<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD Journal Publishing DTD v2.3 20070202//EN" "journalpublishing.dtd">
<article xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" xml:lang="en" article-type="research-article">
<front>
<journal-meta>
<journal-id journal-id-type="nlm-ta">Sensors</journal-id>
<journal-title>Sensors</journal-title>
<issn pub-type="epub">1424-8220</issn>
<publisher>
<publisher-name>Molecular Diversity Preservation International (MDPI)</publisher-name></publisher></journal-meta>
<article-meta>
<article-id pub-id-type="doi">10.3390/s110908813</article-id>
<article-id pub-id-type="publisher-id">sensors-11-08813</article-id>
<article-categories>
<subj-group>
<subject>Article</subject></subj-group></article-categories>
<title-group>
<article-title>Use of a New Ziprasidone-Selective Electrode in Mixed Solvents and Its Application in the Analysis of Pharmaceuticals and Biological Fluids</article-title></title-group>
<contrib-group>
<contrib contrib-type="author">
<name><surname>García</surname><given-names>Mª Soledad</given-names></name><xref ref-type="corresp" rid="c1-sensors-11-08813"><sup>*</sup></xref></contrib>
<contrib contrib-type="author">
<name><surname>Ortuño</surname><given-names>Joaquín A.</given-names></name></contrib>
<contrib contrib-type="author">
<name><surname>Cuartero</surname><given-names>María</given-names></name></contrib>
<contrib contrib-type="author">
<name><surname>Abuherba</surname><given-names>Mustafa Salem</given-names></name></contrib>
<aff id="af1-sensors-11-08813">Department of Analytical Chemistry, Faculty of Chemistry, University of Murcia, E-30100 Murcia, Spain; E-Mails: <email>jortuno@um.es</email> (J.A.O.); <email>mcb2@um.es</email> (M.C.); <email>abumusalem@yahoo.es</email> (M.S.A.)</aff></contrib-group>
<author-notes>
<corresp id="c1-sensors-11-08813">
<label>*</label>Author to whom correspondence should be addressed; E-Mail: <email>msgarcia@um.es</email>; Tel.: +34-868-887-404; Fax: +34-868-887-409.</corresp></author-notes>
<pub-date pub-type="collection">
<year>2011</year></pub-date>
<pub-date pub-type="epub">
<day>13</day>
<month>9</month>
<year>2011</year></pub-date>
<volume>11</volume>
<issue>9</issue>
<fpage>8813</fpage>
<lpage>8825</lpage>
<history>
<date date-type="received">
<day>4</day>
<month>8</month>
<year>2011</year></date>
<date date-type="rev-recd">
<day>7</day>
<month>9</month>
<year>2011</year></date>
<date date-type="accepted">
<day>8</day>
<month>9</month>
<year>2011</year></date></history>
<permissions>
<copyright-statement>© 2011 by the authors; licensee MDPI, Basel, Switzerland.</copyright-statement>
<copyright-year>2011</copyright-year>
<license>
<p>This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/).</p></license></permissions>
<abstract>
<p>The construction and characterization of a new ion-selective electrode for the determination of the antipsychotic ziprasidone in mixed solvents is presented. The electrode contains a plasticized polymeric membrane based on a ziprasidone-tetraphenylborate ion-exchanger. The influence of membrane composition on the electrode response towards ziprasidone in hydroalcoholic solutions was studied. The electrode displayed a stable response in a 2:3 (v/v) methanol/water medium from a ziprasidone concentration of 3 × 10<sup>−6</sup> M with a fast response time of less than 20 s. The electrode also showed good selectivity towards ziprasidone over common inorganic and organic compounds and several species with pharmacological activity. The electrode was successfully applied to the determination of ziprasidone in pharmaceuticals and human urine and serum.</p></abstract>
<kwd-group>
<kwd>ion-selective electrode</kwd>
<kwd>ziprasidone</kwd>
<kwd>mixed solvents</kwd>
<kwd>pharmaceuticals</kwd>
<kwd>urine</kwd>
<kwd>serum</kwd></kwd-group></article-meta></front>
<body>
<sec sec-type="intro">
<label>1.</label>
<title>Introduction</title>
<p>Many drugs are very hydrophobic and consequently poorly soluble in water, resulting in problems that may affect different fields of pharmaceutical science. The use of aqueous mixed solvents to increase the solubility of these drugs is one of the oldest, most powerful and widely used methods to address this issue [<xref ref-type="bibr" rid="b1-sensors-11-08813">1</xref>].</p>
<p>The application of ion selective electrodes in mixed solvents involves a series of problems concerning membrane selection and the measuring technique to be used. However, potentiometric studies in these media have opened up new possibilities for analytical and thermodynamic fields based on the data obtained [<xref ref-type="bibr" rid="b2-sensors-11-08813">2</xref>,<xref ref-type="bibr" rid="b3-sensors-11-08813">3</xref>].</p>
<p>A thorough literature search revealed little information on the use of ion-selective electrodes (ISEs) based on plasticized polymeric membranes in mixed solvents. This is probably because the membrane components should exhibit low leaching into the sample solutions. Despite this, some authors have succeeded in using plasticized polymeric membrane ISEs for determining metal ions [<xref ref-type="bibr" rid="b4-sensors-11-08813">4</xref>–<xref ref-type="bibr" rid="b24-sensors-11-08813">24</xref>], alkaline and alkaline-earth ions [<xref ref-type="bibr" rid="b25-sensors-11-08813">25</xref>], monohydrogen phosphate [<xref ref-type="bibr" rid="b26-sensors-11-08813">26</xref>], chromate ions [<xref ref-type="bibr" rid="b27-sensors-11-08813">27</xref>] and surfactants [<xref ref-type="bibr" rid="b28-sensors-11-08813">28</xref>,<xref ref-type="bibr" rid="b29-sensors-11-08813">29</xref>] in mixed solvents.</p>
<p>Ziprasidone (5-[2-[4-(1,2-benzisothiazol-3-yl)-1-piperazin-yl ethyl]-6-chloro-1,3-dihydro-2H-indol-2-one, ZPD, <xref ref-type="fig" rid="f1a-sensors-11-08813">Figure 1</xref>), is a antipsychotic agent indicated for the treatment of schizophrenia, bipolar disorder and acute mania because of its affinity for serotonin (5HT2A) and dopamine (D2) receptors [<xref ref-type="bibr" rid="b30-sensors-11-08813">30</xref>]. The therapeutic importance of this drug requires the development of sensitive and rapid analytical methods for its determination in pharmaceutical and clinical analysis. A review of the literature showed that several analytical methods have been described in this respect, including spectrophotometric [<xref ref-type="bibr" rid="b31-sensors-11-08813">31</xref>–<xref ref-type="bibr" rid="b33-sensors-11-08813">33</xref>], thin layer densitometry [<xref ref-type="bibr" rid="b34-sensors-11-08813">34</xref>], capillary electrophoresis [<xref ref-type="bibr" rid="b35-sensors-11-08813">35</xref>], voltammetric [<xref ref-type="bibr" rid="b36-sensors-11-08813">36</xref>] and chromatographic [<xref ref-type="bibr" rid="b37-sensors-11-08813">37</xref>–<xref ref-type="bibr" rid="b41-sensors-11-08813">41</xref>] methods. However, to our knowledge, there is no ISE to determine ZPD. A possible reason for this could be the low solubility of ZPD in water, which makes it necessary to use non-aqueous or mixed solvents.</p>
<p>Since potentiometry with ISEs offers many advantages, among them simple design and operation, low cost, fast response and the possibility of automation, the aim of the present work was to develop a potentiometric selective electrode for ZPD determination which could be used in mixed solvents for application in pharmaceutical and clinical analysis.</p></sec>
<sec>
<label>2.</label>
<title>Experimental Section</title>
<sec>
<label>2.1.</label>
<title>Apparatus</title>
<p>Potentials were measured with an Orion 960 Autochemistry System (Thermo-Orion, Cambridge, MA, USA), the recorder output of which was connected to a personal computer via a DGH Corporation 1121 module analogue-to-digital converter (Manchester, UK). An Orion 90–02 double junction silver-silver chloride reference electrode containing a 10% (v/v) solution of KNO<sub>3</sub> in the outer compartment and a Fluka (Munich, Germany) ISE electrode body was used.</p></sec>
<sec>
<label>2.2.</label>
<title>Reagents and Solutions</title>
<p>High molecular weight poly(vinyl chloride) (PVC), 2-nitrophenyl octyl ether (NPOE), bis(2-ethylhexyl) sebacate (DOS), and tetrahydrofuran (THF) were Selectophore products from Fluka (Munich, Germany). Sodium tetraphenylborate (NaTPB) was purchased from Sigma (Munich, Germany). Ziprasidone hydrochloride powder was kindly provided by Pfizer Laboratories (New York, USA). All other reagents used were of analytical reagent grade and Milli-Q water was used throughout. Ac<sup>−</sup>/HAc buffer of pH 4.6 and 2 × 10<sup>−1</sup> M total concentration was used.</p>
