<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD Journal Publishing DTD v2.3 20070202//EN" "journalpublishing.dtd">
<article xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" xml:lang="en" article-type="research-article">
<front>
<journal-meta>
<journal-id journal-id-type="publisher-id">Sensors</journal-id>
<journal-title>Sensors</journal-title>
<issn pub-type="epub">1424-8220</issn>
<publisher>
<publisher-name>Molecular Diversity Preservation International (MDPI)</publisher-name></publisher></journal-meta>
<article-meta>
<article-id pub-id-type="doi">10.3390/s120810980</article-id>
<article-id pub-id-type="publisher-id">sensors-12-10980</article-id>
<article-categories>
<subj-group>
<subject>Article</subject></subj-group></article-categories>
<title-group>
<article-title>Use of a Combined SpO<sub>2</sub>/PtcCO<sub>2</sub> Sensor in the Delivery Room</article-title></title-group>
<contrib-group>
<contrib contrib-type="author">
<name><surname>Rubortone</surname><given-names>Serena Antonia</given-names></name></contrib>
<contrib contrib-type="author">
<name><surname>De Carolis</surname><given-names>Maria Pia</given-names></name><xref ref-type="corresp" rid="c1-sensors-12-10980"><sup>*</sup></xref></contrib>
<contrib contrib-type="author">
<name><surname>Lacerenza</surname><given-names>Serafina</given-names></name></contrib>
<contrib contrib-type="author">
<name><surname>Bersani</surname><given-names>Iliana</given-names></name></contrib>
<contrib contrib-type="author">
<name><surname>Occhipinti</surname><given-names>Federica</given-names></name></contrib>
<contrib contrib-type="author">
<name><surname>Romagnoli</surname><given-names>Costantino</given-names></name></contrib>
<aff id="af1-sensors-12-10980">Division of Neonatology, Department of Paediatrics, Catholic University of Sacred Heart, Largo Gemelli 8, I-00168 Rome, Italy; E-Mails: <email>sererubor@hotmail.it</email> (S.A.R.); <email>saralac@inwind.it</email> (S.L.); <email>ilianabersani@gmail.com</email> (I.B.); <email>federicaocchipinti@hotmail.it</email> (F.O.); <email>cromagnoli@rm.unicatt.it</email> (C.R.)</aff></contrib-group>
<author-notes>
<corresp id="c1-sensors-12-10980">
<label>*</label>Author to whom correspondence should be addressed; E-Mail: <email>mpia.decarolis@rm.unicatt.it</email>; Tel./Fax: +39-06-3015-4357.</corresp></author-notes>
<pub-date pub-type="collection">
<year>2012</year></pub-date>
<pub-date pub-type="epub">
<day>08</day>
<month>08</month>
<year>2012</year></pub-date>
<volume>12</volume>
<issue>8</issue>
<fpage>10980</fpage>
<lpage>10989</lpage>
<history>
<date date-type="received">
<day>12</day>
<month>06</month>
<year>2012</year></date>
<date date-type="rev-recd">
<day>16</day>
<month>07</month>
<year>2012</year></date>
<date date-type="accepted">
<day>31</day>
<month>07</month>
<year>2012</year></date></history>
<permissions>
<copyright-statement>© 2012 by the authors; licensee MDPI, Basel, Switzerland.</copyright-statement>
<copyright-year>2012</copyright-year>
<license>
<p>This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/).</p></license></permissions>
<abstract>
<p>Arterial oxygen saturation (SaO<sub>2</sub>) and partial arterial pressure of carbon dioxide (PaCO<sub>2</sub>) are important respiratory parameters in critically ill neonates. A sensor combining a pulse oximeter with the Stow-Severinghaus electrode, required for the measurement of peripheral oxygen saturation (SpO<sub>2</sub>) and transcutaneous partial pressure of carbon dioxide (PtcCO<sub>2</sub>), respectively, has been recently used in neonatal clinical practice (TOSCA<sup>500Ò</sup>Radiometer). We evaluated TOSCA usability and reliability in the delivery room (DR), throughout three different periods, on term, late-preterm, and preterm neonates. During the first period (<bold><italic>period A</italic></bold>), 30 healthy term neonates were simultaneously monitored with both TOSCA and a MASIMO pulse oximeter. During the second period (<bold><italic>period B</italic></bold>), 10 healthy late-preterm neonates were monitored with both TOSCA and a transcutaneous device measuring PtcCO<sub>2</sub> (TINA<sup>Ò</sup> TCM3, Radiometer). During the third period (<bold><italic>period C</italic></bold>), 15 preterm neonates were monitored with TOSCA and MASIMO after birth, during stabilization, and during transport to the neonatal intensive care unit (NICU). Blood gas analyses were performed to compare transcutaneous and blood gas values. TOSCA resulted easily and safely usable in the DR, allowing reliable noninvasive SaO<sub>2</sub> estimation. Since PtcCO<sub>2</sub> measurements with TOSCA required at least 10 min to be stable and reliable, this parameter was not useful during the early resuscitation immediately after birth. Moreover, PtcCO<sub>2</sub> levels were less precise if compared to the conventional transcutaneous monitoring. However, PtcCO<sub>2</sub> measurement by TOSCA was useful as trend-monitoring after stabilization and during transport to NICU.</p></abstract>
<kwd-group>
<kwd>oxygen saturation</kwd>
<kwd>partial pressure of carbon dioxide</kwd>
<kwd>pulse oximeter</kwd>
<kwd>delivery room</kwd>
<kwd>TOSCA sensor</kwd>
<kwd>neonate</kwd></kwd-group></article-meta></front>
<body>
<sec sec-type="intro">
<label>1.</label>
<title>Introduction</title>
