The prototype of the radiation detection system was designed, implemented, assembled, and experimentally validated by the authors as an extension of the work previously presented in Pavón et al. (2024) [
5]. The developed system integrates the Geiger–Müller tube, high-voltage supply, nuclear pulse conditioning circuit, and data acquisition interface within a compact electronic structure specifically built for this research.
Figure 6 shows the final assembled prototype constructed by the authors and used during the laboratory and field validation tests.
4.1. Detector Response Behavior Within the Geiger Plateau Region
To evaluate the effectiveness of the proposed nuclear pulse conditioning architecture, a first experimental procedure was carried out using a known gamma radiation reference source positioned near the detector. The objective of this experiment was not to perform a metrological calibration, but rather to validate the correct operation of the Geiger–Müller tube, verify the stability of the pulse conditioning stages, and experimentally obtain the detector response behavior within the Geiger plateau region. Since this stage focused on functional validation and pulse analysis, environmental compensation variables such as temperature and humidity were not incorporated into the procedure. The experiment allowed verification of stable pulse generation, proper high-voltage operation, and repeatable count acquisition under controlled exposure conditions.
Figure 7 and
Figure 8 present the experimental setup used during the first validation stage. The detector was exposed to the gamma radiation reference source at controlled distances in order to analyze the detector response and evaluate the pulse counting stability. This setup was also used to experimentally determine the Geiger plateau behavior and verify the effectiveness of the nuclear pulse conditioning process.
To experimentally verify the effectiveness of the proposed nuclear pulse conditioning architecture, oscilloscope measurements were performed at different stages of the signal path. The objective of these measurements was to validate the correct transformation of the high-voltage analog pulses generated by the Geiger–Müller tube into stable TTL-compatible digital signals suitable for pulse counting and real-time acquisition systems. The measurements also allowed verification of signal polarity inversion, pulse standardization, and voltage adaptation throughout the conditioning process.
Figure 9 presents the complete pulse conditioning response. The lower trace corresponds to the original nuclear pulse obtained directly from the Geiger–Müller tube, exhibiting narrow high-voltage analog pulses with variable amplitude. The upper trace shows the final conditioned output generated by the monostable stage, producing standardized TTL pulses with approximately 5 V amplitude and controlled pulse width. The results confirm the successful conversion of unstable analog nuclear events into uniform digital pulses suitable for reliable counting by microcontrollers and SCADA-based acquisition systems.
Figure 10 illustrates the intermediate inversion stage of the conditioning circuit. The lower trace corresponds to the original nuclear pulse, while the upper trace shows the output of the CMOS inverter stage. The observed polarity inversion confirms the correct operation of the logic adaptation stage prior to pulse shaping. This inversion process is necessary to properly trigger the monostable multivibrator and ensure consistent pulse generation in the final output stage.
Figure 11 presents the relationship between the inverter stage and the final monostable output. The lower trace corresponds to the output of the CMOS inverter stage, while the upper trace shows the TTL-compatible pulse generated after pulse shaping. The temporal alignment between both signals confirms the correct triggering sequence of the monostable multivibrator and demonstrates that the inverter stage provides an adequate logic transition for stable pulse generation. This synchronization is essential to ensure reliable pulse counting and avoid false triggering in the digital acquisition stage.
Figure 12 shows the transition between the conditioned inverted nuclear pulse and the final TTL output signal. The lower trace corresponds to the inverted nuclear pulse prior to pulse shaping, while the upper trace shows the standardized TTL pulse produced by the 555 timer stage. The figure demonstrates the successful transformation of narrow and variable analog nuclear events into digitally stable pulses with controlled amplitude and duration. This behavior confirms the effectiveness of the proposed conditioning architecture for converting Geiger–Müller analog pulses into signals compatible with microcontrollers and SCADA-based monitoring systems.
Finally,
Figure 13 presents detailed oscilloscope measurements of the conditioned pulse, including amplitude and pulse-width markers. The results confirm a stable TTL output amplitude close to 5 V and an experimentally measured pulse duration near 1600 μs, consistent with the theoretical timing behavior of the monostable multivibrator implemented using the 555 timer. The measured pulse width validates the selected resistor–capacitor configuration used in the pulse shaping stage and confirms the repeatability and stability of the conditioning process. The standardized pulse duration also establishes the practical counting limit of the system, since excessively high radiation flux may lead to pulse overlap and dead-time effects. Nevertheless, for environmental radiation monitoring applications with moderate count rates, the obtained pulse characteristics ensure stable and reliable operation.