<p>A 4 × 10<sup>−3</sup> M ziprasidone hydrochloride standard solution was prepared by dissolving 0.0467 g of ziprasidone hydrochloride in 25 mL of methanol. Working solutions (8 × 10<sup>−4</sup> M to 1.6 × 10<sup>−6</sup> M) were prepared by diluting appropriate volumes of the standard solution with 15 mL of Ac<sup>−</sup>/HAc buffer of pH 4.6 and methanol to a final volume of 25 mL in a calibrated flask.</p>
<p>For the selectivity study, 4 × 10<sup>−3</sup> M solutions of amoxicillin, carbamazepine, cimetidine, clomipramine hydrochloride, diclofenac sodium, fluoxetine hydrochloride, furosemide, haloperidol, lansoprazol, lorazepan, ranitidine, ofloxacine and tenoxicam were prepared by dissolving suitable amounts of the corresponding drug in methanol.</p>
<p>Dosage forms of ziprasidone for testing: Zeldox capsules (Pfizer Lb., Madrid, Spain) containing 20 mg of ziprasidone hydrochloride with lactose monohydrate, starch, magnesium stereate, gelatine, dioxidum of titanium (E171) and indigotine (E132) up to total capsule weight; Zeldox powder for injectable solution containing 20 mg of ziprasidone mesylate and sodium sulfobutylether-β-cyclo-dextrin up to total vial powder weight.</p></sec>
<sec>
<label>2.3.</label>
<title>Ion-Exchanger Preparation</title>
<p>The ion-exchanger (ZPD-TPB) was prepared by mixing 10 mL of 4 × 10<sup>−3</sup> M ZPD with 10 mL of a 4 × 10<sup>−3</sup> M NaTPB aqueous solution. The mixture was filtered through a porosity-4 sintered glass crucible. The residue was washed with 1:1 (v/v) methanol/water solution until no chloride ion was detected in the washing solution and was then dried at room temperature.</p></sec>
<sec>
<label>2.4.</label>
<title>Construction and Conditioning of the Electrode</title>
<p>The selected membrane was prepared by dissolving 3.0 mg of ZPD-TPB, 100 mg of PVC and 200 mg of NPOE in 3 mL of THF. This solution was poured into a Fluka glass ring (inner diameter 28 mm, height 30 mm) on a Fluka glass plate. The solution was allowed to evaporate overnight. A 7-mm diameter piece was cut out with a Fluka punch for ion-selective membranes and incorporated into a Fluka electrode body ISE containing 1 × 10<sup>−2</sup> M KCl and 8 × 10<sup>−5</sup> M ZPD saturated with AgCl as the internal filling solution. The electrode was conditioned by soaking with constant stirring in 8 × 10<sup>−5</sup> M ZPD working solution until the electrode gave a constant potential. When not in use, the electrode was kept immersed in the same solution.</p></sec>
<sec>
<label>2.5.</label>
<title>Calibration of the Electrode</title>
<p>The ZPD-selective electrode and reference electrode were immersed in any of the working ZPD solutions and the potential was measured, under constant stirring, until it gave a constant value. These potentials were then plotted <italic>vs.</italic> logarithmic values of ZPD concentration and the calibration parameters were calculated by fitting the calibration data to <xref ref-type="disp-formula" rid="FD1">Equation (1)</xref>:
<disp-formula id="FD1">
<label>(1)</label>
<mml:math display="block">
<mml:mrow>
<mml:mi>E</mml:mi>
<mml:mo>=</mml:mo>
<mml:msup>
<mml:mrow>
<mml:mi>E</mml:mi></mml:mrow>
<mml:mn>0</mml:mn></mml:msup>
<mml:mo>+</mml:mo>
<mml:mi mathvariant="italic">Slog</mml:mi>
<mml:mo stretchy="false">(</mml:mo>
<mml:mi mathvariant="italic">LOD</mml:mi>
<mml:mo>+</mml:mo>
<mml:msub>
<mml:mi>C</mml:mi>
<mml:mi mathvariant="italic">ZPD</mml:mi></mml:msub>
<mml:mo stretchy="false">)</mml:mo></mml:mrow></mml:math></disp-formula>where E is the potential of the cell, E<sup>0</sup> is the standard potential, S and LOD are the slope and the limit of detection of the electrode, respectively.</p>
<p>For the dynamic response studies a dynamic calibration of the electrode was made by adding, while stirring, adequate small volumes by micropipette of the ZPD standard solution to 25 mL of a 2:3 (v/v) methanol/pH buffer solution to cover the concentration range from 1.6 × 10<sup>−6</sup> to 1.6 × 10<sup>−4</sup> M.</p></sec>
<sec>
<label>2.6.</label>
<title>Procedure for the Determination of Ziprasidone in Pharmaceuticals</title>
<p>The ZPD content in capsules was determined by analyzing four capsules separately. The powder content of each one was shaken with 15 mL of methanol and then placed in an ultrasonic bath for 15 min and finally diluted with methanol in a 25 mL calibrated flask. A portion of this solution was centrifuged at 3,000 rpm for 5 min and the supernatant was filtered through a filter paper. Different volumes (500–1,000 μL) of the filtrate were taken, to which 15 mL of buffer were added and finally diluted to 25 mL with methanol in a calibrated flask. A similar procedure was followed in the case of ZPD powder for injectable solutions. The ZPD concentration in the sample was determined by comparing the potential value measured with the calibration graph.</p>
<p>Recovery studies were performed by adding different volumes (125–400 μL) of the ZPD standard solution to 1.0 mL of the pharmaceutical sample solution. Then, 15 mL of pH buffer were added and the solution was diluted to 25 mL with methanol in a calibrated flask. The samples were analyzed in triplicate using the procedure described above.</p></sec>
<sec>
<label>2.7.</label>
<title>Procedure for the Determination of Ziprasidone in Human Urine and Serum</title>
<p>For the determination of ZPD in human urine samples the sample pH was adjusted to 4.6 with HCl and NaOH solutions. Serum samples were first deproteinized by adding a volume of 8% trichloroacetic acid solution to the same volume of serum. Then, the solutions were shaken in a vortex mixer for 30 s, allowed to settle for 20 min and centrifuged for 10 min at 1,200 rpm. The clear supernatant layer was filtered through a 0.45 μm Millipore filter and the pH was adjusted to 4.5 by adding 2 M NaOH. In the absence of urine and serum samples containing ZPD, known amounts of ZPD were added. The potential was measured and compared with calibration graphs made in pooled urines and pooled serum, respectively.</p></sec></sec>
<sec sec-type="results|discussion">
<label>3.</label>
<title>Results and Discussion</title>
<sec>
<label>3.1.</label>
<title>Influence of Membrane Composition</title>
<p>Membranes of different compositions (<xref ref-type="table" rid="t1-sensors-11-08813">Table 1</xref>), prepared as described in the Experimental section, were tested to optimize the membrane composition of the ZPD-electrode. Two plasticizers with different dielectric constants were used as membrane solvent, NPOE (ɛ = 23.9) and DOS (ɛ = 2.4), membranes A and B, respectively. The corresponding calibration graphs for ZPD obtained in a 2:3 (v/v) methanol/pH buffer medium are shown in <xref ref-type="fig" rid="f1-sensors-11-08813">Figure 1</xref>. As can be seen, the membrane plasticized with NPOE showed higher potential span and lower detection limit, an effect that agrees with reports in the literature for other drug ISEs [<xref ref-type="bibr" rid="b42-sensors-11-08813">42</xref>,<xref ref-type="bibr" rid="b43-sensors-11-08813">43</xref>]. The difference observed between both membranes can be explained by the fact that the detection limit of ISEs based on dissolved ion-exchangers may be governed by the analyte ion activity present in the sample solution as a result of the distribution equilibrium of the ion-exchanger between the membrane and the solution [<xref ref-type="bibr" rid="b44-sensors-11-08813">44</xref>]. The solubility of the ion-exchanger in the membrane generally increases as the polarity and dielectric constant of the membrane increases [<xref ref-type="bibr" rid="b45-sensors-11-08813">45</xref>], which implies a lower concentration in the sample solution and therefore a lower detection limit. Taking these results into account NPOE was selected as plasticizer for further studies.</p>
<p>Two different ZPD-TPB ion exchanger concentrations in the membrane were tested (membranes A and C, respectively). As can be seen in <xref ref-type="fig" rid="f1-sensors-11-08813">Figure 1</xref>, the membrane containing the higher ion-exchanger concentration displayed a poorer potential response. Taking all this into account, membrane A was selected for further studies.</p></sec>
