chan.oeurn.chey@liu.se</email>; Tel.: +46-11-363-167; Fax: +46-11-363-270.</corresp></author-notes>
<pub-date pub-type="collection">
<year>2012</year></pub-date>
<pub-date pub-type="epub">
<day>06</day>
<month>11</month>
<year>2012</year></pub-date>
<volume>12</volume>
<issue>11</issue>
<fpage>15063</fpage>
<lpage>15077</lpage>
<history>
<date date-type="received">
<day>07</day>
<month>09</month>
<year>2012</year></date>
<date date-type="rev-recd">
<day>17</day>
<month>10</month>
<year>2012</year></date>
<date date-type="accepted">
<day>02</day>
<month>11</month>
<year>2012</year></date></history>
<permissions>
<copyright-statement>© 2012 by the authors; licensee MDPI, Basel, Switzerland.</copyright-statement>
<copyright-year>2012</copyright-year>
<license>
<p>This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/).</p></license></permissions>
<abstract>
<p>A potentiometric glucose biosensor based on immobilization of glucose oxidase (GOD) on ZnO nanorods (ZnO-NRs) has been developed for the indirect determination of environmental mercury ions. The ZnO-NRs were grown on a gold coated glass substrate by using the low temperature aqueous chemical growth (ACG) approach. Glucose oxidase in conjunction with a chitosan membrane and a glutaraldehyde (GA) were immobilized on the surface of the ZnO-NRs using a simple physical adsorption method and then used as a potentiometric working electrode. The potential response of the biosensor between the working electrode and an Ag/AgCl reference electrode was measured in a 1mM phosphate buffer solution (PBS). The detection limit of the mercury ion sensor was found to be 0.5 nM. The experimental results provide two linear ranges of the inhibition from 0.5 × 10<sup>−6</sup> mM to 0.5 × 10<sup>−4</sup> mM, and from 0.5 × 10<sup>−4</sup> mM to 20 mM of mercury ion for fixed 1 mM of glucose concentration in the solution. The linear range of the inhibition from 10<sup>−3</sup> mM to 6 mM of mercury ion was also acquired for a fixed 10 mM of glucose concentration. The working electrode can be reactivated by more than 70% after inhibition by simply dipping the used electrode in a 10 mM PBS solution for 7 min. The electrodes retained their original enzyme activity by about 90% for more than three weeks. The response to mercury ions was highly sensitive, selective, stable, reproducible, and interference resistant, and exhibits a fast response time. The developed glucose biosensor has a great potential for detection of mercury with several advantages such as being inexpensive, requiring minimum hardware and being suitable for unskilled users.</p></abstract>
<kwd-group>
<kwd>potentiometric inhibition biosensor</kwd>
<kwd>mercury</kwd>
<kwd>glucose oxidase</kwd>
<kwd>ZnO nanorods</kwd></kwd-group></article-meta></front>
<body>
<sec sec-type="intro">
<label>1.</label>
<title>Introduction</title>
<p>Among heavy metals, mercury metal, which is found in biological materials, natural water, soil, air, chemicals and waste products, is highly toxic, resulting in interference and disturbance of natural systems as well as the production of damaging effects to the environment [<xref ref-type="bibr" rid="b1-sensors-12-15063">1</xref>]. The toxicity of mercury and its compounds produces harmful effects on the central nervous system and causes neuropsychiatric disorders [<xref ref-type="bibr" rid="b2-sensors-12-15063">2</xref>], so the determination of mercury in environmental samples is therefore very important. There are several techniques which are normally used for the determination of mercury in environmental and biological samples, including atomic absorption spectrometry (AAS), stripping voltammetry, inductively coupled plasma spectrometry, cold vapour atomic fluorescence spectrometry (CVAFS), and cold vapour atomic absorption spectrometry (CVAAS), but these techniques require sample pretreatment, expensive instrumentation, complicated devices and require operation by skilled operators, hence these methods are not suitable for on-site testing and monitoring tasks [<xref ref-type="bibr" rid="b1-sensors-12-15063">1</xref>–<xref ref-type="bibr" rid="b3-sensors-12-15063">3</xref>]. Therefore, a simple new technology is needed with appropriate high capability to monitor mercury with fast response, while being inexpensive and making on-site monitoring possible. Biosensors are useful analytical tools for developing sensors in order to meet these requirements [<xref ref-type="bibr" rid="b4-sensors-12-15063">4</xref>–<xref ref-type="bibr" rid="b6-sensors-12-15063">6</xref>]. There are many types of biosensors for the detection of heavy metals such as purified protein-based, antibody-based, whole-cell-based, and both enzyme inhibition and activation-based methods [<xref ref-type="bibr" rid="b6-sensors-12-15063">6</xref>]. Different electrochemical techniques are used for the detection of trace mercury in different enzyme reactions using different electrodes which have demonstrated great potential as effective tools to determine heavy metals in environment samples. Examples of these are the detection of mercury ion by an amperometric method using glucose oxidase immobilized on a carbon paste electrode [<xref ref-type="bibr" rid="b7-sensors-12-15063">7</xref>], Pt electrode [<xref ref-type="bibr" rid="b8-sensors-12-15063">8</xref>], thiolate self-assembled monolayer [<xref ref-type="bibr" rid="b9-sensors-12-15063">9</xref>], poly-<italic>o</italic>-phenylenediamine (Pt/PPD/GOD) [<xref ref-type="bibr" rid="b10-sensors-12-15063">10</xref>,<xref ref-type="bibr" rid="b11-sensors-12-15063">11</xref>], cylindrical carbon film electrodes [<xref ref-type="bibr" rid="b12-sensors-12-15063">12</xref>] and bienzyme electrodes based on their competitive activities [<xref ref-type="bibr" rid="b13-sensors-12-15063">13</xref>]. Moreover, the determination of mercury ion using urease immobilized on the surface of an iridium oxide pehametric detector, self-assembled gold nanoparticles, nanostructured polyaniline-Nafion/Au/Al<sub>2</sub>O<sub>3</sub> electrode and on polymeric membrane have been reported [<xref ref-type="bibr" rid="b14-sensors-12-15063">14</xref>–<xref ref-type="bibr" rid="b17-sensors-12-15063">17</xref>]. Furthermore, studies of mercury inhibition based on invertase enzyme immobilized on a copper-based electrode and GOD-modified platinum electrode [<xref ref-type="bibr" rid="b18-sensors-12-15063">18</xref>] and a fast spectrometric method based on the inhibition of glucose-oxidase by mercury have been reported [<xref ref-type="bibr" rid="b19-sensors-12-15063">19</xref>].</p>
<p><xref ref-type="table" rid="t1-sensors-12-15063">Table 1</xref> lists some of the mercury determinations using different electrodes with immobilized glucose oxidase. From these reports, it is clearly observed that glucose oxidase is one of the most promising enzymes. It can be used for the indirect determination of mercury ions by different methods. However, there is no report using a ZnO nanomaterial in a study of mercury inhibition of glucose oxidase using a potentiometric technique. An advantage of the potentiometric technique over amperometric techniques is that for living biological samples, no current passes, and only the accumulation of charge is measured.</p>
