Dielectrically-Loaded Cylindrical Resonator-Based Wireless Passive High-Temperature Sensor

1
Key Laboratory of Instrumentation Science & Dynamic Measurement, Ministry of Education, North University of China, Taiyuan 030051, China
2
Science and Technology on Electronic Test and Measurement Laboratory, North University of China, Taiyuan 030051, China
3
Department of Mechanical & Electrical Engineering, Xiamen University, Xiamen 361005, China
*
Authors to whom correspondence should be addressed.
These authors contributed equally to this work.
Sensors 2016, 16(12), 2037; https://doi.org/10.3390/s16122037
Submission received: 14 August 2016 / Revised: 18 November 2016 / Accepted: 24 November 2016 / Published: 1 December 2016
(This article belongs to the Special Issue Resonator Sensors)

Abstract

:
The temperature sensor presented in this paper is based on a microwave dielectric resonator, which uses alumina ceramic as a substrate to survive in harsh environments. The resonant frequency of the resonator is determined by the relative permittivity of the alumina ceramic, which monotonically changes with temperature. A rectangular aperture etched on the surface of the resonator works as both an incentive and a coupling device. A broadband slot antenna fed by a coplanar waveguide is utilized as an interrogation antenna to wirelessly detect the sensor signal using a radio-frequency backscattering technique. Theoretical analysis, software simulation, and experiments verified the feasibility of this temperature-sensing system. The sensor was tested in a metal-enclosed environment, which severely interferes with the extraction of the sensor signal. Therefore, frequency-domain compensation was introduced to filter the background noise and improve the signal-to-noise ratio of the sensor signal. The extracted peak frequency was found to monotonically shift from 2.441 to 2.291 GHz when the temperature was varied from 27 to 800 °C, leading to an average absolute sensitivity of 0.19 MHz/°C.

1. Introduction

Instrumentation is a key generic technology in the gas turbine industry that influences the development cost, efficiency, and competitiveness of gas turbine products. The very hostile gas turbine environment presents unique challenges for instrumentation. The demand for greater efficiency is steadily increasing the temperature and pressure in engines [1,2,3]. The gas turbine output power is increased by 10%, and the efficiency is increased by 1% for every 55.5 °C increase in temperature. The European Virtual Institute for Gas Turbine Instrumentation identified three areas where the lack of adequate instrumentation capability is perceived to be either holding back gas turbine engine development or leading to increased uncertainty in design methods and component life prediction [4]: measurement of gas temperature, pressure, flow, and blade tip clearances at very high temperatures (>1000 °C); measurement of component temperatures in the hottest parts of the engine; and measurement of component vibration on very hot components.
The temperatures inside gas turbines are usually indirectly obtained using temperature sensors, such as thermocouples. By measuring the voltage of the thermocouple and amplifying the voltage signal through a series of amplifier devices, the temperature can be obtained. Nevertheless, thermocouples suffer from the disadvantage of inaccuracy and cannot perform in situ temperature parameter measurements.
Sensors that work inside an engine must survive high temperatures and corrosive substances while being fully enclosed in metal. Active devices are normally unavailable in high-temperature environments because thermo-mechanical fatigue of the package material will cause the device to fail. Traditional wire-testing approaches suffer from ohmic contact failure and unstable wire connections at high temperature [5,6]. Wireless passive sensors are promising in harsh environments because they implement battery-free and contactless measurements. Currently, a wireless passive temperature sensor based on inductive-capacitive (LC) resonance could work at 800 °C using high-temperature co-fired ceramic techniques, but the LC lumped circuit suffers from the disadvantage of having a low quality factor and cannot approach the metal material characteristics, which affects the sensing distance and applications [7,8]. Wireless surface acoustic wave temperature sensors could work up to 900 °C, but chemical instabilities in the piezoelectric substrate at high temperatures limit their use in harsh-environment applications [9,10,11].
A promising approach to realize wireless temperature measurement in a harsh environment is the microwave backscattering technique, which has been utilized to achieve multi-parameter measurements, such as temperature and strain. This temperature sensor type, reported in [12], based on a dielectric resonator antenna, could work at 370 °C with a sensitivity of 307 kHz/K, although it was conducted in an open environment. A wireless passive temperature sensor that uses an integrated cylindrical resonator/antenna could work up to 1300 °C [13]; however, it has a relatively low quality factor at high temperature, which affects the reliability and accuracy of the sensor data.
The wireless passive temperature sensor presented in the present study consists of a dielectrically-loaded resonator and an aperture, which was measured at high temperature and in a metal-enclosed environment, and was demonstrated to be reliable for harsh environment applications owing to its high quality-factor stability in temperatures of up to 800 °C. The wireless signal transmission mechanism is based on radio-frequency backscattering, as shown in Figure 1. The aperture on the resonator works as a response antenna to receive the linearly-polarized wave transmitted from the interrogation antenna and coupled into the resonator. When the frequency component of the transmitted wave matches the resonant frequency of the resonator, the aperture will reflect the other frequency component back to the interrogation antenna because the matched frequency component will resonate in the resonator, which decays with time. By extracting the return loss (S11) of the interrogation antenna, the resonant frequency of the resonator can be obtained as a peak in the spectrum signal. This temperature-sensor type has beneficial characteristics, such as small volume, simple structure, and high quality factor, which can reduce measurement uncertainty for design validation.

