Next Article in Journal
Hybridisation Potential of 1',3'-Di-O-methylaltropyranoside Nucleic Acids
Next Article in Special Issue
Anisotropy in Bone Demineralization Revealed by Polarized Far-IR Spectroscopy
Previous Article in Journal
Volatile Compounds and Antioxidant Capacity of the Bio-Oil Obtained by Pyrolysis of Japanese Red Pine (Pinus Densiflora Siebold and Zucc.)
Previous Article in Special Issue
A Study on the Application of Near Infrared Hyperspectral Chemical Imaging for Monitoring Moisture Content and Water Activity in Low Moisture Systems
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Image Monitoring of Pharmaceutical Blending Processes and the Determination of an End Point by Using a Portable Near-Infrared Imaging Device Based on a Polychromator-Type Near-Infrared Spectrometer with a High-speed and High-Resolution Photo Diode Array Detector

1
Innovation Headquarters, Yokogawa Electric Corporation, 2-9-32 Nakacho, Musashino, Tokyo 180-8750, Japan
2
Graduate School of Agricultural Science, Tohoku University, 1-1 Amamiya, Tsutsumidori, Aobaku, Sendai 981-8555, Japan
3
School of Science and Technology, Kwansei Gakuin University, 2-1 Gakuen, Sanda, Hyogo 669-1337, Japan
4
Faculty of Life and Environmental Sciences, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8572, Japan
*
Authors to whom correspondence should be addressed.
These authors contributed equally to this work.
Molecules 2015, 20(3), 4007-4019; https://doi.org/10.3390/molecules20034007
Submission received: 15 January 2015 / Revised: 17 February 2015 / Accepted: 26 February 2015 / Published: 3 March 2015
(This article belongs to the Special Issue Advances of Vibrational Spectroscopic Technologies in Life Sciences)

Abstract

:
In the present study we have developed a new version (ND-NIRs) of a polychromator-type near-infrared (NIR) spectrometer with a high-resolution photo diode array detector, which we built before (D-NIRs). The new version has four 5 W halogen lamps compared with the three lamps for the older version. The new version also has a condenser lens with a shorter focal point length. The increase in the number of the lamps and the shortening of the focal point of the condenser lens realize high signal-to-noise ratio and high-speed NIR imaging measurement. By using the ND-NIRs we carried out the in-line monitoring of pharmaceutical blending and determined an end point of the blending process. Moreover, to determinate a more accurate end point, a NIR image of the blending sample was acquired by means of a portable NIR imaging device based on ND-NIRs. The imaging result has demonstrated that the mixing time of 8 min is enough for homogeneous mixing. In this way the present study has demonstrated that ND-NIRs and the imaging system based on a ND-NIRs hold considerable promise for process analysis.

