Diagnosis of Intoxication by the Organophosphate VX: Comparison Between an Electrochemical Sensor and Ellman´s Photometric Method

Miroslav Pohanka * email, Martina Hrabinova email and Kamil Kuca email
Centre of Advanced Studies and Department of Toxicology, Faculty of Military Health Sciences, University of Defense / Trebesska 1575, 50001 Hradec Kralove, Czech Republic
* Author to whom correspondence should be addressed.
Received: 22 July 2008; in revised form: 27 August 2008 / Accepted: 28 August 2008 / Published: 1 September 2008
PDF Full-textDownload PDF Full-Text [92 KB, uploaded 16 September 2008 11:02 CET]
Abstract: An electrochemical sensor is introduced as a tool applicable for diagnosis of intoxication by cholinesterase inhibitors caused by the well-known nerve agent VX. The traditional Ellman method was chosen for comparison with the sensor's analytical parameters. Both methods are based on estimation of blood cholinesterase inhibition as a marker of intoxication. While Ellman´s method provided a limit of detection of 5.2´10-7 M for blood containing VX, the electrochemical sensor was able to detect 4.0´10-7 M. Good correlation between both methods was observed (R = 0.92). The electrochemical sensor could be considered a convenient tool for a fast yet accurate method, easily available for field as well as laboratory use. Time and cost savings are key features of the sensor-based assay.
Keywords: Organophosphate; carbamate; intoxication; diagnosis; biosensor

Article Statistics

Click here to load and display the download statistics.

Cite This Article

MDPI and ACS Style

Pohanka, M.; Hrabinova, M.; Kuca, K. Diagnosis of Intoxication by the Organophosphate VX: Comparison Between an Electrochemical Sensor and Ellman´s Photometric Method. Sensors 2008, 8, 5229-5237.

AMA Style

Pohanka M., Hrabinova M., Kuca K. Diagnosis of Intoxication by the Organophosphate VX: Comparison Between an Electrochemical Sensor and Ellman´s Photometric Method. Sensors. 2008; 8(9):5229-5237.

Chicago/Turabian Style

Pohanka, Miroslav; Hrabinova, Martina; Kuca, Kamil. 2008. "Diagnosis of Intoxication by the Organophosphate VX: Comparison Between an Electrochemical Sensor and Ellman´s Photometric Method." Sensors 8, no. 9: 5229-5237.

Sensors EISSN 1424-8220 Published by MDPI Publishing, Basel, Switzerland RSS E-Mail Table of Contents Alert