2. Design and Fabrication of the Temperature Sensor

A dielectrically-loaded cylindrical resonator working in the TM010 mode is employed in this study owing to its stability and simplicity, as well as its concentrated area of electric fields. The resonant frequency of the resonator is determined by the relative permittivity of the alumina ceramic when the dimensions of the resonator are fixed, and it monotonically changes with increasing temperature [14]. The resonant frequency ƒr of the resonator is expressed as [15]:
f r = c ε r λ 0
where c is the speed of light and εr represents the relative permittivity of the substrate. Through electromagnetic theory analysis, when the cylindrical resonator is working in the TM010 mode, the resonant wavelength λ0 could be expressed as:
λ 0 = 2.62 R
where R is the radius of the resonator, set to 14.5 mm. The height of resonator h is set to 5 mm to ensure that the TM010 mode is the dominant mode of the cylindrical resonator and to reduce the sensor size. The relative permittivity of the alumina ceramic is 9.7 at 25 °C. The resonant frequency of the resonator is calculated as 2.54 GHz according to Equations (1) and (2) and is simulated as 2.576 GHz using the ANSYS high-frequency structure simulator (Southpointe 2600 ANSYS Drive, HFSS, ANSYS, Inc., Canonsburg, PA, USA) eigenmode.
The sensor consists of a resonator and an aperture. The coupling between them is essential for the signal sensing system to reduce the reflection and transmission losses. To better understand the coupling process and realize matching between the resonator and aperture, one coaxial line is weakly coupled to the resonator, as shown in Figure 2a. The coaxial probe inserted into the resonator excites the resonator, and the field distribution within a coaxial line is shown in Figure 2b. The resonator is excited through magnetic coupling because the coaxial line and the resonator TM010 mode have the same magnetic field direction.
The simulated vector current density distribution on the surface of the resonator without the aperture in HFSS is shown in Figure 3a, and the simulated resonant frequency is 2.571 GHz, which implies that the probe has little influence on the resonant frequency of the resonator. We can intuitively see that when the aperture cuts the current line, as shown in Figure 3b, the aperture will be excited, and the energy will radiate outward through the aperture.
The aperture works as a response antenna, whereas the resonator works as its excitation source. The equivalent matching circuit model can be roughly described, as shown in Figure 4. Reflection coefficients τ1 and τ2 represent the reflection between the coaxial line and resonator and between the resonator and aperture, respectively. When the position of the coaxial line is fixed, τ1 will be constant. The change in the relative position between the resonator and aperture will lead to a change in reflection coefficient τ2. By checking the return loss of the coaxial line, impedance matching between the resonator and aperture could be obtained.
Three primary parameters are present, namely, aperture length L, width W, and distance d between the aperture and resonator center, that determine the impedance matching between the aperture and resonator. Figure 5 shows the parametric study. The resonant frequency decreases with increased L, W, and d. This can be explained by the fact that the relative position of the aperture leads to a change in inductance La and capacitance Ca of the equivalent circuit and, therefore, changes the resonant frequency. The change in the relative position between the resonator and aperture also affects the coupling coefficient K12, which could also affect the impedance matching between the resonator and aperture. Finally, L = 16 mm, d = 8 mm, and W = 2 mm are chosen as the optimized values.
To form the resonator and aperture, silver paste was coated on the alumina ceramic by screen printing with a film thickness of 25 μm. The thickness and surface roughness of the metal film greatly influence the sensor signal strength [16]. Therefore, the screen-printing process should be smooth to reduce unnecessary loss. After the silver paste was coated on the ceramic, it was leveled for 10 min at room temperature, dried at 125 °C for 15 min, and sintered at 850 °C for 10 min to form a dense metal film. The fabrication process is shown in Figure 6. The broadband coplanar waveguide (CPW)-fed square slot antenna with a widened tuning stub, used as the interrogation antenna, will not be described here in detail because it has been previously discussed [17].

