Effects of Major Components of Synovial Fluid on the Morphology and Wear Rate of Polyetheretherketone (PEEK) Particles under an Accelerated Wear Process

1
Department of Chemical Engineering and Biotechnology, National Taipei University of Technology, 1, Sec. 3, Zhongxiao E. Rd., Taipei 10608, Taiwan
2
Institute of Biomedical Engineering and Nanomedicine, National Health Research Institutes, No. 35, Keyan Road, Zhunan Town, Miaoli County 35053, Taiwan
*
Author to whom correspondence should be addressed.
Received: 2 May 2018 / Revised: 5 June 2018 / Accepted: 6 June 2018 / Published: 8 June 2018
View Full-Text   |   Download PDF [1727 KB, uploaded 8 June 2018]   |  

Abstract

Wear particle-induced biological responses are the major factors for the failure of total joint arthroplasties, but it is possible to improve the lubrication and reduce the wear of an artificial joint system. Polyetheretherketone (PEEK), with ultra-high molecular weight polyethylene, is a suitable bearing material due to its resistance to fatigue strain. However, the effects of major compositions of synovial fluid on the wear of PEEK are unclear. We characterized the effects of three major components of synovial fluid including albumin, globulin, and phospholipids on the morphology and wear rate of PEEK wear particles. Our results demonstrated that the concentrations of albumin and globulin could affect the morphology of PEEK wear particles. In addition, a higher concentration of globulin and phospholipids (12.5 mg/mL) resulted in an increase in the amount of wear particles by 2.8- and 1.7-fold, respectively. In contrast, increasing albumin caused a reduction of wear particle numbers. These results indicate increasing concentration of albumin or reducing concentration of globulin or phospholipids has a better effect on reducing the numbers of wear particles and provides a potential solution of reducing PEEK wear particles, thus it can be more effectively applied in other biomedical systems. View Full-Text
Keywords: polymer; bio-tribology; joint prostheses polymer; bio-tribology; joint prostheses
Figures

Figure 1

This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. (CC BY 4.0).
SciFeed

Share & Cite This Article

MDPI and ACS Style

Su, C.-Y.; Huang, S.-S.; Fang, H.-W. Effects of Major Components of Synovial Fluid on the Morphology and Wear Rate of Polyetheretherketone (PEEK) Particles under an Accelerated Wear Process. Polymers 2018, 10, 635.

Show more citation formats Show less citations formats

Note that from the first issue of 2016, MDPI journals use article numbers instead of page numbers. See further details here.

Related Articles

Article Metrics

Article Access Statistics

1

Comments

[Return to top]
Polymers EISSN 2073-4360 Published by MDPI AG, Basel, Switzerland RSS E-Mail Table of Contents Alert
Back to Top