<p>ZnO is a unique material that forms a diverse family of nanostructures such as nano-combs, nano-rings, nano-helixes, nano-bows, nano-belts, nanowires, and nano-cages and it exhibits multiple semiconducting, piezoelectric, and pyroelectric properties [<xref ref-type="bibr" rid="b20-sensors-12-15063">20</xref>]. Furthermore, one dimensional (1D) ZnO nanostructures exhibit remarkable properties for sensing applications due to their high surface to volume ratio, high catalytic efficiency, non-toxicity, biocompatibility, chemical stability, strong adsorption ability because of the high isoelectric point (IEP ∼ 9.5) [<xref ref-type="bibr" rid="b21-sensors-12-15063">21</xref>], bio-safety and high ionic characteristics (60%). In addition, ZnO does not dissolve at biological pH [<xref ref-type="bibr" rid="b22-sensors-12-15063">22</xref>–<xref ref-type="bibr" rid="b24-sensors-12-15063">24</xref>] and it also provides fast electron transfer properties [<xref ref-type="bibr" rid="b21-sensors-12-15063">21</xref>,<xref ref-type="bibr" rid="b25-sensors-12-15063">25</xref>]. Moreover, the advantages of using ZnO nanostructures for sensing applications are their high sensitivity, and time domain chemical sensing for low concentrations and the possibility of sensing in single cells or molecule detection available in small volumes at low concentration [<xref ref-type="bibr" rid="b26-sensors-12-15063">26</xref>]. Such advantages cannot be achieved simultaneously using large sized sensors. Moreover, ZnO-NRs can be grown on flexible plastic substrates [<xref ref-type="bibr" rid="b27-sensors-12-15063">27</xref>] which have excellent mechanical properties and can be suitable for medical and implantable biosensors. In this work, we have successfully presented the first potentiometric glucose biosensor made by the functionalization of a ZnO-NR array for studying the inhibition of mercury by glucose oxidase and perform simple and rapid determination of Hg<sup>2+</sup> ions. The performance of the proposed sensor for both glucose detection and Hg<sup>2+</sup> ion was monitored in test electrolyte solutions prepared in phosphate buffer solutions (PBS) having a pH of 7.4. Finally, the detection of other metals ions via a process of inhibition of the enzyme activity has been evaluated. In addition to this, the reproducibility of the enzymatic activity of the sensor was also examined for a glucose biosensor application.</p></sec>
<sec sec-type="methods">
<label>2.</label>
<title>Material and Methods</title>
<sec>
<label>2.1.</label>
<title>Reagents</title>
<p>Glucose oxidase from <italic>Aspergillus niger</italic> with activity of 280 units/mg, chitosan (C3646), D-(+)-glucose (99.5%), zinc nitrate hexahydrate, mercury(II) chloride and hexamethylenetetramine (HMT), acetic acid, were purchased from Sigma-Aldrich (Stockholm, Sweden). Phosphate buffer solution (PBS, 10 mM) was prepared by mixing 8 mM of Na<sub>2</sub>HPO<sub>4</sub>, 1.5 mM of KH<sub>2</sub>PO<sub>4</sub>, 0.135 mM of NaCl and 2.7 mM of KCl in deionized water and then the pH was adjusted to 7.4. A stock solution of 100 mM glucose was prepared in PBS and stored at 4 °C and 100 mM of mercury(II) chloride was prepared in deionized water. The low concentration standard solutions of both glucose and the mercury were freshly prepared before the measurements.</p></sec>
<sec>
<label>2.2.</label>
<title>Fabrication of Glucose Biosensor Electrodes</title>
<p>The ZnO-NR array electrodes were prepared on a glass substrate by first evaporating titanium (Ti) as an adhesion layer and then followed by gold (Au) films with a thickness of 10 nm and 100 nm, respectively using an evaporation system (Evaporator Satis CR725). Then these gold coated glass electrodes were ultrasonically cleaned with isopropanol followed by rinsing in deionized water and then dried in air at room temperature. Then, ZnO-NRs were grown on the electrode by using the low temperature ACG approach as described in [<xref ref-type="bibr" rid="b28-sensors-12-15063">28</xref>,<xref ref-type="bibr" rid="b29-sensors-12-15063">29</xref>]. In this paper, the gold coated glass electrodes were spin coated in two steps with a seed solution containing zinc acetate using a speed of 1,000 rpm for 10 s and 2,500 rpm for 20 s, respectively, and then annealed at 120 °C for 10 min. After that the electrodes were placed horizontally in an aqueous solution of 0.025 M zinc nitrate hexahydrate [(Zn(NO)<sub>3</sub>)<sub>2</sub>·6H<sub>2</sub>O)] and 0.025 M hexamethylenetetramine [C<sub>6</sub>H<sub>12</sub>N<sub>4</sub>] and kept in a preheated oven for 5 h at 80 °C. When the growth was completed, the grown ZnO-NRs were cleaned with deionized water and dried at room temperature. It should noticed that at the top of the gold coated glass substrates were partially covered and the covered part was used as contact area.</p></sec>
<sec>
<label>2.3.</label>
<title>Enzyme Immobilization on ZnO-NRs Arrays</title>
<p>The GOD was immobilized on the ZnO-NRs using a mixed solution consisting of 1 mL of GOD solution prepared by dissolving 10 mg/mL of GOD in PBS with a pH of 7.4 and 1 mL of chitosan membrane in order to improve the stability and activity of the enzyme on the electrode surface. The chitosan is used as the matrix for the immobilization of the enzyme due to its good properties, which include an excellent membrane-forming ability, high permeability toward water, good adhesion, biocompatibility, non-toxicity and high mechanical strength [<xref ref-type="bibr" rid="b30-sensors-12-15063">30</xref>]. The chitosan membranes were prepared as described in [<xref ref-type="bibr" rid="b31-sensors-12-15063">31</xref>]. Next 0.1 mL of a glutaraldehyde (GA) solution (2.5%) was added into the GOD-chitosan solution. Then the electrostatic physical adsorption method was applied for the immobilization of GOD on the ZnO-NR electrode due to the fact the chemical structures of GOD and ZnO both have polar atoms which can easily be attracted through electrostatic binding. The enzyme was electrostatically immobilized on the electrodes by dipping the electrodes into the mixture of the above solution for 5 min, then it was dried in air at room temperature for 1 hour. The immobilized electrodes were kept in dry conditions at 4 °C when not in use.</p></sec>
<sec>
<label>2.4.</label>
<title>Electrochemical Measurements of the Proposed Mercury Ion Sensor</title>
<p>All the measurements of the glucose oxidase-based mercury ion sensor, with and without inhibitor, were performed at room temperature by using a pH meter (Model 744, Metrohm). The used Ag/AgCl reference electrode was purchased from Metrohm (3 M KCl). The response time was measured by a Keithly model 2400 series, which can provide the sensed electrochemical output <italic>vs.</italic> time.</p></sec></sec>
<sec sec-type="results|discussion">
<label>3.</label>
<title>Results and Discussion</title>
<sec>
<label>3.1.</label>
<title>Morphology and Chemical Composition of the Electrode</title>
<p>Scanning electron microscope (SEM) images of the grown ZnO-NRs are shown in <xref ref-type="fig" rid="f1-sensors-12-15063">Figure 1(a)</xref>, while the X-ray Diffraction (XRD) pattern of the grown ZnO-NRs is shown in <xref ref-type="fig" rid="f1-sensors-12-15063">Figure 1(b)</xref>.</p>