<p>Arterial oxygen saturation (SaO<sub>2</sub>) and partial arterial pressure of carbon dioxide (PaCO<sub>2</sub>) are two of the most important respiratory parameters in the evaluation of critically ill neonates. SaO<sub>2</sub> is commonly estimated by pulse oximetry (SpO<sub>2</sub>), a non invasive technique widely used in neonatal intensive care units. Partial pressure of carbon dioxide (PCO<sub>2</sub>) can be assessed by sampling blood (PCO<sub>2BGA</sub>) or by noninvasive techniques such as transcutaneous monitoring (PtcCO<sub>2</sub>), which reduces the need for frequent blood drawing and painful punctures and provides a continuous estimation of PCO<sub>2</sub>.</p>
<p>As already described elsewhere [<xref ref-type="bibr" rid="b1-sensors-12-10980">1</xref>], the TOSCA monitor combines a pulse oximeter of new technology (Masimo SET) with the Stow-Severinghaus electrode for PtcCO<sub>2</sub> measurement in a single sensor, which is also equipped with a heating element to increase the local perfusion <xref ref-type="fig" rid="f1-sensors-12-10980">Figure 1(a)</xref>. The sensor is attached to the ear by a clip having an adhesive holder and a reflective element placed onto the inner surface.</p>
<p>Like other similar devices, the TOSCA pulse oximeter determines SaO<sub>2</sub> by measuring the difference in absorption of selected light wavelengths by oxyhemoglobin and deoxyhemoglobin. It utilizes two light-emitting diodes (LEDs): one LED emits red light (LED-R; 658 nm wavelength) selectively adsorbed by deoxyhemoglobin, while the other one emits infrared light (LED-IR; 880 nm wavelength) selectively absorbed by oxyhemoglobin. The sensor also contains a photodiode (photodetector), an electronic device which converts light into an electronic signal proportional to the incident light intensity and which sends it to the TOSCA monitor for calculation. Unlike other pulse oximeters, either the LEDs or the photodiode are positioned on the same surface, since the light emitted is reflected by the clip surface opposite to the sensor place. The TOSCA pulse oximeter is equipped with Masimo SET technology and uses a set of algorithms designed to improve SpO<sub>2</sub> monitoring during motion and low perfusion.</p>
<p>Transcutaneous measurement of CO<sub>2</sub> is based on the observation that this gas has a high tissue solubility and diffusion through the skin, and that the application of local heat dilates blood vessels and enhances skin permeability. The PCO<sub>2</sub> component of the TOSCA consists of a Stow-Severinghaus type electrode, a potentiometric sensor combining a silver/silver chloride reference electrode, and a glass pH electrode. Over the sensor surface there is an electrolyte solution, provided within a thin hydrophilic spacer and coupled to the skin via a highly gas permeable hydrophobic membrane. PCO<sub>2</sub> determines a pH change of this electrolyte solution by diffusing through the skin to the sensor. The equation below illustrates the reaction occurring between carbon dioxide and water:
<disp-formula id="FD1">
<label>(1)</label>
<mml:math id="mm1" display="block">
<mml:semantics id="sm1">
<mml:mrow>
<mml:msub>
<mml:mtext>CO</mml:mtext>
<mml:mn>2</mml:mn></mml:msub>
<mml:mo>+</mml:mo>
<mml:msub>
<mml:mtext>H</mml:mtext>
<mml:mn>2</mml:mn></mml:msub>
<mml:mtext>O</mml:mtext>
<mml:mo>↔</mml:mo>
<mml:msub>
<mml:mtext>H</mml:mtext>
<mml:mn>2</mml:mn></mml:msub>
<mml:msub>
<mml:mtext>CO</mml:mtext>
<mml:mn>3</mml:mn></mml:msub>
<mml:mo>↔</mml:mo>
<mml:msup>
<mml:mtext>H</mml:mtext>
<mml:mo>+</mml:mo></mml:msup>
<mml:mo>+</mml:mo>
<mml:msup>
<mml:mtext>HCO</mml:mtext>
<mml:mrow>
<mml:mn>3</mml:mn>
<mml:mo>−</mml:mo></mml:mrow></mml:msup></mml:mrow></mml:semantics></mml:math></disp-formula></p>
<p>Generation of hydrogen ions at the glass electrode is directly proportional to the amount of carbon dioxide which is present. An increase in the hydrogen ion concentration leads to a fall in pH and this change is proportional to the logarithm of the PCO<sub>2</sub> change.</p>
<p>The heating element, increasing local perfusion, induces the arterialization of the capillary blood flow. However, due to the elevated temperature of the sensor, the transcutaneous PCO<sub>2</sub> results higher than the arterial value mostly due to two factors [<xref ref-type="bibr" rid="b2-sensors-12-10980">2</xref>]: first, the elevated temperature increases local blood and tissue PCO<sub>2</sub> (“<italic>anaerobic factor</italic>”); second, the epidermal cells produce CO<sub>2</sub>, which contributes to the capillary CO<sub>2</sub> concentration by a constant amount (“<italic>metabolic constant</italic>”). As a consequence, in order to provide a monitor readout corresponding as close as possible to PaCO<sub>2</sub>, the device applies an algorithm correction of the transcutaneous values as already specifically described by Severinghaus [<xref ref-type="bibr" rid="b3-sensors-12-10980">3</xref>].</p>
<p>Since the monitor automatically performs calibration, the sensor is steadily ready to use. The small-sized sensor is easily attached to the ear lobe with a low pressure clip, and has been tested in adults [<xref ref-type="bibr" rid="b4-sensors-12-10980">4</xref>–<xref ref-type="bibr" rid="b6-sensors-12-10980">6</xref>], children [<xref ref-type="bibr" rid="b7-sensors-12-10980">7</xref>], and neonates with birth weight (BW) &gt;1,500 g and gestational age (GA) &gt;28 weeks [<xref ref-type="bibr" rid="b8-sensors-12-10980">8</xref>,<xref ref-type="bibr" rid="b9-sensors-12-10980">9</xref>]. TOSCA monitor resulted reliable and applicable also in very low birth weight infants when the sensor was placed at the ear pinna, rather than at the earlobe, because of its small size. <xref ref-type="fig" rid="f1-sensors-12-10980">Figure 1(b)</xref> [<xref ref-type="bibr" rid="b10-sensors-12-10980">10</xref>].</p>