1. Introduction

In modern pharmaceutical industries, the idea of new quality management has been proposed for the delivery of reliable pharmaceutical products. The idea is the quality by design (QbD) approach. QbD is an approach that clarifies the factors which significantly influence the quality properties of the medicine from the stage of the formulation design and the process design, and it carries out the quality control based on a scientific basis [1]. Process analytical technology (PAT) is used to analyze the manufacturing process, determine the factors essential for understanding the manufacturing process, and clarify variable factors that affect quality control [2,3]. PAT is positioned as a powerful technology to achieve the QbD approach. To evaluate the quality of an oral medication, the monitoring of component distribution including that of active pharmaceutical ingredient (API) is essential. It is well known that spectroscopic technology is attractive for the non-destructive monitoring of pharmaceutical tablets [3,4,5,6,7,8,9,10,11]. Evaluation methods for inhomogeneity in a tablet based on ultraviolet-visible (UV-Vis) spectroscopy and high performance liquid chromatography (HPLC) were developed about two decades ago [10,11]. However, the UV-Vis spectroscopy methods are usually based on destructive methods, so there are only a few reports on the use of UV-Vis spectroscopy in the quantitative analysis of tablet components. Recently a number of research groups have been involved in the development of vibrational spectroscopy methods for PAT [4,5,6,7,8,9,10]. Near-infrared (NIR) spectroscopy in particular has attracted keen interest as a powerful PAT method since it has superior features for real time monitoring; it is a non-destructive and non-contact method [4,12]. Moreover, NIR can use optical fibers, allowing remote-sensing. Imaging techniques in the NIR region have also been developing [13,14,15,16,17,18,19].
Our research group has been involved in PAT studies through the development of new NIR instruments, particularly, imaging instruments [15,19,20,21]. We recently developed a distribution type NIR spectrometer (D-NIRs), which is an NIR imaging instrument based on a polychromator-type NIR spectrometer with a high sensitivity and high-resolution photodiode array detector (PDA). The portability of D-NIRs is superior. The size of the imaging unit is only 3 liter volume. D-NIRs enables high speed monitoring (25 ms/pixel) and high wavelength resolution (1.25 nm) of NIR imaging due to the newly developed Indium Gallium Arsenide (InGaAs) detector. The detector consists of a high-density PDA containing 640 elements with 20-μm pitch. InGaAs photodiodes with a wavelength sensitivity of 900–1700 nm (photoreceptive sensitivity: 0.8 at 1550 nm) are used [20].
A NIR image obtained during the water dissolution process of a tablet demonstrated the potential of D-NIRs as a PAT tool [19,22,23,24]. Of course, NIR images of tablets measured by using D-NIRs can clarify the distribution and the relative abundance of chemical components in the tablets [15]. It was successfully demonstrated that D-NIRs is an attractive tool for PAT and that it can be applied for QbD.
However, evaluation methods for pharmaceutical blending process based on NIR imaging are still under development, although the blending process can be monitored by NIR spectra [10,11,25] and it was reported that NIR imaging technique can be utilized for studying the effects of the particle sizes of chemical components on the blending process [26]. In addition, the end point of the blending process was determined by using a NIR imaging technique [27]. However, in this case it took long time to get an image and also the wavelength resolution was very low. Thus, it would be highly desirable to develop a NIR imaging system that enables both high speed and high wavelength resolution measurements in NIR imaging.
Quite recently, as a part of the present study we have built a new type of D-NIRs named ND-NIRs. The new type has four 5 W halogen lamps as light sources, that is, one more halogen lamp than the original type. In addition, in the scanning unit of ND-NIRs the focal length of the condenser lens is shortened to optimize the performance of the optical scanning system. The increase in the number of lamps together with the shortening of the focal length has enabled about seven times faster imaging measurements compared with the original version. The purpose of the present study was to carry out the in-line monitoring of a pharmaceutical blending process and the determination of an end point of this blending process by using a ND-NIRs, polychromator-type NIR spectrometer with high speed and high wavelength resolution. Moreover, to determinate an accurate end point in a short time, a NIR image of the blending sample was acquired by means of a portable NIR imaging device based on ND-NIRs.

2. Results and Discussion

2.1. NIR Spectra of Ascorbic Acid, D-mannitol and Magnesium Stearate

Figure 1 shows: (a) diffuse-reflectance (DR) NIR spectra and (b) the second derivative spectra of powder of ascorbic acid (AsA), D-mannitol and magnesium stearate in the 950–1700 nm region. Narrow bands of AsA at around 1361 and 1458 nm may be due to the first overtones of stretching vibrations of free and intermolecular hydrogen-bonded OH groups, respectively [28]. A broad absorption in the 1150–1250 nm region of AsA may be assigned to the second overtones of CH stretching vibrations [28]. A broad absorption in the 1490–1590 nm region of D-mannitol may be assigned to the first overtones of OH stretching vibrations.