<p>It is seen that the ZnO-NRs exhibit a hexagonal wurtzite crystalline structure. The observed peaks were 002, 100, 101, 102 and 103 planes. The highest intensity peak was the 002, demonstrating that the ZnO-NRs grow along the c-axis and perpendicular to the substrate [<xref ref-type="bibr" rid="b32-sensors-12-15063">32</xref>]. The energy-dispersive X-ray spectroscopy (EDS) analyses for the chemical composition of grown ZnO-NRs is shown in <xref ref-type="fig" rid="f1-sensors-12-15063">Figure 1(c)</xref>. It is clearly indicated that the grown ZnO-NRs are composed of Zn and O atoms only and no other impurity was found. The immobilized ZnO-NRs arrays are shown in <xref ref-type="fig" rid="f1-sensors-12-15063">Figure 1(d)</xref>. After the immobilization of enzyme it is seen that it completely covers the ZnO-NRs providing an extra protection for the ZnO-NRs from dissolving. Moreover, this figure gives evidence for the presence of glucose oxidase on the whole surface of ZnO-NRs. Finally, the SEM image of the used electrode is shown in <xref ref-type="fig" rid="f1-sensors-12-15063">Figure 1(e)</xref>. It was revealed that the enzyme is still covered on the electrode and the ZnO nanorods were not dissolved after the electrodes were used.</p></sec>
<sec>
<label>3.2.</label>
<title>Interaction between Chitosan and Glucose Oxidase</title>
<p>In order to confirm the presence of GOD on the surface of ZnO-NRs, a Fourier transform infrared (FTIR) study was carried out. The most important observed IR peaks for the GOD are the amide I band (around 1,650 cm<sup>−1</sup>) and amide II band (around 1,540 cm<sup>−1</sup>) of the amide group and this spectrum is almost identical to the IR spectrum of chitosan, which shows these peaks at 1,654 cm<sup>−1</sup> and 1,597 cm<sup>−1</sup> [<xref ref-type="bibr" rid="b33-sensors-12-15063">33</xref>]. However two weak peaks also appeared below 1,400 cm<sup>−1</sup> which can assigned to the carboxylate groups in the enzyme. The measured FTIR spectrum of GOD/ZnO-NRs shows infrared bands of amide I at 1,653 cm<sup>−1</sup> (C=O) and amide II at 1,541 cm<sup>−1</sup> (N–H and C–H) as shown in <xref ref-type="fig" rid="f2-sensors-12-15063">Figure 2</xref>. These interesting small changes from the natural spectrum of GOD indicated that the glucose oxidase has formed a good matrix with the chitosan membrane.</p></sec>
<sec>
<label>3.3.</label>
<title>The pH Effect</title>
<p>The pH of the substrate solutions (glucose solution) can affect the overall enzymatic activity. Therefore the investigation of the effect of pH on the performance of the biosensor is very important. In this work, we recorded the potential response for 10 mM of glucose in PBS by varying the pH value from 4 to 9. The response of the sensor to the change of pH is shown in <xref ref-type="fig" rid="f3-sensors-12-15063">Figure 3</xref>. The highest activity of the enzyme was observed around pH 7. Therefore, pH 7.4 was adopted for all the measurements.</p></sec>
<sec>
<label>3.4.</label>
<title>Temperature Effect</title>
<p>The efficiency of the glucose sensor with respect to the temperature can be explained in terms of the optimum temperature at which the sensor electrode would show the maximum performance. The sensor electrode has shown its maximum response around 50 °C (<xref ref-type="fig" rid="f4-sensors-12-15063">Figure 4</xref>) which can be attributed to the agitation of ions inside the testing solution with the increase in temperature. Although the optimum temperature for glucose oxidase is about 40 °C, we did all experiments at room temperature due to simplicity of the measurement and to avoid any possible evaporation processes. Moreover, the sensor electrode has shown a decreasing potentiometric response above 50 °C due to possible denaturation of the protein molecules at high temperature.</p></sec>
<sec>
<label>3.5.</label>
<title>Measurement of Glucose Substrate</title>
<p>For the glucose sensing mechanism based on an enzymatic reaction catalysed by glucose oxidase (GOD) with β <sc>d</sc>–glucose gives the charged products gluconate<sup>−</sup> and a proton (H<sup>+</sup>), according to <xref ref-type="fig" rid="f5-sensors-12-15063">Figure 5</xref>. The electromotive force (EMF) response of ion selective electrodes, the glucose biosensor, can be explained according to Nernst's expression:
<disp-formula id="FD1">
<label>(1)</label>
<mml:math id="mm1" display="block">
<mml:semantics id="sm1">
<mml:mrow>
<mml:mi>E</mml:mi>
<mml:mo>+</mml:mo>
<mml:msup>
<mml:mi>E</mml:mi>
<mml:mn>0</mml:mn></mml:msup>
<mml:mo>+</mml:mo>
<mml:mn>2.3</mml:mn>
<mml:mfrac>
<mml:mrow>
<mml:mtext>RT</mml:mtext></mml:mrow>
<mml:mrow>
<mml:mtext>nF</mml:mtext></mml:mrow></mml:mfrac>
<mml:mo>log</mml:mo>
<mml:mrow>
<mml:mo>(</mml:mo>
<mml:mrow>
<mml:mfrac>
<mml:mrow>
<mml:mtext>Ox</mml:mtext></mml:mrow>
<mml:mrow>
<mml:mtext>Red</mml:mtext></mml:mrow></mml:mfrac></mml:mrow>
<mml:mo>)</mml:mo></mml:mrow></mml:mrow></mml:semantics></mml:math></disp-formula>where <italic>E</italic> is the cell potential at some moment in time, <italic>E</italic>° is the cell potential when the reaction is at standard-state conditions, <italic>R</italic> is the ideal gas constant in units of Joules per mole, <italic>T</italic> is the temperature in Kelvin, <italic>n</italic> is the number of moles of electrons transferred in the balanced equation for the reaction, and <italic>F</italic> is the charge on a mole of electrons.</p>
<p>The EMF response of the glucose biosensor was measured between the working electrode and the reference electrode for different glucose concentrations. The experimental results showed large dynamic ranges over glucose concentrations going from 10<sup>−3</sup> mM to 10 mM with a linear output response <italic>vs.</italic> logarithmic concentrations of glucose with a slope of 41.9 mV/decade as shown in <xref ref-type="fig" rid="f6-sensors-12-15063">Figure 6</xref>.</p></sec>
<sec>
<label>3.6.</label>
<title>Inhibition Studies</title>
<p>The enzyme–inhibitor reaction is often complex. The inhibition reactions between the enzyme and the toxic compound are reversible and irreversible inhibition. The mechanism of glucose oxidase inhibition by heavy metals is reversible and it makes the inhibitor bind at a site other than the active site of the enzyme and causes changes in the shape of the enzyme that lead to a change in the activity and a mixed inhibition cannot be overcome at high substrate concentration [<xref ref-type="bibr" rid="b34-sensors-12-15063">34</xref>,<xref ref-type="bibr" rid="b35-sensors-12-15063">35</xref>]. The degree of inhibition of the glucose oxidase is given by the following expression:
<disp-formula id="FD2">
<label>(2)</label>
<mml:math id="mm2" display="block">
<mml:semantics id="sm2">
<mml:mrow>
<mml:mi>I</mml:mi>
<mml:mo>%</mml:mo>
<mml:mo>=</mml:mo>
<mml:mn>100</mml:mn>
<mml:mrow>
<mml:mo>(</mml:mo>
<mml:mrow>
<mml:mfrac>
<mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn></mml:msub>
<mml:mo>−</mml:mo>
<mml:mi>I</mml:mi></mml:mrow>
<mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn></mml:msub></mml:mrow></mml:mfrac></mml:mrow>
