Abstract
Microfluidic fluorescence immunoassays enable sensitive biomarker detection, but performance is limited by inefficient capture on planar surfaces and background fluorescence from unbound labels. Here, we present chip-based optical modulation biosensing (Chip-OMB), combining three-dimensional magnetic bead-based target capture, bead immobilization, internal fluid flushing, and modulated optical background subtraction. Magnetic beads are concentrated and immobilized in the chamber, while controlled on-chip flushing displaces unbound fluorescent labels without requiring user-performed washing after chip loading. Alternating optical interrogation of the bead cluster and surrounding solution enables background subtraction. Using fluorescently labeled magnetic beads, Chip-OMB achieved limits of detection of 0.1 pM in buffer and 22 pM in undiluted plasma, corresponding to 580-fold and 22-fold lower equivalent fluorophore-concentration limits of detection, respectively, than conventional steady-state bulk fluorescence measurements using a plate reader. Measurements correlated with the laboratory optical modulation biosensing system (R2 = 0.98). In an anti-dengue virus serotype 2 (DENV-2) nonstructural protein 1 (NS1) immunoglobulin G (IgG) immunoassay, Chip-OMB achieved a 3 ng/mL limit of detection with a ~4-log dynamic range. Among 56 sera, 29 DENV-2 cases confirmed by reverse transcription quantitative polymerase chain reaction (RT-qPCR) were Panbio IgG-positive and above the exploratory Chip-OMB cutoff, while all 27 study-defined negative controls were below it.
1. Introduction
Rapid and accurate detection of disease biomarkers in decentralized and resource-limited settings remains an important analytical challenge, particularly when clinically relevant targets must be detected at low concentrations in complex biological samples [1,2,3]. Conventional laboratory immunoassays can provide high analytical performance, but they often require centralized instrumentation, trained personnel, and multistep workflows that limit their use outside well-equipped laboratories [1,2,4]. Lateral flow assays (LFAs) offer a simple and rapid alternative for point-of-care testing [5], but their analytical sensitivity and quantitative capability are often limited, particularly near the detection limit [4,5]. These limitations motivate the development of compact biosensing platforms that combine simplified operation with sensitive and quantitative detection of biomarkers in complex biological samples.
Microfluidic technologies provide a promising route toward compact biosensing platforms by integrating fluid handling, reaction control, and detection within miniaturized analytical systems [6]. Their ability to process small sample volumes and support biomarker detection in complex biological matrices, including whole blood [7], serum or plasma [8], and saliva [9], has enabled their application to a wide range of diagnostic assays [10]. In particular, fluorescence-based immunoassays implemented in microfluidic devices have been widely investigated for infectious disease diagnostics [11], cancer biomarker detection [12], and modeling inflammatory responses [13]. However, miniaturization alone does not ensure high analytical sensitivity. Sensitive fluorescence immunoassays require efficient target capture together with effective suppression of optical background arising from unbound or nonspecifically retained labels and the surrounding sample matrix.
In many microfluidic affinity assays, capture probes, including antibodies or aptamers, are immobilized on planar channel surfaces [14,15]. As samples flow through the channel, target molecules bind to these surface-immobilized probes and are subsequently labeled with fluorescent reporters, enabling quantitative detection based on fluorescence measurements. However, planar implementations present several fundamental limitations. First, capture is confined to a two-dimensional surface, which can limit the available capture interface and mass transport to the immobilized probes, particularly at low target concentrations [16,17]. Second, steady-state fluorescence readouts are susceptible to optical background arising from unbound or nonspecifically retained labels, sample autofluorescence, and scattering, often necessitating washing steps to suppress background signals [18]. These washing steps increase system complexity and assay time and can hinder implementation in low-resource or point-of-care settings [19]. Finally, at low biomarker concentrations, the fluorescent signal generated by the small number of captured target molecules can be weak relative to the surrounding optical background, often requiring advanced optical instrumentation or signal-enhancement strategies to achieve adequate sensitivity [20,21,22]. Together, these limitations motivate the use of alternative capture architectures that can concentrate target-associated labels while enabling effective suppression of optical background.
Previously, we introduced optical modulation biosensing (OMB), a fluorescence detection strategy that enables background subtraction through temporal modulation without conventional washing steps [23]. In OMB, target molecules are captured in a bead-based sandwich assay using superparamagnetic beads and fluorescently labeled probes. The magnetic beads are aggregated and immobilized using a sharp-tipped permanent magnet, concentrating target-bound fluorophores within a confined detection region. By rapidly alternating the excitation laser between the bead aggregate and the surrounding solution, OMB enables spatially resolved background subtraction of signals arising from unbound labels, sample autofluorescence, and Raman scattering, thereby improving detection of the concentrated target-associated fluorescence [23]. A fully automated implementation of this principle, termed OMBi, demonstrated high sensitivity, specificity, and rapid assay times in laboratory-based, high-throughput microplate formats [24]. This OMB-based approach has also been applied to saliva-based molecular detection, toxin detection, and dengue serology [25,26,27]. However, OMBi remains a microplate-based laboratory platform that relies on bulky instrumentation and external liquid handling. Thus, it does not provide a chip-integrated architecture in which magnetic bead capture and concentration, fluidic displacement of unbound labels, bead immobilization, and optical background subtraction are coordinated within a compact microfluidic device.
Here, we develop chip-based optical modulation biosensing (Chip-OMB), a chip-integrated implementation of OMB that combines magnetic bead-based target capture with on-chip bead concentration and immobilization, controlled fluidic displacement of unbound fluorophores, and optical background subtraction (Figure 1). Chip-OMB is designed to address three key limitations of conventional microfluidic fluorescence assays. First, magnetic beads replace planar capture surfaces, providing a three-dimensional capture architecture that enhances target–probe interactions and enables magnetic concentration of target-bound fluorophores into a confined detection region (Figure 1a) [28]. Second, controlled on-chip flushing displaces unbound fluorescent labels after sample loading while the bead cluster remains immobilized, replacing user-performed washing at the chip stage while retaining the sensitivity advantages of heterogeneous bead-based immunoassays (Figure 1b). Third, optical modulation between the bead cluster and the surrounding solution enables local background subtraction, contributing to the high signal-to-background ratio achieved by the integrated Chip-OMB workflow (Figure 1c) [21,23]. Analytical benchmarking using fluorescently labeled magnetic beads demonstrated limits of detection (LoD) of 0.1 pM in buffer and 22 pM in undiluted plasma, corresponding to approximately 580-fold and 22-fold lower equivalent fluorophore-concentration LoDs, respectively, than those obtained with conventional steady-state bulk fluorescence measurements using a plate reader. Clinical feasibility was evaluated in a 56-sample serum cohort comprising 29 samples from patients with reverse-transcription quantitative polymerase chain reaction (RT-qPCR)-confirmed dengue virus serotype 2 (DENV-2) infection and 27 study-defined negative controls. All 29 patient samples were positive for immunoglobulin G (IgG) on the Panbio Dengue IgG assay. Chip-OMB signals from all patient samples exceeded an exploratory cutoff derived from the 27 negative-control sera, whereas all negative-control signals were below it. These findings demonstrate the feasibility of integrating established magnetic bead-based capture, controlled on-chip fluidic background reduction, and OMB-based optical detection within a single microfluidic assay architecture for sensitive fluorescence detection in complex biological samples.