3. Temperature Measurements and Results

To characterize the sensor high-temperature properties, a wireless measurement was set up, as shown in Figure 7. Software was utilized to set the temperature curve and send it to the temperature controller of a Nabertherm LHT 02/16 furnace (Nabertherm GmbH, Lilienthal, Germany), which was used as the heat source in these experiments. The interrogation antenna was connected to a vector network analyzer (Agilent E5061B, Agilent Technologies, Santa Clara, CA, USA) through a coaxial cable to excite the sensor and detect the backscattered spectrum signal of the sensor. The sensor was installed in the heating furnace above the interrogation antenna. We should note that the relative position between the interrogation antenna and aperture has a significant effect on the readout signal because of the directivity and polarization direction of the interrogation antenna and the aperture. The experimental results show that when the aperture was aligned with the tuning stub, maximum sensor signal strength was obtained. An insulation layer was applied to keep the temperature stable inside the furnace.
The metallic environment caused severe interference during the sensor signal extraction. In addition to the expected sensor signal backscatter, the interrogation antenna received unexpected reflections from the environment. The sensor characteristic signal was quite dim and was a small target compared with the background clutter of the signals, which rendered reading erroneous or even impossible. Figure 8 shows the sensor response in the furnace at 27 °C. The sensor characteristic signal is represented as a negative frequency as observed. Nevertheless, it still contains some ripples due to the influence of the metallic environment, which deteriorates the sensor signal at high temperature.
Suppression of the background clutter was necessary to improve the signal-to-noise ratio (SNR) of the sensor signal. Frequency-domain compensation is an available option to reduce the interference of background clutter. Figure 9 shows the process of the frequency-domain compensation. The upper curve with a black line represents the power of the background clutter in the furnace (without the sensor) before compensation. The lower curve with a red line represents the background signal in the furnace (without the sensor) after compensation and is generated using the real-time background clutter data vector substrate in the upper curve, which are stored in the memory of the network analyzer beforehand. The power of the background clutter was significantly reduced from an average of −10 dB to an average of −65 dB, indicating that the frequency-domain compensation could effectively improve the SNR of the sensor signal. Figure 10 shows the signal segment after compensation with the sensor in the furnace. The curve is relatively smooth compared with that shown in Figure 8.
The temperature sensor was investigated using the setup shown in Figure 7, and the result is shown in Figure 11. The sensor resonant frequency f0 changed from 2.441 to 2.291 GHz when the temperature varied from 27 to 800 °C, leading to average absolute sensitivity sf = ∆f0/∆t = 194 kHz/°C and average temperature coefficient τf = ∆f0/(f0*∆t) = 79.5 ppm/°C, as shown in Figure 12a. The dimensional change in the sensor was approximately 0.78% from 27 to 800 °C considering that the 99% alumina ceramic has a maximum coefficient of thermal expansion of 9.8 × 10−6 /°C at 800 °C. According to Equations (1) and (2), the sensor resonant-frequency change due to the thermal expansion was approximately 0.77%, which was much less than the 6.15% resonant-frequency change caused by the temperature. The thermal expansion of the sensor will decrease the dimensions of W, L, and d, which increase the resonant frequency of the sensor according to the HFSS simulation results shown in Figure 5; however, the dimension changes in W, L, and d are rather small and could be neglected. The measured resonant frequency at 27 °C slightly deviates from the simulated resonant frequency, which may be caused by inaccuracies in the fabrication process and the relative permittivity of the alumina ceramic. The negative peak curve gradually flattens as the temperature ranges from 27 to 300 °C, becoming stable in the temperature range from 300–800 °C, meaning that the sensor has a stably-loaded quality factor (Q0 = f0/∆f3dB), and is maintained at approximately 50. Repetitive experiments were conducted to verify the feasibility of this type of sensor. To ensure the same test environment, we keep the position unchanged and set the time interval between each peak to 10 min and the holding time to 20 min. By extracting the frequency peak shown in Figure 12a, a quasi-linear relationship between the temperature and frequency can be obtained. This relationship, shown in Figure 12b for the three subsequent experiments, suggests that the sensor possesses good repeatability. The slight frequency deviation among the three experiments is believed to be mainly due to the inaccuracy of the furnace temperature control and measurement system.