The measured pulse width validates the selected resistor–capacitor configuration used in the monostable stage and confirms the repeatability of the pulse shaping process. The standardized pulse duration also establishes the practical counting limit of the system, since excessively high radiation flux may lead to pulse overlap and dead-time effects. Nevertheless, for environmental radiation monitoring applications with moderate count rates, the obtained pulse characteristics ensure stable and reliable operation.
To assess the operational stability of the Geiger–Müller tube, a real plateau curve was experimentally obtained, as shown in
Figure 14. The curve clearly identifies the ionization region, the Geiger–Müller region, and the onset of continuous discharge. Within the plateau region, the count rate remains relatively stable despite increases in the applied voltage, validating the optimal operating range of the detector. This behavior is essential to ensure that the pulse amplitude and count rate are largely independent of the initial ionization process, thereby enabling consistent and reliable radiation measurements.
The plateau slope was evaluated using the standard Geiger–Müller characterization metric expressed in percentage variation per 100 V, according to Equation (
8), where
and
correspond to the count rates measured at voltages
and
, respectively, and
represents the average count rate within the selected plateau interval. Based on the experimental measurements, the obtained plateau slope was 31.12%/100 V. Although the count rate exhibits a progressive increase with voltage, the detector maintained stable pulse generation within the evaluated operating range, confirming the functionality of the proposed conditioning and high-voltage stages.
4.2. Experimental Calibration and Exposure Response Validation
To evaluate the measurement accuracy, repeatability, and exposure–response behavior of the proposed radiation detection system, a controlled calibration procedure was conducted in 2014 at the Secondary Standards Laboratory of the Ecuadorian nuclear regulatory authority. The experiment was carried out using a certified gamma radiation reference source under controlled laboratory conditions, allowing validation of the detector response in the operational range between 0 and 10 mR/h. Unlike the previous functional validation experiment, this procedure was specifically designed to assess the quantitative response of the system and compare the measured exposure values against certified reference levels.
The calibration process consisted of positioning the detector at different distances from the gamma radiation source while maintaining a fixed attenuation configuration. For each operating point, multiple repeated measurements were acquired in order to evaluate the repeatability and stability of the pulse counting system. The detector response was then compared with the certified reference exposure values provided by the calibration laboratory. Environmental conditions including temperature, pressure, and relative humidity were also documented during the calibration procedure to ensure traceability of the measurements.
Table 4 summarizes the principal equipment, reference instruments, and source characteristics employed during the calibration procedure.
Figure 15 presents the experimental calibration setup used during the laboratory validation tests. The developed prototype was positioned at controlled distances from the certified gamma radiation source in order to obtain exposure measurements under traceable reference conditions. This setup allowed verification of the detector response stability, pulse counting consistency, and exposure estimation capability across different radiation levels.
To improve the statistical consistency and reliability of the calibration analysis, multiple repeated measurements were performed at each exposure level instead of relying on single acquisitions. The calibration procedure evaluated three reference exposure regions corresponding approximately to 25%, 50%, and 75% of the operational scale of the detector. For each operating condition, nine independent measurements were acquired, allowing estimation of the average detector response, analysis of measurement repeatability, and evaluation of statistical dispersion. This approach provides a more representative characterization of the detector behavior and directly addresses the variability associated with pulse-counting systems operating under ionizing radiation exposure.
Table 5 summarizes the repeated measurements obtained during the calibration procedure. The experimental results show relatively low dispersion between repeated acquisitions, particularly at lower exposure levels, indicating stable pulse conditioning and repeatable digital pulse acquisition. Small variations between measurements are mainly attributed to counting statistics inherent to Geiger–Müller detection, minor positioning differences during the experimental setup, and electronic noise associated with the conditioning stages.
The statistical analysis of the calibration results is summarized in
Table 6, where the measured average exposure values are compared against the certified reference exposure levels corresponding to 25%, 50%, and 75% of the radiation source intensity. The obtained results reveal relatively small absolute deviations between measured and reference values, with relative errors ranging from approximately 7.3% to −13%. These deviations are considered acceptable for field-grade environmental radiation monitoring systems and are consistent with the practical limitations of Geiger–Müller-based detectors, including counting statistics, detector sensitivity variations, electronic conditioning tolerances, and geometric alignment uncertainties during calibration.