Figure 1.
Concept and operating principle of the Chip-OMB system. (a) Functionalized magnetic beads are suspended within the microfluidic chamber, providing a three-dimensional capture interface and enabling magnetic concentration of target-bound fluorescent complexes into a localized bead cluster. (b) After bead immobilization, controlled internal microfluidic flushing displaces unbound fluorophores while the bead cluster remains fixed, eliminating user-performed washing at the chip stage. (c) Optical modulation is performed by sequentially positioning a single excitation beam on the bead cluster and on an adjacent background region, enabling background-subtracted fluorescence detection.
2. Materials and Methods
2.1. Ethical Approval
All experiments involving the collection and testing of biological materials from human participants were conducted in accordance with the principles of the Declaration of Helsinki and protocols approved by the Institutional Ethical Review Board of Hanoi Medical University, Vietnam (IRB-VN01.001/IRB00003121/FWA 00004148; Approval No. 615/GCN-HĐĐĐNCYSH-ĐHYHN, approved 7 April 2022), and the Institutional Review Board of Sheba Medical Center, Israel (IRB 4420-17).
2.2. Informed Consent
Written informed consent was obtained from all participants whose serum samples were included in this study.
2.3. System Structure
The Chip-OMB optical system (Figure S1) employs a diode-pumped solid-state 532 nm laser module (CPS532, Thorlabs, Newton, NJ, USA) with a 3.5 mm diameter beam and an output power of 0.5 mW. The laser beam is reshaped using a pair of plano-convex lenses (focal lengths of 200 and 100 mm; LA1708 and LA1509, Thorlabs), redirected by a scanning galvo mirror (GVS211, Thorlabs), and then directed vertically by a dichroic mirror (BrightLine Di02-R532, Semrock, Rochester, NY, USA). The beam is focused by an infinity-corrected objective (MY10X-803, Mitutoyo Plan Apochromatic Objective, 0.28 numerical aperture, 34 mm working distance; Thorlabs) to a 150 μm-diameter spot on the microfluidic reaction chamber containing the biological sample.
Unless otherwise specified, experiments were performed using a 10 μL reaction-chamber chip (#10001268, Topas, Fluidic 585; 10 μL volume, 500 μm depth, and 140 μm lid thickness; microfluidic ChipShop GmbH, Jena, Thüringen, Germany). A 20 μL chip was used for the buffer dose–response experiments, as described below. Magnetic beads conjugated to fluorescently labeled targets were aggregated and immobilized at the detection spot using a cylindrical, axially magnetized neodymium N42 permanent magnet (6.35 × 25.4 mm, D4X0, K&J Magnetics, Pipersville, PA, USA) equipped with a conical tip (6.35 mm diameter, 8.62 mm height; Hiperco 50, Ed Fagan, Kingsbridge, UK) and positioned beneath the microfluidic chip.
Fluorescence emitted from the detection region is collected by the objective, transmitted through the dichroic mirror, and focused by a plano-convex lens (100 mm focal length, LA1509, Thorlabs). The emitted light is then passed through two emission filters (FF01-575/25, Semrock) and detected using a complementary metal-oxide-semiconductor (CMOS) camera (GS3-U3-23S6M, FLIR Systems, Wilsonville, OR, USA) operating at 50 frames per second.
2.4. Image Analysis
For each measurement, 50 frames were acquired over 1 s while the excitation laser alternated between the bead-cluster and background positions on successive frames, resulting in 25 frames with the laser focused on the bead cluster and 25 frames with the laser focused on the surrounding background region. Each frame contained two regions of interest (ROIs): one encompassing the bead cluster and one containing only the surrounding background solution. The ROIs were identified automatically from the acquired image sequence using an image-processing approach adapted from our previously reported spatiotemporal fluorescence analysis method [29]. Briefly, the bead-cluster ROI was assigned to the laser position containing the magnetically immobilized bead aggregate, whereas the background ROI was assigned to the corresponding laser position in the surrounding solution without beads. The same ROIs were maintained throughout all 50 frames of each measurement.
When the laser was focused on the bead cluster, the fluorescence measured within the bead-cluster ROI contained both the bead-associated signal and the local background contribution. When the laser was focused on the background region, the fluorescence measured within the background ROI represented the local background signal. The background-subtracted fluorescence signal was calculated by subtracting the background-region intensity measured during background illumination from the bead-cluster ROI intensity measured during bead-cluster illumination [23,24].
2.5. Dose Response of Fluorescently Labeled Magnetic Beads and Correlation Between Chip-OMB and OMBi
2.5.1. Dose Response in Buffer
To determine the LoD and dynamic range of the Chip-OMB system, approximately 1.5 × 105 streptavidin-coated magnetic beads (2.8 μm; Dynabeads M-280 Streptavidin, Thermo Fisher Scientific, Waltham, MA, USA) were incubated for 5 min with biotinylated Atto 532 (AD 532-71, ATTO-TEC, Siegen, Germany) in 120 μL of reaction buffer consisting of phosphate-buffered saline (PBS) supplemented with 10 mg mL−1 bovine serum albumin (BSA) and 0.05% (v/v) Tween-20. Dye-loading levels corresponded to 0 and 101–107 dye molecules per bead [23]. Each suspension was divided into six equal 20 μL aliquots, containing approximately 2.5 × 104 beads per aliquot. The total number of dye molecules per aliquot therefore ranged from 0 to 2.5 × 1011. Within each independent experiment, the blank preparation was performed twice without added dye, yielding a total of twelve 20 μL blank aliquots.
Four aliquots from each nonzero dye-loading preparation and eight blank aliquots were analyzed using Chip-OMB in a 20 μL microfluidic reaction-chamber chip (#10000560, Topas, Fluidic 559; 20 μL volume, 350 μm depth, and 140 μm lid thickness; microfluidic ChipShop GmbH, Jena, Thüringen, Germany). Beads were immobilized using a conical-tipped permanent magnet, and unbound dye was displaced by flowing 20 μL of reaction buffer through the chamber at 0.75 mL min−1 using a syringe pump (NE-300, New Era Pump Systems Inc., Farmingdale, NY, USA).
To evaluate the correlation between Chip-OMB and the laboratory OMBi system, the two remaining aliquots from each nonzero dye-loading preparation and the four remaining blank aliquots were transferred to a 96-well plate and measured using OMBi after a single buffer replacement, with a final volume of 100 μL per well. Three independent experiments were performed (n = 3), and data are reported as mean ± standard error of the mean (SEM).