4. Conclusions

The temperature sensor presented in this paper uses alumina ceramic and silver paste for wireless sensing, and offers a great potential for harsh environment applications. The theoretical analysis, simulation, and experiments confirm the feasibility of the sensing system. The high-temperature characteristics of the sensor were investigated and discussed. The SNR of the sensor signal has improved due to the frequency-domain compensation. The extracted peak frequency was found to monotonically shift from 2.441 to 2.291 GHz when the temperature changed from 27 to 800 °C, leading to an average absolute sensitivity of 194 kHz/°C and an average temperature coefficient of 79.5 ppm/°C. The sensor was measured at 800 °C to avoid damage to the sensor because silver melts at 950 °C. However, according to the trend of the measured curve, achieving a higher temperature range is possible. Our future work will focus on optimizing the uniform fabrication of this type of sensor and improving the gain and directivity of the interrogation antenna to increase the sensing distance.

Acknowledgments

This work was supported by the National Natural Science Foundation of China (No. 61471324 and No. 51425505), and the Program for the Top Young Academic Leaders of Higher Learning Institutions of Shanxi Province, China, and the Outstanding Youth talents Program of Shanxi Province.

Author Contributions

All works with relation to this paper have been accomplished by all authors’ efforts. The idea and design of the sensor were proposed by Qiulin Tan and Guozhu Wu. The experiments of the sensor were completed with the help from Tanyong Wei. Sanmin Shen and Helei Dong designed the fabrication method of the sensor. At last, every segment relate to this paper is accomplished under the guidance from Jijun Xiong. Dezhi Wu and Wendong Zhang have put forward valuable suggestions for the revision of the manuscript.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Culley, D.; Garg, S.; Hiller, S.J.; Horn, W.; Kumar, A.; Mathews, H.K.; Moustapha, H.; Pfoertner, H.; Rosenfeld, T.; Rybarik, P.; et al. More Intelligent Gas. Turbine Engines; North Atlantic Treaty Organisation: Brussels, Belgium, 2009. [Google Scholar]
  2. Boyce, M.P. Gas Turbine Engineering Handbook; Elsevier: Waltham, MA, USA, 2011. [Google Scholar]
  3. Habib, Z.; Parthasarathy, R.; Gollahalli, S. Performance and emission characteristics of biofuel in a small-scale gas turbine engine. Appl. Energy 2010, 87, 1701–1709. [Google Scholar] [CrossRef]
  4. Accurate High Temperature Engine Aero-Thermal Measurements for Gas-Turbine Life Optimization, Performance and Condition Monitoring. Available online: http://www.transport-research.info/sites/default/files/project/documents/20121017_144252_70354_122807161EN6.pdf (accessed on 28 November 2016).
  5. Kroetz, G.H.; Eickhoff, M.H.; Moeller, H. Silicon compatible materials for harsh environment sensors. Sens. Actuators A Phys. 1999, 74, 182–189. [Google Scholar] [CrossRef]