2.2. Evaluation of Inhomogeneity during Blending Process by In-Line NIR Spectrometer

Figure 2 shows a series of second derivative spectra of the blending sample measured during rotation mixing. The inset of Figure 2 depicts the enlargement of the 1445–1475 nm region. It is noted that the peak intensity at 1458 nm due to the first overtone of the OH stretching vibration mode of AsA changes with time. The band of 1458 nm is not detected in the initial stage of the mixing. However, this band becomes strong as mixing proceeds. The spectrum measured at 0 s reflects D-mannitol only; the band at 1495 nm arises only from the first overtone of the OH stretching mode. The intensity at 1495 nm becomes intense with mixing.
Figure 1. (a) NIR spectra of ascorbic acid, D-mannitol and magnesium stearate; (b) The second derivative of the spectra shown in (a).
Figure 1. (a) NIR spectra of ascorbic acid, D-mannitol and magnesium stearate; (b) The second derivative of the spectra shown in (a).
Molecules 20 04007 g001
Figure 2. The second derivative spectra of mixing sample measured during the blending process.
Figure 2. The second derivative spectra of mixing sample measured during the blending process.
Molecules 20 04007 g002
Figure 3 plots a 5-point moving block standard deviation of the peak intensity at 1458 nm in the second-derivative spectra versus time. We also prepared similar plot by using the band at 1495 nm due to D-mannitol. The 1495 nm band gave similar result to the 1458 nm band. It is noted in Figure 3 that during the initial stage of mixing, the powder sample in the vessel is inhomogeneous, and therefore, the standard deviation (SD) value is high. As mixing proceeds, the sample becomes homogeneous. Then, the spectra of the mixtures of ascorbic acid, D-mannitol and magnesium stearate as the mixing ratio are detected, and thus, the spectral variation decreases, yielding a low SD value. The result in Figure 3 shows that the SD gives a constant value around 168 s after the start of rotation mixing. Thus, it is very likely that the sample in the bottle becomes homogeneous after 168 s. Note that the SD does not exceed 0.00002 after 168 s.
Figure 3. A plot for standard deviation of the second derivative intensity at 1,458 nm versus time. The DR-NIR spectrum was measured with 4 s intervals.
Figure 3. A plot for standard deviation of the second derivative intensity at 1,458 nm versus time. The DR-NIR spectrum was measured with 4 s intervals.
Molecules 20 04007 g003