For comparison with conventional steady-state bulk fluorescence measurements, we initially attempted to measure the same fluorescently labeled magnetic beads using a conventional plate reader after magnetically aggregating the beads in the wells. However, the bead-associated fluorescence was insufficient for reliable quantitative analysis under the plate-reader measurement conditions. We therefore used unconjugated Atto 532 dye in bulk solution as the conventional steady-state fluorescence benchmark. Serial 10-fold dilutions of unconjugated Atto 532 dye (i.e., dye not bound to magnetic beads) were prepared in reaction buffer and measured using a Synergy H1 plate reader (Agilent Technologies, Santa Clara, CA, USA). Unlike Chip-OMB, which combines magnetic concentration with temporally modulated background subtraction, the plate reader measures steady-state fluorescence from the bulk solution under continuous uniform illumination. At each nonzero dye concentration, four 100 μL aliquots were measured; eight 100 μL aliquots were measured for the blank. The total number of dye molecules per aliquot ranged from 0 to 2.5 × 1011, matching the total fluorophore counts used in the corresponding Chip-OMB measurements. Fluorescence was measured using excitation and emission wavelengths of 532 and 560 nm, respectively. Three independent experiments were performed (n = 3), and data are reported as mean ± SEM.
2.5.2. Dose Response in 100% Plasma
To evaluate analytical performance in a complex biological matrix, the Atto 532 dose–response experiment was repeated in 100% plasma. Given the higher viscosity of plasma and potential bead loss or retention on channel surfaces, the number of beads per reaction was increased threefold to promote reproducible coverage of the laser interrogation region by the immobilized bead cluster. Accordingly, approximately 3 × 105 streptavidin-coated magnetic beads (Dynabeads M-280 Streptavidin, Thermo Fisher Scientific) were incubated for 5 min in 40 μL of plasma with biotinylated Atto 532 (AD 532-71, ATTO-TEC) at loading levels corresponding to 0 and 101–108 dye molecules per bead [23]. Each suspension was divided into four equal 10 μL aliquots, containing approximately 7.5 × 104 beads per aliquot. The total number of dye molecules per aliquot therefore ranged from 0 to 7.5 × 1012. Within each independent experiment, the blank preparation was performed twice without added dye, yielding a total of eight 10 μL blank aliquots.
Four aliquots from each nonzero dye-loading preparation and eight blank aliquots were analyzed using Chip-OMB, with the beads immobilized by a conical-tipped permanent magnet. Unbound dye was displaced by flowing 10 μL of 100% plasma through the chamber at 3 mL min−1 using a syringe pump. This higher flow rate was used to facilitate removal of unbound fluorophores in the plasma experiments. Three independent experiments were performed (n = 3), and data are reported as mean ± SEM.
For comparison with conventional steady-state bulk fluorescence measurements in plasma, serial 10-fold dilutions of unconjugated Atto 532 dye (i.e., dye not bound to magnetic beads) were prepared in 100% plasma and measured using a Synergy H1 plate reader. At each nonzero dye concentration, four 100 μL aliquots were measured; eight 100 μL aliquots were measured for the blank. The total number of dye molecules per aliquot ranged from 0 to 7.5 × 1012, matching the total fluorophore counts used in the corresponding Chip-OMB measurements. Three independent experiments were performed (n = 3), and data are reported as mean ± SEM.
2.6. Dose Response of an Antibody as a Biological Target
The analytical performance of Chip-OMB was evaluated using an immunoassay for anti-DENV-2 nonstructural protein 1 (NS1) IgG. For each preparation, approximately 5 × 104 tosyl-activated magnetic beads in 10 μL (M-280, Thermo Fisher Scientific) were functionalized with recombinant DENV-2 NS1 antigen (DENV-X4-NS1-100, The Native Antigen Company, Kidlington, UK) according to the manufacturer’s coupling protocol [30].
The antigen-functionalized beads were incubated for 1 h at room temperature with 20 μL of mouse anti-DENV-2 NS1 IgG (MAB12133-100, The Native Antigen Company) at concentrations ranging from 10−1 to 104 ng mL−1. Blank preparations contained no target antibody (0 ng mL−1) and otherwise underwent the same assay procedure. The beads were then washed to remove unbound target antibodies, resuspended in 20 μL of fresh reaction buffer, and incubated for 30 min at room temperature with Alexa Fluor 532 goat anti-mouse IgG (A-11002, Thermo Fisher Scientific; 2 μg mL−1).
Following incubation, 10 μL aliquots were loaded into the microfluidic chip, where the beads were immobilized by a conical-tipped permanent magnet beneath the chamber. Unbound fluorescently labeled detection antibodies were displaced by flowing 10 μL of buffer through the chamber at 0.75 mL min−1 using a syringe pump. Thus, pre-loading washing removed unbound target antibodies before fluorescent labeling, whereas on-chip flushing removed unbound detection antibodies after bead immobilization.
Bead number, functionalization, incubation conditions, bead dispersion, and magnetic immobilization were kept constant across measurements. Each target-antibody concentration was evaluated in three independent experiments performed on separate days (n = 3). Within each experiment, two independently prepared bead suspensions were each divided into duplicate 10 μL aliquots, yielding four replicate measurements per nonzero target-antibody concentration. For the blank, four independently prepared bead suspensions were each divided into duplicate 10 μL aliquots, yielding eight blank measurements per experiment. Data are reported as mean ± SEM.
2.7. Clinical Feasibility Evaluation of Chip-OMB
2.7.1. Sample Collection
The clinical feasibility of Chip-OMB was evaluated using 56 serum samples, comprising 29 samples from patients with RT-qPCR-confirmed DENV-2 infection and 27 study-defined negative-control samples. The 29 DENV-2-positive samples were collected at Hanoi Medical University from patients with infection confirmed by RT-qPCR. These clinical samples were previously described [27]. For the present study, aliquots of the same samples were re-evaluated for Dengue IgG seropositivity using the Panbio™ Dengue IgG Indirect enzyme-linked immunosorbent assay (ELISA) kit (#01PE30, Abbott, Chicago, IL, USA) at the National Center for Zoonotic Viruses, Central Virology Laboratory, Sheba Medical Center, and the resulting measurements were used to establish the reference classification (Table S1). Chip-OMB measurements of these samples were subsequently performed and analyzed as part of the present study.
The study-defined negative-control cohort consisted of 27 serum samples collected from healthy Israeli donors with no documented history of dengue virus exposure. All specimens were stored at −80 °C, transported on dry ice to Bar-Ilan University, and subjected to a single freeze–thaw cycle before analysis.
2.7.2. Detection of Anti-DENV-2 NS1 IgG in Clinical Samples
The clinical feasibility evaluation of Chip-OMB was conducted using a bead-based serological assay targeting anti-DENV-2 NS1 IgG. Approximately 5 × 104 tosyl-activated magnetic beads (M-280 Tosylactivated, Thermo Fisher Scientific) functionalized with recombinant DENV-2 NS1 antigen (DENV-X4-NS1-100, The Native Antigen Company, UK) were incubated off-chip with 2 μL of clinical serum for 60 min at room temperature to allow binding of anti-DENV-2 NS1 IgG. Following serum incubation, the beads were washed to remove unbound serum components, resuspended in 20 μL of reaction buffer, and incubated for 30 min at room temperature with a phycoerythrin (PE)-labeled detection antibody (donkey F(ab′)2 anti-human IgG H&L-PE, ab7005, Abcam, Cambridge, UK, 2 µg mL−1).