<p>Since the use of pulse oximetry immediately after birth is indicated to titrate oxygen supplementation, as recommended by the Guidelines for Neonatal Resuscitation [<xref ref-type="bibr" rid="b11-sensors-12-10980">11</xref>], the present study was designed to test the usability and reliability of TOSCA in the delivery room (DR) among term, late-preterm, and preterm neonates.</p></sec>
<sec>
<label>2.</label>
<title>Experimental Section</title>
<sec sec-type="materials|methods">
<label>2.1.</label>
<title>Materials and Methods</title>
<p>This prospective observational study was carried out between January and October 2010. It was approved by the ethical committee of our institution and oral informed parental consent was obtained. The study was conducted throughout three different periods among neonates recruited immediately after birth.</p>
<p>During the first period (<bold><italic>period A</italic></bold>), usability and reliability of TOSCA (TOSCA<sup>500®</sup> Radiometer, Copenhagen, Denmark) were evaluated in 30 term healthy neonates. A simultaneous SpO<sub>2</sub> monitoring with TOSCA and pulse oximeter MASIMO Radical-7 SET (MASIMO<sup>®</sup> Corporation, Irvine, CA, USA) was performed. SpO<sub>2</sub> values reported by TOSCA (SpO<sub>2TOSCA</sub>) were compared to those reported by MASIMO (SpO<sub>2MASIMO</sub>) throughout 10 min of monitoring. During this period, also PtcCO<sub>2</sub> levels were recorded with TOSCA (PtcCO<sub>2TOSCA</sub>) and a blood gas analysis (BGA) by capillary sampling was performed at the end of monitoring.</p>
<p>During the second period (<bold><italic>period B</italic></bold>), PtcCO<sub>2TOSCA</sub> levels were recorded in 10 late-preterm neonates and compared to PtcCO<sub>2</sub> values (PtcCO<sub>2TINA</sub>) recorded by a transcutaneous device (TINA<sup>®</sup> TCM3, Radiometer, Copenhagen, Denmark). TINA was equipped with a sensor applied on the abdomen and set at 44 °C. After 10 minutes of monitoring, a BGA was performed from umbilical line.</p>
<p>In the third period (<bold><italic>period C</italic></bold>), 15 preterm neonates were monitored. SpO<sub>2TOSCA</sub> and SpO<sub>2MASIMO</sub> values were recorded immediately after birth and during stabilization and transport to the neonatal intensive care unit (NICU) (total time: 40 min). Also PtcCO<sub>2TOSCA</sub> values were registered during the same period and a BGA was performed after 10 min of monitoring from umbilical line.</p>
<p>TOSCA was always used in QUICKSTART mode (temperature set at 44 °C during the first 20 min, then at 42 °C). The sensor probe cleaned with alcohol and dried before each application was applied to the adhesive clip at the ear pinna, using one drop of contact solution. Since Lacerenza <italic>et al.</italic>, showed that the placement site at the ear pinna facilitated the application of the sensor, it was electively applied at the right ear pinna rather than at the ear lobe [<xref ref-type="bibr" rid="b10-sensors-12-10980">10</xref>]. The MASIMO sensor was placed around the right hand and applied before connection to the oximeter [<xref ref-type="bibr" rid="b12-sensors-12-10980">12</xref>,<xref ref-type="bibr" rid="b13-sensors-12-10980">13</xref>].</p>
<p>Blood samples were analyzed for PCO<sub>2BGA</sub> with a blood gas analyzer (Stat Profile® Critical Care Xpress, Nova Biomedical Corporation, Waltham, MA, USA), routinely calibrated according to the manufacturer's instructions. BGA was performed at 10 min of monitoring since this time is indicated as an adequate time for TOSCA equilibration by the manifacturers themselves [<xref ref-type="bibr" rid="b14-sensors-12-10980">14</xref>,<xref ref-type="bibr" rid="b15-sensors-12-10980">15</xref>].</p>
<p>Patient characteristics (GA, BW, gender) and resuscitative measures required at birth were recorded.</p>
<p>The conduct of the study did not interfere with routine clinical practice: while a neonatologist was assisting the neonate, a nurse placed the pulse oximeter sensor, and a dedicated researcher applied both TOSCA and TINA sensors and collected data. Transcutaneous monitoring began as the sensors were adequately placed. Data were collected every min during the first 5 min, then every 5 min for the remaining time. At the end of the monitoring phase, sensors were removed and the underlying skin was examined to verify the presence of burns.</p></sec>
<sec sec-type="methods">
<label>2.2.</label>
<title>Statistical Analysis</title>
<p>Statistical analysis was performed using the Stata Statistical Software: Release 10 (StataCorp LP, College Station, TX, USA). Results are presented as mean ± standard deviation (SD) for continuous variables or as median (interquartile range) and as number (percentage) for categorical variables. Unpaired Student's <italic>t</italic> test was used for parametric data, Wilcoxon rank-sum test (Mann-Whitney U test) for nonparametric data, and Fisher's exact test for categorical variables. Bland-Altman analysis [<xref ref-type="bibr" rid="b16-sensors-12-10980">16</xref>] was used to evaluate the agreement at 10 min between PtcCO<sub>2TOSCA</sub> and PCO<sub>2TINA</sub>, PtcCO<sub>2TOSCA</sub> and PCO<sub>2BGA</sub>, PCO<sub>2BGA</sub> and PtcCO<sub>2TINA</sub>. Precision was defined as 2 SD of the mean difference (bias). A <italic>p</italic> &lt; <italic>0.05</italic> was considered statistically significant.</p></sec></sec>
<sec sec-type="results">
<label>3.</label>
<title>Results</title>