2.3. Evaluation of Inhomogeneity during Blending Process by NIR Imaging

Figure 4, Figure 5 and Figure 6 depict: (a) images for the distribution of AsA and (b) binary images for mixing times of 1, 8, and 15 min. A binary image is the two color image, it usually consists of white and black color. To determine the color in an image, the arbitrary threshold is defined prior to the visualization from the digital numbers and intensities, etc. of all pixels. Figure 7a–c show the second derivative spectra collected at different points in the two-dimensional image of the blend sample at 1 min after the start of blending. In the measurement of two-dimensional image of blend sample NIR spectra acquisition, second differentiation treatment, and absorbance mapping using specific wavelength were carried out at the same time. The ND-NIRs allows the spectral acquisition in 242 s for a measurement area of 5 mm × 5 mm, spatial resolution 0.1 mm in the wavelength region of 934–1713 nm, with the wavelength resolution of 1 nm. In this system the number of data is 2601 pixels times 780 absorbance data = 202,878 datapoints. Table 1 summarizes the measurement times for various measurement conditions. The results in the table demonstrate that ND-NIRs can provide two-dimensional spectra imaging data at a high speed. Second derivative spectra of one pixel (one point) shown in Figure 7 demonstrate that they catch clearly the specific absorption wavelength of 1458 nm due to AsA. These results suggest that ND-NIRs has high signal-to-noise ratio and high wavelength resolution, enabling the correction of base line deviation and the band separations without the influence of the increase of noise induced by the second derivative treatment. Each point spectrum obtained from the binary images corresponds well to the spectrum of AsA, showing that the sample existing in the points is AsA.
Table 1. Measurement time for each condition by ND-NIRs.
Table 1. Measurement time for each condition by ND-NIRs.
Measurement Area5 mm × 5 mm
Spatial Resolution1 mm0.5 mm0.2 mm0.1 mm
Number of Measurement Point (Number of Spectra Data)361216762061
Measurement Time5 s13 s64 s242 s
Figure 4. (a) The second derivative intensity mapping image of the mixing sample at 1 min after the start of the blending. (b) A binary image of the sample at 1 min after the start of the blending. A binary image developed by arbitrary threshold value.
Figure 4. (a) The second derivative intensity mapping image of the mixing sample at 1 min after the start of the blending. (b) A binary image of the sample at 1 min after the start of the blending. A binary image developed by arbitrary threshold value.
Molecules 20 04007 g004
Figure 5. (a) The second derivative intensity mapping image of the mixing sample at 8 min after the start of the blending. (b) A binary image of the sample at 8 min after the start of the blending. A binary image developed by arbitrary threshold value.
Figure 5. (a) The second derivative intensity mapping image of the mixing sample at 8 min after the start of the blending. (b) A binary image of the sample at 8 min after the start of the blending. A binary image developed by arbitrary threshold value.
Molecules 20 04007 g005
Figure 6. (a) The second derivative intensity mapping image of the mixing sample at 15 min after the start of the blending. (b) A binary image of the sample at 15 min after the start of the blending. A binary image developed by arbitrary threshold value.
Figure 6. (a) The second derivative intensity mapping image of the mixing sample at 15 min after the start of the blending. (b) A binary image of the sample at 15 min after the start of the blending. A binary image developed by arbitrary threshold value.
Molecules 20 04007 g006
Figure 7. (a) The second derivative spectra at X3.8, Y0.5 and X1.4, Y4.7 points of the binary image of the blending at 1 min after the start of the blending. (b) Those at X0.0, Y0.0 and X3.5, Y1.4 points. (c) Those at X0.0, Y3.7 and X2.4, Y1.0 points.
Figure 7. (a) The second derivative spectra at X3.8, Y0.5 and X1.4, Y4.7 points of the binary image of the blending at 1 min after the start of the blending. (b) Those at X0.0, Y0.0 and X3.5, Y1.4 points. (c) Those at X0.0, Y3.7 and X2.4, Y1.0 points.
Molecules 20 04007 g007
We calculated area ratios (pixel ratios) of AsA and non-AsA from the binary images, and then estimated the temporal change in the relative amount of AsA. The area ratio of AsA in the binary images of two-dimensional spectra at the mixing time of 1, 8, and 15 min was 1.49%, 9.57%, and 8.74%, respectively. The relative amount of AsA at the mixing time of 8 and 15 min was almost the same as that the initial sample throw amount (10%). On the other hand, the relative amount of AsA at the mixing time of 1 min was much smaller, indicating that the mixing is insufficient. From the present result we can conclude that the mixing time of 8 min is enough for homogeneous mixing.

3. Experimental Section

3.1. New NIR Imaging Device (ND-NIRs)

A new version of D-NIRs (Yokogawa Electric Co., Tokyo, Japan) has been developed for the collection of two-dimensional DR-NIR spectra data. Its imaging unit consists of galvanic mirrors, condenser lenses, a rectangular prism, a fiber-optic cable, and four halogen lamps as shown in Figure 8 and Figure 9. There are two significant major differences between the original version (D-NIRs) and the new version (ND-NIRs). One is the number of lamps. The new version has four 5 W halogen lamps while the old version had only three. This increase in the number of lamps increased the light power significantly. Another important difference between the two versions is the length of the focal point of the condenser lens. The length has become significantly shorter in the new version (70 mm) compared with that in the old version (100 mm), greatly increasing the signal intensity. The increase in the number of the lamps and shortening of the focal length of the condenser lens achieve high signal-to-noise ratio, enabling high-speed NIR imaging measurement. Figure 10 plots a root mean square (rms) noise of D-NIRs and ND-NIRs at the high-light flux. The rms noise evaluation test between D-NIRs and ND-NIRs that was carried out in accordance with the US Pharmacopeia [29]. The rms noise level of the old version is 0.5 mAbs/0.5 s, while that of the new version is 0.05 mAbs/0.5 s. The measurement speed of the new version (11 s/100 pixels) has become seven times higher compared with that of the original one (80 s/100 pixels).
It is also notable that the size of the new imaging unit is only 220 × 90 × 150 mm due to the contribution of these optimized elements in the imaging unit. The irradiation energy from the four halogen lamps reaches a sample, and the diffuse reflected reaches the galvanic mirror. Finally, this signal is detected by a spectrometer named P-NIRs through the optical fiber. P-NIRs was also developed as a new spectrometer by our research group [20]. The high speed spectra mapping of 13 s per 25 mm2 area of ND-NIRs is due to the P-NIRs. The superior futures of P-NIRs were reported in a previous paper [20,21]. The maximum scan area of ND-NIRs is 10 × 10 mm, and its spatial resolution is 25 μm.