For each serum sample, two independently prepared bead suspensions were generated, and each suspension was divided into duplicate 10 μL aliquots before loading onto the Chip-OMB microfluidic chip. Following sample loading, the beads were magnetically immobilized within the chip, and unbound fluorescent labels were displaced by flowing 10 μL of reaction buffer through the chamber at 0.75 mL min−1. Fluorescence was then measured under the standard Chip-OMB operating conditions.
Each experimental run included two no-serum assay controls, in which serum was replaced with PCR-grade water, and two positive assay controls. An exploratory cutoff of 2.17 normalized signal units was calculated as the mean normalized Chip-OMB signal of the 27 study-defined negative-control sera plus three standard deviations. This cutoff was used descriptively to assess separation between the two study-defined cohorts and was not independently validated.
2.8. Analytical Performance and Statistical Analysis
The LoD for each dose–response experiment was determined using the conventional 3σ criterion. Specifically, the detection threshold was defined as the mean blank signal plus three standard deviations (SD) of the blank measurements, and the LoD was determined as the analyte concentration corresponding to this threshold on the dose–response curve. The dynamic range was defined as the concentration interval over which the assay signal increased above the LoD without reaching saturation.
Same-day (intra-assay) repeatability and day-to-day (inter-assay) reproducibility were evaluated using the coefficient of variation (CV), calculated as the standard deviation divided by the mean signal and expressed as a percentage. Same-day CVs were calculated across replicate aliquots within each experiment for both Chip-OMB and the plate reader, where applicable. Day-to-day CVs were calculated for Chip-OMB across three independent experiments performed on separate days. Reported values represent the maximum CV across the tested concentration range. Unless otherwise stated, data are reported as mean ± SEM from independent experiments.
For the clinical feasibility analysis, an exploratory cutoff was calculated as the mean normalized Chip-OMB signal of the 27 study-defined negative-control sera plus three SDs. The numbers of samples in each cohort with signals above and below this cutoff were reported descriptively. Because the same negative-control cohort was used both to derive the cutoff and to assess group separation, diagnostic sensitivity and specificity were not estimated.
2.9. Use of Generative Artificial Intelligence
OpenAI ChatGPT (GPT-5.6 Sol) was used for language refinement and manuscript editing, and ChatGPT’s AI image-generation tool was used to prepare the Graphical Abstract. These tools were not used for study design, data generation, data analysis, or interpretation.
3. Results
3.1. Dose–Response of Fluorescently Labeled Magnetic Beads in Buffer and Plasma and Correlation with the OMBi System
Figure 2a presents the dose–response curve of fluorescently labeled magnetic beads in buffer measured using the Chip-OMB system. Signal intensity is plotted as a function of the total fluorophore concentration in the assay volume, expressed in picomolar (pM) and calculated from the total number of dye molecules per aliquot and the corresponding assay volume. Chip-OMB exhibited a 4-log dynamic range and achieved an LoD of 0.1 pM for Atto 532. Under the corresponding conventional steady-state bulk fluorescence measurement, the Synergy H1 plate reader yielded an LoD of 58 pM. Same-day repeatability and day-to-day reproducibility, expressed as CV, were below 26% and 24%, respectively, across three independent experiments.
Figure 2.
(a) Dose–response curves of fluorescently labeled magnetic beads measured using Chip-OMB in buffer and of unconjugated Atto 532 measured using a Synergy H1 plate reader. Signal intensity is plotted against the total fluorophore concentration in the assay volume, calculated from the number of dye molecules per aliquot and the corresponding assay volume. Chip-OMB achieved an LoD of 0.1 pM, compared with 58 pM for the plate reader, and exhibited a 4-log dynamic range. The CV values shown in the plot represent same-day repeatability and were below 26% for Chip-OMB and 17% for the plate reader; the day-to-day CV for Chip-OMB was below 24%. (b) Corresponding dose–response measurements in 100% plasma. Chip-OMB achieved an LoD of 22 pM, compared with 483 pM for the plate reader, corresponding to an approximately 22-fold lower LoD. The CV values shown in the plot represent same-day repeatability and were below 16% for Chip-OMB and 18% for the plate reader; the day-to-day CV for Chip-OMB was below 25%. Error bars in panels (a,b) represent the SEM from three independent experiments (n = 3). At some concentrations, the error bars are smaller than the data markers and are therefore not visually distinguishable. (c) Comparison of normalized signals obtained using Chip-OMB and the laboratory OMBi system for fluorescently labeled magnetic beads in buffer (R2 = 0.98).
Figure 2b shows the corresponding dose–response curve measured in 100% plasma. Chip-OMB maintained a 4-log dynamic range in this complex biological matrix and achieved an LoD of 22 pM, whereas the corresponding plate-reader measurement yielded an LoD of 483 pM. These values correspond to approximately a 22-fold lower detection limit for Chip-OMB. Same-day repeatability and day-to-day reproducibility were below 16% and 25%, respectively, across three independent experiments.
To evaluate the consistency of Chip-OMB measurements with the laboratory OMBi platform, normalized signals obtained using the two systems were compared (Figure 2c). A strong linear correlation was observed, with a coefficient of determination (R2) of 0.98.
3.2. Analytical Performance of Chip-OMB for Anti-DENV-2 NS1 IgG Detection
Figure 3 shows the dose–response curve of the Chip-OMB system for anti-DENV-2 NS1 IgG. The system achieved an LoD of 3 ng mL−1 and maintained ~4-log dynamic range. Intra-assay variability, expressed as CV, was below 24%, while inter-assay variability across three independent experiments (n = 3) remained below 13% across the tested concentration range.
Figure 3.
Dose–response curve of the Chip-OMB system for anti-DENV-2 NS1 IgG. Chip-OMB achieved an LoD of 3 ng mL−1 and maintained an approximately four-log dynamic range. Intra-assay and inter-assay CVs were below 24% and 13%, respectively, across the tested concentration range; the CV value shown in the plot represents the inter-assay CV. Error bars represent the SEM from three independent experiments (n = 3).
3.3. Clinical Feasibility Evaluation of Chip-OMB: Detection of Anti-DENV-2 NS1 IgG in Clinical Samples
Figure 4 summarizes the clinical feasibility evaluation of the Chip-OMB system. All 29 sera from patients with RT-qPCR-confirmed DENV-2 infection and Panbio Dengue IgG seropositivity yielded normalized Chip-OMB signals above the exploratory cutoff, whereas all 27 study-defined negative-control sera yielded signals below the cutoff. Thus, complete separation was observed between the two study-defined cohorts. On-chip flushing substantially reduced background fluorescence in the clinical samples, supporting the effectiveness of the integrated flushing step for background reduction (Figure S2).
Figure 4.