<mml:mo>)</mml:mo></mml:mrow></mml:mrow></mml:semantics></mml:math></disp-formula>where I% represents the degree of inhibition, <italic>I</italic><sub>0</sub> represents the response to glucose in the absence of inhibitor and <italic>I</italic> represents the response to glucose in the presence of inhibitor [<xref ref-type="bibr" rid="b12-sensors-12-15063">12</xref>,<xref ref-type="bibr" rid="b34-sensors-12-15063">34</xref>,<xref ref-type="bibr" rid="b35-sensors-12-15063">35</xref>]. In this work, after the stable response of sensor electrode in the standard PBS solution was obtained, an amount of low concentration glucose solution was added to give a final concentration of 1 mM glucose in the solution, and then its stable response was recorded (I<sub>0</sub>) after that solutions of higher concentrations of Hg<sup>2+</sup> ions were added to inhibit the enzyme activity and the change in the output response (I) corresponding to the concentration of inhibitor in solution was recorded. The linear range <italic>vs.</italic> natural logarithm of Hg<sup>2+</sup> ion concentration of the degree of inhibition of the GOD can be divided into two parts, as shown in <xref ref-type="fig" rid="f7-sensors-12-15063">Figure 7(a,b)</xref>. In <xref ref-type="fig" rid="f7-sensors-12-15063">Figure 7(a)</xref>, the linear range is from 0.5 × 10<sup>−6</sup> mM to 0.5 × 10<sup>−4</sup> mM and the correlation coefficient is 0.99. From <xref ref-type="fig" rid="f7-sensors-12-15063">Figure 7(b)</xref>, the linear range is from 0.5 × 10<sup>−4</sup> mM to 20 mM, also with a coefficient of 0.99. In order to study the effect of the substrate concentration on the inhibition of enzyme activity, a higher concentration 10 mM of glucose was also performed. The inhibition degree of the GOD at high concentration of the substrate is given by <xref ref-type="fig" rid="f7-sensors-12-15063">Figure 7(c)</xref>.</p>
<p>From this graph we can see clearly that the linear range at the high substrate concentration is given from 10<sup>−3</sup> mM to 6 mM. From these obtained results we can say that the lower concentration gives the lower limit of detection. Furthermore, the affected electrodes during the Hg<sup>2+</sup> ions measurement were immediately used to measure the response of the glucose solution. <xref ref-type="fig" rid="f7-sensors-12-15063">Figure 7(d)</xref> shows the response in decreasing order while for the glucose concentration it is in increasing order. This is clear evidence that the higher substrate concentration provides a lower inhibition response. In order to perform multiple experiments, the activity of the enzyme at the surface of the electrodes has to be restored. To restore the enzyme activity, the used electrodes were washed with PBS after contact with the heavy metal. After this, for the electrodes used in inhibitor for 1 hour, the regeneration efficiency can reach up to 70% by dipping electrodes in 10 mM of PBS for 5 min. The biosensor activity could be restored more than 70% after dipping the used electrodes in PBS for more than 7 min, which demonstrates that Hg<sup>2+</sup> ions are not strongly bound to the enzyme.</p></sec>
<sec>
<label>3.7.</label>
<title>Selectivity</title>
<p>Selectivity is very important issue for the performance of a biosensor. In this study, we use the same approach as for the mercury for calculating the degree of inhibition. The influences of possible interfering metals such as Cu<sup>2+</sup>, Zn<sup>2+</sup>, Fe<sup>2+</sup> and Co<sup>2+</sup> were investigated in the presence of fixed amounts of 1 mM of glucose solution. The selective coefficient values for interference were calculated by the separation solution method (SSM) [<xref ref-type="bibr" rid="b36-sensors-12-15063">36</xref>], using the following equation:
<disp-formula id="FD3">
<label>(3)</label>
<mml:math id="mm3" display="block">
<mml:semantics id="sm3">
<mml:mrow>
<mml:mo>log</mml:mo>
<mml:msubsup>
<mml:mi>K</mml:mi>
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>B</mml:mi></mml:mrow>
<mml:mrow>
<mml:mtext mathvariant="italic">pot</mml:mtext></mml:mrow></mml:msubsup>
<mml:mo>=</mml:mo>
<mml:mfrac>
<mml:mrow>
<mml:msub>
<mml:mi>E</mml:mi>
<mml:mi>B</mml:mi></mml:msub>
<mml:mo>−</mml:mo>
<mml:msub>
<mml:mi>E</mml:mi>
<mml:mi>A</mml:mi></mml:msub></mml:mrow>
<mml:mrow>
<mml:msub>
<mml:mi>S</mml:mi>
<mml:mi>A</mml:mi></mml:msub></mml:mrow></mml:mfrac>
<mml:mo>+</mml:mo>
<mml:mrow>
<mml:mo>(</mml:mo>
<mml:mrow>
<mml:mn>1</mml:mn>
<mml:mo>−</mml:mo>
<mml:mfrac>
<mml:mrow>
<mml:msub>
<mml:mi>S</mml:mi>
<mml:mi>B</mml:mi></mml:msub></mml:mrow>
<mml:mrow>
<mml:msub>
<mml:mi>S</mml:mi>
<mml:mi>A</mml:mi></mml:msub></mml:mrow></mml:mfrac></mml:mrow>
<mml:mo>)</mml:mo></mml:mrow>
<mml:mo>log</mml:mo>
<mml:mo stretchy="false">(</mml:mo>
<mml:msub>
<mml:mi>a</mml:mi>
<mml:mi>A</mml:mi></mml:msub>
<mml:mo stretchy="false">)</mml:mo></mml:mrow></mml:semantics></mml:math></disp-formula>where A represents for target ion (Hg<sup>2+</sup>) and B represents for interference ions (Cu<sup>2+</sup>, Zn<sup>2+</sup>, Fe<sup>2+</sup> or Co<sup>2+</sup>), E<sub>A</sub> and E<sub>B</sub> represent for EMF response for A and B at same 0.01 mM concentration, respectively, while S<sub>A</sub> and S<sub>B</sub> are the sensitivities of A and B, respectively. Finally, <italic>a<sub>A</sub></italic> is the concentration of A. The calculated selectivity coefficient results are shown in <xref ref-type="table" rid="t2-sensors-12-15063">Table 2</xref>. From <xref ref-type="table" rid="t2-sensors-12-15063">Table 2</xref> the calculated selectivity coefficient values of each interfering ion are fairly constant and the sensitivity to mercury ions is 1,000 times higher than for Cu<sup>2+</sup> and Zn<sup>2+</sup> and 100 times more than Fe<sup>2+</sup> and Co<sup>2+</sup>. These values show that the interference effects are negligible.</p></sec>
<sec>
<label>3.8.</label>
<title>Response Time, Reproducibility and Lifetime</title>
<p><xref ref-type="fig" rid="f8-sensors-12-15063">Figure 8</xref> shows a response time of 8 s for the proposed sensor in the presence of 1 mM of Hg<sup>2+</sup> solution at a fixed concentration of 1 mM substrate concentration. The sensor to sensor reproducibility was evaluated by using five independent electrodes fabricated under the same conditions. The relative standard deviation of the fabricated sensor electrodes in standard glucose solutions was less than ±5% as shown in <xref ref-type="fig" rid="f9-sensors-12-15063">Figure 9</xref>. The stability of the presented biosensor has been investigated after the electrodes were stored for three weeks in dry conditions at 4 °C. It has been found that this biosensor has good storability and maintained 90% of its original sensitivity.</p></sec></sec>
<sec sec-type="conclusions">
<label>4.</label>
<title>Conclusions</title>
<p>A potentiometric glucose biosensor made by immobilization of a glucose oxidase enzyme using physical adsorption in combination with chitosan membrane and GA on a ZnO-NRs electrode has been demonstrated. It has been observed that such an approach is good for measuring the inhibition effect of heavy metal ions on the enzyme activity, especially for measuring Hg<sup>2+</sup> ion. The advantages of using ZnO-NRs, like e.g., suitability for detecting in small volumes with high sensitivity, low detection limit, fast response and ease of regeneration, are among the reasons of the developing such a sensor. The fabrication of this biosensor further has extra advantages like, e.g., being a low-cost approach, simple, requiring minimum hardware and possessing fast response times. Moreover, the presented sensor has demonstrated good sensitivity, reusability, reproducibility and good anti-interferent effects. Such a biosensor is also convenient for assembly into portable chips for chemical sensing. The proposed sensor electrode is also suitable for use by unskilled users and it can be applied to on-site measurement of mercury ions.</p></sec></body>
<back>
<ref-list>
<title>References</title>