3.2. Data Collection

The blending machine was rotated by 15 rpm in 15 min. The spectrometer, P-NIRs, was placed under this machine (Figure 11), and when the machine returned to the original position, the DR-NIR spectrum of blending powder was recorded by P-NIRs. The NIR spectra in the 950–1700 nm region were measured with a 1 nm interval, and they were converted to CSV format immediately. These spectral data were subjected to Savitzky-Golay (SG) smoothing and then to second derivative treatment with 19 point, and 2nd order [30]. Two-dimensional measurements of DR-NIR spectra were carried out several times by using a ND-NIRs. DR-NIR spectra at each pixel were obtained by line scanning method, and the peak at 1458 nm due to the first overtone of OH stretching vibration of AsA identified by the point measurement was used for the NIR mapping.
Figure 8. Overview of the developed NIR imaging device (ND-NIRs). A: X axis galvano mirror, B: Y axis galvano mirror, C: 5 W halogen lamp, D: Condensing lens 1, E: Right angle prism mirror, F: Condensing lens 2, G: Optical fiber cable.
Figure 8. Overview of the developed NIR imaging device (ND-NIRs). A: X axis galvano mirror, B: Y axis galvano mirror, C: 5 W halogen lamp, D: Condensing lens 1, E: Right angle prism mirror, F: Condensing lens 2, G: Optical fiber cable.
Molecules 20 04007 g008
Figure 9. Outline of D-NIRs. A: P-NIRs, B: XY scanning unit, C: control unit, D: Sample plinth with temperature control function.
Figure 9. Outline of D-NIRs. A: P-NIRs, B: XY scanning unit, C: control unit, D: Sample plinth with temperature control function.
Molecules 20 04007 g009
Figure 10. A plot for root mean square noise of D-NIRs and ND-NIRs at high-light flux.
Figure 10. A plot for root mean square noise of D-NIRs and ND-NIRs at high-light flux.
Molecules 20 04007 g010
Figure 11. Outline of the blending test setup. A: P-NIRs for the in-line blending process monitor, B: Vessel-type blending machine.
Figure 11. Outline of the blending test setup. A: P-NIRs for the in-line blending process monitor, B: Vessel-type blending machine.
Molecules 20 04007 g011

3.3. Sample Preparation

Ascorbic acid (AsA), magnesium stearate and D-mannitol were purchased from Kanto Chemical Co., Inc. (Tokyo, Japan), and their powders were prepared for blending. These materials were introduced into a vessel type blending machine (Tsutsui Scientific Instruments Co., Ltd., Taito-ku, Japan). The outline of the blending test setup is shown in Figure 3. To monitor the blending process, each material was added in the order D-mannitol, magnesium stearate and AsA. Finally, the prepared powder consisted of 88% (w/w) D-mannitol, 2% (w/w) magnesium stearate and 10% (w/w) AsA.