<sec>
<label>3.2.</label>
<title>Influence of the pH</title>
<p>The effect of pH on the electrode potential at various ZPD concentrations in the range 8.0 × 10<sup>−6</sup>–4.0 × 10<sup>−5</sup> M, was studied. The pH was varied by adding HCl or NaOH and the results are shown in <xref ref-type="fig" rid="f2-sensors-11-08813">Figure 2</xref>. As can be seen, the electrode potential was somewhat dependent on pH and therefore the use of a pH buffer was thought convenient. A pH of 4.6 adjusted with 2 × 10<sup>−1</sup> M sodium Ac<sup>−</sup>/AcH buffer solution was used for further studies.</p></sec>
<sec>
<label>3.3.</label>
<title>Influence of the Mixed Solvent Composition</title>
<p>Since ZPD is soluble in methanol and ethanol, the performance of the electrode was evaluated in hydroalcoholic media, <italic>i.e.</italic>, methanol-water and ethanol-water mixtures of different ratios. All of them contained buffer solution of pH 4.6. <xref ref-type="fig" rid="f3-sensors-11-08813">Figure 3</xref> shows the ZPD calibration graphs A-D determined in methanol/water 1:4, 2:3 and 3:2 (v/v), and ethanol/water 2:3 (v/v), respectively. By comparing the curves corresponding to different methanol contents, it is obvious that the best response was displayed when a methanol/water 2:3 (v/v) medium was used (curve B). When ethanol was used instead of methanol, the potential response was very poor. Accordingly to all this, a methanol/water 2:3 (v/v) medium was used for further studies.</p></sec>
<sec>
<label>3.4.</label>
<title>Validation of the Analytical Method</title>
<p>The validation of the analytical method for the determination of ziprasidone based on the present ZPD-selective electrode was made following recommendations of the literature [<xref ref-type="bibr" rid="b46-sensors-11-08813">46</xref>,<xref ref-type="bibr" rid="b47-sensors-11-08813">47</xref>].</p>
<sec sec-type="intro">
<label>3.4.1.</label>
<title>Response Characteristics</title>
<p>The slope and the limit of detection (LOD) of the selected electrode were determined by non-linear curve fitting of the corresponding calibration data to <xref ref-type="disp-formula" rid="FD1">Equation (1)</xref>. The values obtained for the range, linearity, slope, limit of detection and accuracy together with the response time of the electrode obtained from a dynamic calibration (<xref ref-type="fig" rid="f4-sensors-11-08813">Figure 4</xref>), are shown in <xref ref-type="table" rid="t2-sensors-11-08813">Table 2</xref>. As can be seen, there was a Nernstian response over almost two concentration decades, with a low detection limit and a rapid response, even at low ZPD concentrations, was obtained.</p></sec>
<sec>
<label>3.4.2.</label>
<title>Repeatability and Reproducibility</title>
<p>The repeatability and reproducibility of the calibration parameters were studied by making successive calibrations with the same membrane on the same day (n = 6), with the same membrane on different days (n = 6) and with different membranes (n = 3). The results obtained for the LOD were 1.8 × 10<sup>−5</sup> ± 2.1 × 10<sup>−6</sup> M, 2.2 × 10<sup>−5</sup> ± 7.5 × 10<sup>−6</sup> M and 2.8 × 10<sup>−5</sup> ± 1.1 × 10<sup>−5</sup> M, respectively, and for the slopes, 59.3 ± 1.8, 58.4 ± 2.7 and 59.2 ± 2.8 mV per decade of concentration, respectively. In all these aspects, a reasonably good reproducibility was obtained.</p></sec>
<sec>
<label>3.4.3.</label>
<title>Electrode Lifetime</title>
<p>The electrode lifetime was obtained by periodically performing calibration graphs for ZPD and calculating the response parameters. The ZPD selective electrode worked for at least 40 days, during which time no appreciable change in the calibration characteristics or response time was observed. After this time, the slope of the electrode started to decrease.</p></sec>
<sec>
<label>3.4.4.</label>
<title>Selectivity</title>
<p>The response of the ZPD-selective electrode was studied toward K<sup>+</sup>, NH<sub>4</sub><sup>+</sup>, Ca<sup>2+</sup>, Mg<sup>2+</sup>, glucose, lactose, saccharose, urea, uric acid, hipuric acid, amoxicillin, cimetidine, ofloxacin, diclofenac, tenoxicam, lansoprazole, loracepam, carbamazepine, fluoxetine, biperidene, haloperidol, and ranitidine. This list includes species frequently present in pharmaceuticals and biological fluids and drugs that could be administered at the same time as ZPD. Solutions of these compounds were prepared in methanol and the corresponding calibration graphs were carried out in the selected medium. The electrode did not respond to K<sup>+</sup>, NH<sub>4</sub><sup>+</sup>, Ca<sup>2+</sup>, Mg<sup>2+</sup>, glucose, lactose, saccharose, urea, uric acid, hipuric acid, amoxiciline, cimetidine, ofloxacin, diclofenac, tenoxicam, lansoprazole, loracepam, carbamazepine and ranitidine at concentrations below 3 × 10<sup>−4</sup> M. The electrode displayed a near Nernstian response to fluoxetine, haloperidol and biperidene and so the selectivity coefficients of these three species were determined by applying the separate solution method [<xref ref-type="bibr" rid="b48-sensors-11-08813">48</xref>]. The values obtained were 1.5, 0.5 and 2.4, respectively.</p></sec></sec>
<sec>
<label>3.5.</label>
<title>Analytical Applications</title>
<p>In order to demonstrate the applicability of the proposed method to the determination of ZPD, the method was applied to its determination in pharmaceuticals and human urine and serum. Two different pharmaceutical dosage forms of ZPD were analysed following the method described in the Experimental (<xref ref-type="table" rid="t3-sensors-11-08813">Table 3</xref>). In the absence of a reference method for the analytical determination of ZPD in the literature, recovery studies were carried out (<xref ref-type="table" rid="t3-sensors-11-08813">Table 3</xref>), providing quantitative recoveries between 99.3 and 101.7% in all the cases.</p>
<p>For the determination of ZPD in human urine, previous calibrations using different urines were made. No significant differences between the corresponding calibration parameters were found. Therefore, a calibration made using pooled urine samples was used for the determination of ZPD in different samples. A similar procedure was used for the determination of ZPD in serum. The results obtained are shown in <xref ref-type="table" rid="t4-sensors-11-08813">Table 4</xref>. Good recoveries were obtained for the determination of ZPD in both types of biological sample.</p>
<p>The systemic bioavailability of ZPD administered intramuscularly is 100% and 60% if it is administered orally. After a single dose intramuscular administration, the peak serum concentration typically occurs at about 60 minutes after the dose is administered, or earlier. In the case of an oral administration, the peak serum concentration is reached from 2 to 6 hours after the administration of therapeutic doses and approximately 44% of the total serum ZPD remains unchanged. The normal therapeutic dose of ZPD recommended for adults is between 40 and 160 mg/day and from 1 to 4% is excreted unchanged in the urine after its administration. Taking into account these data and the results obtained, it can be concluded that the determination of ZPD in human urine and serum is feasible using the present electrode being this method very useful for routine clinical analysis.</p></sec>
<sec>
<label>3.6.</label>
<title>Advantages of the Proposed Method</title>
<p>The potentiometric method proposed shows some advantages over the other electroanalytical method reported in the literature [<xref ref-type="bibr" rid="b36-sensors-11-08813">36</xref>] which makes use of voltammetric techniques. Thus, while the voltammetric method requires electrode polishing before each experiment, no electrode pre-treatment is needed in the present potentiometric method. Variations of the pH of the medium also have less influence in the potentiometric method, which is not affected by the presence of some other drugs and compounds frequently present in pharmaceuticals and biological fluids. Finally, while both methods are useful for pharmaceuticals and serum analysis, the potentiometric method was also applied to urine analysis.</p></sec></sec>
<sec sec-type="conclusions">
<label>4.</label>
<title>Conclusions</title>