  6. Lai, P.T.; Li, B.; Chan, C.L.; Sin, J.K. Spreading-resistance temperature sensor on silicon-on-insulator. IEEE Electron Device Lett. 1999, 20, 589–591. [Google Scholar] [CrossRef] [Green Version]
  7. Tan, Q.; Kang, H.; Qin, L.; Xiong, J.; Zhou, Z.; Zhang, W.; Luo, T.; Xue, C.; Liu, J. Measurement of relative permittivity of LTCC ceramic at different temperatures. AIP Adv. 2014, 4. [Google Scholar] [CrossRef]
  8. Tan, Q.; Luo, T.; Xiong, J.; Kang, H.; Ji, X.; Zhang, Y.; Yang, M.; Wang, X.; Xue, C.; Liu, J.; et al. A harsh environment-oriented wireless passive temperature sensor realized by LTCC technology. Sensors 2014, 14, 4154–4166. [Google Scholar] [CrossRef] [PubMed]
  9. Canabal, A.; Davulis, P.M.; Pollard, T.; da Cunha, M.P. Multi-sensor wireless interrogation of SAW resonators at high temperatures. In Proceedings of the 2010 IEEE International Ultrasonics Symposium, San Diego, CA, USA, 11–14 October 2010.
  10. Kang, A.; Zhang, C.; Ji, X.; Han, T.; Li, R.; Li, X. SAW-RFID enabled temperature sensor. Sens. Actuators A Phys. 2013, 201, 105–113. [Google Scholar] [CrossRef]
  11. Canabal, A.; Davulis, P.M.; Harris, G.M.; Da Cunha, M.P. High-temperature battery-free wireless microwave acoustic resonator sensor system. Electron. Lett. 2010, 46, 471–472. [Google Scholar] [CrossRef]
  12. Kubina, B.; Schüßler, M.; Mandel, C.; Mehmood, A.; Jakoby, R. Wireless high-temperature sensing with a chipless tag based on a dielectric resonator antenna. In Proceedings of the 2013 IEEE SENSORS, Baltimore, MD, USA, 3–6 November 2013.
  13. Cheng, H.; Ren, X.; Ebadi, S.; Chen, Y.; An, L.; Gong, X. Wireless passive temperature sensors using integrated cylindrical resonator/antenna for harsh-environment applications. IEEE Sens. J. 2015, 15, 1453–1462. [Google Scholar] [CrossRef]
  14. Fargeot, S.; Guihard, D.; Lahitte, P. Dielectric characterization at high temperature (1600 °C) for space applications. In Proceedings of the 2011 IEEE International Conference on Microwave Technology & Computational Electromagnetics (ICMTCE), Beijing, China, 22–25 May 2011.
  15. Ren, X.; Ebadi, S.; Chen, Y.; An, L.; Gong, X. Characterization of SiCN ceramic material dielectric properties at high temperatures for harsh environment sensing applications. IEEE Trans. Microw. Theory Tech. 2013, 61, 960–971. [Google Scholar] [CrossRef]
  16. Li, Y.; Yu, Y.; San, H.; Wang, Y.; An, L. Wireless passive polymer-derived SiCN ceramic sensor with integrated resonator/antenna. Appl. Phys. Lett. 2013, 103, 163505. [Google Scholar] [CrossRef]
  17. Tan, Q.; Wei, T.; Chen, X.; Luo, T.; Wu, G.; Li, C.; Xiong, J. Antenna-resonator integrated wireless passive temperature sensor based on low-temperature co-fired ceramic for harsh environment. Sens. Actuators A Phys. 2015, 236, 299–308. [Google Scholar] [CrossRef]
Figure 1. Schematic of the temperature sensor signal transmission mechanism.
Figure 1. Schematic of the temperature sensor signal transmission mechanism.
Sensors 16 02037 g001