4. Conclusions

We have developed a new version of a polychromator-type NIR spectrometer with a high-resolution photo diode array detector, which we built before (D-NIRs). The new version has achieved a high signal-to-noise ratio and a high speed. The noise level of the new version is 10 times lower than that of the old version, and the measurement speed of the new version is seven times higher than that of the original one. By using the new version we performed an in-line evaluation of inhomogeneity during a blending process and determined the end point for homogeneous mixing. The evaluation of inhomogeneity during the blending process was also carried out by NIR imaging. Two-dimensional spectra of mixing samples were measured at different times. The NIR imaging revealed that the distribution of AsA in the blend sample changes with time. It has turned out that the mixing time of 8 min is sufficient for homogeneous mixing. The present study has demonstrated that NIR imaging is very useful for further understanding of the result of in-line monitoring of blending process.

Acknowledgments

The study was supported by “Innovation Promotion Program” of New Energy and Industrial Technology Development Organization (NEDO), Ministry of Economy, Trade and Industry, Tokyo, Japan.

Author Contributions

K.M., D.I. and T.G. designed research; K.M., H.S., T.G. and M.K. performed the experimental work; K.M., D.I. and Y.O. analyzed data; K.M. and Y.O. wrote the paper.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use. ICH Harmonized Tripartite Guideline Pharmaceutical Development Q8 (R2). Available online: http://www.ich.org/fileadmin/Public_Web_Site/ICH_Products/Guidelines/Quality/Q8_R1/Step4/Q8_R2_Guideline.pdf (accessed on 15 January 2015).
  2. US Food Drug Administration. Guidance for Industry, PAT: A Framework for Innovative Pharmaceutical Development, Manufacturing, and Quality Assurance. Available online: http://www.fda.gov/downloads/Drugs/Guidances/ucm070305.pdf (accessed on 15 January 2015).
  3. Bakeev, K.A. Process Analytical Technology, 1st ed.; Blackwell Publishing: Oxford, UK, 2005; pp. 13–38. [Google Scholar]
  4. Ozaki, Y.; Amari, T. Near-Infrared Spectroscopy in Chemical Process Analysis; Sheffield Academic Press: Sheffield, UK, 2000; pp. 53–95. [Google Scholar]
  5. Andersson, M.; Folestad, S.; Gottfries, J.; Johansson, M.O.; Josefson, M.; Wahlund, K.G. Quantitative analysis of film coating in a fluidized bed process by in-line NIR spectrometry and multivariate batch calibration. Anal. Chem. 2000, 72, 2099–2108. [Google Scholar] [CrossRef] [PubMed]
  6. Kogermann, K.; Aaltonen, J.; Strachan, C.J.; Pollanen, K.; Heinamaki, J.; Yliruusi, J.; Rantanen, J. Establishing quantitative in-line analysis of multiple solid state transformations during dehydration. J. Pharm. Sci. 2008, 97, 4983–4999. [Google Scholar] [CrossRef] [PubMed]
  7. Li, W.; Worosila, G.D.; Wang, W.; Mascaro, T. Determination of polymorph conversion of an active pharmaceutical ingredient in wet granulation using NIR calibration models generated from the premix blends. J. Pharm. Sci. 2005, 94, 2800–2806. [Google Scholar] [CrossRef] [PubMed]
  8. Trafford, A.D.; Jee, R.D.; Moffatand, A.C.; Graham, P. A rapid quantitative assay of intact paracetamol tablets by reflectance near-infrared spectroscopy. Analyst 1999, 124, 163–167. [Google Scholar] [CrossRef] [PubMed]
  9. Rantanen, J.; Räsänen, E.; Tenhunen, J.; Känsäkoski, M.; Mannermaa, J.P.; Yliruusi, J. In-line moisture measurement during granulation with a four-wavelength near infrared sensor: An evaluation of particle size and binder effect. Eur. J. Pharm. Biopharm. 2000, 50, 271–276. [Google Scholar] [CrossRef] [PubMed]
  10. Sekulic, S.S.; Ward, H.W., II; Brannegan, D.R.; Stanley, E.D.; Evans, C.L.; Sciavolino, S.T.; Hailey, P.A.; Aldridge, P.K. On-Line Monitoring of Powder Blend Homogeneity by Near-Infrared Spectroscopy. Anal. Chem. 1996, 68, 509–513. [Google Scholar] [CrossRef] [PubMed]
  11. El-Hagrasy, A.S.; Delgado-Lopez, M.; Drennen, J.K., III. A Process Analytical Technology approach to near-infrared process control of pharmaceutical powder blending: Part II: Qualitative near-infrared models for prediction of blend homogeneity. J. Pharm. Sci. 2006, 95, 407–421. [Google Scholar] [CrossRef] [PubMed]
  12. Ozaki, Y. Near-infrared spectroscopy—Its versatility in analytical chemistry. Anal. Sci. 2012, 28, 545–563. [Google Scholar] [CrossRef] [PubMed]