<p>The presented ISE provides a rapid, sensitive, reproducible and precise method for the potentiometric determination of ziprasidone in mixed solvents. The new electrode could be very useful in pharmaceutical analysis in the absence of a existing reference method for ziprasidone determination. The electrode can also be used to determine ziprasidone in human urine and serum.</p></sec></body>
<back>
<ack>
<p>We gratefully acknowledge the Ministerio de Ciencia y Tecnología, Spain (Project CTQ2008-04806/BQU) for the financial support of this work. M.C. thanks University of Murcia for a grant.</p></ack>
<ref-list>
<title>References</title>
<ref id="b1-sensors-11-08813"><label>1.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Millard</surname><given-names>JW</given-names></name><name><surname>Alvarez-Núñez</surname><given-names>FA</given-names></name><name><surname>Yalkowsky</surname><given-names>SH</given-names></name></person-group><article-title>Solubilization by cosolvents. Establishing useful constants for the log-linear model</article-title><source>Int. J. Pharm</source><year>2002</year><volume>245</volume><fpage>153</fpage><lpage>166</lpage><pub-id pub-id-type="doi">10.1016/S0378-5173(02)00334-4</pub-id><pub-id pub-id-type="pmid">12270252</pub-id></citation></ref>
<ref id="b2-sensors-11-08813"><label>2.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Pungor</surname><given-names>E</given-names></name><name><surname>Toth</surname><given-names>K</given-names></name><name><surname>Klatsmanyl</surname><given-names>PG</given-names></name><name><surname>Izutsu</surname><given-names>K</given-names></name></person-group><article-title>Applications of ion-selective electrodes in nonaqueous and mixed-solvents</article-title><source>Pure Appl. Chem</source><year>1983</year><volume>55</volume><fpage>2029</fpage><lpage>2065</lpage><pub-id pub-id-type="doi">10.1351/pac198355122029</pub-id></citation></ref>
<ref id="b3-sensors-11-08813"><label>3.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Coetzee</surname><given-names>JF</given-names></name><name><surname>Deshmukh</surname><given-names>BK</given-names></name><name><surname>Liao</surname><given-names>CC</given-names></name></person-group><article-title>Applications of potentiometric ion sensors in the characterization of nonaqueous solvents</article-title><source>Chem. Rev</source><year>1990</year><volume>90</volume><fpage>827</fpage><lpage>835</lpage><pub-id pub-id-type="doi">10.1021/cr00103a009</pub-id></citation></ref>
<ref id="b4-sensors-11-08813"><label>4.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Jain</surname><given-names>AK</given-names></name><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Singh</surname><given-names>LP</given-names></name></person-group><article-title>Neutral carrier and organic resin based membranes as sensors for uranyl ions</article-title><source>Anal. Proc</source><year>1995</year><volume>32</volume><fpage>263</fpage><lpage>266</lpage></citation></ref>
<ref id="b5-sensors-11-08813"><label>5.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Jain</surname><given-names>AK</given-names></name><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Sahoo</surname><given-names>BN</given-names></name><name><surname>Singh</surname><given-names>LP</given-names></name></person-group><article-title>Copper(II)-selective electrodes based on macrocyclic compounds</article-title><source>Anal. Proc</source><year>1995</year><volume>32</volume><fpage>99</fpage><lpage>101</lpage></citation></ref>
<ref id="b6-sensors-11-08813"><label>6.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Srivastava</surname><given-names>SK</given-names></name><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Jain</surname><given-names>S</given-names></name></person-group><article-title>A PVC-Based Benzo-15-Crown-5 membrane sensor for cadmium</article-title><source>Electroanalysis</source><year>1996</year><volume>8</volume><fpage>938</fpage><lpage>940</lpage><pub-id pub-id-type="doi">10.1002/elan.1140081017</pub-id></citation></ref>
<ref id="b7-sensors-11-08813"><label>7.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Srivastava</surname><given-names>SK</given-names></name><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Jain</surname><given-names>S</given-names></name></person-group><article-title>PVC-based 2,2,2-cryptand sensor for zinc ions</article-title><source>Anal. Chem</source><year>1996</year><volume>68</volume><fpage>1272</fpage><lpage>1275</lpage><pub-id pub-id-type="doi">10.1021/ac9507000</pub-id><pub-id pub-id-type="pmid">21619161</pub-id></citation></ref>
<ref id="b8-sensors-11-08813"><label>8.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Jain</surname><given-names>AK</given-names></name><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Singh</surname><given-names>LP</given-names></name><name><surname>Khurana</surname><given-names>U</given-names></name></person-group><article-title>Macrocycle based membrane sensors for the determination of cobalt(II) ions</article-title><source>Analyst</source><year>1997</year><volume>122</volume><fpage>583</fpage><lpage>586</lpage><pub-id pub-id-type="doi">10.1039/a608421d</pub-id></citation></ref>
<ref id="b9-sensors-11-08813"><label>9.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Jain</surname><given-names>AK</given-names></name><name><surname>Singh</surname><given-names>LP</given-names></name><name><surname>Khurana</surname><given-names>U</given-names></name></person-group><article-title>Porphyrins as carrier in PVC based membrane potentiometric sensors for nickel(II)</article-title><source>Anal. Chim. Acta</source><year>1997</year><volume>355</volume><fpage>33</fpage><lpage>41</lpage><pub-id pub-id-type="doi">10.1016/S0003-2670(97)81609-1</pub-id></citation></ref>
<ref id="b10-sensors-11-08813"><label>10.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Mangla</surname><given-names>R</given-names></name><name><surname>Khurana</surname><given-names>U</given-names></name><name><surname>Kumar</surname><given-names>P</given-names></name></person-group><article-title>Determination of uranyl ions using poly(vinyl chloride) based 4-tert-butylcalix[6]arene membrane sensor</article-title><source>Electroanalysis</source><year>1999</year><volume>11</volume><fpage>573</fpage><lpage>576</lpage><pub-id pub-id-type="doi">10.1002/(SICI)1521-4109(199906)11:8&lt;573::AID-ELAN573&gt;3.0.CO;2-Z</pub-id></citation></ref>
<ref id="b11-sensors-11-08813"><label>11.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Singh</surname><given-names>AK</given-names></name><name><surname>Panwar</surname><given-names>A</given-names></name><name><surname>Kumar</surname><given-names>S</given-names></name><name><surname>Baniwal</surname><given-names>S</given-names></name></person-group><article-title>Chromium(III)-selective electrode based on a macrocyclic compound</article-title><source>Analyst</source><year>1999</year><volume>124</volume><fpage>521</fpage><lpage>525</lpage><pub-id pub-id-type="doi">10.1039/a901162e</pub-id></citation></ref>
<ref id="b12-sensors-11-08813"><label>12.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Kumar</surname><given-names>P</given-names></name></person-group><article-title>Cadmium(II)-selective sensors based on dibenzo-24-crown-8 in PVC matrix</article-title><source>Anal. Chim. Acta</source><year>1999</year><volume>389</volume><fpage>205</fpage><lpage>212</lpage><pub-id pub-id-type="doi">10.1016/S0003-2670(99)00154-3</pub-id></citation></ref>
<ref id="b13-sensors-11-08813"><label>13.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Mangla</surname><given-names>R</given-names></name><name><surname>Agarwal</surname><given-names>S</given-names></name></person-group><article-title>Pb(II) selective potentiometric sensor based on 4-tert-butylcalix[4]arene in PVC matrix</article-title><source>Electroanalysis</source><year>2002</year><volume>14</volume><fpage>1127</fpage><lpage>1132</lpage><pub-id pub-id-type="doi">10.1002/1521-4109(200208)14:15/16&lt;1127::AID-ELAN1127&gt;3.0.CO;2-7</pub-id></citation></ref>
<ref id="b14-sensors-11-08813"><label>14.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mittal</surname><given-names>SK</given-names></name><name><surname>Sharma</surname><given-names>HK</given-names></name><name><surname>Kumar</surname><given-names>ASK</given-names></name></person-group><article-title>Samarium (III) selective membrane sensor based on tin (IV) boratophosphate</article-title><source>Sensors</source><year>2004</year><volume>4</volume><fpage>125</fpage><lpage>135</lpage><pub-id pub-id-type="doi">10.3390/s40800125</pub-id></citation></ref>
<ref id="b15-sensors-11-08813"><label>15.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mittal</surname><given-names>SK</given-names></name><name><surname>Kumar</surname><given-names>ASK</given-names></name><name><surname>Sharma</surname><given-names>HK</given-names></name></person-group><article-title>PVC-based dicyclohexano-18-crown-6 sensor for La(III) ions</article-title><source>Talanta</source><year>2004</year><volume>62</volume><fpage>801</fpage><lpage>805</lpage><pub-id pub-id-type="doi">10.1016/j.talanta.2003.10.001</pub-id><pub-id pub-id-type="pmid">18969365</pub-id></citation></ref>
<ref id="b16-sensors-11-08813"><label>16.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Moghimi</surname><given-names>M</given-names></name><name><surname>Bagherinia</surname><given-names>MA</given-names></name><name><surname>Arvand</surname><given-names>M</given-names></name><name><surname>Zanjanchi</surname><given-names>MA</given-names></name></person-group><article-title>Polymeric membrane sensor for potentiometric determination of vanadyl ions</article-title><source>Anal. Chim. Acta</source><year>2004</year><volume>527</volume><fpage>169</fpage><lpage>175</lpage><pub-id pub-id-type="doi">10.1016/j.aca.2004.06.059</pub-id></citation></ref>
<ref id="b17-sensors-11-08813"><label>17.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Jain</surname><given-names>AK</given-names></name><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Ganeshpure</surname><given-names>PA</given-names></name><name><surname>Raisoni</surname><given-names>JR</given-names></name></person-group><article-title>Ni(II)-selective ion sensors of salen type Schiff base chelates</article-title><source>Anal. Chim. Acta</source><year>2005</year><volume>553</volume><fpage>177</fpage><lpage>184</lpage><pub-id pub-id-type="doi">10.1016/j.aca.2005.08.016</pub-id></citation></ref>
<ref id="b18-sensors-11-08813"><label>18.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Arvand</surname><given-names>M</given-names></name><name><surname>Moghimi</surname><given-names>AM</given-names></name><name><surname>Afshari</surname><given-names>A</given-names></name><name><surname>Mahmoodi</surname><given-names>N</given-names></name></person-group><article-title>Potentiometric membrane sensor based on 6-(4-nitrophenyl)-2,4-diphenyl-3,5-diaza-bicyclo[3.1.0]hex-2-ene for detection of Sn(II) in real samples</article-title><source>Anal. Chim. Acta</source><year>2006</year><volume>579</volume><fpage>102</fpage><lpage>108</lpage><pub-id pub-id-type="doi">10.1016/j.aca.2006.07.022</pub-id><pub-id pub-id-type="pmid">17723734</pub-id></citation></ref>
<ref id="b19-sensors-11-08813"><label>19.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Jain</surname><given-names>AK</given-names></name><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Singh</surname><given-names>LP</given-names></name><name><surname>Raisoni</surname><given-names>JR</given-names></name></person-group><article-title>A comparative study of Pb<sup>2+</sup> selective sensors based on derivatized tetrapyrazole and calix[4]arene receptors</article-title><source>Electrochim. Acta</source><year>2006</year><volume>51</volume><fpage>2547</fpage><lpage>2553</lpage><pub-id pub-id-type="doi">10.1016/j.electacta.2005.07.040</pub-id></citation></ref>
<ref id="b20-sensors-11-08813"><label>20.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Singh</surname><given-names>AK</given-names></name><name><surname>Gupta</surname><given-names>B</given-names></name></person-group><article-title>A cerium(III) selective polyvinyl chloride membrane sensor based on a Schiff base complex of N,N′-bis[2-(salicylideneamino)ethyl]ethane-1,2-diamine</article-title><source>Anal. Chim. Acta</source><year>2006</year><volume>575</volume><fpage>198</fpage><lpage>204</lpage><pub-id pub-id-type="doi">10.1016/j.aca.2006.05.090</pub-id><pub-id pub-id-type="pmid">17723591</pub-id></citation></ref>
<ref id="b21-sensors-11-08813"><label>21.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Singh</surname><given-names>AK</given-names></name><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Gupta</surname><given-names>B</given-names></name></person-group><article-title>Chromium(III) selective membrane sensors based on Schiff bases as chelating ionophores</article-title><source>Anal. Chim. Acta</source><year>2007</year><volume>585</volume><fpage>171</fpage><lpage>178</lpage><pub-id pub-id-type="doi">10.1016/j.aca.2006.11.074</pub-id><pub-id pub-id-type="pmid">17386662</pub-id></citation></ref>
<ref id="b22-sensors-11-08813"><label>22.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Singh</surname><given-names>AK</given-names></name><name><surname>Gupta</surname><given-names>B</given-names></name></person-group><article-title>Schiff bases as cadmium(II) selective ionophores in polymeric membrane electrodes</article-title><source>Anal. Chim. Acta</source><year>2007</year><volume>583</volume><fpage>340</fpage><lpage>348</lpage><pub-id pub-id-type="doi">10.1016/j.aca.2006.10.039</pub-id><pub-id pub-id-type="pmid">17386565</pub-id></citation></ref>
<ref id="b23-sensors-11-08813"><label>23.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Jain</surname><given-names>AK</given-names></name><name><surname>Ishtaiwi</surname><given-names>Z</given-names></name><name><surname>Lang</surname><given-names>H</given-names></name><name><surname>Maheshwari</surname><given-names>G</given-names></name></person-group><article-title>Ni<sup>2+</sup> selective sensors based on meso-tetrakis-{4-[tris-(4-allyldimethylsilylphenyl)silyl]phenyl}porphyrin and (sal)(2)triene in poly(vinyl chloride) matrix</article-title><source>Talanta</source><year>2007</year><volume>73</volume><fpage>803</fpage><lpage>811</lpage><pub-id pub-id-type="doi">10.1016/j.talanta.2007.04.049</pub-id><pub-id pub-id-type="pmid">19073105</pub-id></citation></ref>
<ref id="b24-sensors-11-08813"><label>24.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Singh</surname><given-names>AK</given-names></name><name><surname>Pal</surname><given-names>MK</given-names></name></person-group><article-title>Ni(II) selective sensors based on Schiff bases membranes in poly(vinyl chloride)</article-title><source>Anal. Chim. Acta</source><year>2008</year><volume>624</volume><fpage>223</fpage><lpage>231</lpage><pub-id pub-id-type="doi">10.1016/j.aca.2008.06.054</pub-id><pub-id pub-id-type="pmid">18706328</pub-id></citation></ref>
<ref id="b25-sensors-11-08813"><label>25.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Nakamura</surname><given-names>T</given-names></name></person-group><article-title>Development and application of ion-selective electrodes in nonaqueous solutions</article-title><source>Anal. Sci</source><year>2009</year><volume>25</volume><fpage>33</fpage><lpage>40</lpage><pub-id pub-id-type="doi">10.2116/analsci.25.33</pub-id><pub-id pub-id-type="pmid">19139570</pub-id></citation></ref>
<ref id="b26-sensors-11-08813"><label>26.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Ludwigb</surname><given-names>R</given-names></name><name><surname>Agarwal</surname><given-names>S</given-names></name></person-group><article-title>Anion recognition through modified calixarenes: A highly selective sensor for monohydrogen phosphate</article-title><source>Anal. Chim. Acta</source><year>2005</year><volume>538</volume><fpage>213</fpage><lpage>218</lpage><pub-id pub-id-type="doi">10.1016/j.aca.2005.02.008</pub-id></citation></ref>
<ref id="b27-sensors-11-08813"><label>27.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Jaina</surname><given-names>AK</given-names></name><name><surname>Gupta</surname><given-names>VK</given-names></name><name><surname>Singhb</surname><given-names>LP</given-names></name><name><surname>Srivastavac</surname><given-names>P</given-names></name><name><surname>Raisonia</surname><given-names>JR</given-names></name></person-group><article-title>Anion recognition through novel C-thiophenecalix[4]resorcinarene: PVC based sensor for chromate ions</article-title><source>Talanta</source><year>2005</year><volume>65</volume><fpage>716</fpage><lpage>721</lpage><pub-id pub-id-type="doi">10.1016/j.talanta.2004.07.041</pub-id><pub-id pub-id-type="pmid">18969858</pub-id></citation></ref>
<ref id="b28-sensors-11-08813"><label>28.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Patil</surname><given-names>SR</given-names></name><name><surname>Rakshit</surname><given-names>AK</given-names></name></person-group><article-title>Membrane electrode sensitive to a cationic surfactant in aquo-organic media</article-title><source>Anal. Chim. Acta</source><year>2004</year><volume>518</volume><fpage>87</fpage><lpage>91</lpage><pub-id pub-id-type="doi">10.1016/j.aca.2004.05.017</pub-id></citation></ref>
<ref id="b29-sensors-11-08813"><label>29.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Dowle</surname><given-names>CJ</given-names></name><name><surname>Cooksey</surname><given-names>BG</given-names></name><name><surname>Mottaway</surname><given-names>J</given-names></name><name><surname>Campbell</surname><given-names>WC</given-names></name></person-group><article-title>Development of ion-selective electrodes for use in the titration of ionic surfactants in mixed-solvent systems</article-title><source>Analyst</source><year>1987</year><volume>112</volume><fpage>1299</fpage><lpage>1302</lpage><pub-id pub-id-type="doi">10.1039/an9871201299</pub-id></citation></ref>
<ref id="b30-sensors-11-08813"><label>30.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Fort</surname><given-names>J</given-names></name></person-group><article-title>Celecoxib, a cox-2-specific inhibitor: The clinical data</article-title><source>Am. J. Orthop</source><year>1999</year><volume>28</volume><fpage>13</fpage><lpage>18</lpage><pub-id pub-id-type="pmid">10193998</pub-id></citation></ref>
<ref id="b31-sensors-11-08813"><label>31.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Reddy</surname><given-names>P</given-names></name><name><surname>Rani</surname><given-names>B</given-names></name><name><surname>Babu</surname><given-names>G</given-names></name><name><surname>Rao</surname><given-names>J</given-names></name></person-group><article-title>Spectrophotometric methods for the determination of ziprasidone hydrochloride monohydrate in bulk and capsules</article-title><source>Acta Ciencia Indica Chem</source><year>2005</year><volume>31</volume><fpage>223</fpage><lpage>225</lpage></citation></ref>
<ref id="b32-sensors-11-08813"><label>32.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Srinubabu</surname><given-names>G</given-names></name><name><surname>Rani</surname><given-names>B</given-names></name><name><surname>Rao</surname><given-names>JS</given-names></name></person-group><article-title>Spectrophotometric determination of Ziprasidone in pharmaceutical formulations</article-title><source>E-J. Chem</source><year>2006</year><volume>3</volume><fpage>9</fpage><lpage>12</lpage><pub-id pub-id-type="doi">10.1155/2006/643624</pub-id></citation></ref>
<ref id="b33-sensors-11-08813"><label>33.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Chauhan</surname><given-names>C</given-names></name><name><surname>Choudhury</surname><given-names>P</given-names></name></person-group><article-title>UV spectrophotometric determination of ziprasidone hydrochloride in pore and pharmaceutical formulation</article-title><source>Asian J. Chem</source><year>2007</year><volume>19</volume><fpage>819</fpage><lpage>820</lpage></citation></ref>
<ref id="b34-sensors-11-08813"><label>34.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>El-Sherif</surname><given-names>ZA</given-names></name><name><surname>El-Zeany</surname><given-names>B</given-names></name><name><surname>El-Houssini</surname><given-names>OM</given-names></name><name><surname>Rashed</surname><given-names>MS</given-names></name><name><surname>Aboul-Enein</surname><given-names>HY</given-names></name></person-group><article-title>Stability indicating reversed-phase high-performance liquid chromatographic and thin layer densitometric methods for the determination of ziprasidone in bulk powder and in pharmaceutical formulations</article-title><source>Biomed. Chromatogr</source><year>2004</year><volume>18</volume><fpage>143</fpage><lpage>149</lpage><pub-id pub-id-type="doi">10.1002/bmc.299</pub-id><pub-id pub-id-type="pmid">15103698</pub-id></citation></ref>
<ref id="b35-sensors-11-08813"><label>35.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Farin</surname><given-names>C</given-names></name><name><surname>Kremser</surname><given-names>L</given-names></name><name><surname>Raggi</surname><given-names>M</given-names></name><name><surname>Kenndler</surname><given-names>E</given-names></name></person-group><article-title>Determination of ziprasidone in pharmaceutical formulations by capillary zone electrophoresis</article-title><source>J. Pharm. Biomed. Anal</source><year>2008</year><volume>46</volume><fpage>471</fpage><lpage>476</lpage><pub-id pub-id-type="doi">10.1016/j.jpba.2007.11.010</pub-id><pub-id pub-id-type="pmid">18155379</pub-id></citation></ref>
<ref id="b36-sensors-11-08813"><label>36.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Kul</surname><given-names>D</given-names></name><name><surname>Gumustas</surname><given-names>M</given-names></name><name><surname>Uslu</surname><given-names>B</given-names></name><name><surname>Ozkan</surname><given-names>SA</given-names></name></person-group><article-title>Electroanalytical characteristics of antipsychotic drug ziprasidone and its determination in pharmaceuticals and serum samples on solid electrodes</article-title><source>Talanta</source><year>2010</year><volume>82</volume><fpage>286</fpage><lpage>295</lpage><pub-id pub-id-type="doi">10.1016/j.talanta.2010.04.036</pub-id><pub-id pub-id-type="pmid">20685469</pub-id></citation></ref>
<ref id="b37-sensors-11-08813"><label>37.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Janiszewski</surname><given-names>JS</given-names></name><name><surname>Fouda</surname><given-names>HG</given-names></name><name><surname>Cole</surname><given-names>RO</given-names></name></person-group><article-title>Development and validation of a high-sensitivity assay for an antipsychotic agent, CP-88,059, with solid-phase extraction and narrow-bore highperformance liquid chromatography</article-title><source>J. Chromatogr. B</source><year>1995</year><volume>668</volume><fpage>133</fpage><lpage>139</lpage><pub-id pub-id-type="doi">10.1016/0378-4347(95)00071-P</pub-id></citation></ref>
<ref id="b38-sensors-11-08813"><label>38.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Suckow</surname><given-names>RF</given-names></name><name><surname>Fein</surname><given-names>M</given-names></name><name><surname>Correll</surname><given-names>CU</given-names></name><name><surname>Cooper</surname><given-names>TB</given-names></name></person-group><article-title>Determination of ziprasidone using liquid chromatography with fluorescence detection</article-title><source>J. Chromatogr. B</source><year>2004</year><volume>799</volume><fpage>201</fpage><lpage>208</lpage><pub-id pub-id-type="doi">10.1016/j.jchromb.2003.10.027</pub-id></citation></ref>
<ref id="b39-sensors-11-08813"><label>39.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Sachse</surname><given-names>J</given-names></name><name><surname>Haertter</surname><given-names>S</given-names></name><name><surname>Hiemke</surname><given-names>C</given-names></name></person-group><article-title>Automated determination of ziprasidone by HPLC with column switching and spectrophotometric detection</article-title><source>Ther. Drug Monit</source><year>2005</year><volume>27</volume><fpage>158</fpage><lpage>162</lpage><pub-id pub-id-type="doi">10.1097/01.ftd.0000150879.36296.4d</pub-id><pub-id pub-id-type="pmid">15795645</pub-id></citation></ref>
<ref id="b40-sensors-11-08813"><label>40.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Aldirbashi</surname><given-names>O</given-names></name><name><surname>Aboul-Enein</surname><given-names>H</given-names></name><name><surname>Alodaib</surname><given-names>A</given-names></name><name><surname>Jacob</surname><given-names>M</given-names></name><name><surname>Rashed</surname><given-names>M</given-names></name></person-group><article-title>Rapid liquid chromatography-tandem mass spectrometry method for quantification of ziprasidone in human plasma</article-title><source>Biomed. Chromatogr</source><year>2006</year><volume>20</volume><fpage>365</fpage><lpage>368</lpage><pub-id pub-id-type="doi">10.1002/bmc.571</pub-id><pub-id pub-id-type="pmid">16167302</pub-id></citation></ref>
<ref id="b41-sensors-11-08813"><label>41.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Aravagiri</surname><given-names>M</given-names></name><name><surname>Marder</surname><given-names>S</given-names></name><name><surname>Pollock</surname><given-names>B</given-names></name></person-group><article-title>Determination of ziprasidone in human plasma by liquid chromatography-electrospray tandem mass spectrometry and its application to plasma level determination in schizophrenia patients</article-title><source>J. Chromatogr. B</source><year>2007</year><volume>847</volume><fpage>237</fpage><lpage>244</lpage><pub-id pub-id-type="doi">10.1016/j.jchromb.2006.10.024</pub-id></citation></ref>
<ref id="b42-sensors-11-08813"><label>42.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>García</surname><given-names>MS</given-names></name><name><surname>Ortuño</surname><given-names>JA</given-names></name><name><surname>Albero</surname><given-names>MI</given-names></name><name><surname>Cuartero</surname><given-names>M</given-names></name></person-group><article-title>Application of a trazodone-selective electrode to pharmaceutical quality control and urine analyses</article-title><source>Anal. Bioanal. Chem</source><year>2009</year><volume>394</volume><fpage>1563</fpage><lpage>1567</lpage><pub-id pub-id-type="doi">10.1007/s00216-009-2699-7</pub-id><pub-id pub-id-type="pmid">19238364</pub-id></citation></ref>
<ref id="b43-sensors-11-08813"><label>43.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Hassouna</surname><given-names>MEM</given-names></name><name><surname>Elsuccary</surname><given-names>SAA</given-names></name></person-group><article-title>PVC membrane electrode for the potentiometric determination of Ipratropium bromide using batch and flow injection techniques</article-title><source>Talanta</source><year>2008</year><volume>75</volume><fpage>1175</fpage><lpage>1183</lpage><pub-id pub-id-type="doi">10.1016/j.talanta.2008.01.014</pub-id><pub-id pub-id-type="pmid">18585199</pub-id></citation></ref>
<ref id="b44-sensors-11-08813"><label>44.</label><citation citation-type="book"><person-group person-group-type="author"><name><surname>Koryta</surname><given-names>J</given-names></name><name><surname>Stulik</surname><given-names>K</given-names></name></person-group><source>Ion-Selective Electrodes</source><edition>2nd ed</edition><publisher-name>Cambridge University Press</publisher-name><publisher-loc>Cambridge, UK</publisher-loc><year>1983</year></citation></ref>
<ref id="b45-sensors-11-08813"><label>45.</label><citation citation-type="book"><person-group person-group-type="author"><name><surname>Senike</surname><given-names>T</given-names></name><name><surname>Hasegawa</surname><given-names>Y</given-names></name></person-group><source>Solvent Extraction Chemistry Fundamentals and Applications</source><publisher-name>Marcel Dekker</publisher-name><publisher-loc>New York, NY, USA</publisher-loc><year>1977</year></citation></ref>
<ref id="b46-sensors-11-08813"><label>46.</label><citation citation-type="book"><person-group person-group-type="author"><name><surname>Swartz</surname><given-names>ME</given-names></name><name><surname>Krull</surname><given-names>IS</given-names></name></person-group><source>Analytical Method Development and Validation</source><publisher-name>Marcel Dekker</publisher-name><publisher-loc>New York, NY, USA</publisher-loc><year>1997</year></citation></ref>
<ref id="b47-sensors-11-08813"><label>47.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Gumustas</surname><given-names>M</given-names></name><name><surname>Ozkan</surname><given-names>SA</given-names></name></person-group><article-title>The role of and the place of method validation in drug analysis using electroanalytical techniques</article-title><source>Open Anal. Chem. J</source><year>2011</year><volume>5</volume><fpage>1</fpage><lpage>21</lpage></citation></ref>
<ref id="b48-sensors-11-08813"><label>48.</label><citation citation-type="journal"><person-group person-group-type="author"><name><surname>Umezawa</surname><given-names>Y</given-names></name><name><surname>Bühlmann</surname><given-names>P</given-names></name><name><surname>Umezawa</surname><given-names>K</given-names></name><name><surname>Tohda</surname><given-names>K</given-names></name><name><surname>Amemiya</surname><given-names>S</given-names></name></person-group><article-title>Selectivity coefficients of ion-selective electrodes</article-title><source>Pure Appl. Chem</source><year>2000</year><volume>72</volume><fpage>1851</fpage><lpage>2082</lpage><pub-id pub-id-type="doi">10.1351/pac200072101851</pub-id></citation></ref></ref-list>
<sec sec-type="display-objects">
<title>Figures and Tables</title>
<fig id="f1a-sensors-11-08813" position="float">
<label>Figure 1.</label>
<caption>
<p>Ziprasidone.</p></caption>
<graphic xlink:href="sensors-11-08813f1a.gif"/></fig>
<fig id="f1-sensors-11-08813" position="float">
<label>Figure 1.</label>
<caption>
<p>Calibration graphs of ziprasidone obtained with membranes A, B and C.</p></caption>
<graphic xlink:href="sensors-11-08813f1.gif"/></fig>
<fig id="f2-sensors-11-08813" position="float">
<label>Figure 2.</label>
<caption>
<p>Influence of the pH on the electrode potential for different ziprasidone concentrations.</p></caption>
<graphic xlink:href="sensors-11-08813f2.gif"/></fig>
<fig id="f3-sensors-11-08813" position="float">
<label>Figure 3.</label>
<caption>
<p>Calibration graphs in different media. (<bold>A</bold>) 1:4 (v/v) methanol/water; (<bold>B</bold>) 2:3 (v/v) methanol/water; (<bold>C</bold>) 3:2 (v/v) methanol/water; and (<bold>D</bold>) 2:3 (v/v) ethanol/water.</p></caption>
<graphic xlink:href="sensors-11-08813f3.gif"/></fig>
<fig id="f4-sensors-11-08813" position="float">
<label>Figure 4.</label>
<caption>
<p>Dynamic response of the electrode for ziprasidone concentrations of (1) 0 M; (2) 1.6 × 10<sup>−6</sup>; (3) 3.2 × 10<sup>−6</sup> M; (4) 7.2 × 10<sup>−6</sup> M; (5) 1.1 × 10<sup>−5</sup> M; (6) 1.9 × 10<sup>−5</sup> M; (7) 2.7 × 10<sup>−5</sup> M; (8) 4.3 × 10<sup>−5</sup> M; (9) 5.8 × 10<sup>−5</sup> M; (10) 9.7 × 10<sup>−5</sup> M; and (11) 1.6 × 10<sup>−4</sup> M.</p></caption>
<graphic xlink:href="sensors-11-08813f4.gif"/></fig>
<table-wrap id="t1-sensors-11-08813" position="float">
<label>Table 1.</label>
<caption>
<p>Composition of the membranes assayed.</p></caption>
<table frame="hsides" rules="groups">
<thead>
<tr>
<th align="center" valign="bottom"/>
<th colspan="4" align="center" valign="bottom"><bold>Percentage (w/w) of components in membranes</bold></th></tr>
<tr>
<th colspan="5" align="center" valign="bottom">
<hr/></th></tr>
<tr>
<th align="center" valign="bottom"><bold>Membrane</bold></th>
<th align="center" valign="bottom">PVC</th>
<th align="center" valign="bottom">NPOE</th>
<th align="center" valign="bottom">DOS</th>
<th align="center" valign="bottom">ZPD-TPB</th></tr></thead>
<tbody>
<tr>
<td align="center" valign="bottom">A</td>
<td align="center" valign="bottom">33.0</td>
<td align="center" valign="bottom">66.0</td>
<td align="center" valign="bottom"/>
<td align="center" valign="bottom">1.0</td></tr>
<tr>
<td align="center" valign="bottom">B</td>
<td align="center" valign="bottom">33.0</td>
<td align="center" valign="bottom"/>
<td align="center" valign="bottom">66.0</td>
<td align="center" valign="bottom">1.0</td></tr>
<tr>
<td align="center" valign="bottom">C</td>
<td align="center" valign="bottom">32.3</td>
<td align="center" valign="bottom">64.7</td>
<td align="center" valign="bottom"/>
<td align="center" valign="bottom">3.0</td></tr></tbody></table></table-wrap>
<table-wrap id="t2-sensors-11-08813" position="float">
<label>Table 2.</label>
<caption>
<p>Response characteristics of the ziprasidone selective electrode.</p></caption>
<table frame="hsides" rules="groups">
<thead>
<tr>
<th align="center" valign="bottom"><bold>Parameter</bold></th>
<th align="center" valign="bottom"><bold>Value</bold></th></tr></thead>
<tbody>
<tr>
<td align="left" valign="bottom">Range (M)</td>
<td align="center" valign="bottom">1.6 × 10<sup>−6</sup>–1.6 × 10<sup>−4</sup></td></tr>
<tr>
<td align="left" valign="bottom">Linearity (M)</td>
<td align="center" valign="bottom">2 × 10<sup>−5</sup>–8 × 10<sup>−4</sup></td></tr>
<tr>
<td align="left" valign="bottom">Slope (mV decade<sup>−1</sup>) ± RSD <xref ref-type="table-fn" rid="tfn1-sensors-11-08813"><sup>a</sup></xref></td>
<td align="center" valign="bottom">59.3 ± 1.8</td></tr>
<tr>
<td align="left" valign="bottom">Limit of detection (M) ± RSD <xref ref-type="table-fn" rid="tfn1-sensors-11-08813"><sup>a</sup></xref></td>