<p>Of the 30 neonates born during <bold><italic>period A</italic></bold> (GA: 37.9 ± 1 weeks; BW: 3138 ± 459 g, 13 males and 17 females), none needed respiratory support at birth. No differences between SpO<sub>2TOSCA</sub> and SpO<sub>2MASIMO</sub> were detected at all time points (<xref ref-type="fig" rid="f2-sensors-12-10980">Figure 2</xref>(top)). For both monitors, data were available 60 s after sensor placement. The clipping of the TOSCA sensor was easy. Sensor repositioning was necessary in 16/30 registrations (53%) with TOSCA and in 5/30 registrations (16.6%) with MASIMO (<italic>p</italic> = <italic>0.003</italic>). After sensor repositioning, time required to obtain a stable reading was shorter for TOSCA than for MASIMO (about 10 <italic>vs.</italic> 15 s). PtcCO<sub>2TOSCA</sub> progressively increased during the whole monitoring (<xref ref-type="fig" rid="f2-sensors-12-10980">Figure 2(bottom)</xref>).</p>
<p>PCO<sub>2BGA</sub> at 10 min was 61.5 ± 10.8 mmHg, significantly higher than PtcCO<sub>2TOSCA</sub> (48.6 ± 11.7 mmHg; <italic>p</italic> &lt; <italic>0.005</italic>). At 10 min, the bias (precision) between PtcCO<sub>2TOSCA</sub> and PCO<sub>2BGA</sub> values was 12.9 (26.4). No signs of skin erythema or burn at the ear pinna were found at the end of monitoring.</p>
<p>None of the 10 neonates (GA: 36.1 ± 1.4 weeks; BW: 2743 ± 714 g; five males and five females) born during <bold><italic>period B</italic></bold> needed respiratory support. As shown in <xref ref-type="fig" rid="f3-sensors-12-10980">Figure 3</xref>, during the first 5 min PtcCO<sub>2TOSCA</sub> values resulted significantly lower than PtcCO<sub>2TINA</sub> values, while at 10 min similar values were recorded (PtcCO<sub>2TOSCA</sub> 52 ± 17.6 mmHg <italic>vs.</italic> PtcCO<sub>2TINA</sub> 58.6 ± 20.2 mmHg; <italic>p</italic> = <italic>ns</italic>).</p>
<p>PCO<sub>2BGA</sub> at 10 min was 59.4 ± 17.3 mmHg; the bias (precision) between PtcCO<sub>2TOSCA</sub> and PCO<sub>2BGA</sub> values was 7.42 (39.8), between PCO<sub>2BGA</sub> and PtcCO<sub>2TINA</sub> 0.82 (31.9), and between PtcCO<sub>2TOSCA</sub> and PtcCO<sub>2TINA</sub> 6.6 (43.4). Sensor repositioning was required 3 times (30%) for TOSCA.</p>
<p>Fifteen preterm neonates (GA: 31.6 ± 2.0 weeks; BW: 1622 ± 569 grams; nine males and six females) born during <bold><italic>period C</italic></bold> were monitored for 40 min with TOSCA and MASIMO. Two neonates required resuscitation and were transferred to NICU mechanically ventilated; six neonates were assisted with nasal continuous positive airway pressure. Sensor repositioning was required six times (40%) for TOSCA and never for MASIMO. SpO<sub>2TOSCA</sub> and SpO<sub>2MASIMO</sub> values were similar during the whole study period (<xref ref-type="fig" rid="f4-sensors-12-10980">Figure 4</xref>(top)). PtcCO<sub>2TOSCA</sub> values progressively increased until 10 min, then became stable (<xref ref-type="fig" rid="f4-sensors-12-10980">Figure 4(bottom)</xref>).</p>
<p>PCO<sub>2BGA</sub> at 10 min was 63.1 ± 20.3 mmHg and the bias (precision) between PtcCO<sub>2TOSCA</sub> and PCO<sub>2BGA</sub> was 11.1 (43.4).</p></sec>
<sec sec-type="discussion">
<label>4.</label>
<title>Discussion</title>
<p>The Guidelines on Neonatal Resuscitation recommend the use of pulse oximetry to monitor neonates' oxygenation status in the DR [<xref ref-type="bibr" rid="b11-sensors-12-10980">11</xref>], since skin color is only partially reliable [<xref ref-type="bibr" rid="b17-sensors-12-10980">17</xref>]. Moreover, awareness about neonatal oxygenation status is helpful to titrate oxygen supplementation, if requested.</p>
<p>In case of frequent motion artifacts or low perfusion, often occurring in DR, pulse oximeters with MASIMO Signal Extraction Technology are indicated [<xref ref-type="bibr" rid="b18-sensors-12-10980">18</xref>].</p>
<p>TOSCA monitor records SpO<sub>2</sub> with MASIMO technology and simultaneously detects PtcCO<sub>2</sub>, allowing a combined assessment of both parameters. The present study <italic>investigated for the first time the possible use of TOSCA monitor in neonates immediately after birth</italic>, a period in which the use of excessive pressure or volume (overstretching) and oxidative aggression (reactive oxygen and nitrogen species) may cause acute damage to the lung and other organs, in particular in preterm babies. The simultaneous measurement of SpO<sub>2</sub> and PtcCO<sub>2</sub> could be helpful both to verify the physiologic changes occurring during transition from the intra- to the extrauterine life and to determine the appropriate grade of respiratory support when resuscitation is required.</p>
<p>Concerning usability, our results showed that TOSCA sensor can be easily applied, even in the DR immediately after birth, to the ear pinna of both term and preterm neonates. As previously demonstrated [<xref ref-type="bibr" rid="b10-sensors-12-10980">10</xref>], sensor application to the ear pinna rather than to the ear lobe did not alter data displaying. The size of the clipped sensor is well suited for the thin, flat, and soft ear pinna of the newborn, while it may result too large compared to the small ear lobe. Nevertheless, dislocations occurred more often using TOSCA than MASIMO. In this study, repositioning of the TOSCA sensor was more frequently required among term than among late-preterm or preterm neonates, presumably because of their greater mobility. However, the viability of TOSCA monitoring was not impaired thanks both to the shorter time required for sensor clipping and to its faster reading, due to the high sensor's stability [<xref ref-type="bibr" rid="b1-sensors-12-10980">1</xref>,<xref ref-type="bibr" rid="b2-sensors-12-10980">2</xref>].</p>
<p>No burns underneath the TOSCA sensor were identified in any of the infants, including premature neonates monitored for longer period (40 min) in period C. This was presumably due to a lower electrode temperature after 20 min of monitoring in the QUICKSTART mode. In fact, this modality allows a long term (up to 12 h) application of the sensor without adverse events, although measurements of PtCO<sub>2</sub> may result less reliable [<xref ref-type="bibr" rid="b9-sensors-12-10980">9</xref>]. This could be due to the lower sensor temperature which is not able to provide an adequate grade of arterialization in order to compensate the greater CO<sub>2</sub> local production induced by temperature itself. However, this occurrence has only limited implications in our study, since we performed a “long time evaluation” of the PtcCO<sub>2TOSCA</sub> exclusively in Period C, throughout a monitoring period which is much shorter than the one described by other authors [<xref ref-type="bibr" rid="b9-sensors-12-10980">9</xref>].</p>
<p>Concerning the reliability of SpO<sub>2TOSCA</sub>, the recorded values were comparable to those detected by SpO<sub>2MASIMO</sub>. As formerly demonstrated [<xref ref-type="bibr" rid="b10-sensors-12-10980">10</xref>], the different placement site of both sensors, always preductal, did not affect SpO<sub>2</sub> detection.</p>
<p>Regarding PtcCO<sub>2TOSCA</sub> measurement in DR, the estimation accuracy of TOSCA resulted generally acceptable compared to blood samples collected at 10 min of use. Although this single BGA, representing a punctual evaluation, may be a limit in the estimation of PtcCO<sub>2</sub> values, a higher number of blood collection was avoided to prevent painful punctures. However, the agreement between PtcCO<sub>2TOSCA</sub> and PCO<sub>2BGA</sub> was more accentuated when blood samples were taken by the umbilical line (period B and C) compared to the capillary sampling (period A). A possible explanation of these results may be that capillary PCO<sub>2BGA</sub> is comparable to arterial PCO<sub>2BGA</sub> only after the first hours of life [<xref ref-type="bibr" rid="b19-sensors-12-10980">19</xref>]. Nevertheless, after 10 min of registration, the TOSCA monitor reveals less accurate PtcCO<sub>2</sub> values than the TINA one compared to the umbilical PCO<sub>2BGA</sub> (period B), confirming that the measurement of PtCO<sub>2</sub> with TOSCA was less precise than the conventional monitoring [<xref ref-type="bibr" rid="b10-sensors-12-10980">10</xref>].</p>
<p>The results achieved throughout <bold><italic>period B</italic></bold> demonstrated that during the first minutes of registration the PtcCO<sub>2TOSCA</sub> values were lower than those recorded by TINA and became comparable only after 5 min. This is in agreement with previously reported data achieved among preterm neonates admitted to NICU, and was probably dependent on the placement site of the TOSCA sensor [<xref ref-type="bibr" rid="b8-sensors-12-10980">8</xref>]. The thin dermal layer and the poor capillary bed of the ear pinna may induce a lower grade of arterialization than the one obtained in other sites [<xref ref-type="bibr" rid="b10-sensors-12-10980">10</xref>].</p>
<p>A similar slowly increasing trend of the PtcCO<sub>2TOSCA</sub> was also detected among preterm infants monitored after birth and during transport to the NICU (period C), and was followed by value stabilization after 10 min of monitoring.</p></sec>
<sec sec-type="conclusions">
<label>5.</label>
<title>Conclusions</title>
<p>We can assert that TOSCA was easily and safely usable in DR both in term and preterm neonates, allowing a reliable and noninvasive estimation of SpO<sub>2</sub>. Considering that at least 10 min were required to obtain reliable and stable values of PtcCO<sub>2TOSCA</sub>, TOSCA reliability results limited during the first minutes of life, when resuscitation procedures may be necessary. However, even if the optimal level of PCO<sub>2</sub> during neonatal resuscitation is not known yet [<xref ref-type="bibr" rid="b20-sensors-12-10980">20</xref>], PtcCO<sub>2</sub> measurement with TOSCA could be useful as trend monitoring after neonatal stabilization and during transport to the NICU to prevent lung injury, since its beginning may occur soon after birth among neonates requiring positive-pressure ventilation. We believe that our study may represent a stimulating starting point for further investigations focusing on the development of respiratory function monitoring devices to be used in DR [<xref ref-type="bibr" rid="b21-sensors-12-10980">21</xref>]. In the near future, an efficient technology may enable the individualization of the specific ventilatory and oxygenation requirements, thus minimizing physical and oxygen-derived damage [<xref ref-type="bibr" rid="b22-sensors-12-10980">22</xref>].</p></sec></body>
<back>
<ref-list>
<title>References</title>
<ref id="b1-sensors-12-10980"><label>1.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Eberhard</surname><given-names>P.</given-names></name><name><surname>Gisiger</surname><given-names>P.A.</given-names></name><name><surname>Gardaz</surname><given-names>J.P.</given-names></name><name><surname>Spahn</surname><given-names>D.R.</given-names></name></person-group><article-title>Combining transcutaneous blood gas measurement and pulse oximetry</article-title><source>Anesth. Analg.</source><year>2002</year><volume>94</volume><fpage>S76</fpage><lpage>S80</lpage><pub-id pub-id-type="pmid">11900043</pub-id></citation></ref>