Figure 2. (a) Top and 3-D view of the integrated aperture/resonator with coaxial line; (b) Field distribution and configuration of the coaxial line.
Figure 2. (a) Top and 3-D view of the integrated aperture/resonator with coaxial line; (b) Field distribution and configuration of the coaxial line.
Sensors 16 02037 g002
Figure 3. Coaxial line-feed dielectrically-loaded cylindrical resonator in the TM010 mode (a) without the aperture and (b) with the aperture.
Figure 3. Coaxial line-feed dielectrically-loaded cylindrical resonator in the TM010 mode (a) without the aperture and (b) with the aperture.
Sensors 16 02037 g003
Figure 4. Equivalent circuit of the matching model.
Figure 4. Equivalent circuit of the matching model.
Sensors 16 02037 g004
Figure 5. HFSS simulation results of (a) position d; (b) aperture length L; and (c) aperture width W.
Figure 5. HFSS simulation results of (a) position d; (b) aperture length L; and (c) aperture width W.
Sensors 16 02037 g005
Figure 6. Sensor fabrication process. (a) Alumina ceramic; (b) Fabricated temperature sensor.
Figure 6. Sensor fabrication process. (a) Alumina ceramic; (b) Fabricated temperature sensor.
Sensors 16 02037 g006
Figure 7. Schematic of the high-temperature measurement setup using a furnace.
Figure 7. Schematic of the high-temperature measurement setup using a furnace.
Sensors 16 02037 g007
Figure 8. Detected original sensor response curve.
Figure 8. Detected original sensor response curve.
Sensors 16 02037 g008
Figure 9. Power of the background clutter before and after frequency-domain compensation in the metal-sealed chamber.
Figure 9. Power of the background clutter before and after frequency-domain compensation in the metal-sealed chamber.
Sensors 16 02037 g009
Figure 10. Received signal after compensation in the frequency domain.
Figure 10. Received signal after compensation in the frequency domain.
Sensors 16 02037 g010
Figure 11. Temperature measurement setup.
Figure 11. Temperature measurement setup.
Sensors 16 02037 g011
Figure 12. (a) Measured return loss curve versus frequency; (b) Extracted frequency versus temperature.
Figure 12. (a) Measured return loss curve versus frequency; (b) Extracted frequency versus temperature.
Sensors 16 02037 g012

Share and Cite

MDPI and ACS Style

Xiong, J.; Wu, G.; Tan, Q.; Wei, T.; Wu, D.; Shen, S.; Dong, H.; Zhang, W. Dielectrically-Loaded Cylindrical Resonator-Based Wireless Passive High-Temperature Sensor. Sensors 2016, 16, 2037. https://doi.org/10.3390/s16122037

AMA Style

Xiong J, Wu G, Tan Q, Wei T, Wu D, Shen S, Dong H, Zhang W. Dielectrically-Loaded Cylindrical Resonator-Based Wireless Passive High-Temperature Sensor. Sensors. 2016; 16(12):2037. https://doi.org/10.3390/s16122037

Chicago/Turabian Style

Xiong, Jijun, Guozhu Wu, Qiulin Tan, Tanyong Wei, Dezhi Wu, Sanmin Shen, Helei Dong, and Wendong Zhang. 2016. "Dielectrically-Loaded Cylindrical Resonator-Based Wireless Passive High-Temperature Sensor" Sensors 16, no. 12: 2037. https://doi.org/10.3390/s16122037

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop