A Cuprous Oxide Thin Film Non-Enzymatic Glucose Sensor Using Differential Pulse Voltammetry and Other Voltammetry Methods and a Comparison to Different Thin Film Electrodes on the Detection of Glucose in an Alkaline Solution

1
Department of Chemical and Biomolecular Engineering and Electronics Design Center, Case Western Reserve University, 10900 Euclid Ave, Cleveland, OH 44106, USA
2
Dipartimento di Chimica Materiali ed Ingegneria Chimica “Giulio Natta”, Politecnico di Milano, Via Mancinelli 7, 20131 Milano, Italy
3
Swagelok Center for Surface Analysis of Materials (SCSAM), Case Western Reserve University, 10900 Euclid Ave, Cleveland, OH 44106, USA
*
Author to whom correspondence should be addressed.
Received: 8 November 2017 / Revised: 2 January 2018 / Accepted: 4 January 2018 / Published: 6 January 2018
(This article belongs to the Special Issue Continuous Glucose Monitoring)
View Full-Text   |   Download PDF [4022 KB, uploaded 6 January 2018]   |  

Abstract

A cuprous oxide (Cu2O) thin layer served as the base for a non-enzymatic glucose sensor in an alkaline medium, 0.1 NaOH solution, with a linear range of 50–200 mg/dL using differential pulse voltammetry (DPV) measurement. An X-ray photoelectron spectroscopy (XPS) study confirmed the formation of the cuprous oxide layer on the thin gold film sensor prototype. Quantitative detection of glucose in both phosphate-buffered saline (PBS) and undiluted human serum was carried out. Neither ascorbic acid nor uric acid, even at a relatively high concentration level (100 mg/dL in serum), interfered with the glucose detection, demonstrating the excellent selectivity of this non-enzymatic cuprous oxide thin layer-based glucose sensor. Chronoamperometry and single potential amperometric voltammetry were used to verify the measurements obtained by DPV, and the positive results validated that the detection of glucose in a 0.1 M NaOH alkaline medium by DPV measurement was effective. Nickel, platinum, and copper are commonly used metals for non-enzymatic glucose detection. The performance of these metal-based sensors for glucose detection using DPV were also evaluated. The cuprous oxide (Cu2O) thin layer-based sensor showed the best sensitivity for glucose detection among the sensors evaluated. View Full-Text
Keywords: cuprous oxide; non-enzymatic glucose sensor; alkaline solution; differential pulse voltammetry cuprous oxide; non-enzymatic glucose sensor; alkaline solution; differential pulse voltammetry
Figures

Figure 1

This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. (CC BY 4.0).

Share & Cite This Article

MDPI and ACS Style

Dai, Y.; Molazemhosseini, A.; Abbasi, K.; Liu, C.C. A Cuprous Oxide Thin Film Non-Enzymatic Glucose Sensor Using Differential Pulse Voltammetry and Other Voltammetry Methods and a Comparison to Different Thin Film Electrodes on the Detection of Glucose in an Alkaline Solution. Biosensors 2018, 8, 4.

Show more citation formats Show less citations formats

Note that from the first issue of 2016, MDPI journals use article numbers instead of page numbers. See further details here.

Related Articles

Article Metrics

Article Access Statistics

1

Comments

[Return to top]
Biosensors EISSN 2079-6374 Published by MDPI AG, Basel, Switzerland RSS E-Mail Table of Contents Alert
Back to Top