Clinical feasibility evaluation of Chip-OMB for anti-DENV-2 NS1 IgG detection. Normalized signals are shown for 29 sera from patients with RT-qPCR-confirmed DENV-2 infection and Panbio Dengue IgG seropositivity and 27 study-defined negative-control sera. The exploratory cutoff (dotted line) was calculated as the mean normalized signal of the negative-control cohort plus three standard deviations. All samples in the DENV-2/Panbio IgG-positive cohort yielded signals above the cutoff, whereas all study-defined negative-control sera yielded signals below it.
4. Discussion
Microfluidic immunoassays offer an attractive approach for decentralized biomarker detection, but miniaturization alone does not overcome the sensitivity limitations of fluorescence-based readout [31,32,33]. In many surface-based microfluidic assays, target capture is confined to planar channel surfaces [16,17,22], while unbound fluorescent labels, sample autofluorescence, and optical scattering contribute to background and can necessitate washing steps or more complex optical instrumentation [18,19]. These limitations become particularly important at low analyte concentrations, where only a small number of target-bound fluorophores are present within the optical interrogation volume [20,21]. Thus, a key challenge in microfluidic fluorescence biosensing is to integrate efficient target capture and concentration with effective background suppression in a simple and reproducible workflow.
Chip-OMB builds on the OMB principle, in which fluorescence from the bead cluster is distinguished from background fluorescence by alternating illumination between the bead cluster and the surrounding solution and subtracting the corresponding background ROI signal [23]. We extend this principle from a laboratory microplate format to a microfluidic architecture that integrates magnetic bead-based capture and concentration, internal microfluidic flushing, and optical modulation within a chip-scale detection chamber. The magnetic beads provide a three-dimensional capture interface and enable concentration of target-bound fluorophores into a confined detection volume. Following sample loading, the beads are magnetically immobilized within the microfluidic chamber, allowing controlled on-chip flushing to displace unbound fluorescent labels while retaining the bead cluster. This fluidic separation reduces background without requiring user-performed washing after chip loading. Optical modulation between the bead cluster and the surrounding solution then enables subtraction of the background fluorescence measured in the surrounding solution from the fluorescence measured at the bead cluster [23]. Thus, optical, magnetic, and fluidic operations are coordinated within a single enclosed microfluidic workflow, making the chip an integrated assay environment rather than simply a miniaturized sample holder.
The contribution of Chip-OMB lies in both the coordinated implementation of established sensing operations and the adaptation of the OMB measurement geometry to a shallow microfluidic chamber. This study demonstrates a microfluidic implementation combining OMB-based optical background subtraction with magnetic bead concentration and immobilization during controlled on-chip flushing. This integration addresses the coupled requirements of aggregating and retaining beads during fluid flow, including in undiluted plasma, and optically interrogating both the bead aggregate and an adjacent background region within a confined chamber. Compared with the approximately 3 mm sample depth in the laboratory microplate configuration at a working volume of 100 µL, the 350–500 µm-deep microfluidic chambers reduce the sample depth traversed by the excitation light and collected fluorescence. This geometry may reduce optical attenuation in absorbing or scattering matrices and facilitate magnetic bead aggregation in viscous samples by shortening the vertical distance over which beads are attracted toward the magnet. In the plasma experiments, approximately 75,000 beads per chamber were used, compared with approximately 25,000 in buffer, to promote sufficient coverage of the optical interrogation region. Although the individual technologies are established, this study demonstrates their coordinated operation within a shallow microfluidic chamber, enabling magnetic bead retention during controlled flushing and sensitive modulation-based fluorescence detection in buffer and undiluted plasma.
Analytically, Chip-OMB achieved an LoD of 0.1 pM for fluorescently labeled magnetic beads in buffer, compared with 58 pM for the conventional steady-state bulk fluorescence benchmark, corresponding to an approximately 580-fold lower LoD. In undiluted plasma, Chip-OMB maintained a 4-log dynamic range and achieved an LoD of 22 pM, compared with 483 pM for the corresponding bulk fluorescence benchmark, corresponding to an approximately 22-fold lower LoD. The smaller difference observed in plasma is consistent with the increased optical background and other matrix-related effects that can affect fluorescence measurements in complex biological samples [8,18]. The ability to obtain a measurable dose response and picomolar-level LoD in undiluted plasma demonstrates that the modulation-based readout can be applied in a complex biological matrix.
Cell-free hemoglobin, particularly in hemolyzed plasma, may attenuate the 532 nm excitation and fluorescence detected around 575 nm through absorption, despite its weak intrinsic fluorescence [34]. Bilirubin may introduce wavelength-dependent absorption, whereas lipoprotein particles can increase turbidity and light scattering. The magnitude of these effects depends on constituent concentrations, spectral overlap, and optical path length. Although optical modulation enables subtraction of additive background signals, it does not necessarily compensate for attenuation of bead-associated fluorescence through absorption or scattering, or for fluorophore-specific quenching. The fluorescent-bead experiments in undiluted plasma demonstrate performance in the tested matrix but did not independently vary hemoglobin, bilirubin, or lipid levels and therefore do not establish quantitative interference tolerance. Controlled studies varying each constituent separately will be needed to quantify signal bias and establish assay-specific tolerance limits [35].
To place Chip-OMB in the context of related magnetic-particle-based microfluidic fluorescence immunoassays, representative platforms are compared in Table S2. Hsu et al. [36] used digital microfluidics to perform incubation, labeling, and washing with a fixed number of magnetic beads, followed by bead aggregation for fluorescence detection. Delshadi et al. [37] localized magnetic nanoparticles on micromagnet arrays and used spatial fluorescence background subtraction to enable detection without washing. Chip-OMB combines bead concentration and immobilization using a conical-tipped magnet with controlled on-chip flushing and alternating optical interrogation of the bead cluster and surrounding solution. This comparison highlights differences in measurement geometry, background reduction, and workflow integration, consistent with the present study’s focus on translating OMB into a microfluidic format.
The comparison with the laboratory OMBi system supports preservation of the OMB readout response in the microfluidic format, as indicated by the strong linear relationship between measurements of fluorescently labeled beads in buffer (R2 = 0.98) [24]. Beyond this agreement, Chip-OMB demonstrated fluorescence measurement directly in undiluted plasma, achieving a fluorophore-concentration LoD of 22 pM and a four-log dynamic range. Previous OMBi work demonstrated ricin detection following target capture in 50% plasma, with washing and buffer replacement before fluorescence measurement [26]. Thus, the present fluorescent-bead experiments extend the demonstrated operating conditions of modulation-based readout to an undiluted plasma measurement environment. To promote coverage of the optical interrogation region in plasma, approximately 75,000 beads per chamber were used, compared with 25,000 in buffer. The lower fluorophore-concentration detection limits relative to conventional steady-state bulk fluorescence measurements reflect the combined effects of bead-based fluorophore capture, magnetic concentration, on-chip fluid handling, and modulated background subtraction; they do not isolate the contribution of chip integration or any individual component. The microfluidic implementation provides an enclosed measurement environment that retains the bead cluster during controlled flushing and offers a pathway toward cartridge-based operation [31,32,33]. Although the current prototype retains off-chip assay preparation, external fluidic control, and benchtop optics, it establishes a practical architecture for future fully integrated Chip-OMB systems.
Because Chip-OMB relies on magnetic immobilization of the beads within a confined microfluidic chamber, variations in bead number, cluster density, or cluster geometry could affect the amount and spatial distribution of bead-associated fluorescence within the optical interrogation region and thereby contribute to measurement variability. However, the 150 µm laser spot interrogates only a portion of the larger bead cluster, following the geometry of the previously reported OMBi platform [23,24]. In the buffer experiments, approximately 25,000 beads were loaded per chamber, while the illuminated population is estimated to comprise approximately 4000 beads. A modest reduction in the total retained bead number after labeling is therefore not expected to cause a proportional signal decrease, provided that the illuminated region maintains similar bead density and thickness. Interrogating several thousand beads also provides ensemble averaging of independent bead-to-bead differences in fluorophore loading. These factors were controlled as far as was practical in the present experiments by maintaining a consistent bead concentration, using the same magnetic immobilization configuration and flow conditions, and performing measurements with independently prepared bead suspensions and replicate aliquots. For the anti-DENV-2 NS1 IgG assay, the intra-assay CV was below 24% and the inter-assay CV was below 13%. These measurements indicate the reproducibility of the complete assay under the tested conditions, but they do not independently quantify the contribution of bead-cluster morphology, local bead density, or bead aggregation to the measured fluorescence. Variations in magnetic aggregation therefore remain a potential source of measurement variability. Future studies could address this contribution directly by characterizing bead-cluster size, density, or morphology and correlating these parameters with the corresponding fluorescence measurements.
The analytical performance of Chip-OMB was further demonstrated using an anti-DENV-2 NS1 IgG immunoassay. In the dose–response experiments, Chip-OMB achieved an LoD of 3 ng mL−1 and maintained ~4-log dynamic range, supporting quantitative detection over a broad antibody concentration range. This performance is consistent with previous OMB-based anti-DENV-2 NS1 serological assays, which demonstrated sensitive antibody detection while reducing background from unbound labels and biological matrices [27]. The present study extends these assays into a chip-based microfluidic format while preserving sensitive modulation-based fluorescence detection.
Clinical feasibility was evaluated in a 56-sample serum cohort. All 29 sera from patients with RT-qPCR-confirmed DENV-2 infection and Panbio Dengue IgG seropositivity yielded signals above the exploratory cutoff, whereas all 27 study-defined negative-control sera yielded signals below it, demonstrating complete separation between the two study-defined cohorts. This finding should be interpreted as exploratory because the cutoff was derived from the same 27 study-defined negative-control sera used to assess group separation and was not independently validated. Accordingly, the present data do not provide independently validated estimates of diagnostic sensitivity or specificity. The present study also did not evaluate serotype specificity or cross-reactivity with other flavivirus infections, which are important considerations in DENV serology [38,39,40]. Our previous OMB study evaluated cross-reactivity of the DENV-2 assay with DENV-1-positive sera and of related DENV-1 and DENV-3 assays with other flavivirus infections [27]; these findings provide relevant context but do not establish the cross-reactivity profile of the present Chip-OMB implementation. Thus, the current study does not address all aspects of dengue serological interpretation but demonstrates sensitive fluorescence-based antibody detection in a closed microfluidic format with reduced manual processing. The clear separation observed between positive and negative samples supports the potential of Chip-OMB for serological testing in settings where sensitive detection and simplified workflows are desirable.
Several limitations should be acknowledged. First, broader clinical studies using prospectively collected samples, different stages of infection, clinically relevant comparator assays, and samples from other flavivirus infections will be needed to establish diagnostic performance and evaluate serotype specificity and cross-reactivity. Second, although the workflow eliminates user-performed washing after chip loading, pre-chip assay steps, including sample incubation with functionalized beads and detection reagents, are not yet integrated into a single cartridge-based workflow. Third, the analytical experiments provide proof-of-concept characterization based on three independent runs performed on different days, with replicate measurements at each concentration. These experiments enabled assessment of dose response, estimated detection limits, within-day repeatability, and between-day precision. More extensive studies would be required for comprehensive analytical validation and establishment of a validated quantitative range. Fourth, the present Chip-OMB system uses a benchtop optical configuration, including a diode-pumped solid-state laser, galvanometric beam scanning, microscope objective, optical filters, and a CMOS camera. Although this configuration provides controlled optical excitation and imaging for the present proof-of-concept study, these components contribute to the size and complexity of the current platform. Future miniaturization could replace the benchtop optical components with compact semiconductor light sources, simplified illumination or beam-steering architectures, and miniature image sensors or photodetectors. A further route to simplifying the optical configuration is to incorporate chopped optical biosensing (COB), which measures background fluorescence before bead aggregation and target-plus-background fluorescence after aggregation without beam scanning [41]. Adapting this approach to Chip-OMB could eliminate the galvanometric scanning mirror and its associated control electronics, facilitating future miniaturization. Finally, the present system should be viewed as a prototype microfluidic biosensing architecture. Integration of sample and reagent handling, fluidic control, magnetic actuation, and compact optical detection into a cartridge-based platform represents an important engineering step toward translation and decentralized testing rather than a conceptual limitation of the sensing principle [31,32,33].
Future development should focus on cartridge-level automation while preserving the controlled on-chip flushing and optical modulation principles demonstrated here. Expansion to multiplexed assays using antigen panels, spectrally encoded beads, or spatially separated reaction chambers could broaden the platform beyond single-target detection and support more comprehensive serological characterization.
5. Conclusions
Chip-OMB translates optical modulation biosensing into a microfluidic format by combining magnetic bead-based capture and concentration, internal chip flushing, and optical background subtraction without user-performed washing after chip loading. The platform achieved high analytical sensitivity in buffer and undiluted plasma, showed strong linear correlation with the laboratory OMBi system, and demonstrated complete signal separation between the two study-defined cohorts in an exploratory 56-sample clinical feasibility evaluation. These findings support Chip-OMB as a promising foundation for future compact, high-sensitivity serological diagnostics and represent an important step toward translating modulation-based biosensing from laboratory platforms to integrated microfluidic systems for decentralized testing. Future development will focus on cartridge-level integration, miniaturized optical detection, automated sample and reagent handling, multiplexed assays, and broader clinical validation.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/bios16110623/s1. Figure S1: 3D schematic of the Chip-OMB optical system, Figure S2: Effect of on-chip flushing on fluorescence background in clinical serum samples, and Table S1: Panbio Dengue IgG ELISA results and Chip-OMB measurements for DENV-2-positive clinical serum samples. Table S2: Comparison of Chip-OMB with representative microfluidic fluorescence immunoassay platforms.
Author Contributions
Conceptualization, S.B. and A.D.; methodology, S.B., A.D., M.C., R.A., S.T., R.K., V.I., Y.L., and E.S.; investigation, S.B., M.C., R.A., S.T., R.K., V.I., T.M.H., T.T.K.T., M.T.H., N.M.H., L.T.H., N.V.T., and E.S.; formal analysis, S.B. and R.A.; validation, S.B., M.C., R.A., S.T., R.K., V.I., T.M.H., T.T.K.T., M.T.H., and N.M.H.; resources, R.K., V.I., E.S., T.M.H., T.T.K.T., M.T.H., N.M.H., L.T.H., and N.V.T.; supervision, A.D., Y.L., L.T.H., and N.V.T.; funding acquisition, A.D., L.T.H., and N.V.T.; writing—original draft preparation, S.B.; writing—review and editing, A.D. All authors have read and agreed to the published version of the manuscript.
Funding
This work was supported by the Israeli Ministry of Science and Technology [grant numbers 006016 and 103343] and by the Israel Innovation Authority, Israel National Authority for Technological Innovation [grant number 87618]. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Institutional Review Board Statement
The study was conducted in accordance with the principles of the Declaration of Helsinki and approved by the Institutional Ethical Review Board of Hanoi Medical University, Vietnam (IRB-VN01.001/IRB00003121/FWA 00004148; Approval No. 615/GCN-HĐĐĐNCYSH-ĐHYHN, approved 7 April 2022), and the Institutional Review Board of Sheba Medical Center, Israel (IRB 4420-17).
Informed Consent Statement
Written informed consent was obtained from all participants whose serum samples were included in this study.
Data Availability Statement
The data presented in this study are available on request from the corresponding author.
Acknowledgments
During the preparation of this manuscript, the authors used OpenAI ChatGPT (GPT-5.6 Sol) for language refinement and manuscript editing, and ChatGPT’s AI image-generation tool to prepare the Graphical Abstract. The authors reviewed and edited the generated text and image and take full responsibility for the content of this publication.
Conflicts of Interest
The authors declare no conflicts of interest.
Abbreviations
The following abbreviations are used in this manuscript:
| OMB | Optical modulation biosensing |
| Chip-OMB | Chip-based optical modulation biosensing |
| COB | Chopped optical biosensing |
| LoD | Limit of detection |
| CV | Coefficient of variation |
| SD | Standard deviation |
| SEM | Standard error of the mean |
| DENV-2 | Dengue virus serotype 2 |
| NS1 | Nonstructural protein 1 |
| IgG | Immunoglobulin G |
| RT-qPCR | Reverse transcription quantitative polymerase chain reaction |
| CMOS | Complementary metal-oxide-semiconductor |
| ROI | Region of interest |
| ELISA | Enzyme-linked immunosorbent assay |
References
- Yager, P.; Edwards, T.; Fu, E.; Helton, K.; Nelson, K.; Tam, M.R.; Weigl, B.H. Microfluidic diagnostic technologies for global public health. Nature 2006, 442, 412–418. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Pai, N.P.; Vadnais, C.; Denkinger, C.; Engel, N.; Pai, M. Point-of-care testing for infectious diseases: Diversity, complexity, and barriers in low- and middle-income countries. PLoS Med. 2012, 9, e1001306. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yager, P.; Domingo, G.J.; Gerdes, J. Point-of-care diagnostics for global health. Annu Rev. Biomed Eng. 2008, 10, 107–144. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kozel, T.R.; Burnham-Marusich, A.R. Point-of-Care Testing for Infectious Diseases: Past, Present, and Future. J. Clin. Microbiol. 2017, 55, 2313–2320. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Yetisen, A.K.; Akram, M.S.; Lowe, C.R. Paper-based microfluidic point-of-care diagnostic devices. Lab Chip 2013, 13, 2210–2251. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Behera, P.P.; Kumar, N.; Kumari, M.; Kumar, S.; Mondal, P.K.; Arun, R.K. Integrated microfluidic devices for point-of-care detection of bio-analytes and disease. Sens. Diagn. 2023, 2, 1437–1459. [Google Scholar] [CrossRef] [Scilit]
- Qin, L.; Vermesh, O.; Shi, Q.; Heath, J.R. Self-powered microfluidic chips for multiplexed protein assays from whole blood. Lab Chip 2009, 9, 2016–2020. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Chavez-Pineda, O.G.; Rodriguez-Moncayo, R.; Cedillo-Alcantar, D.F.; Guevara-Pantoja, P.E.; Amador-Hernandez, J.U.; Garcia-Cordero, J.L. Microfluidic systems for the analysis of blood-derived molecular biomarkers. Electrophoresis 2022, 43, 1667–1700. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Herr, A.E.; Hatch, A.V.; Throckmorton, D.J.; Tran, H.M.; Brennan, J.S.; Giannobile, W.V.; Singh, A.K. Microfluidic immunoassays as rapid saliva-based clinical diagnostics. Proc. Natl. Acad. Sci. USA 2007, 104, 5268–5273. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Wu, K.; He, X.; Wang, J.; Pan, T.; He, R.; Kong, F.; Cao, Z.; Ju, F.; Huang, Z.; Nie, L. Recent progress of microfluidic chips in immunoassay. Front Bioeng. Biotechnol. 2022, 10, 1112327. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Lehnert, T.; Gijs, M.A.M. Microfluidic systems for infectious disease diagnostics. Lab Chip 2024, 24, 1441–1493. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Gong, Y.X.; Huang, X.F.; Liang, J.; Xie, J.; Qing, L.S. Aptamer-based microfluidics for the detection of cancer biomarkers. Anal. Bioanal. Chem. 2025, 417, 3529–3550. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nam, U.; Kim, S.; Park, J.; Jeon, J.S. Lipopolysaccharide-Induced Vascular Inflammation Model on Microfluidic Chip. Micromachines 2020, 11, 747. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Zhong, H.; Li, Y.; Liu, G.; Xu, T.; Suo, Y.; Wang, Z. Study on covalent coupling process and flow characteristics of antibody on the surface of immunoassay microfluidic chip. Prep. Biochem Biotechnol. 2022, 52, 424–432. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nandimandalam, M.; Costantini, F.; Lovecchio, N.; Iannascoli, L.; Nascetti, A.; de Cesare, G.; Caputo, D.; Manetti, C. Split Aptamers Immobilized on Polymer Brushes Integrated in a Lab-on-Chip System Based on an Array of Amorphous Silicon Photosensors: A Novel Sensor Assay. Materials 2021, 14, 7210. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Choi, S.; Lee, W.I.; Lee, G.H.; Yoo, Y.E. Analysis of the Binding of Analyte-Receptor in a Micro-Fluidic Channel for a Biosensor based on Brownian Motion. Micromachines 2020, 11, 570. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kuo, J.S.; Chiu, D.T. Controlling mass transport in microfluidic devices. Annu Rev. Anal. Chem. 2011, 4, 275–296. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Torul, H.; Arslan, Z.C.; Tezcan, T.; Kayis, E.C.; Calimci, M.; Gumustas, A.; Yildirim, E.; Kulah, H.; Tamer, U. Microfluidic-based blood immunoassays. J. Pharm. BioMed Anal. 2023, 228, 115313. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Park, J.; Sunkara, V.; Kim, T.-H.; Hwang, H.; Cho, Y.-K. Lab-on-a-Disc for Fully Integrated Multiplex Immunoassays. Anal. Chem. 2012, 84, 2133–2140, Correction in Anal. Chem. 2012, 84, 4634. https://doi.org/10.1021/ac300979f. [Google Scholar] [CrossRef] [Scilit]