<ref id="b2-sensors-12-10980"><label>2.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Eberhard</surname><given-names>P.</given-names></name></person-group><article-title>The design, use, and results of transcutaneous carbon dioxide analysis: Current and future directions</article-title><source>Anesth. Analg.</source><year>2007</year><volume>105</volume><fpage>S48</fpage><lpage>S52</lpage><pub-id pub-id-type="doi">10.1213/01.ane.0000278642.16117.f8</pub-id><pub-id pub-id-type="pmid">18048898</pub-id></citation></ref>
<ref id="b3-sensors-12-10980"><label>3.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Severinghaus</surname><given-names>J.W.</given-names></name></person-group><article-title>Transcutaneous blood gas analysis</article-title><source>Respir. Care</source><year>1982</year><volume>27</volume><fpage>152</fpage><lpage>159</lpage></citation></ref>
<ref id="b4-sensors-12-10980"><label>4.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Senn</surname><given-names>O.</given-names></name><name><surname>Clarenbach</surname><given-names>C.F.</given-names></name><name><surname>Kaplan</surname><given-names>V.</given-names></name><name><surname>Maggiorini</surname><given-names>M.</given-names></name><name><surname>Bloch</surname><given-names>K.E.</given-names></name></person-group><article-title>Monitoring carbon dioxide tension and arterial oxygen saturation by a single earlobe sensor in patients with critical illness or sleep apnea</article-title><source>Chest</source><year>2005</year><volume>128</volume><fpage>1291</fpage><lpage>1296</lpage><pub-id pub-id-type="doi">10.1378/chest.128.3.1291</pub-id><pub-id pub-id-type="pmid">16162720</pub-id></citation></ref>
<ref id="b5-sensors-12-10980"><label>5.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Bendjelid</surname><given-names>K.</given-names></name><name><surname>Schutz</surname><given-names>N.</given-names></name><name><surname>Stotz</surname><given-names>M.</given-names></name><name><surname>Gerard</surname><given-names>I.</given-names></name><name><surname>Suter</surname><given-names>P.M.</given-names></name><name><surname>Romand</surname><given-names>J.A.</given-names></name></person-group><article-title>Transcutaneous PCO<sub>2</sub> monitoring in critically ill adults: Clinical evaluation of a new sensor</article-title><source>Crit. Care Med.</source><year>2005</year><volume>33</volume><fpage>2203</fpage><lpage>2206</lpage><pub-id pub-id-type="doi">10.1097/01.CCM.0000181734.26070.26</pub-id><pub-id pub-id-type="pmid">16215371</pub-id></citation></ref>
<ref id="b6-sensors-12-10980"><label>6.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Parker</surname><given-names>S.M.</given-names></name><name><surname>Gibson</surname><given-names>G.J.</given-names></name></person-group><article-title>Evaluation of a transcutaneous carbon dioxide monitor (“TOSCA”) in adult patients in routine respiratory practice</article-title><source>Resp. Med.</source><year>2007</year><volume>101</volume><fpage>261</fpage><lpage>264</lpage><pub-id pub-id-type="doi">10.1016/j.rmed.2006.05.011</pub-id></citation></ref>
<ref id="b7-sensors-12-10980"><label>7.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Dullenkopf</surname><given-names>A.</given-names></name><name><surname>Di Bernardo</surname><given-names>S.</given-names></name><name><surname>Berger</surname><given-names>F.</given-names></name><name><surname>Fasnacht</surname><given-names>M.</given-names></name><name><surname>Gerber</surname><given-names>A.C.</given-names></name><name><surname>Weiss</surname><given-names>M.</given-names></name></person-group><article-title>Evaluation of a new combined SpO<sub>2</sub>/PtcCO<sub>2</sub> sensor in anaesthetized paediatric patients</article-title><source>Paediatr. Anaesth.</source><year>2003</year><volume>13</volume><fpage>777</fpage><lpage>784</lpage><pub-id pub-id-type="doi">10.1046/j.1460-9592.2003.01146.x</pub-id><pub-id pub-id-type="pmid">14617118</pub-id></citation></ref>
<ref id="b8-sensors-12-10980"><label>8.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Bernet-Buettiker</surname><given-names>V.</given-names></name><name><surname>Ugarte</surname><given-names>M.J.</given-names></name><name><surname>Frey</surname><given-names>B.</given-names></name><name><surname>Hug</surname><given-names>M.I.</given-names></name><name><surname>Baenziger</surname><given-names>O.</given-names></name><name><surname>Weiss</surname><given-names>M.</given-names></name></person-group><article-title>Evaluation of a new combined transcutaneous measurement of PCO<sub>2</sub>/pulse oximetry oxygen saturation ear sensor in newborn patients</article-title><source>Pediatrics</source><year>2005</year><volume>115</volume><fpage>e64</fpage><lpage>e68</lpage><pub-id pub-id-type="pmid">15601814</pub-id></citation></ref>
<ref id="b9-sensors-12-10980"><label>9.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Bernet</surname><given-names>V.</given-names></name><name><surname>Döll</surname><given-names>C.</given-names></name><name><surname>Cannizzaro</surname><given-names>V.</given-names></name><name><surname>Ersch</surname><given-names>J.</given-names></name><name><surname>Frey</surname><given-names>B.</given-names></name><name><surname>Weiss</surname><given-names>M.</given-names></name></person-group><article-title>Longtime performance and reliability of two different PtcCO2 and SpO2 sensors in neonates</article-title><source>Paediatr. Anaesth.</source><year>2008</year><volume>18</volume><fpage>872</fpage><lpage>877</lpage><pub-id pub-id-type="doi">10.1111/j.1460-9592.2008.02661.x</pub-id><pub-id pub-id-type="pmid">18768047</pub-id></citation></ref>