The results also demonstrate that the proposed conditioning architecture and pulse counting system maintain stable operation across different exposure levels while preserving repeatability in the acquired measurements. Although no complete uncertainty propagation model was implemented, the repeated experimental acquisitions provide a representative estimation of the variability associated with the detector response under controlled laboratory conditions.
To further analyze the consistency and statistical distribution of the experimental measurements,
Figure 16 presents a violin plot of the exposure values obtained at different operating conditions. The distribution shape and spread provide a graphical representation of the variability associated with repeated pulse-counting measurements. The relatively narrow distributions observed at lower exposure levels indicate low measurement dispersion and stable detector behavior, while slightly wider distributions at higher exposure levels reflect the increased variability commonly associated with Geiger–Müller counting statistics and experimental positioning tolerances. Overall, the results confirm the repeatability and robustness of the proposed radiation detection system for environmental monitoring applications.
The exposure–distance behavior obtained during calibration is summarized in
Figure 17. The figure compares the experimentally measured exposure values against the theoretical response expected from the inverse-square radiation propagation model. As expected, the measured exposure decreases as the detector is positioned farther from the source. Although small deviations are observed between experimental and theoretical values, the measured response follows the expected inverse trend with acceptable agreement throughout the evaluated operating range. These deviations are associated with practical measurement uncertainties including counting statistics, detector sensitivity variations, geometric alignment, and electronic conditioning tolerances.
A complementary representation of the calibration results is shown in
Figure 18, where the average measured exposure values are presented together with their corresponding experimental variability. The relatively small dispersion observed in the repeated measurements confirms the repeatability and stability of the proposed pulse conditioning architecture and pulse counting algorithm. These results demonstrate that the system is capable of producing consistent exposure measurements under controlled radiation conditions while maintaining stable signal standardization and digital pulse acquisition.
Finally,
Figure 19 presents the experimental validation of the detector response as a function of the distance to the radioactive source. In contrast to a linear approximation, the measured exposure follows the expected physical behavior of ionizing radiation propagation, which is governed by an inverse power-law relationship. The experimental data were therefore fitted using the model expressed in Equation (
9), where
represents the measured exposure at distance
d, while
a and
b are experimentally determined fitting parameters. The resulting fit yielded the relationship
with a coefficient of determination of
, demonstrating excellent agreement between the experimental measurements and the theoretical attenuation trend. The obtained exponent value close to the ideal inverse-square behavior confirms the consistency of the radiation measurements and validates the detector response over the evaluated operating range. Minor deviations from the theoretical exponent of
can be attributed to environmental scattering effects, detector geometry, source positioning uncertainties, and practical limitations associated with low-cost Geiger–Müller instrumentation. These results confirm that the proposed system accurately reproduces the expected decrease in radiation exposure with increasing distance, supporting its applicability for environmental radiation monitoring and field measurements.
Although the proposed system was intentionally designed as a low-cost and electronically simple radiation monitoring solution, several engineering trade-offs were consciously accepted during development. The architecture prioritizes accessibility, ease of implementation, low power consumption, and stable pulse acquisition over high-precision spectrometric performance typically associated with more complex radiation instrumentation. Consequently, the system provides reliable pulse counting and exposure estimation capabilities for environmental monitoring applications, while accepting moderate measurement deviations associated with Geiger–Müller counting statistics, simplified conditioning electronics, and practical calibration limitations. In addition, the use of a Geiger–Müller tube inherently limits the possibility of energy discrimination, since the detector generates pulses with nearly uniform amplitude independent of the incident radiation energy. Therefore, the proposed architecture is primarily intended for radiation presence detection and exposure monitoring rather than spectroscopic analysis. Nevertheless, the modular structure of the conditioning and acquisition stages facilitates future integration with embedded processing and communication platforms. The standardized TTL-compatible output generated by the conditioning circuit enables straightforward incorporation into microcontroller-, IoT-, and SCADA-based distributed monitoring systems for real-time environmental sensing applications. Future developments may incorporate digital filtering, adaptive dead-time correction, statistical compensation algorithms, and wireless communication interfaces through dedicated microcontroller integration, potentially improving measurement stability, scalability, and real-time processing capabilities under more demanding operating conditions.