  13. Šašić, S.; Ozaki, Y. Raman, Infrared, and Near-Infrared Chemical Imaging; John Wiley & Sons Ltd.: Hoboken, NJ, USA, 2009. [Google Scholar]
  14. Shinzawa, H.; Awa, K.; Ozaki, Y. Compression induced morphological and molecular structural changes of cellulose tablets probed with near infrared imaging. J. Near Infrared Spectrosc. 2011, 19, 15–22. [Google Scholar] [CrossRef]
  15. Ishikawa, D.; Murayama, K.; Genkawa, T.; Awa, K.; Komiyama, M.; Ozaki, Y. Development of compact near infrared imaging device with high-speed and portability for pharmaceutical process monitoring. NIR News 2012, 23, 14–17. [Google Scholar] [CrossRef]
  16. Unger, M.; Ozaki, Y.; Siesler, H.W. Variable-Temperature Fourier Transform Near-Infrared (FT-NIR) Imaging Spectroscopy of the diffusion process of Butanol (OD) into Polyamide 11. Appl. Spectrosc. 2011, 65, 1051–1055. [Google Scholar] [CrossRef] [PubMed]
  17. Neil Lewis, E.; Schoppelrei, J.; Lee, E. Near-infrared Chemical Imaging and the PAT Initiative-NIR-CI adds a completely new dimension to conventional NIR spectroscopy. Spectroscopy 2004, 19, 26–36. [Google Scholar]
  18. Gao, P.; Meury, R.H. Swelling of hydroxypropyl methylcellulose matrix tablets. 1. Characterization of swelling using a novel optical imaging method. J. Pharm. Sci. 1996, 85, 725–731. [Google Scholar] [CrossRef] [PubMed]
  19. Ishikawa, D.; Murayama, K.; Genkawa, T.; Awa, K.; Komiyama, M.; Sergei, K.; Ozaki, Y. Application of a newly developed portable NIR imaging device to dissolution process monitoring of tablets. Anal. Bioanal. Chem. 2013, 405, 9401–9409. [Google Scholar] [CrossRef] [PubMed]
  20. Murayama, K.; Genkawa, T.; Ishikawa, D.; Komiyama, M.; Ozaki, Y. A polychromator-type near-infrared spectrometer with a high-sensitivity and high-resolution photodiode array detector for pharmaceutical process monitoring on the millisecond time scale. Rev. Sci. Instrum. 2013, 84, 023104. [Google Scholar] [CrossRef] [Green Version]
  21. Ishikawa, D.; Genkawa, T.; Murayama, K.; Komiyama, M.; Ozaki, Y. Feasibility study of diffuse reflectance and transmittance near infrared spectroscopy for rapid analysis of ascorbic acid concentration in bilayer tablets using a high-speed polychromator-type spectrometer. J. Near Infrared Spectrosc. 2014, 22, 189–197. [Google Scholar] [CrossRef]
  22. Otsuka, M.; Tanabe, H.; Osaki, K.; Otsuka, K.; Ozaki, Y. Chemoinformetrical Evaluation of Dissolution Property of Indomethacin Tablets by Near-Infrared Spectroscopy. J. Pharm. Sci. 2007, 96, 788–801. [Google Scholar] [CrossRef] [PubMed]
  23. Van der Weerd, J.; Kazarian, S.G. Combined approach of FTIR imaging and conventional dissolution tests applied to drug release. J. Control. Release 2004, 98, 295–305. [Google Scholar] [CrossRef] [PubMed]
  24. Kazarian, S.G.; van der Weerd, J. Simultaneous FTIR Spectroscopic Imaging and Visible Photography to Monitor Tablet Dissolution and Drug Release. Pharm. Res. 2004, 25, 853–860. [Google Scholar] [CrossRef]
  25. Ufret, C.; Morris, K. Modeling of powder blending using on-line near infrared measurements. Drug Dev. Ind. Pharm. 2001, 27, 719–729. [Google Scholar] [CrossRef] [PubMed]
  26. Li, W.; Woldu, A.; Kelly, R.; McCool, J.; Bruce, R.; Rasmussen, H.; Cunningham, J.; Winstead, D. Measurement of drug agglomerates in powder blending simulation samples by near infrared chemical imaging. Int. J. Pharm. 2008, 350, 369–373. [Google Scholar] [CrossRef] [PubMed]
  27. El-Hagrasy, A.S.; Morris, H.R.; D’amico, F.; Lodder, R.A.; Drennen, J.K., III. Near-Infrared Spectroscopy and Imaging for the Monitoring of Powder Blend Homogeneity. J. Pharm. Sci. 2001, 90, 1298–1307. [Google Scholar] [CrossRef] [PubMed]
  28. Lui, H.; Xiang, B.; Qu, L. Structure analysis of ascorbic acid using near-infrared spectroscopy and generalized two dimensional correlation spectroscopy. J. Mol. Struct. 2006, 794, 12–17. [Google Scholar] [CrossRef]
  29. Cogdill, R.P.; Anderson, C.A.; Delgado-Lopez, M.; Molseed, D.; Chisholm, R.; Bolton, R.; Herkert, T.; Afnan, A.M.; Drennen, J.K., III. Process Analytical Technology Case Study Part I: Feasibility Studies for Quantitative Near-Infrared Method Development. AASP Pharm. Sci. Tech. 2005, 6, 262–272. [Google Scholar] [CrossRef]
  30. Savitzky, A.; Golay, M.J.E. Smoothing and differentiation of data by simplified least squares procedures. J. Anal. Chem. 1964, 36, 1627–1639. [Google Scholar] [CrossRef]
  • Sample Availability: Not available.