<ref id="b1-sensors-12-15063"><label>1.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Compagnone</surname><given-names>D.</given-names></name><name><surname>Palleschi</surname><given-names>G.</given-names></name><name><surname>Varallo</surname><given-names>G.</given-names></name><name><surname>Imperiali</surname><given-names>P.L.</given-names></name></person-group><article-title>Amperometric biosensors for the determination of heavy metals</article-title><source>Proc. SPIE</source><year>1995</year><volume>2504</volume><fpage>141</fpage><lpage>152</lpage></citation></ref>
<ref id="b2-sensors-12-15063"><label>2.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mohammadi</surname><given-names>H.</given-names></name><name><surname>Rhazi</surname><given-names>M.E.</given-names></name><name><surname>Amine</surname><given-names>A.</given-names></name><name><surname>Brettband</surname><given-names>A.M.O.</given-names></name><name><surname>Brett</surname><given-names>C.M.A.</given-names></name></person-group><article-title>Determination of mercury(II) by invertase enzyme inhibition coupled with batch injection analysis</article-title><source>Analyst</source><year>2002</year><volume>127</volume><fpage>1088</fpage><lpage>1093</lpage><pub-id pub-id-type="doi">10.1039/b202887e</pub-id><pub-id pub-id-type="pmid">12195951</pub-id></citation></ref>
<ref id="b3-sensors-12-15063"><label>3.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mozaz</surname><given-names>S.R.</given-names></name><name><surname>Alda</surname><given-names>M.J.L.</given-names></name><name><surname>Barceló</surname><given-names>D.</given-names></name></person-group><article-title>Biosensors as useful tools for environmental analysis and monitoring</article-title><source>Anal. Bioanal. Chem.</source><year>2006</year><volume>386</volume><fpage>1025</fpage><lpage>1041</lpage><pub-id pub-id-type="doi">10.1007/s00216-006-0574-3</pub-id><pub-id pub-id-type="pmid">16807703</pub-id></citation></ref>
<ref id="b4-sensors-12-15063"><label>4.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mozaz</surname><given-names>S.R.</given-names></name><name><surname>Marco</surname><given-names>M.P.</given-names></name><name><surname>Alda</surname><given-names>M.J.L.</given-names></name><name><surname>Barceló</surname><given-names>D.</given-names></name></person-group><article-title>Biosensors for environmental applications: Future development trends</article-title><source>Pure Appl. Chem.</source><year>2004</year><volume>76</volume><fpage>723</fpage><lpage>752</lpage><pub-id pub-id-type="doi">10.1351/pac200476040723</pub-id></citation></ref>
<ref id="b5-sensors-12-15063"><label>5.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mozaz</surname><given-names>S.R.</given-names></name><name><surname>Alda</surname><given-names>M.J.L.</given-names></name><name><surname>Marco</surname><given-names>M.P.</given-names></name><name><surname>Barceló</surname><given-names>D.</given-names></name></person-group><article-title>Biosensors for environmental monitoring: A global perspective</article-title><source>Talanta</source><year>2005</year><volume>65</volume><fpage>291</fpage><lpage>297</lpage><pub-id pub-id-type="pmid">18969798</pub-id></citation></ref>
<ref id="b6-sensors-12-15063"><label>6.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Verma</surname><given-names>N.</given-names></name><name><surname>Singh</surname><given-names>M.</given-names></name></person-group><article-title>Biosensors for heavy metals</article-title><source>BioMetals</source><year>2005</year><volume>18</volume><fpage>121</fpage><lpage>129</lpage><pub-id pub-id-type="doi">10.1007/s10534-004-5787-3</pub-id><pub-id pub-id-type="pmid">15954738</pub-id></citation></ref>
<ref id="b7-sensors-12-15063"><label>7.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Samphao</surname><given-names>A.</given-names></name><name><surname>Rerkchai</surname><given-names>H.</given-names></name><name><surname>Jitcharoen</surname><given-names>J.</given-names></name><name><surname>Nacapricha</surname><given-names>D.</given-names></name><name><surname>Kalcher</surname><given-names>K.</given-names></name></person-group><article-title>Indirect determination of mercury by inhibition of glucose oxidase immobilized on a carbon paste electrode</article-title><source>Int. J. Electrochem. Sci.</source><year>2012</year><volume>7</volume><fpage>1001</fpage><lpage>1010</lpage></citation></ref>
<ref id="b8-sensors-12-15063"><label>8.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Liu</surname><given-names>J.</given-names></name><name><surname>Xu</surname><given-names>X.</given-names></name><name><surname>Tang</surname><given-names>L.</given-names></name><name><surname>Zeng</surname><given-names>G.</given-names></name></person-group><article-title>Determination of trace mercury in compost extract by inhibition based glucose oxidase biosensor</article-title><source>Trans. Nonferrous Met. Soc. China</source><year>2009</year><volume>19</volume><fpage>235</fpage><lpage>240</lpage><pub-id pub-id-type="doi">10.1016/S1003-6326(08)60258-7</pub-id></citation></ref>
<ref id="b9-sensors-12-15063"><label>9.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Alexanderand</surname><given-names>P.W.</given-names></name><name><surname>Rechnitz</surname><given-names>G.A.</given-names></name></person-group><article-title>Enzyme inhibition assays with an amperometric glucose biosensor based on a thiolate self-assembled monolayer</article-title><source>Electroanalysis</source><year>2000</year><volume>12</volume><fpage>343</fpage><lpage>350</lpage><pub-id pub-id-type="doi">10.1002/(SICI)1521-4109(20000301)12:5&lt;343::AID-ELAN343&gt;3.0.CO;2-E</pub-id></citation></ref>
<ref id="b10-sensors-12-15063"><label>10.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Guascito</surname><given-names>M.R.</given-names></name><name><surname>Malitesta</surname><given-names>C.</given-names></name><name><surname>Mazzotta</surname><given-names>E.</given-names></name><name><surname>Turco</surname><given-names>A.</given-names></name></person-group><article-title>Inhibitive determination of metal ions by an amperometric glucose oxidase biosensor study of the effect of hydrogen peroxide decomposition</article-title><source>Sens. Actuators B</source><year>2008</year><volume>131</volume><fpage>394</fpage><lpage>402</lpage><pub-id pub-id-type="doi">10.1016/j.snb.2007.11.049</pub-id></citation></ref>
<ref id="b11-sensors-12-15063"><label>11.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Malitesta</surname><given-names>C.</given-names></name><name><surname>Guascito</surname><given-names>M.R.</given-names></name></person-group><article-title>Heavy metal determination by biosensors based on enzyme immobilised by electropolymerisation</article-title><source>Biosens. Bioelectr.</source><year>2005</year><volume>20</volume><fpage>1643</fpage><lpage>1647</lpage><pub-id pub-id-type="doi">10.1016/j.bios.2004.08.003</pub-id></citation></ref>
<ref id="b12-sensors-12-15063"><label>12.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Ghica</surname><given-names>M.E.</given-names></name><name><surname>Brett</surname><given-names>C.M.A.</given-names></name></person-group><article-title>Glucose oxidase inhibition in poly (neutral red) mediated enzyme biosensors for heavy metal determination</article-title><source>Microchim. Acta</source><year>2008</year><volume>163</volume><fpage>185</fpage><lpage>193</lpage><pub-id pub-id-type="doi">10.1007/s00604-008-0018-1</pub-id></citation></ref>