<ref id="b10-sensors-12-10980"><label>10.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Lacerenza</surname><given-names>S.</given-names></name><name><surname>De Carolis</surname><given-names>M.P.</given-names></name><name><surname>Fusco</surname><given-names>F.P.</given-names></name><name><surname>La Torre</surname><given-names>G.</given-names></name><name><surname>Chiaradia</surname><given-names>G.</given-names></name><name><surname>Romagnoli</surname><given-names>C.</given-names></name></person-group><article-title>An evaluation of a new combined SpO2/PtcCO2 sensor in very low birth weight infants</article-title><source>Anesth. Analg.</source><year>2008</year><volume>107</volume><fpage>125</fpage><lpage>129</lpage><pub-id pub-id-type="doi">10.1213/ane.0b013e3181733e47</pub-id><pub-id pub-id-type="pmid">18635477</pub-id></citation></ref>
<ref id="b11-sensors-12-10980"><label>11.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Kattwinkel</surname><given-names>J.</given-names></name><name><surname>Perlman</surname><given-names>J.M.</given-names></name><name><surname>Aziz</surname><given-names>K.</given-names></name><name><surname>Colby</surname><given-names>C.</given-names></name><name><surname>Fairchild</surname><given-names>K.</given-names></name><name><surname>Gallagher</surname><given-names>J.</given-names></name><name><surname>Hazinski</surname><given-names>M.F.</given-names></name><name><surname>Halamek</surname><given-names>L.P.</given-names></name><name><surname>Kumar</surname><given-names>P.</given-names></name><name><surname>Little</surname><given-names>G.</given-names></name><etal/></person-group><article-title>American heart association. Neonatal resuscitation: 2010 American heart association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care</article-title><source>Pediatrics</source><year>2010</year><volume>126</volume><fpage>e1400</fpage><lpage>e1413</lpage><pub-id pub-id-type="doi">10.1542/peds.2010-2972E</pub-id><pub-id pub-id-type="pmid">20956432</pub-id></citation></ref>
<ref id="b12-sensors-12-10980"><label>12.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>O'Donnell</surname><given-names>C.P.</given-names></name><name><surname>Kamlin</surname><given-names>C.O.</given-names></name><name><surname>Davis</surname><given-names>P.G.</given-names></name><name><surname>Morley</surname><given-names>C.J.</given-names></name></person-group><article-title>Feasibility of and delay in obtaining pulse oximetry during neonatal resuscitation</article-title><source>J. Pediatr.</source><year>2005</year><volume>147</volume><fpage>698</fpage><lpage>699</lpage><pub-id pub-id-type="doi">10.1016/j.jpeds.2005.07.025</pub-id><pub-id pub-id-type="pmid">16291367</pub-id></citation></ref>
<ref id="b13-sensors-12-10980"><label>13.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>O'Donnell</surname><given-names>C.P.</given-names></name><name><surname>Kamlin</surname><given-names>C.O.</given-names></name><name><surname>Davis</surname><given-names>P.G.</given-names></name><name><surname>Morley</surname><given-names>C.J.</given-names></name></person-group><article-title>Obtaining pulse oximetry data in neonates: A randomised crossover study of sensor application techniques</article-title><source>Arch. Dis. Child Fetal Neonatal Ed.</source><year>2005</year><volume>90</volume><fpage>84</fpage><lpage>85</lpage><pub-id pub-id-type="doi">10.1136/adc.2004.058925</pub-id></citation></ref>
<ref id="b14-sensors-12-10980"><label>14.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Kelly</surname><given-names>A.M.</given-names></name><name><surname>Klim</surname><given-names>S.</given-names></name></person-group><article-title>Agreement between arterial and transcutaneous PCO2 in patients undergoing non-invasive ventilation</article-title><source>Respir. Med.</source><year>2011</year><volume>105</volume><fpage>226</fpage><lpage>229</lpage><pub-id pub-id-type="doi">10.1016/j.rmed.2010.11.010</pub-id><pub-id pub-id-type="pmid">21131188</pub-id></citation></ref>
<ref id="b15-sensors-12-10980"><label>15.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>De Oliveira</surname><given-names>G.S.</given-names><suffix>Jr.</suffix></name><name><surname>Ahmad</surname><given-names>S.</given-names></name><name><surname>Fitzgerald</surname><given-names>P.C.</given-names></name><name><surname>McCarthy</surname><given-names>R.J.</given-names></name></person-group><article-title>Detection of hypoventilation during deep sedation in patients undergoing ambulatory gynaecological hysteroscopy: A comparison between transcutaneous and nasal end-tidal carbon dioxide measurements</article-title><source>Br. J. Anaesth.</source><year>2010</year><volume>104</volume><fpage>774</fpage><lpage>778</lpage><pub-id pub-id-type="doi">10.1093/bja/aeq092</pub-id><pub-id pub-id-type="pmid">20418266</pub-id></citation></ref>
<ref id="b16-sensors-12-10980"><label>16.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Bland</surname><given-names>M.J.</given-names></name><name><surname>Altman</surname><given-names>D.G.</given-names></name></person-group><article-title>Statistical methods for assessing agreement between two methods of clinical measurements</article-title><source>Lancet</source><year>1986</year><volume>1</volume><fpage>307</fpage><lpage>310</lpage><pub-id pub-id-type="pmid">2868172</pub-id></citation></ref>