Share and Cite

MDPI and ACS Style

Murayama, K.; Ishikawa, D.; Genkawa, T.; Sugino, H.; Komiyama, M.; Ozaki, Y. Image Monitoring of Pharmaceutical Blending Processes and the Determination of an End Point by Using a Portable Near-Infrared Imaging Device Based on a Polychromator-Type Near-Infrared Spectrometer with a High-speed and High-Resolution Photo Diode Array Detector. Molecules 2015, 20, 4007-4019. https://doi.org/10.3390/molecules20034007

AMA Style

Murayama K, Ishikawa D, Genkawa T, Sugino H, Komiyama M, Ozaki Y. Image Monitoring of Pharmaceutical Blending Processes and the Determination of an End Point by Using a Portable Near-Infrared Imaging Device Based on a Polychromator-Type Near-Infrared Spectrometer with a High-speed and High-Resolution Photo Diode Array Detector. Molecules. 2015; 20(3):4007-4019. https://doi.org/10.3390/molecules20034007

Chicago/Turabian Style

Murayama, Kodai, Daitaro Ishikawa, Takuma Genkawa, Hiroyuki Sugino, Makoto Komiyama, and Yukihiro Ozaki. 2015. "Image Monitoring of Pharmaceutical Blending Processes and the Determination of an End Point by Using a Portable Near-Infrared Imaging Device Based on a Polychromator-Type Near-Infrared Spectrometer with a High-speed and High-Resolution Photo Diode Array Detector" Molecules 20, no. 3: 4007-4019. https://doi.org/10.3390/molecules20034007

Article Metrics

Back to TopTop