<ref id="b13-sensors-12-15063"><label>13.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Cosnier</surname><given-names>S.</given-names></name><name><surname>Mousty</surname><given-names>C.</given-names></name><name><surname>Guelorget</surname><given-names>A.</given-names></name><name><surname>Lopez</surname><given-names>M.S.P.</given-names></name><name><surname>Shan</surname><given-names>D.</given-names></name></person-group><article-title>A fast and direct amperometric determination of Hg<sup>2+</sup> by a bienzyme electrode based on the competitive activities of glucose oxidase and laccase</article-title><source>Electroanalysis</source><year>2011</year><volume>23</volume><fpage>1776</fpage><lpage>1779</lpage><pub-id pub-id-type="doi">10.1002/elan.201100200</pub-id></citation></ref>
<ref id="b14-sensors-12-15063"><label>14.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Krawczyk</surname><given-names>T.K.</given-names></name><name><surname>Moszczynska</surname><given-names>M.</given-names></name><name><surname>Trojanowicz</surname><given-names>M.</given-names></name></person-group><article-title>Inhibitive determination of mercury and other metal ions by potentiometric urea biosensor</article-title><source>Biosens. Bioelectr.</source><year>2000</year><volume>15</volume><fpage>681</fpage><lpage>691</lpage><pub-id pub-id-type="doi">10.1016/S0956-5663(00)00085-3</pub-id></citation></ref>
<ref id="b15-sensors-12-15063"><label>15.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Yang</surname><given-names>Y.</given-names></name><name><surname>Wang</surname><given-names>Z.</given-names></name><name><surname>Yang</surname><given-names>M.</given-names></name><name><surname>Guo</surname><given-names>M.</given-names></name><name><surname>Wu</surname><given-names>Z.</given-names></name><name><surname>Shen</surname><given-names>G.</given-names></name><name><surname>Yu</surname><given-names>R.</given-names></name></person-group><article-title>Inhibitive determination of mercury ion using a renewable urea biosensor based on self-assembled gold nanoparticles</article-title><source>Sens. Actuators B</source><year>2006</year><volume>114</volume><fpage>1</fpage><lpage>8</lpage><pub-id pub-id-type="doi">10.1016/j.snb.2005.04.005</pub-id></citation></ref>
<ref id="b16-sensors-12-15063"><label>16.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Do</surname><given-names>J.S.</given-names></name><name><surname>Lin</surname><given-names>K.H.</given-names></name><name><surname>Ohara</surname><given-names>R.</given-names></name></person-group><article-title>Preparation of urease/nano-structured polyaniline-Nafion/Au/Al<sub>2</sub>O<sub>3</sub> electrode for inhibitive detection of mercury ion</article-title><source>J. Taiwan Inst. Chem. Eng.</source><year>2011</year><volume>42</volume><fpage>662</fpage><lpage>668</lpage><pub-id pub-id-type="doi">10.1016/j.jtice.2010.11.007</pub-id></citation></ref>
<ref id="b17-sensors-12-15063"><label>17.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Hamlaouia</surname><given-names>M.L.</given-names></name><name><surname>Reybier</surname><given-names>K.</given-names></name><name><surname>Marrakchi</surname><given-names>M.</given-names></name><name><surname>Renault</surname><given-names>N.J.</given-names></name><name><surname>Martelet</surname><given-names>C.</given-names></name><name><surname>Kherrat</surname><given-names>R.</given-names></name><name><surname>Walcarius</surname><given-names>A.</given-names></name></person-group><article-title>Development of a urea biosensor based on a polymeric membrane including zeolite</article-title><source>Anal. Chim. Acta</source><year>2002</year><volume>466</volume><fpage>39</fpage><lpage>45</lpage><pub-id pub-id-type="doi">10.1016/S0003-2670(02)00548-2</pub-id></citation></ref>
<ref id="b18-sensors-12-15063"><label>18.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mohammadi</surname><given-names>H.</given-names></name><name><surname>Rhazi</surname><given-names>M.E.</given-names></name><name><surname>Amine</surname><given-names>A.</given-names></name><name><surname>Brettband</surname><given-names>A.M.O.</given-names></name><name><surname>Brett</surname><given-names>C.M.A.</given-names></name></person-group><article-title>Determination of mercury(II) by invertase enzyme inhibition coupled with batch injection analysis</article-title><source>Analyst</source><year>2002</year><volume>127</volume><fpage>1088</fpage><lpage>1093</lpage><pub-id pub-id-type="doi">10.1039/b202887e</pub-id><pub-id pub-id-type="pmid">12195951</pub-id></citation></ref>
<ref id="b19-sensors-12-15063"><label>19.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Ciucu</surname><given-names>A.</given-names></name></person-group><article-title>Fast spectrometric method for Mercury (II) determination based on glucose-oxidase inhibition</article-title><source>Anal. Lett.</source><year>2010</year><volume>43</volume><fpage>1377</fpage><lpage>1386</lpage><pub-id pub-id-type="doi">10.1080/00032710903518781</pub-id></citation></ref>
<ref id="b20-sensors-12-15063"><label>20.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Wang</surname><given-names>Z.L.</given-names></name></person-group><article-title>Zinc oxide nanostructures: Growth, properties and applications</article-title><source>J. Phys. Condens. Matter</source><year>2004</year><volume>16</volume><fpage>829</fpage><lpage>858</lpage></citation></ref>
<ref id="b21-sensors-12-15063"><label>21.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Solanki</surname><given-names>P.R.</given-names></name><name><surname>Kaushik</surname><given-names>A.</given-names></name><name><surname>Ansari</surname><given-names>A.A.</given-names></name><name><surname>Sumana</surname><given-names>G.</given-names></name><name><surname>Malhotra</surname><given-names>B.D.</given-names></name></person-group><article-title>Zinc oxide-chitosan nanobiocomposite for urea sensor</article-title><source>Appl. Phys. Lett.</source><year>2008</year><volume>93</volume><pub-id pub-id-type="doi">10.1063/1.2980448</pub-id></citation></ref>
<ref id="b22-sensors-12-15063"><label>22.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Wei</surname><given-names>A.</given-names></name><name><surname>Sun</surname><given-names>X.W.</given-names></name><name><surname>Wang</surname><given-names>J.X.</given-names></name><name><surname>Lei</surname><given-names>Y.</given-names></name><name><surname>Cai</surname><given-names>X.P.</given-names></name><name><surname>Li</surname><given-names>C.M.</given-names></name><name><surname>Dong</surname><given-names>Z.L.</given-names></name><name><surname>Huang</surname><given-names>W.</given-names></name></person-group><article-title>Enzymatic glucose biosensor based on ZnO nanorod array grown by hydrothermal decomposition</article-title><source>Appl. Phys. Lett.</source><year>2006</year><volume>89</volume><pub-id pub-id-type="doi">10.1063/1.2356307</pub-id></citation></ref>
<ref id="b23-sensors-12-15063"><label>23.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Solanki</surname><given-names>P.R.</given-names></name><name><surname>Kaushik</surname><given-names>A.</given-names></name><name><surname>Agrawal</surname><given-names>V.V.</given-names></name><name><surname>Malhotra</surname><given-names>B.D.</given-names></name></person-group><article-title>Nanostructured metal oxide-based biosensors</article-title><source>NPG Asia Mater.</source><year>2011</year><volume>3</volume><fpage>17</fpage><lpage>24</lpage><pub-id pub-id-type="doi">10.1038/asiamat.2010.137</pub-id></citation></ref>