<td align="center" valign="bottom">1.8 × 10<sup>−5</sup> ± 2.1 × 10<sup>−6</sup></td></tr>
<tr>
<td align="left" valign="bottom">Accuracy <xref ref-type="table-fn" rid="tfn3-sensors-11-08813"><sup>c</sup></xref> (%)</td>
<td align="center" valign="bottom">99.38 ± 0.37</td></tr>
<tr>
<td align="left" valign="bottom">Response time (s) <xref ref-type="table-fn" rid="tfn2-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="bottom">t<sub>95%</sub> ≤ 15</td></tr>
<tr>
<td align="left" valign="bottom">Working pH range <xref ref-type="table-fn" rid="tfn2-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="bottom">3.5–5.5</td></tr>
<tr>
<td align="left" valign="bottom">Operational lifetime (days)</td>
<td align="center" valign="bottom">≥ 40</td></tr></tbody></table>
<table-wrap-foot><fn id="tfn1-sensors-11-08813">
<label>a</label>
<p>mean of six calibrations.</p></fn><fn id="tfn2-sensors-11-08813">
<label>b</label>
<p>ZPD concentration range of 1.6 × 10<sup>−6</sup>–1.6 × 10<sup>−4</sup> M.</p></fn><fn id="tfn3-sensors-11-08813">
<label>c</label>
<p>Recovery ± RSD (n = 10).</p></fn></table-wrap-foot></table-wrap>
<table-wrap id="t3-sensors-11-08813" position="float">
<label>Table 3.</label>
<caption>
<p>Determination of ziprasidone in pharmaceuticals.</p></caption>
<table frame="hsides" rules="groups">
<thead>
<tr>
<th align="center" valign="middle" rowspan="2"><bold>Sample</bold></th>
<th align="center" valign="middle" rowspan="2"><bold>Labelled<xref ref-type="table-fn" rid="tfn4-sensors-11-08813"><sup>a</sup></xref></bold></th>
<th align="center" valign="middle" rowspan="2"><bold>Amount found<xref ref-type="table-fn" rid="tfn4-sensors-11-08813"><sup>a</sup></xref></bold></th>
<th align="center" valign="middle" rowspan="2"><bold>% Bias</bold></th>
<th colspan="4" align="center" valign="middle"><bold>Recovery studies</bold>
<hr/></th></tr>
<tr>
<th align="center" valign="middle"><bold>Added<xref ref-type="table-fn" rid="tfn4-sensors-11-08813"><sup>a</sup></xref></bold></th>
<th align="center" valign="middle"><bold>Found<xref ref-type="table-fn" rid="tfn4-sensors-11-08813"><sup>a</sup></xref></bold></th>
<th align="center" valign="middle"><bold>Recovery<xref ref-type="table-fn" rid="tfn6-sensors-11-08813"><sup>c</sup></xref> (%)</bold></th>
<th align="center" valign="middle"><bold>% Bias</bold></th></tr></thead>
<tbody>
<tr>
<td align="center" valign="middle" rowspan="3">Zeldox capsules</td>
<td align="center" valign="middle" rowspan="3">20.00</td>
<td align="center" valign="middle" rowspan="3">20.02 ± 0.54 <xref ref-type="table-fn" rid="tfn5-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="middle" rowspan="3">0.10</td>
<td align="center" valign="middle">5.86</td>
<td align="center" valign="middle">25.82 <xref ref-type="table-fn" rid="tfn5-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="middle">99.3 ± 0.5</td>
<td align="center" valign="middle">−0.68</td></tr>
<tr>
<td align="center" valign="middle">11.73</td>
<td align="center" valign="middle">31.93 <xref ref-type="table-fn" rid="tfn5-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="middle">101.7 ± 4.1</td>
<td align="center" valign="middle">1.70</td></tr>
<tr>
<td align="center" valign="middle">18.70</td>
<td align="center" valign="middle">38.68 <xref ref-type="table-fn" rid="tfn5-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="middle">99.9 ± 3.5</td>
<td align="center" valign="middle">−0.11</td></tr>
<tr>
<td colspan="8" align="left" valign="middle">
<hr/></td></tr>
<tr>
<td align="center" valign="middle" rowspan="3">Zeldox injectables</td>
<td align="center" valign="middle" rowspan="3">20.00</td>
<td align="center" valign="middle" rowspan="3">20.39 ± 1.09 <xref ref-type="table-fn" rid="tfn5-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="middle" rowspan="3">1.90</td>
<td align="center" valign="middle">5.86</td>
<td align="center" valign="middle">25.87 <xref ref-type="table-fn" rid="tfn5-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="middle">100.2 ± 0.7</td>
<td align="center" valign="middle">0.17</td></tr>
<tr>
<td align="center" valign="middle">11.73</td>
<td align="center" valign="middle">31.80 <xref ref-type="table-fn" rid="tfn5-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="middle">100.6 ± 0.2</td>
<td align="center" valign="middle">0.59</td></tr>
<tr>
<td align="center" valign="middle">18.70</td>
<td align="center" valign="middle">38.66 <xref ref-type="table-fn" rid="tfn5-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="middle">99.8 ± 0.1</td>
<td align="center" valign="middle">−0.21</td></tr></tbody></table>
<table-wrap-foot><fn id="tfn4-sensors-11-08813">
<label>a</label>
<p>ZPD mg/capsule or injectable.</p></fn><fn id="tfn5-sensors-11-08813">
<label>b</label>
<p>Average of three experiments.</p></fn><fn id="tfn6-sensors-11-08813">
<label>c</label>
<p>Mean ± RSD (n = 4).</p></fn></table-wrap-foot></table-wrap>
<table-wrap id="t4-sensors-11-08813" position="float">
<label>Table 4.</label>
<caption>
<p>Determination of ziprasidone in human urine and serum.</p></caption>
<table frame="hsides" rules="groups">
<thead>
<tr>
<th align="center" valign="bottom"><bold>Sample</bold></th>
<th align="center" valign="bottom"><bold>Amount added</bold></th>
<th align="center" valign="bottom"><bold>Found<xref ref-type="table-fn" rid="tfn9-sensors-11-08813"><sup>c</sup></xref></bold></th>
<th align="center" valign="bottom"><bold>Recovery<xref ref-type="table-fn" rid="tfn9-sensors-11-08813"><sup>c</sup></xref> ± RSD (%)</bold></th>
<th align="center" valign="bottom"><bold>% Bias</bold></th></tr></thead>
<tbody>
<tr>
<td align="center" valign="bottom">Urine 1</td>
<td align="center" valign="bottom">2.80<xref ref-type="table-fn" rid="tfn7-sensors-11-08813"><sup>a</sup></xref></td>
<td align="center" valign="bottom">2.70</td>
<td align="center" valign="bottom">96.4 ± 1.2</td>
<td align="center" valign="bottom">−3.57</td></tr>
<tr>
<td align="center" valign="bottom">Urine 2</td>
<td align="center" valign="bottom">4.70<xref ref-type="table-fn" rid="tfn7-sensors-11-08813"><sup>a</sup></xref></td>
<td align="center" valign="bottom">4.73</td>
<td align="center" valign="bottom">100.6 ± 0.5</td>
<td align="center" valign="bottom">0.64</td></tr>
<tr>
<td align="center" valign="bottom">Urine 3</td>
<td align="center" valign="bottom">8.30<xref ref-type="table-fn" rid="tfn7-sensors-11-08813"><sup>a</sup></xref></td>
<td align="center" valign="bottom">8.08</td>
<td align="center" valign="bottom">97.4 ± 2.0</td>
<td align="center" valign="bottom">−2.65</td></tr>
<tr>
<td align="center" valign="bottom">Urine 4</td>
<td align="center" valign="bottom">12.0<xref ref-type="table-fn" rid="tfn7-sensors-11-08813"><sup>a</sup></xref></td>
<td align="center" valign="bottom">12.3</td>
<td align="center" valign="bottom">102.5 ± 1.4</td>
<td align="center" valign="bottom">2.50</td></tr>
<tr>
<td align="center" valign="bottom">Serum 1</td>
<td align="center" valign="bottom">0.90<xref ref-type="table-fn" rid="tfn8-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="bottom">0.93</td>
<td align="center" valign="bottom">103.3 ± 1.1</td>
<td align="center" valign="bottom">3.33</td></tr>
<tr>
<td align="center" valign="bottom">Serum 2</td>
<td align="center" valign="bottom">2.80<xref ref-type="table-fn" rid="tfn8-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="bottom">2.79</td>
<td align="center" valign="bottom">99.6 ± 0.3</td>
<td align="center" valign="bottom">−0.36</td></tr>
<tr>
<td align="center" valign="bottom">Serum 3</td>
<td align="center" valign="bottom">6.50<xref ref-type="table-fn" rid="tfn8-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="bottom">6.49</td>
<td align="center" valign="bottom">99.8 ± 0.2</td>
<td align="center" valign="bottom">−0.15</td></tr>
<tr>
<td align="center" valign="bottom">Serum 4</td>
<td align="center" valign="bottom">15.6<xref ref-type="table-fn" rid="tfn8-sensors-11-08813"><sup>b</sup></xref></td>
<td align="center" valign="bottom">15.88</td>
<td align="center" valign="bottom">101.8 ± 0.9</td>
<td align="center" valign="bottom">1.79</td></tr></tbody></table>
<table-wrap-foot><fn id="tfn7-sensors-11-08813">
<label>a</label>
<p>ZPD mg/L urine.</p></fn><fn id="tfn8-sensors-11-08813">
<label>b</label>
<p>ZPD μg/mL serum.</p></fn><fn id="tfn9-sensors-11-08813">
<label>c</label>
<p>Average of five experiments.</p></fn></table-wrap-foot></table-wrap></sec></back></article>