<ref id="b17-sensors-12-10980"><label>17.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>O'Donnell</surname><given-names>C.P.</given-names></name><name><surname>Kamlin</surname><given-names>C.O.</given-names></name><name><surname>Davis</surname><given-names>P.G.</given-names></name><name><surname>Carlin</surname><given-names>J.B.</given-names></name><name><surname>Morley</surname><given-names>C.J.</given-names></name></person-group><article-title>Clinical assessment of infant colour at delivery</article-title><source>Arch. Dis. Child-Fetal Neonatal.</source><year>2007</year><volume>92</volume><fpage>465</fpage><lpage>467</lpage><pub-id pub-id-type="doi">10.1136/adc.2007.120634</pub-id></citation></ref>
<ref id="b18-sensors-12-10980"><label>18.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Hay</surname><given-names>W.W.</given-names><suffix>Jr.</suffix></name><name><surname>Rodden</surname><given-names>D.J.</given-names></name><name><surname>Collins</surname><given-names>S.M.</given-names></name><name><surname>Melara</surname><given-names>D.L.</given-names></name><name><surname>Hale</surname><given-names>K.A.</given-names></name><name><surname>Fashaw</surname><given-names>L.M.</given-names></name></person-group><article-title>Reliability of conventional and new pulse oximetry in neonatal patients</article-title><source>J. Perinatol.</source><year>2002</year><volume>22</volume><fpage>360</fpage><lpage>366</lpage><pub-id pub-id-type="doi">10.1038/sj.jp.7210740</pub-id><pub-id pub-id-type="pmid">12082469</pub-id></citation></ref>
<ref id="b19-sensors-12-10980"><label>19.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Koch</surname><given-names>G.</given-names></name><name><surname>Wendel</surname><given-names>H.</given-names></name></person-group><article-title>Comparison of pH, carbon dioxide tension, standard bicarbonate and oxygen tension in arterial and capillary blood in the newborn period</article-title><source>Acta Paediatr. Scand.</source><year>1967</year><volume>56</volume><fpage>10</fpage><lpage>16</lpage></citation></ref>
<ref id="b20-sensors-12-10980"><label>20.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Saugstad</surname><given-names>O.D.</given-names></name></person-group><article-title>New guidelines for newborn resuscitation</article-title><source>Acta Paediatr.</source><year>2007</year><volume>96</volume><fpage>333</fpage><lpage>337</lpage><pub-id pub-id-type="doi">10.1111/j.1651-2227.2006.00181.x</pub-id><pub-id pub-id-type="pmid">17407451</pub-id></citation></ref>
<ref id="b21-sensors-12-10980"><label>21.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Perlman</surname><given-names>J.</given-names></name><name><surname>Kattwinkel</surname><given-names>J.</given-names></name><name><surname>Wyllie</surname><given-names>J.</given-names></name><name><surname>Guinsburg</surname><given-names>R.</given-names></name><name><surname>Velaphi</surname><given-names>S.</given-names></name></person-group><article-title>Nalini singhal for the neonatal ILCOR task force group. Neonatal resuscitation: In pursuit of evidence gaps in knowledge</article-title><source>Resuscitation</source><year>2012</year><volume>83</volume><fpage>545</fpage><lpage>550</lpage><pub-id pub-id-type="doi">10.1016/j.resuscitation.2012.01.003</pub-id><pub-id pub-id-type="pmid">22245921</pub-id></citation></ref>
<ref id="b22-sensors-12-10980"><label>22.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Vento</surname><given-names>M.</given-names></name><name><surname>Cheung</surname><given-names>P.Y.</given-names></name><name><surname>Aguar</surname><given-names>M.</given-names></name></person-group><article-title>The first golden minutes of the extremely-low-gestational-age neonate: A gentle approach</article-title><source>Neonatology</source><year>2009</year><volume>95</volume><fpage>286</fpage><lpage>298</lpage><pub-id pub-id-type="doi">10.1159/000178770</pub-id><pub-id pub-id-type="pmid">19052475</pub-id></citation></ref></ref-list>
<sec sec-type="display-objects">
<title>Figures</title>
<fig id="f1-sensors-12-10980" position="float">
<label>Figure 1.</label>
<caption>
<p>(<bold>a</bold>) TOSCA sensor combines an optical sensor for the measurement of SpO<sub>2</sub> (coloured in red) with a Stow-Severinghaus type CO<sub>2</sub> sensor (coloured in blue), and is also equipped with a heating element (coloured in black); image from Eberhard P. [<xref ref-type="bibr" rid="b2-sensors-12-10980">2</xref>]. (<bold>b</bold>) Application of the TOSCA sensor at the ear pinna; image from Lacerenza S. [<xref ref-type="bibr" rid="b10-sensors-12-10980">10</xref>].</p></caption>
<graphic xlink:href="sensors-12-10980f1.gif"/></fig>
<fig id="f2-sensors-12-10980" position="float">
<label>Figure 2.</label>
<caption>
<p>Trend of SpO<sub>2MASIMO</sub> (white diamonds) and SpO<sub>2TOSCA</sub> (black squares) (<bold>top</bold>), and trend of PtcCO<sub>2TOSCA</sub>; (<bold>bottom</bold>) during period A.</p></caption>
<graphic xlink:href="sensors-12-10980f2.gif"/></fig>
<fig id="f3-sensors-12-10980" position="float">
<label>Figure 3.</label>
<caption>
<p>Trend of PtcCO<sub>2TOSCA</sub> (black dots) and PtcCO<sub>2 TINA</sub> (white triangles) during <bold><italic>period B</italic></bold>. * <italic>p</italic> &lt; 0.05.</p></caption>
<graphic xlink:href="sensors-12-10980f3.gif"/></fig>
<fig id="f4-sensors-12-10980" position="float">
<label>Figure 4.</label>
<caption>
<p>Trend of SpO<sub>2MASIMO</sub> (white diamonds) and SpO<sub>2TOSCA</sub> (black squares) (<bold>top</bold>), and trend of PtcCO<sub>2TOSCA</sub> (<bold>bottom</bold>) during period C.</p></caption>
<graphic xlink:href="sensors-12-10980f4.gif"/></fig></sec></back></article>