<ref id="b24-sensors-12-15063"><label>24.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Ibupoto</surname><given-names>Z.H.</given-names></name><name><surname>Ali</surname><given-names>S.M.U.</given-names></name><name><surname>Chey</surname><given-names>C.O.</given-names></name><name><surname>Khun</surname><given-names>K.</given-names></name><name><surname>Nur</surname><given-names>O.</given-names></name><name><surname>Willander</surname><given-names>M.</given-names></name></person-group><article-title>Selective zinc ion detection by functionalised ZnO nanorods with Ionophore</article-title><source>J. Appl. Phys.</source><year>2011</year><volume>110</volume><pub-id pub-id-type="doi">10.1063/1.3662107</pub-id></citation></ref>
<ref id="b25-sensors-12-15063"><label>25.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Ibupoto</surname><given-names>Z.H.</given-names></name><name><surname>Ali</surname><given-names>S.M.U.</given-names></name><name><surname>Khun</surname><given-names>K.</given-names></name><name><surname>Chey</surname><given-names>C.O.</given-names></name><name><surname>Nur</surname><given-names>O.</given-names></name><name><surname>Willander</surname><given-names>M.</given-names></name></person-group><article-title>ZnO nanorods based enzymatic biosensor for selective determination of Penicillin</article-title><source>Biosensors</source><year>2011</year><volume>1</volume><fpage>153</fpage><lpage>163</lpage><pub-id pub-id-type="doi">10.3390/bios1040153</pub-id></citation></ref>
<ref id="b26-sensors-12-15063"><label>26.</label><citation citation-type="book"><person-group person-group-type="author"><name><surname>Lee</surname><given-names>J.W.</given-names></name><name><surname>Foot</surname><given-names>R.S.</given-names></name></person-group><source>Micro and Nano Technologies in Bioanalysis Methods and Protocols</source><publisher-name>Springer</publisher-name><publisher-loc>New York, NY, USA</publisher-loc><year>2009</year><fpage>168</fpage></citation></ref>
<ref id="b27-sensors-12-15063"><label>27.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Willander</surname><given-names>M.</given-names></name><name><surname>Nur</surname><given-names>O.</given-names></name><name><surname>Zaman</surname><given-names>S.</given-names></name><name><surname>Zainelabdin</surname><given-names>A.</given-names></name><name><surname>Bano</surname><given-names>N.</given-names></name><name><surname>Hussain</surname><given-names>I.</given-names></name></person-group><article-title>Zinc oxide nanorods/polymer hybrid heterojunctions for white light emitting diodes</article-title><source>J. Phys. D Appl. Phys.</source><year>2011</year><volume>44</volume><pub-id pub-id-type="doi">10.1088/0022-3727/44/22/224017</pub-id></citation></ref>
<ref id="b28-sensors-12-15063"><label>28.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Greene</surname><given-names>L.E.</given-names></name><name><surname>Law</surname><given-names>M.</given-names></name><name><surname>Goldberger</surname><given-names>J.</given-names></name><name><surname>Kim</surname><given-names>F.</given-names></name><name><surname>Johnson</surname><given-names>J.C.</given-names></name><name><surname>Zhang</surname><given-names>Y.</given-names></name><name><surname>Saykally</surname><given-names>R.J.</given-names></name><name><surname>Yang</surname><given-names>P.</given-names></name></person-group><article-title>Low-temperature wafer-scale production of ZnO nanowire arrays</article-title><source>Angew. Chem. Int. Ed.</source><year>2003</year><volume>42</volume><fpage>3031</fpage><lpage>3034</lpage><pub-id pub-id-type="doi">10.1002/anie.200351461</pub-id></citation></ref>
<ref id="b29-sensors-12-15063"><label>29.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Li</surname><given-names>Q.</given-names></name><name><surname>Kumar</surname><given-names>V.</given-names></name><name><surname>Li</surname><given-names>Y.</given-names></name><name><surname>Zhang</surname><given-names>H.</given-names></name><name><surname>Marks</surname><given-names>T.J.</given-names></name><name><surname>Chang</surname><given-names>R.P.H.</given-names></name></person-group><article-title>Fabrication of ZnO nanorods and nanotubes in aqueous solutions</article-title><source>Chem. Mater.</source><year>2005</year><volume>17</volume><fpage>1001</fpage><lpage>1006</lpage><pub-id pub-id-type="doi">10.1021/cm048144q</pub-id></citation></ref>
<ref id="b30-sensors-12-15063"><label>30.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Wang</surname><given-names>H.S.</given-names></name><name><surname>Pan</surname><given-names>Q.X.</given-names></name><name><surname>Wang</surname><given-names>G.X.</given-names></name></person-group><article-title>A biosensor based on immobilization of horseradish peroxidase in chitosan matrix cross-linked with glyoxal for amperometric determination of hydrogen peroxide</article-title><source>Sensors</source><year>2005</year><volume>5</volume><fpage>266</fpage><lpage>276</lpage><pub-id pub-id-type="doi">10.3390/s5040266</pub-id></citation></ref>
<ref id="b31-sensors-12-15063"><label>31.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Chey</surname><given-names>C.O.</given-names></name><name><surname>Ali</surname><given-names>S.M.U.</given-names></name><name><surname>Ibupoto</surname><given-names>Z.H.</given-names></name><name><surname>Khun</surname><given-names>K.</given-names></name><name><surname>Nur</surname><given-names>O.</given-names></name><name><surname>Willander</surname><given-names>M.</given-names></name></person-group><article-title>Potentiometric creatinine biosensor based on ZnO nanowires</article-title><source>J. Nanosci. Lett.</source><year>2012</year><volume>2</volume><fpage>24</fpage></citation></ref>
<ref id="b32-sensors-12-15063"><label>32.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Willander</surname><given-names>M.</given-names></name><name><surname>Yang</surname><given-names>L.L.</given-names></name><name><surname>Wadeasa</surname><given-names>A.</given-names></name><name><surname>Ali</surname><given-names>S.U.</given-names></name><name><surname>Asif</surname><given-names>M.H.</given-names></name><name><surname>Zhao</surname><given-names>Q.X.</given-names></name><name><surname>Nur</surname><given-names>O.</given-names></name></person-group><article-title>Zinc oxide nanowires: Controlled low temperature growth and some electrochemical and optical nano-devices</article-title><source>J. Mater. Chem.</source><year>2009</year><volume>19</volume><fpage>1006</fpage><lpage>1018</lpage><pub-id pub-id-type="doi">10.1039/b816619f</pub-id></citation></ref>
<ref id="b33-sensors-12-15063"><label>33.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Hui</surname><given-names>Y.</given-names></name><name><surname>Nan</surname><given-names>L.</given-names></name><name><surname>Zhong</surname><given-names>X.J.</given-names></name><name><surname>Jie</surname><given-names>Z.J.</given-names></name></person-group><article-title>A glucose biosensor based on immobilization of glucose oxidase in chitosan network matrix</article-title><source>Chin. J. Chem.</source><year>2005</year><volume>23</volume><fpage>275</fpage><lpage>279</lpage><pub-id pub-id-type="doi">10.1002/cjoc.200590275</pub-id></citation></ref>