- Banerjee, A.; Shuai, Y.; Dixit, R.; Papautsky, I.; Klotzkin, D. Concentration dependence of fluorescence signal in a microfluidic fluorescence detector. J. Lumin. 2010, 130, 1095–1100. [Google Scholar] [CrossRef] [Scilit]
- Liu, X.; Zhu, H.; Sabo, J.; Lansky, Z.; Neuzil, P. Improvement of the signal to noise ratio for fluorescent imaging in microfluidic chips. Sci. Rep. 2022, 12, 18911. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ryu, G.; Huang, J.; Hofmann, O.; Walshe, C.A.; Sze, J.Y.; McClean, G.D.; Mosley, A.; Rattle, S.J.; deMello, J.C.; deMello, A.J.; et al. Highly sensitive fluorescence detection system for microfluidic lab-on-a-chip. Lab Chip 2011, 11, 1664–1670. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Burg, S.; Roth, S.; Cohen, M.; Avivi-Mintz, S.; Margulis, M.; Rohana, H.; Peretz, A.; Danielli, A. High throughput optical modulation biosensing for highly sensitive and rapid detection of biomarkers. Talanta 2022, 248, 123624. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Burg, S.; Margulis, M.; Danielli, A. From Concept to Commercialization: High-Throughput Optical Modulation Biosensing for Detecting Low Concentrations of Biomarkers. In Current Developments in Biosensor Applications and Smart Strategies; Karakuş, S., Ed.; IntechOpen: Rijeka, Croatia, 2025. [Google Scholar]
- Margulis, M.; Rohana, H.; Erster, O.; Mandelboim, M.; Biber, A.; Schwartz, E.; Peretz, A.; Danielli, A. Highly sensitive extraction-free saliva-based molecular assay for rapid diagnosis of SARS-CoV-2. J. Clin. Microbiol. 2024, 62, e0060024. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Levy, E.; Golani-Zaidie, L.; Burg, S.; Makdasi, E.; Alcalay, R.; Falach, R.; Schuster, O.; Danielli, A. Rapid and Highly Sensitive Detection of Ricin in Biological Fluids Using Optical Modulation Biosensing. Biosensors 2025, 15, 295. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Terenteva, S.; Golani-Zaidie, L.; Avivi, S.; Lustig, Y.; Indenbaum, V.; Koren, R.; Hoa, T.M.; Tuyen, T.T.K.; Huyen, M.T.; Hoan, N.M.; et al. Sensitivity and Cross-Reactivity Analysis of Serotype-Specific Anti-NS1 Serological Assays for Dengue Virus Using Optical Modulation Biosensing. Biosensors 2025, 15, 453. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Cai, G.; Yang, Z.; Chen, Y.C.; Huang, Y.; Liang, L.; Feng, S.; Zhao, J. Magnetic Bead Manipulation in Microfluidic Chips for Biological Application. Cyborg Bionic Syst. 2023, 4, 0023. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Kremer, R.; Roth, S.; Bross, A.; Danielli, A.; Noam, Y. Using Temporally and Spatially Resolved Measurements to Improve the Sensitivity of Fluorescence-Based Immunoassays. Biosensors 2024, 14, 220. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Dynabeads® M-280 Tosylactivated, Product Specifications, Thermo Fisher Life Technologies Corporation. Available online: https://documents.thermofisher.com/TFS-Assets/LSG/manuals/dynabeads_m280tosylactivated_man.pdf (accessed on 19 October 2026).
- Chin, C.D.; Linder, V.; Sia, S.K. Commercialization of microfluidic point-of-care diagnostic devices. Lab Chip 2012, 12, 2118–2134. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Whitesides, G.M. The origins and the future of microfluidics. Nature 2006, 442, 368–373. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Sackmann, E.K.; Fulton, A.L.; Beebe, D.J. The present and future role of microfluidics in biomedical research. Nature 2014, 507, 181–189. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Meng, F.; Alayash, A.I. Determination of extinction coefficients of human hemoglobin in various redox states. Anal. Biochem. 2017, 521, 11–19. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Ho, C.K.M.; Chen, C.; Setoh, J.W.S.; Yap, W.W.T.; Hawkins, R.C.W. Optimization of hemolysis, icterus and lipemia interference thresholds for 35 clinical chemistry assays. Pract. Lab Med. 2021, 25, e00232. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Hsu, W.; Shih, Y.T.; Lee, M.S.; Huang, H.Y.; Wu, W.N. Bead Number Effect in a Magnetic-Beads-Based Digital Microfluidic Immunoassay. Biosensors 2022, 12, 340. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Delshadi, S.; Fratzl, M.; Ramel, O.; Bigotte, P.; Kauffmann, P.; Kirk, D.; Masse, V.; Brenier-Pinchart, M.P.; Fricker-Hidalgo, H.; Pelloux, H.; et al. Magnetically localized and wash-free fluorescence immunoassay (MLFIA): Proof of concept and clinical applications. Lab Chip 2023, 23, 645–658. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Tyson, J.; Tsai, W.Y.; Tsai, J.J.; Brites, C.; Mässgård, L.; Ha Youn, H.; Pedroso, C.; Drexler, J.F.; Stramer, S.L.; Balmaseda, A.; et al. Combination of Nonstructural Protein 1-Based Enzyme-Linked Immunosorbent Assays Can Detect and Distinguish Various Dengue Virus and Zika Virus Infections. J. Clin. Microbiol. 2019, 57, e01464-18. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Nascimento, E.J.M.; George, J.K.; Velasco, M.; Bonaparte, M.I.; Zheng, L.; DiazGranados, C.A.; Marques, E.T.A.; Huleatt, J.W. Development of an anti-dengue NS1 IgG ELISA to evaluate exposure to dengue virus. J. Virol. Methods 2018, 257, 48–57. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Galula, J.U.; Salem, G.M.; Destura, R.V.; Remenyi, R.; Chao, D.-Y. Comparable Accuracies of Nonstructural Protein 1- and Envelope Protein-Based Enzyme-Linked Immunosorbent Assays in Detecting Anti-Dengue Immunoglobulin G Antibodies. Diagnostics 2021, 11, 741. [Google Scholar] [CrossRef] [Scilit] [PubMed]
- Burg, S.; Teboul, Y.Y.; Cohen, M.; Ashkenazy, R.; Golani-Zaidie, L.; Terenteva, S.; Lustig, Y.; Indenbaum, V.; Koren, R.; Hoa, T.M.; et al. Chopped optical biosensing enables temporal background subtraction for high-sensitivity fluorescence immunoassays. Talanta 2026, 305, 129666. [Google Scholar] [CrossRef] [Scilit] [PubMed]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.