<ref id="b34-sensors-12-15063"><label>34.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Amine</surname><given-names>A.</given-names></name><name><surname>Mohammadi</surname><given-names>H.</given-names></name><name><surname>Bourais</surname><given-names>I.</given-names></name><name><surname>Palleschi</surname><given-names>G.</given-names></name></person-group><article-title>Enzyme inhibition-based biosensors for food safety and environmental monitoring</article-title><source>Biosens. Bioelectr.</source><year>2006</year><volume>21</volume><fpage>1405</fpage><lpage>1423</lpage><pub-id pub-id-type="doi">10.1016/j.bios.2005.07.012</pub-id></citation></ref>
<ref id="b35-sensors-12-15063"><label>35.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Turdean</surname><given-names>G.L.</given-names></name></person-group><article-title>Design and development of biosensors for the detection of heavy metal toxicity</article-title><source>Int. J. Electrochem.</source><year>2011</year><pub-id pub-id-type="doi">10.4061/2011/343125</pub-id></citation></ref>
<ref id="b36-sensors-12-15063"><label>36.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Wang</surname><given-names>H.X.</given-names></name><name><surname>Pu</surname><given-names>M.</given-names></name></person-group><article-title>A method of determining selectivity coefficients based on the practical slope of ion selective electrodes</article-title><source>Chin. Chem. Lett.</source><year>2002</year><volume>13</volume><fpage>355</fpage><lpage>358</lpage></citation></ref></ref-list>
<sec sec-type="display-objects">
<title>Figures and Tables</title>
<fig id="f1-sensors-12-15063" position="float">
<label>Figure 1.</label>
<caption>
<p>(<bold>a</bold>) SEM image of ZnO-NRs (<bold>b</bold>) XRD spectra of ZnO-NRs (<bold>c</bold>) EDS spectra of ZnO-NRs, (<bold>d</bold>) SEM image of the immobilized GOD on ZnO-NRs, (<bold>e</bold>) SEM image of the used electrode.</p></caption>
<graphic xlink:href="sensors-12-15063f1a.gif"/>
<graphic xlink:href="sensors-12-15063f1b.gif"/></fig>
<fig id="f2-sensors-12-15063" position="float">
<label>Figure 2.</label>
<caption>
<p>Fourier transform infrared (FTIR) spectrum of glucose oxidase immobilized on the ZnO-NRs.</p></caption>
<graphic xlink:href="sensors-12-15063f2.gif"/></fig>
<fig id="f3-sensors-12-15063" position="float">
<label>Figure 3.</label>
<caption>
<p>The pH effect on the potentiometric response of biosensor electrode.</p></caption>
<graphic xlink:href="sensors-12-15063f3.gif"/></fig>
<fig id="f4-sensors-12-15063" position="float">
<label>Figure 4.</label>
<caption>
<p>The temperature effect on the potentiometric response of biosensor electrode.</p></caption>
<graphic xlink:href="sensors-12-15063f4.gif"/></fig>
<fig id="f5-sensors-12-15063" position="float">
<label>Figure 5.</label>
<caption>
<p>Schematic diagram of the sensing mechanism.</p></caption>
<graphic xlink:href="sensors-12-15063f5.gif"/></fig>
<fig id="f6-sensors-12-15063" position="float">
<label>Figure 6.</label>
<caption>
<p>The calibration curve for glucose concentrations.</p></caption>
<graphic xlink:href="sensors-12-15063f6.gif"/></fig>
<fig id="f7-sensors-12-15063" position="float">
<label>Figure 7.</label>
<caption>
<p>(<bold>a</bold>) A calibration curve for mercury ion inhibition at low glucose concentration and (<bold>b</bold>), (<bold>c</bold>) at high glucose concentration; (<bold>d</bold>) the effect of glucose concentration on degree of inhibition.</p></caption>
<graphic xlink:href="sensors-12-15063f7a.gif"/>
<graphic xlink:href="sensors-12-15063f7b.gif"/></fig>
<fig id="f8-sensors-12-15063" position="float">
<label>Figure 8.</label>
<caption>
<p>The response time of the proposed sensor.</p></caption>
<graphic xlink:href="sensors-12-15063f8.gif"/></fig>
<fig id="f9-sensors-12-15063" position="float">
<label>Figure 9.</label>
<caption>
<p>Reproducibility of the sensor electrodes.</p></caption>
<graphic xlink:href="sensors-12-15063f9.gif"/></fig>
<table-wrap id="t1-sensors-12-15063" position="float">
<label>Table 1.</label>
<caption>
<p>The performance of some glucose biosensors for mercury determination.</p></caption>
<table frame="box" rules="all">
<thead>
<tr>
<th align="center" valign="top"><bold>Electrodes</bold></th>
<th align="center" valign="top"><bold>Low Detection</bold></th>
<th align="center" valign="top"><bold>Range (μM)</bold></th>
<th align="center" valign="top"><bold>Response (s)</bold></th>
<th align="center" valign="top"><bold>Anti-interference</bold></th>
<th align="center" valign="top"><bold>Reference</bold></th></tr></thead>
<tbody>
<tr>
<td align="center" valign="top">Carbon Paste Electrode</td>
<td align="center" valign="top">0.5 mg/L</td>
<td align="center" valign="top">2.0–32.5 mg/L</td>
<td align="center" valign="top">N/A</td>
<td align="center" valign="top">Cr<sup>3+</sup> and Zn<sup>2+</sup></td>
<td align="center" valign="top">[<xref ref-type="bibr" rid="b7-sensors-12-15063">7</xref>]</td></tr>
<tr>
<td align="center" valign="middle">Pt electrode</td>
<td align="center" valign="middle">0.49 μg/L</td>
<td align="center" valign="middle">0.49–783.21 μg/L and 783.21 μg/L–25.55 mg/L</td>
<td align="center" valign="middle">60 s</td>
<td align="center" valign="middle">Cr<sup>3+</sup>, Pb<sup>2+</sup>, Cu<sup>2+</sup> and Cd<sup>2+</sup></td>
<td align="center" valign="middle">[<xref ref-type="bibr" rid="b8-sensors-12-15063">8</xref>]</td></tr>
<tr>
<td align="center" valign="middle">Thiolate self-assembled monolayer</td>
<td align="center" valign="middle">0.2 ppb</td>
<td align="center" valign="middle">1–100 ppb</td>
<td align="center" valign="middle">N/A</td>
<td align="center" valign="middle">Ascorbic acid, H<sub>2</sub>O<sub>2</sub>, Cu<sup>2+</sup> and Zn<sup>2+</sup></td>
<td align="center" valign="middle">[<xref ref-type="bibr" rid="b9-sensors-12-15063">9</xref>]</td></tr>
<tr>
<td align="center" valign="top">Pt/PPD/GOD</td>
<td align="center" valign="top">2.5 μM</td>
<td align="center" valign="top">5–180 μM</td>
<td align="center" valign="top">100 s</td>
<td align="center" valign="top">Cu<sup>2+</sup>, Cd<sup>2+</sup>, Co<sup>2+</sup> and Ni<sup>2+</sup></td>
<td align="center" valign="top">[<xref ref-type="bibr" rid="b10-sensors-12-15063">10</xref>]</td></tr></tbody></table></table-wrap>
<table-wrap id="t2-sensors-12-15063" position="float">
<label>Table 2.</label>
<caption>
<p>The summary of calculated selective coefficient values.</p></caption>
<table frame="hsides" rules="groups">
<thead>
<tr>
<th align="center" valign="middle"><bold>Interferences</bold></th>
<th align="center" valign="middle">
<mml:math id="mm4" display="inline">
<mml:semantics id="sm4">
<mml:mrow>
<mml:mo>log</mml:mo>
<mml:msubsup>
<mml:mi>K</mml:mi>
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>B</mml:mi></mml:mrow>
<mml:mrow>
<mml:mtext mathvariant="italic">pot</mml:mtext></mml:mrow></mml:msubsup></mml:mrow></mml:semantics></mml:math></th></tr></thead>
<tbody>
<tr>
<td align="center" valign="top">Cu<sup>2+</sup></td>
<td align="center" valign="top">−3.05226</td></tr>
<tr>
<td align="center" valign="top">Zn<sup>2+</sup></td>
<td align="center" valign="top">−3.05009</td></tr>
<tr>
<td align="center" valign="top">Fe<sup>2+</sup></td>
<td align="center" valign="top">−2.60417</td></tr>
<tr>
<td align="center" valign="top">Co<sup>2+</sup></td>
<td align="center" valign="top">−2.86609</td></tr></tbody></table></table-wrap></sec></back></article>